AWB 210 Water Bath AMOSAWB210SM20140309
---2---
take down the power source module from the main board by hand,if it is
difficult to take down the power source module with hand,you can use the
flat apparatus as slotted screwdriver to take down the power source module
by inserting the flat apparatus in the slot and tipping it up slightly, and
then take it down and replace a new power source module.
B、It shows incorrect temperature(offset the set temperature)
1):Digital sensor is broken
Maintenance : Loosens the screw with the screwdriver, and you will see
the digital sensor as it showed in the picture, and then Loosens the screw
in the sensor to replace the sensor.
2):IC is broken
Maintenance: Loosens the screw with the screwdriver, and you will see the
IC as it showed in the picture. you can use the flat apparatus as slotted
screwdriver to take down the IC by inserting the flat apparatus in the slot
and tipping it up slightly and then take it down to replace a new one.
C、There is continuous ring in buzzer(the interval is 10 seconds)
It means that the digital sensor is broken, and need to replace.