Crowcon Gas-Pro User manual

M07998/Eng
Issue 3 July 2020
Pump setting: This function, which is only present if the Gas-Pro has
an internal pump, allows authorised users to turn the pump on or off.
Sounder volume: This function allows authorised users to change the
sounder volume between 2 levels.
IN USE
+ve Safety™
+ve Safety™ (Positive Safety) provides
positive confi rmation of detector status prior
to deployment in the fi eld or on return from
site. The front mount tri-colour LED gives
the Safety Manager or Supervisor unrivalled
visibility of the status of the Operator’s
detector.
Charging
• To charge, insert the charger connected into the socket at the back of
the Gas-Pro.
• The display will show the battery fi lling and RED LED’s will fl ash red
until the battery is full, when they will fl ash green.
Gas Testing
Crowcon recommends regular gas tests (also known as bump tests) to
confi rm sensor operation. A number of fl exible and simple solutions are
available:
Q-Test Quick and simple in-fi led gas test and calibration solution.
I-Test Intelligent wall or desk mount, powered data management
solution.
S-Test Service and maintenance solution for up to 5 detectors.
If confi gured to do so, test and calibration can also be completed manually
by using the non-pumped fl ow plate and following prompts on the screen. If
a test should fail or calibration is due and the Gas-Pro has been manually
zeroed in the last 15 minutes a calibration can also be completed.
See the Gas-Pro User and Operator manual for details.
Safety Information
• Gas-Pro is a hazardous area certifi ed gas detector and as such must
be operated and maintained in strict accordance with the instructions,
warnings and label information.
• Read and understand all instructions in the operation section of “Gas-Pro
User & Operator Manual”, P/N M07995 prior to use.
• The Gas-Pro re-chargeable battery must only be charged in non-
hazardous (safe) areas.
• Gas-Pro is certifi ed for use in ambient temperatures in the range -20°C to
+55°C (-4 to 131°F).
Certifi cation Information
IECEx: Ex db ia IIC T4 Gb Tamb -20°C to +55°C
IECEx ULD 11.0004X
ATEX: II 2 G Ex db ia IIC T4 Gb Tamb -20°C to +55°C
DEMKO 11 ATEX 1031772X
UL: Gas detector use in hazardous locations Class 1 Division 1,
Groups A, B, C and D only as to intrinsic safety.
Contact Crowcon:
UK: 172 Brook Drive, Milton Park, Abingdon, Oxfordshire OX14 4SD
Tel: +44 (0) 1235 557700 Email: [email protected]
US: 1455 Jamike Ave, Suite 100, Erlanger, KY 41018
Tel: +1 859 957 1039 Email: [email protected]
SG: Block 194 Pandan Loop, #06-20 Pantech Industrial Complex,
Singapore, 128383
Tel: +65 6745 2936 Email: [email protected]
CN: Unit 316, Area 1, Tower B, Chuangxin Building, 12 Hongda North Road, Beijing
Economic & Technological Development Area, Beijing, China 100176
Tel: +86 10 6787 0335 Email: [email protected]
www.crowcon.com
5 6
Multi-Gas Monitor
Gas-Pro
Quick Start Guide
SET-UP
Prior to use
Before use, the Gas-Pro should always be checked for any signs of physical
damage.
i Read and understand the User and Operator manual prior to use.
Gas-Pro orientation
Figure 1: Gas-Pro
D-ring Operator button +ve Safety™ indicator  Charging cable
 Alarm bars Sensor apertures Dual colour LCD display Certifi cation label
Sounder  Pump inlet/outlet  Alligator clip
DISPLAY
Home Screen
• Gas-Pro simultaneously displays up
to 5 gases.
• The dual colour backlight aids
recognition of alarm events.
• The home icon will invert to show normal operation.
• The backlight is green and changes to red during an alarm highlighting
the gas in alarm.









OPERATION
Switching On/Off
• In clean air turn on the Gas-Pro by holding down the operator button
for 3 audible blips. The Gas-Pro will warm up, going through a series of
automatic processes.
• To turn off, press and hold the operator button. During the 4 second
countdown hold the button down until the countdown has fi nished.
Quick View
• A momentary press of the operator button allows the user to see the gas
type and unit, battery level, user and +ve Safety™ status.
Alarms
• Gas-Pro offers audible, visual and vibrate alarms as standard.
• Each gas has 2 instantaneous alarms and toxics also have TWA alarms.
The operator has the option to Resume TWA alarm monitoring after
powering-off (software version 1V25 and higher).
• When in alarm the display backlight will change from green to red and
highlight the gas in alarm.
• When the Gas-Pro warns of a low battery, the sounder will emit an audible
double blip every 5 seconds and the +ve Safety™ LED will go red. This
means the battery has at least 30 minutes of battery life remaining. After 30
minutes the Gas-Pro will enter full alarm state and the battery icon will fl ash
empty as shut down is imminent.
Backlight
• The backlight is confi gurable but set as standard to come on at the press
of the operator button or during alarm.
Pump
Please refer to the Safety Information in section 2.4.2 of the Gas-Pro User &
Operator Manual before using pumped/sampling mode.
• If the internal pump option has been chosen, the pump is activated by
attaching the pumped fl ow plate.
• Ensure the gasket is clean and undamaged prior to placement.
• The pump test will activate every time the fl ow plate is attached the
user will be prompted to cover the inlet to ensure a tight seal. Gas-Pro
will indicate a pass or fail, if failed the fl ow plate should be removed,
checked and reapplied.
USER MENU
• Access the user menu by double clicking the operator button, then
scroll through the menu using single clicks. Once the correct function
is highlighted, double click to choose.
Zero
• During start-up the operator has the option to perform an autozero, or
leave the countdown if an autozero is not required/possible.
• A manual zero can be performed at any time following start up and is
needed less than 15 minutes prior to running a manual calibration.
Time Weighted Average (TWA)
• This function allows the 8 hour TWA to be reviewed.
Pre- Entry Check (PEC)
• This function is intended for sampling air of unknown quality before
gaining access to it (e.g. going under ground through a manhole
cover) thereby avoiding unnecessary exposure.
Peak Review
• This function allows the highest level of each gas detected to be
reviewed during the last 8 hours, 12 hours or since turned on.
Flammable Correction factor. Only available for
pellistors.
• This option selects the fl ammable gas correction factor to be used for
fl ammable (pellistor) sensors with respect to the base calibration of
methane.
Settings
User setting: Up to 5 different users can be loaded into Gas-Pro using
the Portables-Pro PC application.
2 3 4
Fold 1 (in)
Fold 1 (in)
Fold 2 (in)
Fold 2 (in)

Other manuals for Gas-Pro

8

Other Crowcon Gas Detector manuals

Crowcon Gas-Pro User manual

Crowcon

Crowcon Gas-Pro User manual

Crowcon Xsafe Bright Manual

Crowcon

Crowcon Xsafe Bright Manual

Crowcon IRmax Assembly instructions

Crowcon

Crowcon IRmax Assembly instructions

Crowcon TXgard-Plus User manual

Crowcon

Crowcon TXgard-Plus User manual

Crowcon Xgard Bright User manual

Crowcon

Crowcon Xgard Bright User manual

Crowcon TETRA 4 User manual

Crowcon

Crowcon TETRA 4 User manual

Crowcon AGP 45 User manual

Crowcon

Crowcon AGP 45 User manual

Crowcon Xgard Bright Manual

Crowcon

Crowcon Xgard Bright Manual

Crowcon Flamgard-4/20 User manual

Crowcon

Crowcon Flamgard-4/20 User manual

Crowcon Detective Net User manual

Crowcon

Crowcon Detective Net User manual

Crowcon AGM 55 User manual

Crowcon

Crowcon AGM 55 User manual

Crowcon TXgard-IS+ Manual

Crowcon

Crowcon TXgard-IS+ Manual

Crowcon Eikon User manual

Crowcon

Crowcon Eikon User manual

Crowcon Xgard Setup guide

Crowcon

Crowcon Xgard Setup guide

Crowcon Gas-Pro User manual

Crowcon

Crowcon Gas-Pro User manual

Crowcon Hydra256 User manual

Crowcon

Crowcon Hydra256 User manual

Crowcon Vortex FP Compact User manual

Crowcon

Crowcon Vortex FP Compact User manual

Crowcon T4 User manual

Crowcon

Crowcon T4 User manual

Crowcon SF6 Assembly instructions

Crowcon

Crowcon SF6 Assembly instructions

Crowcon Xgard IR Assembly instructions

Crowcon

Crowcon Xgard IR Assembly instructions

Crowcon Clip O2 User manual

Crowcon

Crowcon Clip O2 User manual

Crowcon Xgard Bright Manual

Crowcon

Crowcon Xgard Bright Manual

Crowcon IRmax User manual

Crowcon

Crowcon IRmax User manual

Crowcon Gasman II Parts list manual

Crowcon

Crowcon Gasman II Parts list manual

Popular Gas Detector manuals by other brands

RKI Instruments SC-04 Operator's manual

RKI Instruments

RKI Instruments SC-04 Operator's manual

AGS Mini Merlin CH4CO v2-TX Installation, operation & maintenance manual

AGS

AGS Mini Merlin CH4CO v2-TX Installation, operation & maintenance manual

Riken Keiki SD-1D-AS operating manual

Riken Keiki

Riken Keiki SD-1D-AS operating manual

Riken Keiki GD-D58-AC-GH operating manual

Riken Keiki

Riken Keiki GD-D58-AC-GH operating manual

SENKO SI-H100 operating manual

SENKO

SENKO SI-H100 operating manual

RKI Instruments GX-82A instruction manual

RKI Instruments

RKI Instruments GX-82A instruction manual

Helio MAP 50 operating manual

Helio

Helio MAP 50 operating manual

WatchGas POLI quick start guide

WatchGas

WatchGas POLI quick start guide

Riken Keiki SP-220 Series operating manual

Riken Keiki

Riken Keiki SP-220 Series operating manual

Bühler technologies GAS 222.35 AMEX Brief instructions

Bühler technologies

Bühler technologies GAS 222.35 AMEX Brief instructions

Comelit SE237EO user manual

Comelit

Comelit SE237EO user manual

Sensitron SMART3 NC instruction manual

Sensitron

Sensitron SMART3 NC instruction manual

Oldham MX62 user manual

Oldham

Oldham MX62 user manual

Riken Keiki GD-F3A-SC-A operating manual

Riken Keiki

Riken Keiki GD-F3A-SC-A operating manual

RefineAir SENSE CO2 user manual

RefineAir

RefineAir SENSE CO2 user manual

Det-Tronics FlexSight LS2000 instructions

Det-Tronics

Det-Tronics FlexSight LS2000 instructions

Sensitron PATROL manual

Sensitron

Sensitron PATROL manual

Tecnogas 12030 Operation manual

Tecnogas

Tecnogas 12030 Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.