
7
5. PROGRAMMING & CALIBRATION
The BA534D is programmed and calibrated via
four push-buttons which are located behind the
instrument control cover. If frequent access to
the push-buttons is required, the rate totaliser can
be supplied with duplicate membrane push-
buttons mounted on the outside of the control
cover.
All the programming functions are contained in an
easy to use menu that is shown diagramatically in
Fig 6. Each function is summarised in section 5.3
which includes references to more detailed
information. Although this simple menu driven
system enables most adjustments to be made
without repeated reference to this manual, it is
recommended that at least the summary of the
programmable functions in section 5.3 is read
before attempting programming or recalibration.
When the BA534D is fitted with alarms, a pulse
output or a 4/20mA output, the basic menu is
expanded to include the accessories. Section 8
of this manual explains how to programme these
additional functions.
Note: While the instrument is being programmed
or calibrated, totalisation continues so that any
flow occurring during this time is recorded.
5.1 Calibration structure
Fig 5 shows the BA534D calibration structure.
The rate and total display calibration functions are
totally independent allowing the displays to have
different engineering units.
For the rate display the pulse input is divided by
SCALE-r to represent engineering units and
multiplied by the timebase t-bASE to calculate the
rate of flow per second, minute or per hour.
Similarly, to calculate the total flow in engineering
units the number of input pulses is divided by the
total scale factor SCALE-t. Total flow is
continuously added to the grand total.
The BA534D displays ‘real’ decimal points.
Moving the position of a decimal point in a scale
factor will therefore affect the instrument
calibration.
The BA534D updates its total flow calculation
twice per second. If the instrument is used for
batching applications, this resolution may define
the accuracy of the system.
Note: The total and rate displays may be updated
less frequently depending upon the display update
time selected.
Fig 5 Calibration structure
5.2 Accessing programming functions
Throughout this manual push-buttons are shown
in italics e.g. P or Up push-button, and legends
displayed by the instrument are shown within
inverted commas e.g. 'CAL' and ' ALr2'.
Access to the programme menu is obtained by
operating the Pand Epush-buttons
simultaneously. If the instrument is not protected
by a security code the first parameter 'UPdAtE' will
be displayed. If a security code other than the
default code 0000 has already been entered, the
instrument will display 'COdE'. Press Pto clear
this prompt and enter the security code for the
instrument using the Up or Down push-buttons to
adjust each digit, and the P push-button to move
control to the next digit. If the correct code has
been entered pressing Ewill cause the first
parameter 'UPdAtE' to be displayed. If an
incorrect code is entered, or a push-button is not
operated within ten seconds, the instrument will
automatically return to the display mode.
Apart from defining the position of the decimal
point in the rate display all programme functions
and prompts are shown on the large eight digit
display.
Once within the main programme menu the
required parameter can be reached by scrolling
through the menu using the Up or Down push-
buttons as shown by the programme structure in
Fig 6.