Handpi c S l ctor:
Selects high-speed (HP1) or low-speed
(HP2) handpiece operation.
Wat r On/Off Toggl :
Controls the on/off flow of water coolant to
the high-speed handpiece.
Wat r Flow Control:
Adjusts the water coolant flow to the high-
speed handpiece.
Pr ssur Gaug :
Gives a visual indication of the drive air
pressure to the handpiece in use.
Driv Air Pr ssur Control:
Adjusts the drive air pressure to the high
and low-speed handpieces.
Figure 3 DEPRESS
LOCKING COLLAR
Foot Control:
Controls the drive air pressure to the
active handpiece. Apply light foot
pressure to any part of the foot control
cover.
3-Way Air /Wat r Syring :
Delivers air, water, or a combination mist.
Supplied with autoclavable tip.
NWS-7 Wat r Syst m (Optional):
White 1-liter bottle dispenses water
through the 3-Way Air/Water Syringe and
high-speed handpiece.
AA-61 Wat r Conn ctor:
Provides water via standard 1/4” quick
disconnect to independent Cavitron or
scaler units requiring a water supply.
OPERATION FUNCTIONS:
(Refer to Fig. 2)
UNIT OPERATION:
Handpi c s - Connect the proper high-speed
and low-speed handpieces to their appropriate
connector. Switch the handpiece selector tog-
gle to the desired handpiece. Depress foot
control for operation.
3-Way Air/Wat r Syring - Depress the right
button for air operation, and the left button for
water operation. Depressing both buttons will
create a mist. The syringe features quick-
change autoclavable tips: To remove a tip,
press on the locking collar surrounding the tip
socket and pull the used tip straight out of the
socket (see Fig 3). To insert a new tip, press
locking collar and push tip into socket as far
as it will go. Release collar and gently tug on
tip before using to ensure that tip is securely
locked into socket.
CLEANING
Syring Tip St rilization:
1) Remove contaminated syringe tip.
2) Remove all visible signs of contamination
before autoclaving.
3) Autoclave tip at 132° C (270° F) for ten minutes.
4) Sterilize between each patient use.
NOTE: Since only the tips can be autoclaved,
it is recommended that the air/water syringe
be bagged with a disposable, single-use
plastic sleeve between each patient use.
Wat r Lin s: Disinfect the water lines weekly.
Prepare a 1:10 bleach solution (1 part
household bleach to 10 parts water). Remove
water reservoir and discard residual water.
Replace empty water supply tank and air
purge all waterlines. Fill water supply tank with
bleach solution. Run bleach solution through
all lines. Allow bleach solution to stand in lines
for 10 minutes. Remove water supply tank and
discard bleach. Flush water supply tank and
all lines thoroughly with clean water. Air purge
and leave lines dry until next clinical use.