(G) Peak Picker Setting Three Pickers are available in Pickers setting menu
(Auto, tCL, Smart) & user can switch on any as per requirement, one at a time.
(G1) Auto “Auto” mode is used for discontinuous measuring task, such as
object being transported on a conveyer belt in such a case the maximum
value for each object has to be indicated. when the object passes the
measuring beam of the pyrometer, the maximum value is stored until a new
hot object appears in the measuring beam. The temperature which has to
be recognized as “hot” is defined by the low limit of the adjusted sub range. The stored maximum value will be
deleted when the temperature of the new hot object exceeds the low limit “from” of the sub range by at least
1°C. If a lower limit is not entered, the maximum value storage will be deleted whenever the low level of the full
measuring has been exceeded.
(G2) Clear time (tCL) If the peak picker is switched on, the highest last temperature value will always be displayed
and stored. As such, it may be beneficial to periodically clear and reset the stored values in order to obtain new
temperature readings.
Example : If we set the tCl a s “6 sec” the highest last temperature value will be display for 6.0 sec to 12 sec then it
capture next pick.
Clear Time feature is particularly useful when object temperature is not uniform across its dimension or the
pyrometer is not constantly viewing an object to be measured. The peak picker works on two buffer memory to
find maximum value over a defined interval. With the first memory, the highest measured value is held and is
deleted alternately in the time interval set (clear time). The other memory retains the maximum value throughout
the next time interval. The disadvantages of fluctuations in the display with the clock frequency are thereby
eliminated.
The following setting are possible :
OFF : At clear time “OFF” the maximum value storage is switched off and only momentary values are measured.
tCL (10msec...25sec) : Clear Time tCL can be set between 10msec and 25sec. When set, estimates the maximum
values and holds it in two buffer memory. After the entered time, the storage will be deleted.
Note:
The maximum value storage coincides with adjustments made to response time. Therefore:
(I) Clear time<= the adjusted response time is useless
(ii) Clear times must be at least 5 times longer than the response time.
(iii) Only maxima with full maximum value can be recorded, which appear at least 5 times longer than response
time.