POWERING UP - FIRST TIME USE - WITHOUT INSULATING TAB
Use this procedure the first time you power up a TW-DM-01/BL; the detector has not been supplied with the insulating tab.
1) Make sure the Link / program switch is set on “ON”.
2) Insert the two supplied batteries into their device’s lodgments.
POWERING UP - DEVICE LINKED TO THE SYSTEM
Use this procedure when a TW-DM-01/BL is successfully linked to its Taurus system and you have to extract one or both batteries (e.g.
batteries substitution).
1) Reinsert the battery or both batteries into their lodgments.
If performing a batteries substitution, use two brand new batteries and substitute both of them.
Do not touch the Link / program switch.
POWERING UP - RECOVERY
Use this procedure when you fail to link successfully a TW-DM-01/BL or you want to link it again.
1) Move alternatively the Link / program switch 5 times.
2) Set the Link / program switch on “ON”.
3) Insert the two supplied batteries into their device’s lodgments.
LINKING - WAKE-UP - WITHOUT INSULATING TAB
“Wake-up” linking consists in associating one or more child devices to the Taurus system altogether in a single operation.
Wake-up is performed either through the TauREX software or the TW-MTI-01 / TW-MEC-01 keyboard-screen interface; it CANNOT be done
through TW-ME-01 devices.
1) Create the “virtual model” of the TW-DM-01/BL either on TauREX or on the TW-MTI-01 / TW-MEC-01.
2) Power-up the detector (either “first time use” or “recovery”).
3) Set the Link / program switch OPPOSITE to “ON”.
4) Trigger the wake-up procedure either from TauREX or from the TW-MTI-01 / TW-MEC-01.
5) Wait the end of the “wake-up” linking procedure.
6) Check on TauREX or from TW-MTI-01 / TW-MEC-01 for linking success. Consult their user manual.
LINKING - ONE-BY-ONE - WITHOUT INSULATING TAB
“One-by-one” linking consists in associating one child device at a time to the Taurus system.
This operation is performed either through the TauREX software or the TW-MTI-01 / TW-MEC-01 keyboard-screen interface; it CANNOT be
done through TW-ME-01 devices.
1) Create the “virtual model” of the child device either on TauREX or on the TW-MTI-01 / TW-MEC-01.
2) Trigger the linking procedure either from TauREX or from the TW-MTI-01 / TW-MEC-01.
3) Power-up the child device (either “first time use” or “recovery”).
4) Set the child device’s Link / program switch OPPOSITE to “ON”.
5) Wait the end of the “one-by-one” linking procedure.
6) Check on TauREX or from TW-MTI-01 / TW-MEC-01 for linking success. Consult their user manual.
TESTING
Magnet test
1) Activate test mode.
2) Apply again the magnet in correspondence of the “magnet test activation area”.
3) LED indicators signal “Alarm activation”.
Aerosol test
1) Activate test mode.
2) Apply the aerosol test device to the detector.
3) Wait a few seconds.
4) LED indicators signal “Alarm activation”.
Heat test
1) Activate test mode.
2) Apply the heat test device to the detector.
3) Wait a few seconds.
4) LED indicators signal “Alarm activation”.
BATTERY FAULTS AND BATTERY SUBSTITUTION PROCEDURE
When one or both batteries are low in charge, a specific fault message is routed to the control panel. If such event occurs:
1) Remove the safety screw.
2) Remove the detector from its base.
3) Remove the batteries cover.
4) Extract both batteries.
5) Insert both new batteries into their holders, oriented as per polarity marks.
See POWERING UP - DEVICE LINKED TO THE SYSTEM.
6) Reinstall the batteries cover.
7) Reinstall the detector.
8) Reinstall the safety screw.
Local safety standards may require you to
test these devices on a regular basis.
Use only suitable aerosol testers / heat test devic-
es supplied by approved manufacturers. Follow
their specific use instructions.
Before testing every TW-DM-01/BL, always
activate test mode. This is done by holding
a suitable magnet in the “magnet test
activation area”. When activated, LED indicators
signal “Test mode active”.
When a low battery condition is indicated,
both batteries must be changed altogether.
Batteries must be brand new.
Do not touch the Link / program switch.
Ensure that the batteries are installed properly,
with their polarities matching the indications on
the device.
Always ensure that the batteries are
installed properly, with their polarities
matching the indications on the device.
ARGUS SECURITY S.R.L. - Via del Canneto, 14 - 34015 - Muggia (TS) - Italy www.argussecurity.it
3
MAINTENANCE - CLEANING
1) Remove the safety screw.
2) Remove the detector from its base.
3) Smoke entry areas and thermistor area: use a small, soft bristle brush to dislodge any obvious contaminants such as insects, spider webs,
hairs, etc.
4) Smoke entry areas and thermistor area: use a small vacuum tube or dry, clean, compressed air to suck up or blow any remaining small
particles away.
5) Wipe the exterior housing of the detector with a clean, damp, lint-free cloth to remove any surface film that can later attract airborne
contaminants.
6) Install the detector onto its base again.
7) Test the detector.
8) Reinstall the safety screw.
TECHNICAL SPECIFICATIONS *
*See TDS-TWDMX technical specification document for further technical data.
BATTERIES SPECIFICATIONS
*Batteries lifespan depends by environmental conditions, default monitor settings and link quality.
SMOKE SENSITIVITY SPECIFICATIONS
*Detector’s smoke sensitivity can be set through TauREX.
Specification Value
Communication range with TW-MTI-01, TW-MEC-01 or TW-ME-01
network devices 200 m (in open space)
Wireless frequency band 868 MHz
Number of wireless channels 66
Radiated power 14 dBm (25 mW)
Temperature alarm threshold (static) 58 °C
Operating temperature range -10 °C to 55 °C
Maximum humidity (non condensing) 95% RH
IP rating 40
Specification Value
Batteries type CR123A (3 V, 1.25 Ah)
Batteries lifespan * 10 years
Low battery threshold value (nominal) 2.850 V
Detector’s sensitivity setting * Obscuration threshold value for alarm activation
High sensitivity 0.12 dB/m
Medium sensitivity (default setting) 0.15 dB/m
Low sensitivity 0.18 dB/m
Table 2
Table 3
Table 4
WARNINGS AND LIMITATIONS
Our devices use high quality electronic components and plastic materials that are highly resistant to environ-
mental deterioration. However, after 10 years of continuous operation, it is advisable to replace the devices in
order to minimize the risk of reduced performance caused by external factors. Ensure that this device is only
used with compatible control panels. Detection systems must be checked, serviced and maintained on a
regular basis to confirm correct operation. Smoke detectors may respond differently to various kinds of
smoke particles, thus application advice should be sought for special risks. Detectors cannot respond correct-
ly if barriers exist between them and the fire location and may be affected by special environmental condi-
tions. Refer to and follow national codes of practice and other internationally recognized fire engineering
standards. Appropriate risk assessment should be carried out initially to determine correct design criteria and
updated periodically.
Use only in Taurus fire detection and alarm systems.
WARRANTY
All devices are supplied with the benefit of a limited 5 years warranty relating to faulty materials or manufac-
turing defects, effective from the production date indicated on each product. This warranty is invalidated by
mechanical or electrical damage caused in the field by incorrect handling or usage. Product must be returned
via your authorized supplier for repair or replacement together with full information on any problem identified.
Full details on our warranty and product’s returns policy can be obtained upon request.
ARGUS SECURITY S.R.L.
Via del Canneto, 14
34015 Muggia (TS)
Italy
TW-DM-01/BL
EN 54-25:2008
EN 54-7:2018
EN 54-5:2017 + A1:2018
Class A1R
4
0051
22
TU0210CPR