BD PureWick PW100 User manual

1
PureWick
™
Urine Collection System
USER GUIDE AND
LIMITED PRODUCT WARRANTY
For use with
PureWick
™
External Catheters
PW100 (corded only)
PW200 (with lithium-ion battery)
Welcome
Thank you for purchasing the PureWick™ Urine Collection System, the system to be
used with PureWick™ External Catheters. The device pulls urine through tubing that is
connected to a collection canister. Urine is stored in the collection canister for proper
disposal. This product may be used in both home environments and professional
care facilities.
Please read all operating instructions and warnings before the first use of this product.
No special skills or additional training is required.
Intended Users
The PureWick™ Urine Collection System is intended to be operated by:
• User/Patient
• Caregiver/Healthcare professional
All functions can be safely performed by the individuals above.
Indications for Use
The PureWick™ Urine Collection System is to be used with PureWick™ External Catheters
which are intended for non-invasive urine output management.
Contraindications for Use
Do not use the PureWick™ Urine Collection System with PureWick™ External Catheters
on individuals with urinary retention.
Safety and Warnings
Always unplug PureWick™ Urine Collection System before cleaning or when not in use.
DO NOT IMMERSE THE PUREWICK™ URINE COLLECTION SYSTEM IN WATER.
As with most electrical devices, electrical parts in this system are electrically live
even when the power is o. To reduce the risk of electric shock, if the PureWick™
Urine Collection System falls into water, unplug immediately. DO NOT REACH
INTO THE WATER TO RETRIEVE IT.
WARNING: Contains small parts that may cause choking. Keep out of reach of children.
Keep cords and tubing out of the reach of children to avoid the risk of strangulation.
Discontinue use if an allergic reaction occurs. Not recommended for users who are
experiencing skin irritation or skin breakdown in device contact areas.
WARNING: This device should not be used in oxygen rich environments or in conjunction
with flammable anesthetics.

2
If the PureWick™ Urine Collection System is dropped or tipped over spilling urine, unplug
the unit and carefully inspect for loose or damaged parts before resuming use. Contact
Customer Support at 1-888-201-1586 if any damage is observed.
The pump tubing connecting the PureWick™ Urine Collection System to the collection
canister may experience light condensation inside the tube. This is not unusual and does
not aect function. However, if urine or water has streamed into the pump, discontinue
use and contact Customer Support at 1-888-201-1586.
Although the canister can hold up to 2000cc (mL), the urine should be emptied regularly
from the collection canister before volume reaches 1800cc (mL). Failure to empty
canister before urine overflow may cause damage to the PureWick™ Urine Collection
System and is not covered under the warranty.
It is important that the port connections be connected correctly for proper operation of
the PureWick™ Urine Collection System.
Use of this equipment next to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
Portable radio frequency (RF) communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 12 inches
(30cm) to any part of the PureWick™ Urine Collection System, including cables specified
by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
Use only PureWick™ Urine Collection System accessories with this device. Incompatible
parts or accessories can result in degraded performance and will void the warranty.
Do not use accessories past expiration date indicated on package labeling.
Use only the PureWick™ Urine Collection System A/C power cord with the device. Use
of an alternate consumer style A/C power adapter or an extension cord may cause
damage to device and electrical shock or injury and will void the warranty.
Do not place PureWick™ Urine Collection System or its cord across walkways creating a
tripping hazard.
For PW200: The battery cells used in this device may present a fire or chemical hazard.
To minimize the risk of damaging the battery inside the PureWick™ Urine Collection
System, do not use the device outside the indicated operating temperature range of
41°F - 104°F (5°C - 40°C). The battery is not intended to be replaced; replacement could
result in a hazard.
TO REDUCE THE RISK OF ELECTRICAL SHOCK OR INJURY, DO NOT DISASSEMBLE
THIS UNIT. NO MODIFICATION OF THIS EQUIPMENT IS ALLOWED.
KIT CONTENTS
Single-Patient Use Only Items (Accessories):
• 2000cc (mL) Collection Canister with Lid (1)
• Collector Tubing with Elbow Connector (1)
• Pump Tubing (1)
• Privacy Covers (2)
Reusable Items:
• PureWick
TM
Urine Collection System,
with or without battery (1)
• A/C Power Cord (1)
• User Guide (1)

3
Initial Setup
1. Inspect the package for damage. If any damage is noticed, contact Customer
Support at 1-888-201-1586 immediately.
2. Carefully remove all contents from the package, including the packing material
and inspect prior to use. If any damage is noticed, contact Customer Support at
1-888-201-1586 immediately.
3. Place the PureWick™ Urine Collection System (A) next to the bed or chair where it will
be used. It must be close enough for the collector tubing with elbow connector (G)
to easily reach the user and should be placed on a stable, even surface. For optimal
results, position the PureWick™ Urine Collection System either on the floor or as close
to the floor as your facility’s protocol will allow.
Note: The PureWick™ Urine Collection System should be stationary while in use.
Device is not designed to be used while in transport.
CAUTION: Avoid creating a tripping hazard with the collector tubing or power cord.
4. Plug the PureWick™ Urine Collection System power cord into device outlet (B) and
into an A/C power outlet.
Note: The device should be positioned for easy access to the A/C inlet and power cord.
Note: Make sure the power switch is turned o before connecting the power cord.
• Battery Power Operation (PW200 Only)
The built-in battery recharges any time the device is plugged into an outlet.
To fully charge the battery, the system must be plugged in for 12 hours. The
battery will supply 8 hours of backup power before requiring recharging. The
PureWick™ Urine Collection System remains fully functional while the device
is recharging.
5. Place the collection canister (C) in the PureWick™ Urine Collection System base and
press down firmly on the lid making sure the lid is sealed. Optional (prior to sealing
canister lid): While operating the PureWick™ Urine Collection System, ammonia odor
for up to 1800 mL of urine can be eliminated by adding 2 teaspoons (10 grams)
of vitamin C (ascorbic acid) powder into the collection canister before use. At least
99.93% of pure vitamin C (ascorbic acid) is recommended. If using the privacy cover
(J), slip privacy cover onto canister (similar to a coee cup sleeve) prior to placing
canister into the base. Attach the pump tubing (D) to the PureWick™ Urine Collection
System connector port (F) and the connector port (E) on the collection canister lid.
Battery Life Status - Green LED Light
No Light Device is powered o
Solid Device is on and fully charged
Slow Blink Device is charging
Fast Blink Operating on battery; battery less than 20%
Note: If it is too cold or hot, charging time may be longer or battery may not fully charge.

4
6. Attach the collector tubing (G) to the connector port (H) on the collection canister
lid. Connect the other end of the collector tubing securely to a PureWick™ External
Catheter (I).
CAUTION: It is important that the port connections be connected correctly and
securely for proper operation of the PureWick™ Urine Collection System.
7. Turn on the PureWick™ Urine Collection System by pressing the ON/OFF switch. The
PureWick™ Urine Collection System is working when the switch lights up green and
the device makes a soft humming sound. The system is now ready for use.
Operating Instructions
1. After the PureWick™ Urine Collection System has been set up, place the PureWick™
External Catheter according to the PureWick™ External Catheter’s instructions for use.
2. When the canister is ready to be emptied, remove the PureWick™ External Catheter
according to the PureWick™ External Catheter’s instructions for use and throw away.
Note: Keep the PureWick™ Urine Collection System ON to make sure all urine has
been drawn out of the collector tubing and into the collection canister before
removing the PureWick™ External Catheter.
Note: Although the canister can hold up to 2000cc (mL), to prevent overflow empty
urine from the collection canister regularly or before volume reaches 1800cc (mL).
3. Turn o the PureWick™ Urine Collection System by pressing the ON/OFF switch.
Disconnect the A/C power cord from the power outlet and from the device.
Disconnect collector tubing and pump tubing from the canister.
4. Lift collection canister from PureWick™ Urine Collection System. Do not lift
canister by the lid. Take canister into an area appropriate for disposal e.g. a
bathroom, carefully remove the lid, and dispose of urine in a proper receptacle
e.g. a toilet or according to facility protocol. If using a privacy cover and it gets
wet, remove and discard.
5. Clean collector tubing, pump tubing, canister, PureWick™ Urine Collection
System, and power cord according to cleaning instructions in the Cleaning and
Maintenance section.
PureWick™ Urine
Collection System
Device electrical
cord port
2000cc Collection
canister with lid
Pump tubing
Pump tubing port
to canister
Connector port for
pump tubing
Collector tubing with
elbow connector
Connector port for
collector tubing
Attach PureWick™
External Catheter
(sold separately)
A
G
H
E
F
C
D
B
I
Privacy Cover
(optional)
slides onto
Collection
Canister
as shown

5
6. Reassemble the PureWick™ Urine Collection System according to the Initial Setup
instructions when the system is ready to be reused.
When the PureWick™ Urine Collection System is not in use, unplug the power cord from
its outlet. Make sure the unit is cleaned and disinfected prior to storage. Do not store
when parts are wet or damp.
Cleaning and Maintenance
Inspect the PureWick™ Urine Collection System and accessories for damage or wear
prior to use and replace as necessary.
DO NOT ATTEMPT TO OPEN OR DISMANTLE THE PUREWICK™ URINE
COLLECTION SYSTEM. THIS MAY AFFECT PERFORMANCE, CREATE A POTENTIAL
SAFETY HAZARD, CAUSE PERSONAL INJURY, AND WILL VOID THE WARRANTY.
ALWAYS TURN OFF THE DEVICE AND DISCONNECT FROM POWER SOURCE PRIOR
TO CLEANING.
THE PRIVACY COVER COMPONENTS ARE SINGLE USE ONLY AND SHOULD NOT BE
CLEANED OR DISINFECTED.
Collection Canister
and Lid
Collector Tubing
with elbow connector
PureWick™ Urine
Collection System base
Pump Tubing Power Cord
Frequency:
Daily
P Clean
P Disinfect
P Clean
P Disinfect
P Clean
P Disinfect
P Clean
P Disinfect
P Clean
P Disinfect
USE CASE Single Patient
Use Only Single Patient
Use Only Reusable Single Patient
Use Only Reusable
Point of Use Cleaning: After the completion of each use, the collection canister, canister
lid, collector tubing, and pump tubing should be rinsed with cool tap water to remove
any residual debris or dirt. The PureWick™ Urine Collection System base should be wiped
with a clean low lint cloth dampened with cool tap water.
The collection canister, canister lid, collector tubing, pump tubing, and PureWick™ Urine
Collection System base should be cleaned and disinfected at the time of each use, or at
a minimum, daily. The power cord should be cleaned and disinfected at the time of each
use, or at a minimum daily.
Note: Gloves should be worn when handling soiled or dirty accessories.
Canister and Canister Lid
• Initial Rinse: Rinse the canister and lid thoroughly with cool tap water. While rinsing,
remove any excess dirt by wiping the canister and lid with disposable low lint wipes.
Manual Cleaning Instructions for Single User (Home Care):

6
• Cleaning: Prepare a soapy solution by mixing 1 teaspoon (approximately 5 mL)
of dish soap with 1 gallon (approximately 4 L) of cool tap water. Fully submerge
canister and lid in the solution. Allow it to sit for a minimum of ten (10) minutes.
While submerged, use a soft brush (e.g. toothbrush) to brush all accessible areas of
the canister and lid for a minimum of one (1) minute to remove any visible signs of
debris or dirt.
• Rinse: Rinse the canister and the lid thoroughly with cool tap water until there is no
visible sign of the cleaning solution.
• Visual inspection: Inspect the canister and the lid to ensure that all debris and dirt
have been removed. If there is any sign of debris or dirt, repeat steps above until
there is no sign of debris or dirt remaining on the canister and lid.
• Disinfection: Fully submerge the canister and lid in 70% isopropyl alcohol (IPA).
Allow it to sit for a minimum of ten (10) minutes.
• Rinse: Rinse the canister and the lid thoroughly with cool tap water.
• Drying: Dry the canister and lid with a clean low lint towel or cloth.
Collector Tubing (with elbow connector) and Pump Tubing
• Initial Rinse: Disconnect the collector tubing from the elbow connector and
disconnect the elbow connector from the lid. Rinse the disconnected tubing
thoroughly by running cool tap water through the inside of the tubing. While
rinsing, remove any excess dirt by wiping the outside of the tubing and elbow
connector with low lint disposable wipes. While rinsing, flush the tubing with cool
tap water to remove any excess dirt.
• Cleaning: Prepare a soapy solution by mixing 1 teaspoon (approximately 5 mL) of
dish soap with 1 gallon (approximately 4 L) of cool tap water. Completely submerge
the tubing and elbow connector in the solution and allow it to soak for a minimum of
ten (10) minutes. At the beginning of the soak time, flush the inside of the tubing with
the prepared solution. While submerged, use a soft brush (e.g. toothbrush) to brush all
accessible areas of the tubing and elbow connector for a minimum of one (1) minute
to remove any visible signs of debris or dirt.
• Rinse: Rinse the tubing and elbow connector thoroughly with cool tap water until
there is no visible sign of the soap cleaning solution. While rinsing, flush the inside
of the tubing with the tap water.
• Visual inspection: Inspect the tubing to ensure that all debris and dirt have been
removed. If there is any sign of debris or dirt, repeat the steps above until there is
no sign of debris or dirt remaining on or inside the tubing.
• Disinfection: Fully submerge the tubing and elbow connector in 70% isopropyl
alcohol (IPA). Allow it to soak for a minimum of ten (10) minutes. Flush the IPA
through the tubing, prior to starting the ten (10) minute time.
• Rinse: Rinse the tubing and elbow connector thoroughly with cool tap water,
ensuring that the inside of the tubing is flushed with water.
• Drying: Dry the tubing and elbow connector outer surface with a clean low lint
towel or cloth and allow the inside of the tubing to air dry for a minimum of thirty
(30) minutes at room temperature.

7
PureWick™ Urine Collection System Base with power cord
• Initial Wipe: Wipe the PureWickTM Urine Collection System base and power cord
thoroughly with a clean low lint cloth dampened with cool tap water. Remove any
excess dirt by wiping the device with disposable low lint wipes.
• Cleaning: Prepare a soapy solution by mixing 1 teaspoon (approximately 5 mL)
of dish soap with 1 gallon (approximately 4 L) of cool tap water. Completely wet a
clean low lint cloth with the solution and wipe down the PureWickTM Urine Collection
System base and power cord. Allow the cleaning solution to remain on the device
and power cord for a minimum of ten (10) minutes, ensuring that the device and
power cord remains wet for the ten (10) minutes. Using a soft brush (e.g. toothbrush),
brush all accessible areas of the PureWickTM Urine Collection System base and power
cord for a minimum of one (1) minute to remove any visible signs of debris or dirt.
• Rinse: Using a new clean low lint cloth, wipe the PureWickTM Urine Collection
System base and power cord thoroughly with cool tap water to remove the
soap cleaning solution. Perform this step until there is no visible sign of the soap
cleaning solution.
• Visual inspection: Inspect the PureWickTM Urine Collection System base and
power cord to ensure that all debris and dirt have been removed. If there is any
sign of debris or dirt, repeat the steps above until there is no sign of debris or dirt
remaining on the PureWickTM Urine Collection System base and power cord.
• Disinfection: Use 70% isopropyl alcohol (IPA) to completely wet a clean low lint
cloth and wipe down the PureWickTM Urine Collection System base and power
cord. Allow the disinfecting solution to remain on the device and power cord for
a minimum of ten (10) minutes, ensuring the unit is visibly wet with IPA for ten
(10) minutes.
• Rinse: Using a new clean low lint cloth, wipe the PureWickTM Urine Collection
System base and power cord thoroughly with cool tap water to remove the
disinfecting solution.
• Drying: Dry the PureWickTM Urine Collection System base and power cord with a clean
low lint towel or cloth.
Contamination should be assumed prior to use with a new user; therefore, the PureWick™
Urine Collection System base and power cord must be fully cleaned with a disinfectant
before reuse. The collection canister and tubing are not suitable for reuse in a multiple user
environment. A new collection canister and new tubing must be used for each new user.
Canister and Canister Lid (Single User)
• Initial Rinse: Rinse the canister and lid thoroughly with cool tap water. While rinsing,
remove any excess dirt by wiping the canister and lid with disposable low lint wipes.
• Cleaning: Prepare enzymatic cleaner solution (e.g. Enzol Enzymatic Cleaner or
equivalent) following the manufacturer’s recommendation of 1 oz. (approximately
29 mL) per gallon (approximately 4 L) of warm tap water (ratio and water
temperature per manufacturer's recommendation if applicable).
Manual Cleaning Instructions for Hospitals and Long-Term Care Facilities:

8
Completely submerge the canister and lid in the solution and allow it to soak for a
minimum of ten (10) minutes. While the canister and lid are submerged, use a soft
bristle brush (e.g. toothbrush) to brush all accessible areas of the canister and lid for
a minimum of one (1) minute to remove any visible signs of debris or dirt.
• Rinse: Rinse the canister and the lid thoroughly with cool tap water until there is no
visible sign of the enzymatic cleaning solution.
• Visual inspection: Inspect the canister and the lid to ensure that all debris and dirt
have been removed. If there is any sign of debris or dirt, repeat the steps above
until there is no sign of debris or dirt remaining on the canister and lid.
• Drying: Dry the canister and lid with a clean low lint towel or cloth.
• Disinfection: Fully submerge the canister and lid in 70% isopropyl alcohol (IPA).
Allow it to sit for a minimum of ten (10) minutes.
• Rinse: Rinse the canister and the lid thoroughly with cool tap water.
• Drying: Dry the canister and lid with a clean low lint towel or cloth.
Collector Tubing (with elbow connector) and Pump Tubing (Single User)
• Initial Rinse: Disconnect the collector tubing from the elbow connector and
disconnect the elbow connector from the lid. Rinse the disconnected tubing
thoroughly by running cool tap water through the inside of the tubing. While
rinising, remove any excess dirt by wiping the outside of the tubing and elbow
connector with low lint disposable wipes. While rinsing, flush the tubing with cool
tap water to remove any excess dirt.
• Cleaning: Prepare enzymatic cleaner solution (e.g. Enzol Enzymatic Cleaner or
equivalent) following the manufacturer’s recommendation of 1 oz. (approximately 29
mL) per gallon (approximately 4 L) of warm tap water (ratio and water temperature
per manufacturer's recommendation if applicable). Completely submerge the
tubing and elbow connector in the solution and allow it to soak for a minimum of
ten (10) minutes. At the beginning of the soak time, flush the inside of the tubing
with the prepared solution. While the tubing is submerged, use a soft bristle brush
(e.g. toothbrush) to brush all accessible areas of the tubing and elbow connector for
a minimum of one (1) minute to remove any visible signs of debris or dirt.
• Rinse: Rinse the tubing and elbow connector thoroughly with cool tap water until
there is no visible sign of the enzymatic cleaning solution. While rinsing, flush the
inside of the tubing with the cool tap water.
• Visual inspection: Inspect the tubing to ensure that all debris and dirt have been
removed. If there is any sign of debris or dirt, repeat the steps above until there is
no sign of debris or dirt remaining on or inside the tubing.
• Disinfection: Fully submerge the tubing and elbow connector in 70% isopropyl
alcohol (IPA). Allow it to sit for a minimum of ten (10) minutes. Flush the IPA
through the tubing, prior to starting the ten (10) minute time.
• Rinse: Rinse the tubing and elbow connector thoroughly with cool tap water,
ensuring that the inside of the tubing is flushed with water.
• Drying: Dry the tubing and elbow connector outer surface with a clean low lint
towel or cloth and allow the inside of the tubing to air dry for a minimum of thirty
(30) minutes at room temperature.

9
PureWick™ Urine Collection System Base with power cord (Multiple Users)
• Initial Wipe: Wipe the PureWick™ Urine Collection System base and power cord
thoroughly with a clean low lint cloth dampened with cool tap water. Remove any
excess dirt by wiping the unit with disposable low lint wipes.
• Cleaning: Prepare enzymatic cleaner solution (e.g. Enzol Enzymatic Cleaner or
equivalent) following the manufacturer’s recommendation of 1 oz. (approximately
29 mL) per gallon (approximately 4 L) of warm tap water (ratio and water
temperature per manufacturer's recommendation if applicable). Wipe the device
and power cord with a low lint cloth dampened with the solution. Allow the device
and power cord to remain wet for a minimum of ten (10) minutes, ensuring that the
device and power cord are visibly wet for the ten (10) minutes. Using a soft bristle
brush (e.g. toothbrush), brush all accessible areas of the PureWick™ Urine Collection
System base and power cord for a minimum of one (1) minute to remove any
visible signs of debris or dirt.
• Rinse: Using a new clean low lint cloth, wipe the PureWick™ Urine Collection
System base and power cord thoroughly with cool tap water to remove the
enzymatic cleaning solution. Perform this step until there is no visible sign of the
cleaning solution.
• Visual inspection: Inspect the PureWick™ Urine Collection System base and
power cord to ensure that all debris and dirt have been removed. If there is any
sign of debris or dirt, repeat the steps above until there is no sign of debris or dirt
remaining on the PureWick™ Urine Collection System base and power cord.
• Disinfection: Use 70% isopropyl alcohol (IPA) to completely wet a clean low lint
cloth and wipe down the device and power cord. Allow the disinfecting solution
to remain on the PureWick™ Urine Collection System base and power cord for
a minimum of ten (10) minutes, ensuring the unit is visibly wet with IPA for ten
(10) minutes.
• Rinse: Using a new clean low lint cloth, wipe the PureWick™ Urine Collection
System base and power cord thoroughly with cool tap water to remove the
disinfecting solution.
• Drying: Dry the PureWick™ Urine Collection System base and power cord with a
clean low lint towel or cloth.
Replacing Canister and Tubing
Replace the canister and tubing for each patient per facility protocol or at least every 60
days, whichever is sooner. If you notice any of the following, replace immediately:
• Canister or tubing has residual urine build-up
• Canister or tubing appear cloudy
• Canister or tubing appear discolored
• Canister becomes cracked
• Tubing becomes torn
• PureWick™ External Catheter no longer securely connects to collector tubing
Failure to clean and/or replace accessories may aect the performance of the system.
The useful life of the collection canister and tubing is 60 days.

10
Troubleshooting
Symptom Check the following
• The device does not
turn on when external
power is used
• The power switch
does not glow when
turned on
1. Check that the power cord is plugged into the
PureWick™ Urine Collection System and to an
electrical outlet.
2. Ensure the electrical outlet is functioning by
plugging in another electrical device.
• The device is on but is
not suctioning properly 1. Ensure tubing connections are connected
properly.
2. Check collector tubing for blockage or flow
restriction such as pinched or kinked tubing.
3. Ensure overflow stop valve in collection canister
lid is open. The valve floats to the top when the
collection canister is full. The stop valve may
close if the lid or canister is tipped sideways
or upside down. Disconnect tubing and gently
shake the lid to reset the valve down to the
open position.
4. Ensure collection canister is sealed with the lid
tightly closed.
5. Verify suction by disconnecting PureWick™
External Catheter from the collector tubing and
placing the end of the collector tubing into a cup
of water. If water easily flows into the collection
canister replace the PureWick™ External Catheter.
There are no serviceable parts in the PureWick™ Urine Collection System. Adjustments
or modifications made by anyone other than an authorized representative of BD will
void the Warranty.
If the PureWick™ Urine Collection System is not functioning properly after performing
the troubleshooting steps, the device may have reached the end of its useful life. With
normal use, this device shall remain safe for the useful life. Do not use the device if it is
damaged or defective, including, but not limited to the following:
• Cracks
• Separated housing
• Not suctioning properly or no suction
• Damaged power cord
If damage is noticed and the device is still under warranty, contact Customer Support at
1-888-201-1586 immediately.

11
Disposal
The PureWick™ Urine Collection System and accessories may be a potential biohazard.
Clean and disinfect the PureWick™ Urine Collection System and accessories before
disposal. Handle in accordance with accepted medical practice and applicable local and
federal regulations. The PureWick™ Urine Collection System (PW100 and PW200) contains
electrical and/or electronic equipment, and the PW200 product contains a lithium-ion
battery that must be disposed of per local laws and regulations.
Transporting the PureWick™ Urine Collection System (PW200 only):
The PureWick™ Urine Collection System with battery contains a lithium-ion battery
which is considered a hazardous material. Transporting it is governed by federal and
local laws.

12
Specifications (PW100 and PW200)
Class II
Type BF Applied Part (the collector tubing is the applied part)
IP21
Operating suction range: Approximately 120 mmHg – 140 mmHg (2.3 – 2.7 psi)
Maximum suction level: 182 mmHg (3.5 psi)
Electrical Specifications
AC input voltage: 100 –240VAC
AC frequency: 50 – 60Hz 10A
AC connector: NEMA 1-15P
Environmental Specifications
PW100 PW200
Operating Temperature
Limit
41°F–104°F (5°C–40°C) 41°F–104°F (5°C–40°C)
Storage
Temperature Limit
-13°F–158°F (-25°C–70°C) -13°F–104°F (-25°C–40°C)
Transport
Temperature Limit
-13°F–158°F (-25°C–70°C) -13°F–158°F (-25°C–70°C)
Humidity
Limitation
5 –90%, non-condensing 5 –90%, non-condensing
Operating Atmospheric
Pressure Limitation
70 kPa - 106 kPa 70 kPa - 106 kPa
Storage/Transport
Atmospheric Pressure
Limitation
50 kPa - 106 kPa 50 kPa - 106 kPa
When the
PureWick™ Urine Collection System has been stored in a very cold
environment (down to -13°F or -25°C), please allow the device to sit for approximately
90 minutes at room temperature (approximately 68°F or 20°C) before use.
When the
PureWick™ Urine Collection System has been stored in a very warm
environment (up to 158°F or 70°C), please allow the device to sit for approximately 90
minutes at room temperature (approximately 68°F or 20°C) before use.
Physical Specifications
Dimensions with canister: 9” x 6.5” x 11” (22.86cm x 16.51cm x 27.94cm)
Dimensions without canister: 8.75” x 6” x 7.5” (22.22cm x 15.24cm x 19.05cm)
Cord length: 12’ (3.65m)
Battery Specifications (PW200 only)
Battery: Lithium-ion battery
Charging Time: 12 hours

13
Customer Support
Liberator Medical Supply, Inc. Phone: 1-888-201-1586
2979 SE Gran Park Way Email: PureWick@liberatormedical.com
Stuart, FL 34997 www.liberatormedical.com/purewick
PureWickTM Urine Collection System Limited Product Warranty
C. R. Bard, Inc., warrants to you, the original purchaser of this product, that this product (but
not its battery if one is included) will be free from defects in material and workmanship for a
period of three (3) years from your date of purchase. For the purposes of this warranty, the
original purchaser of this product is the first purchaser of this product from C. R. Bard, Inc. or
an authorized retailer. Any limited warranties are not transferable.
The liability of C. R. Bard, Inc. under this limited product warranty will be limited, at its
option, to replacement or repair of the defective product, or to refund your net price paid at
C. R. Bard, Inc.'s discretion.
For all warranty issues, please call 1-888-201-1586 Monday – Friday from 9:00 AM to 5:00
PM EST. Before receiving warranty service, C. R. Bard may require that you furnish proof of
purchase details and respond to questions designed to assist with diagnosing potential
issues for obtaining warranty service.
If the product proves to be defective, C. R. Bard will send you prepaid packaging so that you
may ship your defective PureWick™ Urine Collection System back to the manufacturer. C. R.
Bard will replace and ship your new PureWick™ Urine Collection System back to you within
90 days of receiving your defective equipment. A replaced PureWick™ Urine Collection
System is subject to the same limited warranty as the original device.
Wear and tear from normal use or defects resulting from any abuse, misuse, improper
storage, alteration, further manufacture, packaging, or processing of this product are not
covered by this limited warranty.
This limited warranty does NOT cover any loss of functionality of the battery (if one is
included), including a battery which no longer holds a charge, or the perceived slowness
of charging.
TO THE EXTENT ALLOWABLE BY APPLICABLE LAW, THIS LIMITED PRODUCT WARRANTY
IS IN LIEU OF ALL OTHER WARRANTIES RESPECTING THIS PRODUCT, WHETHER
EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. ANY IMPLIED
WARRANTIES REQUIRED BY APPLICABLE LAW ARE LIMITED IN DURATION TO THE THREE
(3) YEARS TERM OF THIS LIMITED WARRANTY.
IN NO EVENT WILL C. R. BARD, INC. BE LIABLE TO YOU FOR ANY INDIRECT, INCIDENTAL
OR CONSEQUENTIAL DAMAGES UNDER THIS WARRANTY.
C. R. Bard, Inc.'s total liability will not exceed the purchase price paid for the PureWick™
Urine Collection System. In no event will C. R. Bard, Inc. be liable for procurement of
substitute products.
Some states do not allow an exclusion or limitation of implied warranties, incidental or
consequential damages, so the above limitation may not apply to you. You may be entitled
to additional remedies under state law.
If you have any questions regarding how to use your PureWick
TM
Urine Collection System,
please contact Liberator Medical Supply, Inc. at 1-888-201-1586.

14
Table 1: Electromagnetic emissions
This equipment is intended for use in the electromagnetic environment specified below.
The user of the equipment should assure that it is used in such an environment
Emissions Test Compliance
Radiated emissions
EN 55011
Complies (Group B, Group 1)
Conducted emissions
CISPR 11 / EN 55011
Complies
AC Power-line harmonics
IEC 61000-3-2
Complies
AC Power-line Flicker
IEC 61000-3-3
Complies
Table 2: Electromagnetic immunity
This equipment is intended for use in the electromagnetic environment specified below.
The user of the equipment should assure that it is used in such an environment.
Immunity Test IEC 60601 test level Compliance level
Electrostatic discharge
IEC 61000-4-2 + 8 kV Contact, HCP & VCP
+2 kV, +4 kV, +8 kV, + 15 kV Air + 8 kV Contact, HCP & VCP
+2 kV, +4 kV, +8 kV, + 15 kV Air
Radiated RF Immunity
IEC 61000-4-3 80 MHz - 2.7 GHz, 10 V/m,
80% 1 kHz AM 80 MHz - 2.7 GHz, 10 V/m,
80% 1 kHz AM
Electrical Fast
Transient/Burst
IEC 61000-4-4
+ 2 kV (100 kHz rep rate)
AC mains
+ 1 kV (100 kHz rep rate) I/O > 3m
+ 2 kV (100 kHz rep rate)
AC mains
+ 1 kV (100 kHz rep rate) I/O > 3m
Surge Immunity
IEC 61000-4-5 +0.5 kV & + 1 kV line-line,
AC mains
+0.5 kV, + 1 kV, + 2 kV lineground,
AC mains
+0.5 kV & + 1 kV line-line,
AC mains
+0.5 kV, + 1 kV, + 2 kV lineground,
AC mains
Conducted RF
immunity
IEC 61000-4-6
150 kHz to 80 MHz, 3 Vrms, 80%
1 kHz AM, power and I/O
6 Vrms (+modulation) for
ISM frequencies
150 kHz to 80 MHz, 3 Vrms, 80%
1 kHz AM, power and I/O
6 Vrms (+modulation) for
ISM frequencies
Power Frequency
H-field
IEC 61000-4-8
50 and 60 Hz, 30 A/m,
x-, y-, and z- axes 50 and 60 Hz, 30 A/m,
x-, y-, and z- axes
Voltage Dips,
Interrupts
IEC 61000-4-11
100% reduction for 0.5 cycles
100% reduction for 1.0 cycle
30% reduction for 25/30 cycles
100% reduction for 250/300 cycles
100% reduction for 0.5 cycles
100% reduction for 1.0 cycle
30% reduction for 25/30 cycles
100% reduction for
250/300
cycles
Guidance and Manufacturer's Declarations

15
The PureWick
™
Urine Collection System bears the following symbols:
WARNING: This product1can expose you to lead, which is known to the State
of California to cause cancer and birth defects or other reproductive harm. For more
information go to: https://www.P65Warnings.ca.gov.
1The product is the A/C Power Cord.
IP21 Classification – Protection of equipment against ingress of solid foreign objects which
are greater than or equal to 12.5mm in diameter and protection of equipment against
harmful eects of ingress from vertically dripping water.
Units
Catalog
Number
Lot
Number
Use By
date
Do not use
if package
is damaged
Serial
Number
Consult
Instructions
For Use
Manufacturer
Type BF
Applied Part
Class II
Caution
Date of
Manufacture
Humidity
Limitation
Caution, Risk
of Electric
Shock
Degree of
Protection
against harmful ingress
of water and particulate
matter.
Temperature
Limit
Atmospheric
Pressure
Limitation
Intertek ETL
Non-sterile
Certied To CSA STD C22.2 NO. 60601-1
Conforms To Medical Electrical Equipment
STDS 60601-1, 60601-1-6 and 60601-1-11

16

17
FRANÇAIS
PureWick
™
Système de collecte
des urines
GUIDE D’UTILISATION ET GARANTIE
LIMITÉE DU PRODUIT
À utiliser avec les sondes
externes PureWick™ PW100
(avec cordon
d’alimentation uniquement)
PW200
(avec batterie
au lithium-ion)
Bienvenue
Nous vous remercions d’avoir acheté le système de collecte des urines PureWick™, le
système à utiliser avec les sondes externes PureWick™. L’appareil aspire l’urine par le
biais d’une tubulure qui est reliée à un contenant de collecte. L’urine est stockée dans
le contenant de collecte afin de pouvoir être éliminée correctement. Ce produit peut
être utilisé à la fois à domicile et dans les établissements de soins professionnels.
Avant d’utiliser ce produit pour la première fois, veuillez lire toutes les instructions
d’utilisation ainsi que tous les avertissements. Aucune compétence particulière ni
formation supplémentaire n’est requise.
Utilisateurs visés
Le système de collecte des urines PureWick™ est destiné à être utilisé par:
• les utilisatrices/patientes
• les aidants/professionnels de santé
Les personnes citées ci-dessus peuvent réaliser l’ensemble des opérations en toute sécurité.
Indications d’utilisation
Le système de collecte des urines PureWick™ doit être utilisé avec les sondes externes
PureWick™ destinées à la prise en charge non invasive du débit urinaire.
Contre-indications
Ne pas utiliser le système de collecte des urines PureWick™ avec les sondes externes
PureWick™ sur des personnes sourant de rétention urinaire.
Sécurité et avertissements
Débrancher systématiquement le système de collecte des urines PureWick™ avant de
le nettoyer ou lorsqu’il n’est pas utilisé.
NE PAS IMMERGER LE SYSTÈME DE COLLECTE DES URINES PUREWICK™ DANS DE
L’EAU. Comme la plupart des appareils électriques, les parties électriques
de ce système sont sous tension même lorsque l’alimentation est coupée.
Pour réduire le risque de choc électrique, si le système de collecte des urines
PureWick™ tombe dans l’eau, le débrancher immédiatement.
NE PAS PLONGER LA MAIN DANS L’EAU POUR LE RÉCUPÉRER.
AVERTISSEMENT: Contient de petites pièces susceptibles de provoquer un
étouement. Conserver hors de la portée des enfants.

18
Conserver le cordon d’alimentation et les tubulures hors de la portée des enfants pour
éviter tout risque d’étranglement.
Cesser l’utilisation en cas de réaction allergique. Usage déconseillé aux utilisatrices
qui présentent une irritation ou une lésion cutanée au niveau des zones en contact
avec l’appareil.
AVERTISSEMENT: Cet appareil ne doit pas être utilisé dans un environnement riche
en oxygène ou en association avec des anesthésiques inflammables.
Si le système de collecte des urines PureWick™ tombe ou se renverse, répandant ainsi
l’urine, débrancher l’appareil et l’inspecter soigneusement à la recherche de toute
pièce desserrée ou endommagée avant de reprendre l’utilisation. Si des dommages
sont constatés, communiquer avec le service à la clientèle au 1-888-201-1586.
L’intérieur de la tubulure de la pompe reliant le système de collecte des urines
PureWick™ au contenant de collecte peut présenter une légère condensation. Cela n’a
rien d’inhabituel et n’a pas d’incidence sur le fonctionnement. Cependant, si de l’urine
ou de l’eau s’est infiltrée dans la pompe, cesser d’utiliser l’appareil et communiquer
avec le service à la clientèle au 1-888-201-1586.
Bien que le contenant puisse contenir jusqu’à 2000cc (ml) d’urine, le contenant de
collecte doit être vidé régulièrement avant que le volume n’atteigne 1800cc (ml). Ne
pas vider le contenant avant que l’urine ne déborde risque d’endommager le système
de collecte des urines PureWick™ et ceci n’est pas couvert par la garantie.
Il est important que les raccordements aux ports soient correctement réalisés pour
garantir un bon fonctionnement du système de collecte des urines PureWick™.
Éviter d’utiliser cet appareil à côté d’autres appareils ou empilé sur d’autres appareils,
au risque d’entraîner un mauvais fonctionnement de celui-ci. Si une telle configuration
est nécessaire, cet appareil ainsi que les autres appareils devront être surveillés pour
s’assurer qu’ils fonctionnement normalement.
Ne pas utiliser d’appareils de communication de radiofréquences (y compris les
périphériques tels que les câbles d’antenne et les antennes externes) à moins de
30cm (12pouces) de toute partie du système de collecte des urines PureWick™, y
compris les câbles spécifiés par le fabricant. À défaut de procéder ainsi, il pourrait en
résulter une dégradation des performances de l’appareil.
N’utiliser que les accessoires du système de collecte des urines PureWick™ avec cet
appareil. L’utilisation de pièces ou d’accessoires incompatibles risquent d’entraîner une
dégradation des performances et annulera la garantie.
Ne pas utiliser les accessoires au-delà de la date de péremption indiquée sur
l’étiquette de l’emballage.
N’utiliser que le cordon d’alimentation CA du système de collecte des urines PureWick™
avec cet appareil. L’utilisation d’un adaptateur CA ou d’une rallonge peut endommager
l’appareil et provoquer un choc électrique ou des blessures et annulera la garantie.
Ne pas placer le système de collecte des urines PureWick™ ou son cordon
d’alimentation dans des zones de passage, sous peine de créer un risque de
trébuchement.

19
Pour le modèle PW200: Les cellules de la batterie utilisées dans cet appareil peuvent
présenter un risque d’incendie ou un risque chimique. Pour minimiser le risque
d’endommager la batterie à l’intérieur du système de collecte des urines PureWick™,
utiliser uniquement l’appareil dans la plage de température de fonctionnement
indiquée de 5°C à 40°C (41°F à 104°F). La batterie n’est pas destinée à être
remplacée, car son remplacement pourrait présenter un danger.
POUR RÉDUIRE LE RISQUE DE CHOC ÉLECTRIQUE OU DE BLESSURE, NE PAS
DÉMONTER CET APPAREIL. AUCUNE MODIFICATION DE CET ÉQUIPEMENT
N’EST AUTORISÉE.
CONTENU DE LA TROUSSE
Préparation initiale
1. Inspecter l’emballage pour vérifier qu’il n’est pas endommagé. En cas de
détection de dommages, communiquer immédiatement avec le service à la
clientèle au 1-888-201-1586.
2. Retirer soigneusement tout le contenu de l’emballage, y compris le matériel
d’emballage, et l’inspecter avant utilisation. En cas de détection de dommages,
communiquer immédiatement avec le service à la clientèle au 1-888-201-1586.
3. Placer le système de collecte des urines PureWick™ (A) à côté du lit ou de la chaise
où il sera utilisé. Il doit être susamment proche pour que la tubulure de collecte
avec raccord coudé (G) puisse atteindre facilement l’utilisatrice et il doit être placé
sur une surface stable et plane. Pour obtenir des résultats optimaux, placer le
système de collecte des urines PureWick™ soit sur le sol, soit aussi près du sol que
le protocole de votre établissement le permet.
Remarque: Le système de collecte des urines PureWick™ doit être stationnaire
pendant l’utilisation. L’appareil n’est pas conçu pour être utilisé pendant le transport.
AVERTISSEMENT: Éviter que la tubulure de collecte ou le cordon d’alimentation
ne crée un risque de trébuchement.
4. Brancher le cordon d’alimentation du système de collecte des urines PureWick™
sur la prise de l’appareil (B) et sur une prise CA.
Remarque: L’appareil doit être positionné de manière à pouvoir accéder
facilement à la prise CA et au cordon d’alimentation.
Remarque: S’assurer que l’interrupteur est éteint avant de brancher le cordon
d’alimentation.
• Fonctionnement sur batterie (PW200 uniquement)
La batterie intégrée se recharge à chaque fois que l’appareil est branché sur
une prise de courant. Pour charger complètement la batterie, il faut brancher
le système pendant 12heures. La batterie fournira 8heures d’alimentation
Articles à usage unique (accessoires):
• Contenant de collecte de 2000cc (ml) avec
couvercle (1)
• Tubulure de collecte avec raccord coudé (1)
• Tubulure de la pompe (1)
• Emballages de protection de l’intimité (2)
Articles réutilisables:
• Système de collecte des urines
PureWickTM avec ou sans pile (1)
• Cordon d’alimentation CA (1)
• Guide d’utilisation (1)

20
de secours avant de devoir être rechargée. Le système de collecte des urines
PureWick™ reste entièrement fonctionnel pendant la recharge de l’appareil.
5. Placer le contenant de collecte (C) dans la base du système de collecte des urines
PureWick™ et appuyer fermement sur le couvercle pour s’assurer qu’il est bien
fermé. Facultatif (avant de fermer le couvercle du contenant) : lors de l’utilisation
du système de collecte d’urine PureWickTM, l’odeur d’ammoniac pour 1 800 ml
d’urine maximum peut être éliminée en ajoutant 2 cuillères à thé (10 grammes)
de poudre de vitamine C (acide ascorbique) au contenant de collecte avant
utilisation. Il est recommandé d’utiliser de la vitamine C (acide ascorbique) d’une
pureté d’au moins 99,93 %. En cas d’utilisation de l’emballage de protection de
l’intimité (J), glisser celui-ci sur le contenant (comme pour un manchon de tasse
à café) avant de placer le contenant dans la base. Raccorder la tubulure de la
pompe (D) au port de connexion du système de collecte des urines PureWick™ (F)
et au port de connexion (E) sur le couvercle du contenant de collecte.
6. Raccorder la tubulure de collecte (G) au port de connexion (H) sur le couvercle du
contenant de collecte. Raccorder solidement l’autre extrémité de la tubulure de
collecte à une sonde externe PureWick™ (I).
AVERTISSEMENT: Pour garantir un bon fonctionnement du système de collecte
des urines PureWick™, il est important que les raccordements aux ports soient
réalisés correctement et de manière sécurisée.
État de charge de la batterie – Voyant DEL vert
Éteint L’appareil est éteint
Allumé en continu L’appareil est allumé et complètement chargé
Clignotement lent L’appareil est en cours de charge
Clignotement rapide Fonctionnement sur batterie; charge de la
batterie inférieure à 20%
Remarque : S’il fait trop froid ou trop chaud, le temps de charge peut être plus long ou il
se peut que la batterie ne se charge pas complètement.
Système de collecte
des urines PureWick™
Prise pour le cordon
d’alimentation de
l’appareil
Contenant de collecte de
2000cc avec couvercle
Tubulure de la pompe
Port de connexion de la
tubulure de la pompe
au contenant
Port de connexion pour
la tubulure de la pompe
Tubulure de collecte
avec raccord coudé
Port de connexion pour
la tubulure de collecte
Raccorder la sonde
externe PureWick™
(vendue séparément)
A
G
H
E
F
C
D
B
I
Emballage de
protection de
l’intimité (en
option) qui
se glisse sur
le contenant
de collecte,
comme illustré
This manual suits for next models
1
Table of contents
Languages:
Other BD Medical Equipment manuals

BD
BD Kiestra Track User manual

BD
BD FACSAria Manual

BD
BD EleVation User manual

BD
BD PureWick PW200E Operating and maintenance manual

BD
BD Alaris VP Plus Guardrails User manual

BD
BD Alaris neXus GPneXus1 Manual

BD
BD Vacutainer Pronto Reference guide

BD
BD EnCor Enspire User manual

BD
BD EnCor Enspire User manual

BD
BD Halo One Series User manual