AST IR-CAST 2C User manual

AST IR-CAST 2C
USER MANUAL
A S Tccurate ensors echnologies
Non-contact Infrared Pyrometers
We measure accurate temperature in extreme conditions
AST - Accurate Sensors Technologies
Misgav Industrial Park
Tel: +972-4-9990025, Fax: +972-4-9990031
Email: info@accuratesensors.com
Web: www.accuratesensors.com
, Misgav 20173, Israel

Accurate Sensors Technologies
14
Index
1. Chapter - 1
General information
2. Chapter - 2 ....................................................................................................................... 2
Introduction
2.1 Application, range and working principle
2.2 Technical specification
2.3 Optics
3. Chapter - 3 ....................................................................................................................... 6
Installation of pyrometer
3.1 Installation
3.1.1 Location selection
3.1.2 Correct positioning of pyrometer
3.1.3 Mounting of pyrometer
3.1.4 Ambient temperature
3.1.5 Atmospheric condition
3.2 Mechanical accessories
3.3 Electrical installation
3.3.1 Connector pin assignment on the back of the pyrometer
3.3.2 USB 2.0 Connection
3.3.3 Converter RS-232 ↔ RS-485
3.3.4 Display instrument
3.3.5 Display & parameterizer
4. Chapter - 4 ....................................................................................................................... 12
Software installation
4.1 Installation
4.2 Parameters in main screen
4.2.1 Communication
` 4.2.2 Temperature
4.2.3 Parameter Setting
4.2.4 Device Information
4.2.5 Record
5. Chapter - 5 ....................................................................................................................... 19
Calculate Spot size
6. Chapter - 6 ....................................................................................................................... 20
Serial communication protocol
....................................................................................................................... 1

Accurate Sensors Technologies
1
Congratulations on choosing this high quality and highly efficient AST pyrometer for non contact
temperature measurement.
Please read this user manual carefully, step by step, including all notes of security, operation and
maintenance before installing the pyrometer. This manual contains all the necessary instructions for set up and
operation of the pyrometer. This section provides an overview about important safety regulations.
Some Important Safety Regulations Given Below:
1. Safety Precaution :
Each person working with pyrometer must read the user manual before operation. The pyrometer has only to be
used for the purpose described in manual. The pyrometer works only with a potential free low voltage of range
24VDC. This voltage is not harmful for user. The pyrometer may contain harmful material and hence it should not be
disposed of with normal waste.
2. Packaging and Storage :
Always use a shock proof package for shipment of pyrometer. It should be sealed to protect it against humidity. Also
protect the lens of pyrometer with cover. They should be stored at the temperature range from -20°c to 70°c.
3. Limit of Liability and Warranty
All general information and notes for handling, maintenance and cleaning of this instrument are offered according to
the best of our knowledge and experience.
AST reserves the right to revise this document and to make change from time to time in the content hereof without
obligation to notify any person or persons of such revisions or changes
AST instruments have a warranty of two year from the invoice date. This warranty covers manufacturing defects and
faults which arise during operation only if they are the results of defects caused by AST.
AST does not accept liability for any damages or losses which might occur, including consequential damages and
financial losses, as a result of use of the instrument.
4. Copyright
All rights reserved . this document may contain proprietary information and shall be respected as a proprietary
document to AST with permission for review and usage given only to the rightful owner of the instrument with which
this document is associated.
Chapter - 1
General Information

Accurate Sensors Technologies
2
Chapter-2
Introduction
AST IR-CAST 2C are specially designed highly accurate digital two-color rectangular measuring field IR
pyrometers for non-contact temperature measurement specially used in foundries for molten metal.
2.1 Application, Range and Working Principle
The digital AST IR-CAST 2C pyrometer use ratio method in which 2 adjacent wave lengths are used for
measurement of temperature. They are suitable for high temperature measurement ranging from 700°C to 1700°C.
AST IR-CAST 2C pyrometers are suitable for industrial purpose due to following advantages:
ŸThe pyrometer gives accurate reading even when metal stream moves with in rectangular area.
ŸMeasurement of temperature is not affected by emissivity of object.
ŸMeasurement of object temperature is possible where dust, moisture & other contaminants are present in
surrounding environment.
ŸThe measuring object can be smaller than spot size.
ŸThe pyrometer can be switched between 1-colour & 2-colour mode.
These pyrometers have solid body in stainless steel housing which provides high operation safety even in rough
industrial environment, and large variety of optics with fixed focus which can be easily used in all industrial areas.
AST IR-CAST 2C has a fast response time of 20msec. Pyrometer have RS-232, RS-485 & USB 2.0 outputs.
Response time, emissivity, sub range and peak picker selection can be preset ex works or adjusted through available
software.
The pyrometer temperature measurement method utilizes the fact that objects emit thermal radiation in an
amount that directly corresponds to their own temperature and surface emissivity. Two color pyrometer involves
measuring thermal radiance at two different wavelengths and inferring the temperature from the ratio of these
spectral radiances. Radiance ratio thermometers can be significantly more accurate than single band thermometers
in many applications.
The application in which AST pyrometer can be used is:
Ÿ Molten Metal
If the target is within this rectangular area, it gives an accurate temperature reading even when liquid moves within
this area.
118.00 mm
34.50 mm
Ø49.50 mm
Binder
Connector
AST IR-CAST 2C
14.00 mm
199.00 mm
47.50 mm
Ø 25.00 mm
11.00 mm
Measuring Distance
Measuring Field
Field of View
Metal Stream

Accurate Sensors Technologies
3
2.2 Technical Specifications
Standard Item supplied with :-
þ
þConnection cable 12-core (length - 5 m)
þDigital cable 3-core (length - 1.5 m)
þUSB cable
þAST software CD
þManual
þCertificate of calibration
AST IR-CAST 2C
AST pyrometer
Measured Temperature Range
(Analog sub range adjustable)
Spectral Range
Emissivity Range
Model AST IR-CAST 2C
700° - 1700°C
single colour (1.0....1.15µm) two colour (0.7 )....1.15µm & 1.0...1.15µm
0.1….1 adjustable (for single color mode)
Emissivity slop 0.75….1.25 slop adjustable (for two color mode)
Response Time
Accuracy
20msec. Adjustable upto 10 sec
± 0.5% +1°C of measured value
Distance to Spot Size Ratio
Digital Output
Analog Output
USB 2.0, RS-232 & RS-485 (Isolated) User Selectable
4-20 mA or 0-20mA or 0-10 V User selectable
Sighting Through the lens sighting (TL)
Operating temperature range 0°C......70°C, 0°C......200°C(with cooling jacket)
Protection class IP 65
Isolation Power supply and digital output and analog output are galvanically
isolated against each other
Operating humidity
Dimensions/Weight
Adjustable Parameters via software
10 - 95% non - condensing conditions
Dia. = Ø49.5 mm; Length = 118 mm / Weight = 0.6 kg
Emissivity, Emissivity Slope, Analog output, Address, Switch off limit,
Response time, Peak picker, Analog output sub range, switches b\w 2-
color or single color
D = 166:1 (V = Vertical)
V
D = 33:1 (H = Horizontal)
H
Power Supply 24V DC
Photodetector Type Si / Si
Repeatability 0.1% of reading in °C +1°C
Power Consumption Max. 2.5 Watt.
Note : 1. After power supply initialization, keep pyrometer under stable temperature condition for 15 Min. to get
above stated accuracy.
2. Pyrometer instrument should be properly earthed before giving the power supply.

Accurate Sensors Technologies
4
2.3 Optics
has to be measured. This heat radiation is passes through the lens to the sensor and is then converted
to an electrical signal. The farther the measured object is from the pyrometer, the larger the area that will be
measured by the pyrometer. Depending on customer need, the pyrometer is designed with fixed optics.
The pyrometer measure temperature by receiving heat radiation from the object whose temperature
Some fixed optics focus is as below:-
For Installed spot size calculation.
If the pyrometer is not installed at manufactured working distance (WD) then spot size at actual installed
distance should be calculated.
* Manufactured working distance (WD) mentioned on the pyrometer.
Manufactured working distances
WD (mm)
Spot Sizes (mm)
350
1000
Measuring Field
2.1
6
FOV 166:1
V
500 3
700 4.2
10.5
30
15
21
FOV 33:1
H
700° - 1700°C
Aperture (A)
2000 12 60
5000 30
3000 18
150
90
4
Case-I: If installed working distance is greater than manufactured working distance
Case-II: If installed working distance is smaller than manufactured working distance

Accurate Sensors Technologies
5
For Installed spot size calculation, horizontal & vertical spot size should be calculated separately by putting the
parameters from the fixed optics table.
(A) Vertical Spot Size Calculation
CAST 2C(700° - 1700°C) then spot size is 30mm (as given in table). If user installed this pyrometer at 5000mm then
spot size is not 150mm (as given in table), user should have to calculate as given method.
Installed spot size at 5000mm = (5000/1000) x (6 + 4) - 4 = 46mm
for example, if factory made working distance is 1000mm & pyrometer is AST IR-
(B) Horizontal Spot Size Calculation
IR-CAST 2C(700° - 1700°C) then spot size is 6mm (as given in table). If user installed this pyrometer at 5000mm then
spot size is not 30mm (as given in table), user should have to calculate as given method..
Installed spot size at 5000mm = (5000/1000) x (30 + 4) - 4 = 166mm
Final spot size should be
For example, if factory made working distance is 1000mm & pyrometer is AST
46mm
166mm

Accurate Sensors Technologies
6
Chapter - 3
Installation of the Pyrometer
3.1 Installation
3.1.1 Location selection
Qualified operating personnel should do the installation. Location should be good enough so that pyrometer
should get continuous infrared radiation.
Pyrometer distance from object is according to below points:-
1.
2. Know your pyrometer spot size according to point 2.3.
3. Pyrometer ambient temperature should be within 0° to 70°C (read 3.1.4).
3.1.2 Correct Positioning of the pyrometer
With through the lens sighting (TL)
In case of through the lens sighting a reticule rectangle marks the position of the measuring area. This
rectangle is true-sided and parallax-free.
Note: The reticule rectangle (TL) is only for indication of measuring area, not exactly shows the measuring
area.
3.1.3 Mounting of pyrometer
To install the pyrometer at the place of measurement a mounting support is supplied as an accessory, after
losing the clamp screws, it can be fastened correctly.
3.1.4 Ambient temperature
The allowed operation temperature for the pyrometer is 0°C to 70°C. Therefore if pyrometer is to be used
above 70°C upto 200°C water cooling jacket with built in air purge unit is used otherwise it may damage the
pyrometer. The ambient temperature is dependent on the temperature and flow rate of cooling water. Details of air
purge & water cooling jacket is given in point 3.2.
3.1.5 Atmospheric conditions
The pyrometer cannot receive the full infrared energy for the measurement if atmospheric conditions like
smoke, dust or steam are present and hence it result in measuring error. An air purge unit can be helpful to avoid
contamination such as dust and humidity on the lens. The air supplied should be at normal temperature with oil &
moisture free. The air purge generates an air stream shaped like a cone and blows particles from the lens area.
The minimum width of the pouring stream must be 33% of the width of the horizontal measuring width.
Measuring Area
Working distance(WD) (mm)
350
500
700
1000
2000
3000
5000
Width of the measuring area (mm)
10.5
15.0
21.0
30.0
60.0
90.0
150.0
Minimum width of the pouring stream (mm)
3.5
5.0
7.0
10.0
20.0
30.0
50.0
Minimum Width of the Pouring Stream
A B C
For Molten Metal pouring

Accurate Sensors Technologies
7
Air purge unit
(Reference no: 8000-04)
3.2 Mechanical accessories
Water pressure : < 10 bar
Air pressure : < 0.5 bar
Air consumption : 2...3 m³/h
Ambient temperature : < 200°C
Metal : Stainless steel
Weight : 2.75 Kg
Water Cooling Jacket with Adjustable Flange PL
(Reference no: 8000-02)
Water Cooling Jacket with Adjustable Flange TL
(Reference no: 8000-06)
M27 x 0.75
M40 x 1.5
60
48
156
80
287
54
75
18
24.50
M27 x 0.75
M40 x 1.5
175
292
65
80
80
75
48
18
24.50
60
Ø68
Ø63
Ø20
30.60
GRUB SCREW FOR
FIXING PYROMETER
FOR 6 mm INSIDE
DIAMETER HOSE
Adjustable Mounting Stand
(Reference no: 8000-07)
60°
R30
75
60
5
110
60
66
86
3 x M6 THRU AT 120°
Air pressure : < 0.5 bar
Air consumption : 2...3 m³/h
Metal : Aluminium
Weight : 0.15 Kg
Dry, clean air (Oil and dust free)
Metal : Stainless Steel
Weight : 0.9 Kg

Accurate Sensors Technologies
8
Adjustable mounting support
(Reference no: 8000-05)
3.3. Electrical Accessories
Power supply unit
(Reference no: 9000-02)
AST IR CAST 2C is powered by 24V DC (well stabilized ripple
max 50mV).The input power supply is 110/230v AC check the
polarity before connecting the device.
Device has following specifications
Power supply I/P : 100 - 240VAC, 0.9A
50/60 Hz
Power supply O/P : +24V DC, 1.5 A
3.3.1 Connector pin assignment on the back of the pyrometer
RS-232 Connection
12-Core cable
(Reference no: 7002-02)
110
70
50
70
3
Metal : Stainless Steel
Weight : 0.45 Kg
Pink
Blue
Black
Ref. No.: 9000-01
Temperature Indicator
230V, AC
AST 12-core cable
AST Pyrometer with
Digital Output RS-232
White
Red
Violet
Yellow
GND
+24 V DC
20
4-20 mA(+)
17
(-)
9
10
Female
RS-232 Connector
Computer
AST Digital Cable

Accurate Sensors Technologies
9
3.3.2
When pyrometer is connected to USB 2.0 cable, it is powered through USB 2.0. At this stage only USB 2.0
output is provided. For Analog output and Laser to work DC supply (+24V DC) must given to pyrometer.
While USB 2.0 is connected RS-232 / RS-485 communication is not possible. LED light indicates USB 2.0
operation.
USB 2.0 Connection:
Note: If user requires IP65, then USB 2.0 cap must be screw fixed at the back.
12-pin Binder connector RS-232 communication connection
Male
Female
L
C
BA
J
K
G
H
M
F
E
D
LC
J
B
A
K
G
H
M
FE
D
D - Type 9 pin Female
Connector
LC
J
B
A
K
G
H
M
FE
D8
3
7
2
4
9
6
1
5
RxD
DGND
TxD
H
Pin
A
G
C
D
F
H
M
J
K
B
Colour
Red
White
Blue
Pink
Black
Yellow
Violet
Brown
Green
Yellow
Grey
Indication
+ 24 V DC In
Ground
TxD (RS-232) / D- (RS-485)
RxD (RS-232) / D+ (RS-485)
RS-232 communication
(-) 4-20mA / 0-20mA
(+) 4-20mA / 0-20mA
Short / Open both wires for laser
ON/OFF
- (0-10V)
+ (0-10V)
Used for
Power supply
RS-232/ communicationRS-485
Analog current output
Laser ON/OFF
Analog voltage output
E Blue-Yellow N/A
L Red-Grey N/A
Not used
Open screw for USB 2.0
LED Light

Accurate Sensors Technologies
10
Power ON Pyrometer:
1. Connect the 12-core cable (Supplied with pyrometer) with pyrometer 12-pin binder connector given at the
back side of pyrometer.
2. Connect the 12-core cable supply wire with +24V DC as given in table above.
3. Pyrometer instrument should be properly Earthed before giving the power supply.
4. Connect remaining wires as per your requirement (details given in table above).
5. Provide insulation for not used end points of 12-core cable.
6. Now, the pyrometer can be switch ON.
3.3.3 Converter RS-232 ↔ RS-485 :
The pyrometer can communicate with PC using RS-232 or RS-485.
RS-232 is used only for short distances.
RS-485 is well suited for long distance transmission. Standard on PC
is RS-232, so a converter is used which converts RS-485 to RS-232.
Converter RS-232 ↔RS-485
9000-03)
(Reference no:
9 8 7 6
5432 1
9876
5
4
3
21
*Serial Communication Cable
Male/Female
(RS-232 Communication Cable)
Female
RS-232 Connector Male
RS-232 Connector
Ref. No.: 9000-01
Temperature Indicator
230V, AC
AST 12-core cable
White
Red
Violet
Yellow
GND
+24 V DC
20
4-20 mA
17
(-)
9
10
Pink
Blue
Black
(+)
AST Pyrometer with
Digital Output RS-232
1
2
3
4
5
5 4 3 2 1
9876
Female
RS-232 Connector
Ref. No.: 9000-03
*Serial Communication
Cable Male/Female
(RS-232 Communication
Cable)
5
4
3
2
1
98
7
6
Female
RS-232 Connector
Ref. No.: 9000-03
1
2
3
4
5D-
D+
Earth
230 V, AC
Male
RS-232 Connector
RS-485 to RS-232 Connection
RS-232 to RS-485 to RS-232 Connection
Ref. No.: 9000-01
Temperature Indicator
230V, AC
AST 12-core cable
AST Pyrometer with
Digital Output RS-485
White
Red
Violet
Yellow
GND
+24 V DC
20
4-20 mA(+) 17
(-)
9
10 1
2
3
4
5
Earth
D+
D-
Pink
Blue
5 4 3 2 1
9876
Female
RS-232 Connector
Ref. No.: 9000-03
Female
RS-232 Connector
*Serial Communication
Cable Male/Female
(RS-232 Communication
Cable)
Computer
Female
RS-232
Connector
Male
RS-232 Connector
Computer
Male
RS-232
Connector

Accurate Sensors Technologies
11
3.3.4 Display instrument
Temperature indicator
(Reference no: 9000-01)
Display & Parameterizer
P-120
(Reference no: 9001-01)
3.3.5 Display & parameterizer
To display the measured temperature 7 segment digital
indicator is used. Device has following specifications:
Power supply I/P : 85 to 265 V, AC/DC
Analog I/P : 4….20 mA
Retransmission : 4….20 mA
Power supply O/P : 24V, DC
Display : 4 Digits
Alarm : 2
AST P-120 is a high precision LED indicator for non contact
temperature measurement. With indication of measured
temperature user can easily parameterize a connected AST digital
pyrometer without any PC.
Device has following specification:
Power supply : 100 – 240V, AC or 24V, DC
DC analog O/P : 0-20 mA or 4-20 mA
Digital interface : RS-232 or RS-485
Pilot light : Pilot light On/Off with keys
Multi-Pyrometer Connection
AST 12-core cable
AST Pyrometer with
Digital Output RS-485
Ref. No.: 9000-01
Temperature Indicator
230V, AC
White
Red
Violet
Yellow
GND
+24 V DC
20
4-20 mA
17
(-)
9
10
(+)
AST 12-core cable
AST Pyrometer with
Digital Output RS-485
Ref. No.: 9000-01
Temperature Indicator
230V, AC
White
Red
Violet
Yellow
GND
+24 V DC
20
4-20 mA
17
(-)
9
10
(+)
Pink
Blue
Pink
Blue
Ref. No.: 9000-01
Temperature Indicator
230V, AC
AST 12-core cable
AST Pyrometer with
Digital Output RS-485
White
Red
Violet
Yellow
GND
+24 V DC
20
4-20 mA(+) 17
(-)
9
10 1
2
3
4
5
Earth
D+
D-
Pink
Blue
5 4 3 2 1
9 8 7 6
Female
RS-232 Connector
Ref. No.: 9000-03
Female
RS-232
Connector
*Serial Communication
Cable Male/Female
Note: For Multiple Pyrometer Communication,
Each pyrometer must have different address.
Computer
Male
RS-232
Connector

Accurate Sensors Technologies
12
Chapter - 4
Software Installation
The provided AST software “InfraSoft” offers digital PC interface RS-232 RS-485. Using this software we can
set all the parameters like response time, analog scale, emissivity, clear time, communication mode. This software
provide all necessary information about pyrometer.
4.1 Installation
Install the pyrometer software using the installation guide file on CD ROM & restart your PC as per guidelines
provided for installation. After installation of the software; Double click the application. It will open the screen of
software.
&
4.2 Parameters in main screen
4.2.1 Communication
Communication between the AST pyrometer and
the software is implemented via a cable connected
between the pyrometer and the PC serial port. This enables
the acquisition and recording of data, as well as the transfer
of commands from the software application to the AST
pyrometer. Communication can be done by clicking on
“Communication panel" and select correct COM Port
address where pyrometer is connected. Also user has to
select address of the pyrometer(Example: Default 01/
printed on the pyrometer sticker). Then click on CONNECT
Button.

Accurate Sensors Technologies
13
(A) Emissivity settings The emissivity can be set by clicking on
"Parameters” and select or type in the desired emissivity
directly in the description field. The emissivity value will be
transferred to pyrometer by hitting the "TAB " button.
(B) Match This parameter is used to match the known target
temperature value to be displayed by the pyrometer. The User
can enter the required temperature in window provided. This
value can be set by clicking on the “Match Temp.” parameter in
the software. The pyrometer automatically adjusts the
emissivity so that the temperature display indicates the same
temperature.
(C) Response time The desired response time can be chosen in
the panel Parameter by clicking the appropriate list box (as per
the values available in the drop box of response time). This
parameter is use to set the analog response time of pyrometer.
(D) Sub Range User can change the sub range of pyrometer in
the panel Parameter. Sub range must be within the basic range
of pyrometer, the minimum span between higher & lower range
is 51. Analog output will be automatically set according to the
sub-range by hitting "TAB" button.
(E) Sensor Type It shows pyrometer sensor type. User can
change sensor type from two color to single color and vice versa
(only applicable with two color pyrometer).
4.2.2 Temperature
It shows the temperature measured by the pyrometer
4.2.3 Parameter Setting
All user selectable device parameters can be set by using the software in the Panel "Parameter"
For communication of multiple pyrometers, select different com port and address.

Accurate Sensors Technologies
14
(E)Switch off level% ( for two color pyrometer) The switch of level is the function that is used to avoid measurement
errors caused by signals, which are too low. Although factory default is set to 15 %, the switch off limit can be
adjusted between 2 and 50%.
(F)Unit User can change the measuring unit of temperature from “Centigrade" to “Fahrenheit" and vice versa.
(G)Peak Picker If the peak picker is switched on, the highest last temperature value will always be displayed and
stored. This feature is particularly useful when object temperature is not uniform across its dimension or the
pyrometer is not constantly viewing an object to be measured.
Peak Picker Functions :
Peak picker can be turn ON & OFF by using the software. When Peak picker is ON, the peak picker menu is enabled for
setting of the parameters like decay rate function, reset below temperature and peak picker delay.
(I)Decay rate The Decay rate range is 0.00 to 166.66°C/sec. or 0.00
to 300°F/sec. depending upon °F/°C unit’s selection. The slowest
Decay rate is 0 degrees per sec. This feature helps to eliminate
erratic measurements and allows the peaked value to decay down
to lower process temperature values as they occur. Decay rate is set
to retain peak measured temperature value and ignore momentary
decreases in measured temperature.
(ll)Delay Time This function set the delay time in sec. before peak
picker function starts. The delay time is selectable in the range 0.02
to 10.00 sec. Zero (0) turns delay time OFF. This function is used to
delay the start of peaking action for upto 10 sec. following the
detection of leading edge of a new target.
Temp
Time
With Decay RateWithout Decay Rate
Temp
Time
Decay rate in
°C/sec.
Temp
Time
Delay Time (sec.)

Accurate Sensors Technologies
15
4.2.5 Record
Record is for continuous data logging. It records the measured temperature, emissivity with current date &
time.To start data logging click on start button. If user wants to record emissivity, click on record emissivity button.
After Clicking Start button window appears where user can specify the file name & location.
This screen shows the Model, basic range, serial number,
version, Head temperature, internal temperature, working
distance, spot size-apperture.
(H)Relative energy(for two color pyrometer) The relative energy shows a signal weakening which can be caused by
contaminations of the optics or a viewing window or by dust in the field of view or a too small measuring object.
Relative energy shows the measured intensity compared to the intensity, a black body radiation source would have
at a determined ratio temperature of the pyrometer.
(I)Analog Range User can select the analog range from the option 4-20mA, 0-20 mA & 0-10V.
(J)Comm. Mode User can select the communication mode as per requirement [RS-232 / RS-485].
Note: For connection diagram from RS-232 to RS-485 & RS-485 to RS-232 Refer Page No:-9
To view parameters of multiple devices select the pyrometer name from the drop down list that appears at the top of
the screen.
4.2.4 Device information
Pyrometer specific information will be displayed in the Info Panel
(lll) Reset Below Temperature(RBT) The user can set RBT within the limit of pyrometer sub range. This function sets
the temperature above which peak picker action starts. When the target temperature matches or is below the
selected value, the sensor indicates temperature without picking action.
Temp
Time
Without RBT
Temp
Time
RBT
RBT with Decay Rate

Accurate Sensors Technologies
16
If user wants file in Spreadsheet format, user can export by choosing Excel Spreadsheet in file menu.
To record emissivity, click on Record Emissivity button.
Record will be saved as .txt format and the name of file will be user define.

Accurate Sensors Technologies
17
To see previous record open the file by clicking on menu File open.
Screen containing historical trend & historical data will appear.
File will be stored in .xls format named as “export”. This “export.xls” file will be saved where the software is installed.

Accurate Sensors Technologies
18
Note:
After connecting the pyrometer right click on the graph screen and choose "Set to default" option from pop-up
menu.
Temperature graph of connected pyrometer is displayed on right side of the screen. This shows the measured
temperature corresponding to the time.
Table of contents
Other AST Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Avtech
Avtech Room Alert S Series User's guide & reference manual

Bio-Logic
Bio-Logic VMP-300-based Installation and configuration manual

Unit
Unit UTE9800+ Series user manual

AutoTest
AutoTest Battery Analyzer user manual

Delta OHM
Delta OHM LPPYRA02 operating manual

GreenValley
GreenValley LiAirX3-H user guide