
For your safety and that of your patients
8 Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
Device combinations
This device can be operated in combination with
other Dräger devices or with devices from other
manufacturers. Observe the accompanying docu-
ments of the individual devices.
If a device combination is not approved by Dräger,
the safety and the functional state of the individual
devices can be compromised. The operating orga-
nization must ensure that the device combination
complies with the applicable editions of the relevant
standards for medical devices.
Device combinations approved by Dräger meet the
requirements of the following standards:
Third edition
– IEC 60601-1, third edition (general require-
ments for safety, device combinations, soft-
ware-controlled functions)
– IEC 60601-1-2 (electromagnetic compatibility)
– IEC 60601-1-8 (alarm systems)
Second edition
– IEC 60601-1, second edition (general require-
ments for safety)
– IEC 60601-1-1 (device combinations)
– IEC 60601-1-2 (electromagnetic compatibility)
– IEC 60601-1-4 (software-controlled functions)
– IEC 60601-1-8 (alarm systems)
If a device combination is not approved by Dräger,
proper operation of the devices can be compro-
mised.
The operator must ensure that the device combina-
tion meets the applicable standards.
Strictly observe instructions for use and assembly
instructions of all connected devices.
Strictly follow the assembly instructions and instruc-
tions for use for each connected device.
WARNING
Connecting devices to a multiple socket-out-
let (power strip) can result in increased elec-
tric leakage current and expose the user to
current beyond permissible values.
The leakage current also increases when mul-
tiple medical devices are connected to a pa-
tient. Make sure that the electrical isolation of
each device is suitable for the intended appli-
cation. Connect only equipment to the analog
and digital signal inputs and outputs that are
setup and tested according to IEC standards.
To protect the patient from possible injury due
to electrical shock, peripheral devices should
only be connected to a monitor within the
same room. The installer or service provider
should verify that the leakage current of the
interconnected system meets the electrical
safety requirements of IEC 60601-1. CAUTION
Combinations of Dräger devices and third-party
devices that are not approved by Dräger may ad-
versely affect operation of those devices and may
put the patient at greater risk of injury.
CAUTION
The medical device must only be used with soft-
ware tested and approved by Dräger. Any modifi-
cations of the operating system settings can im-
pair operating safety. Responsibility for any such
modifications lies with the owner.