
English - 4 English - 5
Introduction
Dentsply Sirona is an ISO 13485 registered company. All
Dentsply Sirona medical devices sold in Europe are
CE marked in conformance with Council Directive
93/42/EEC.
Caution: United States Federal Law restricts this device to
sale by or on the order of, a licensed dental professional.
For dental use only.
Product Overview
The Cavitron Select®SPS™ Scaling System operates by
converting ordinary AC house current into high frequency
current. the ultrasonic system consists of two parts: an
insert, and the SPS™ (Sustained Performance System)
electronic system. The system incorporates two closed
loops. One loop provides automatic tuning (operating
frequency is adjusted to be at resonance for each insert),
the second loop automatically controls the tip stroke over
different working conditions. The scaler produces 30,000
microscopically small strokes per second at the insert’s
working tip. this combined with acoustic effects of the
coolant water, produces a synergistic action that literally
“powers away” the heaviest calculus deposits while
providing exceptional operator and patient comfort.
The Pump/Reservoir accessory for the Cavitron Select®
SPS™ is an optional lavage dispenser. Footswitch plug,
handpiece plug, water tube and power supply plug are
removed from the Scaler before mating with the Pump/
Reservoir. The Scaler is snapped together with the Pump/
Reservoir. The footswitch plug, handpiece plug and power
supply plug are inserted into mating receptacles at the
rear of the Pump/Reservoir. Ultrasonic signal, footswitch
and DC power connections between the units are made
via an edgecard connector at the rear of the Scaler
and the front of the Pump module. The Reservoir has a
capacity of 450 ml. Fluid from the Reservoir is propelled
from the Reservoir Bottle with 15–19 psig compressed air
generated by an internal diaphragm pump. ON/OFF of
the lavage is controlled through an internal solenoid valve.
Fluid to the Handpiece passes directly from the Pump/
Reservoir to the Handpiece. No fluid passes through the
Cavitron Select®SPS™ when used with this accessory.
Flow through any insert in the Handpiece will be greater
than 40 cc/m water when adjusted to maximum.
Technical Support
For technical support and repair assistance in the U.S., call
the Dentsply Sirona Technical Support at 1-800-989-8826,
Monday through Friday, 8:00 A.M. to 5:00 P.M. (Eastern
Time). For other areas, visit www.dentsplysirona.com/eifu
and search "Cavitron Repair Centers" for repair centers or
contact your local Dentsply Sirona representative.
Supplies & Replacement Parts
To order supplies or replacement parts in the U.S., contact
your local Dentsply Sirona Distributor or call 1-800-989-
8826, Monday through Friday, 8:00 A.M. to 5:00 P.M.
(Eastern Time). For other areas, please contact your local
Dentsply Sirona representative.
1: Indications For Use
Designed for use in general prophylaxis and periodontal
treatments and other areas of operative dentistry. It
is used to debride light to heavy calculus and plaque
deposits from tooth and root surfaces.
2: Contraindications
• For best performance, use only ultrasonic inserts
manufactured by Dentsply Sirona.
• Ultrasonic Systems should not be used for restorative
dental procedures involving the condensation of
amalgam.
3: Warnings
• The use of High Volume Evacuation (HVE) to
manage aerosols during the procedure is highly
recommended.
• The effectiveness of pre-procedural rinses has
yet to be fully determined in clinical research.
However, the use of mouth rinses containing:
Chlorhexidine Gluconate, Hydrogen peroxide,
povidone iodine and essential oils, may reduce the
number of microorganisms during aerosol generating
procedures.
• It is the responsibility of the Dental Healthcare
Professional to determine the appropriate uses of this
product and to understand:
– The health of each patient,
– The dental procedures being undertaken,
– And applicable industry and governmental agency
recommendations for infection control in dental
healthcare settings,
– Requirements, and regulations for safe practice of
dentistry; and
– These Instructions for Use in their entirety, including
Section 4 Precautions, Section 7 Infection Control,
and Section 11 System Care.
• Where asepsis is required or deemed appropriate
in the best professional judgment of the Dental
Healthcare Professional.
• Handle Cavitron insert with care. Improper handling
of insert, specifically the insert tip, may result in injury
and/or cross contamination.
• Failure to follow properly validated sterilization
processes and approved aseptic techniques for
Cavitron inserts may result in cross contamination.
• Persons fitted with cardiac pacemakers, defibrillators
and other active implanted medical devices, have
been cautioned that some types of electronic
equipment might interfere with the operation of the
device. Although no instance of interference has ever
been reported to Dentsply Sirona, we recommend
that the handpiece and cables be kept 6 to 9 inches
(15 to 23 cm) away from any device and their leads
* data on file
during use. There are a variety of pacemakers and
other medically implanted devices on the market.
Clinicians should contact the device manufacturer or
the patient's physician for specific recommendations.
This unit complies with IEC 60601 Medical Device
Standards.
• Insufficient water flow could result in elevated
water and tip temperature. When operated at
the input water temperature specified in the
Waterline Requirements Section 8.1 and with
sufficient water flow, the water and tip temperature
should not exceed 50oC (122oF). Failure to follow
recommendations for environmental operating
conditions, including input water temperature, could
result in injury to patients or users. If temperature is
elevated, increase water flow. If temperature remains
elevated, discontinue use.
• During boil-water advisories, this product should not
be operated as an open water system (e.g. connected
to a public water system). A Dental Healthcare
Professional should disconnect the system from
the central water source. The Cavitron DualSelect
system can be attached to this unit and operated
as a closed system until the advisory is cancelled.
When the advisory is cancelled, flush all incoming
waterlines from the public water system (e.g. faucets,
waterlines and dental equipment) in accordance with
the manufacturer’s instructions for a minimum of 5
minutes.
4: Precautions
4.1 System Precautions
• Close manual shut-off valve on the dental office water
supply every night before leaving the office.
• Do not place the system on or next to a radiator or
other heat source. Excessive heat may damage the
system’s electronics. Place the system where air is free
to circulate on all sides and beneath it.
• The system is portable, but must be handled with care
when moving.
• Equipment flushing and dental water supply system
maintenance are strongly recommended. See Section
11: System Care.
• Never operate system without fluid flowing through
the handpiece.
• Always ensure that the electrical connections on the
handpiece cable and the Steri-Mate®handpiece are
clean and dry before assembling them for use.
4.2 Procedural Precautions
General
• As with all dental procedures, use standard
precautions (i.e., utilize appropriate personal
protective equipment such as mask, protective
eyewear or face shield, protective gown and gloves).
• The power supply shipped with this unit is
• approved for use with devices and must be used to
operate this system.
Ultrasonics
• The Cavitron Select SPS unit works with Cavitron
inserts as a system, and was designed and tested to
deliver maximum performance for all currently available
Cavitron brand ultrasonic inserts. Companies that
manufacture, repair or modify inserts carry the sole
responsibility for proving the efficacy and performance
of their products when used as a part of this system.
Users are cautioned to understand the operating limits
of their insert before using in a clinical setting.
• Like bristles of a toothbrush, ultrasonic inserts “wear”
with use. Inserts with just 2 mm of wear lose about 50%
of their scaling efficiency. In general it is recommended
that ultrasonic inserts be discarded and replaced after
one year of use to maintain optimal efficiency and avoid
breakage.
• If excessive wear is noted, or the insert has been bent,
reshaped or otherwise damaged, discard the insert
immediately.
• Ultrasonic insert tips that have been bent, damaged, or
reshaped are susceptible to in-use breakage and should
be discarded and replaced immediately.
• Retract the lips, cheeks and tongue to prevent contact
with the insert tip whenever it is placed in the patient’s
mouth.
Reservoir System
• Do not place the System on or next to a radiator or
other heat source. Excessive heat may damage the
System’s electronics. Place the System where air is free
to circulate on all sides and beneath it.
• The System is portable, but must be handled with care
when moving.
• Pour liquid into the cap opening of the reservoir bottle.
To prevent potential leaking at the cap, do not overfill
bottle with liquid into the neck of the bottle. Ensure cap
is tightly sealed.
• Turn unit off before refilling reservoir bottle to prevent
potential leaking of fluid. Turn cap a 1/4 revolution
to release air pressure and then remove the cap
completely.
5: Adverse Reactions
None Known.
6: Residual Risks
• Failure to not properly remove water source from
system while no longer in use water damage may occur.
• If unit is not properly assembled and installed in
accordance with these instructions, could cause
potential water damage and/or an overheated
handpiece.
• If the presence of water flow from the insert tip is not
confirmed during operation, an overheated handpiece
may occur.
• The use and maintenance of the water filter is key for
optimum systems operation and will help to prevent
potential safety risks, due to water blockage – causing
overheated handpiece.