EN - 4 A-7314
SET-UP & OPERATION
Fully charge battery for 10-17
HOURS. (7314P series only) Insert container into holder. Disposable Container
Connection: Attach 43/8”
tubing from lter cartridge to
tubing connector on unit.
Attach 6’ patient tubing to
container lid at outlet labeled
<Patient>.
Turn the unit “on”.
If using AC or DC power, plug
the small connector into the
DC power input on the side of
unit.
Plug the other end into an AC
wall outlet or DC receptacle.
NOTE– The AC adapter may
become warm to the touch
during charging or running of
the unit. This is normal.
Adjust the suction level.
Verify suction level.
NOTE - Always verify suction
level before beginning by
occluding open end of patient
tubing while observing gauge.
Adjust knob to desired level.
WARNING
If the unit is not receiving power from an external source or the battery was not recharged, the low battery indicator light will remain
on and the performance of the unit will drop off rapidly. Switch to another power source immediately after the low battery light appears
to avoid an interrupted suction procedure.
NOTE– Gauge is for reference only. If the unit sustains a severe drop, accuracy of the gauge must be checked.
CAUTION
damage to the vacuum pump.
CAUTION
result.
1
Ensure power switch is “off”.
52 3A 4
7314P - Select desired power
source. (Skip steps 6B if using
internal battery power.)
6A 7 9 10
11
NOTE–Inspect suction tubing and container for leaks, cracks, etc. and assure that all connections are secure and without leaks before using.
8
ACCESSORY/REPLACEMENT ITEMS
The following items can be purchased separately as accessories or replacement items for your 7314 Series DeVilbiss Suction Unit:
Description Part No. Description Part No.
6’ patient tubing 6305D-611 12V DC power cord (1 each) 7304D-619
Collection Container Kit (Filter cartridge, 800 ml container, 43/8”and 6’ tubing package) 7305D-633 AC to DC adapter/charger (see Specications for
manufacturer information) 7314P-613
800 ml disposable container with lter cartridge and 43/8” tubing (48 each) 7305D-632 Power cord for US DV51D-606
Filter cartridge (12 pack) (For Disposable Container) 7305D-635 Power cord for Continental Europe DV51D-607
Collection Container Kit (1200 ml reusable container, bacteria lter, elbow, 43/8” tubing) 7314D-603 Power cord for UK DV51D-608
1200 ml reusable container (bacteria lter, elbow, 43/8”tubing) (6 pack) 7314D-604 Power cord for Australia DV51D-609
Bacteria lter (non-sterile) (12 pack) (For Reusable Container) 7305D-608 Power cord for Japan DV51D-613
Carrying case 7314D-606
NOTE
from the product or decreased electromagnetic immunity of the product.
Reusable Container
Connection: Ensure that the
clear side of the bacteria lter
is toward elbow and bottle
when installing. Do not reverse
direction of lter. The bacteria
lter should then be connected
to the 90˚ elbow connection,
and the elbow should be
connected to the top of the
container lid where it says
<Vacuum>.
3B
6B
NOTE– The 7314D series is not factory equipped with an
internal rechargeable battery. 7314P series is factory equipped
with an internal rechargeable battery and all information
regarding battery operation in this guide is applicable.
7314D series (non-battery
label)