Page 10
1. Verify fuse values. Fuses should be 1.6A 250V SLO-
BLO (830040).
2. Check the console for physical imperfections. Verify
that the Foot Switch and Motor connectors are tight by
connecting the relevant cable and attempting to twist
the connector to the left and right by hand.
3. Perform lectrical Safety Tests: Refer to page 17 for
complete instructions on performing the Electrical
Safety Tests on the AEU-7000L and AEU-7000E
units.
4. Connect field programmer AFP-01 to Programming
Port 1 at the rear of the console. Insert memory card
containing Master Code Firmware PN 890048 (7000L)
or PN 890022 (7000 ) into the AFP-01. Turn on the
AFP-01 and wait for the programming process to
complete. (Refer to pages 19 & 20 for complete
Master Code Firmware programming instructions.)
5. Connect field programmer AFP-01 to Programming
Port 2 at the rear of the console. Insert memory card
containing Slave Code Firmware PN 890049 (7000L)
or PN 890023 (7000 ) into the AFP-01. Turn on the
AFP-01 and wait for the programming process to
complete. (Refer to pages 19 & 20 for complete
Slave Code Firmware programming instructions.)
6. Insert System Software memory card PN MC-7000L or
PN MC-7000 into the Memory Card Port at the rear
of the console. Connect a power cord to the console.
Turn the console power switch on and follow the on-
screen prompts to program the console. Verify that the
programming process completed successfully and that
the software version displayed at console startup was
correct. (Refer to page 18 for complete System
Software programming instructions.)
7. Verify that the motor and calibration systems are
functional:
7.1. Attach an A -230L-40 (7000L) or A -230M-40
(7000 ) motor with a 20:1 handpiece and an A -
7PM foot control to the console.
7.2. Use Ratio keys to set Ratio to 20:1 on the display.
Depress the footswitch and verify that the motor
runs. Press the CAL button and follow the on-
screen prompts to complete the calibration
process, verifying that the calibration completed
successfully. If calibration fails, refer to page 12
for Calibration Error Testing instructions.
7.3. Press the S L CT button to enter “ ndo Mode”.
Use the Ratio keys to set the handpiece ratio to
8:1. Attach an 8:1 handpiece to the motor. Press
the CAL button and follow the on-screen prompts
to complete the calibration process, verifying that
the calibration completed successfully. If
calibration fails, refer to page 12 for Calibration
Error Testing instructions. Press the S L CT
button again to return to “Implant Mode”.
8. Connect an A -70V2 foot control and depress each of
the buttons. Verify that the function of each switch is
(from lower left, clockwise): Preset Increment, Motor
direction, Torque Increment, Flow ON/OFF. Select
Preset 1. Press on the variable pedal and verify that
the motor operates in proportion to the pedal position.
(Refer to Operator’s Manual for complete
instructions on installing and using the AE-70V2
foot control.)
9. Change the motor direction to reverse. Press and hold
the footswitch for several seconds and verify that the
motor runs and that the console audible indicator
beeps consistently.
10.Open and close the pump door several times and
verify that the door opens, closes, and latches
smoothly. Insert a length of pump tubing into the
peristaltic pump and verify that the pump latches
securely. With the flow set to 100% depress the
footswitch and verify that the pump runs smoothly.
11.7000L UNIT ONLY: Verify operation of the hand motor
light and associated user interface functionality:
11.1. Press the LIGHT button to turn the keypad light
L D off. Press the footpedal and verify that the
hand motor light stays off.
11.2. Press the LIGHT button to turn the keypad light
L D on. Press the footpedal and verify that the
hand motor light turns on.
12.Verify operation and tactile feedback of the keypad
buttons: (Refer to Operator’s Manual for complete
instructions on using the unit’s control panel).
12.1. Press the RATIO UP and RATIO DOWN buttons,
verifying that the displayed ratio changes with
each press.
12.2. Press the SP D UP and SP D DOWN
buttons, verifying that the displayed speed
changes with each press.
12.3. Press the FWD/R V button repeatedly and verify
that the FWD and R V L Ds change with each
press.
12.4. Press the TORQU UP and TORQU DOWN
buttons, verifying that the displayed torque
changes with each press.
12.5. Press the FLOW ON/OFF button repeatedly to
turn the pump off and then back on, and verify
that the green FLOW ON L D illuminates. Press
the FLOW UP and FLOW DOWN buttons,
verifying that the displayed flow setting changes
with each press.
12.6. Set Ratio to 20:1. Press the AS/MAX button
repeatedly and verify that the A-S and MAX
L Ds illuminate.
12.7. Press each of the preset buttons and verify that
the L D for each preset button illuminates.
FINAL TEST VERIFICATION