manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Dräger
  6. •
  7. Respiratory Product
  8. •
  9. Dräger PAS Micro CFH User manual

Dräger PAS Micro CFH User manual

This manual suits for next models

1

Other Dräger Respiratory Product manuals

Dräger PAS Colt Series User manual

Dräger

Dräger PAS Colt Series User manual

Dräger PSS 5000 Series User manual

Dräger

Dräger PSS 5000 Series User manual

Dräger X-plore 1300 Series User manual

Dräger

Dräger X-plore 1300 Series User manual

Dräger Panorama Nova RA User manual

Dräger

Dräger Panorama Nova RA User manual

Dräger Panorama Nova S-RP User manual

Dräger

Dräger Panorama Nova S-RP User manual

Dräger FPS-COM 7000 User manual

Dräger

Dräger FPS-COM 7000 User manual

Dräger X-plore 6000 User manual

Dräger

Dräger X-plore 6000 User manual

Dräger PSS Series User manual

Dräger

Dräger PSS Series User manual

Dräger X-plore Air FFP2 NR User manual

Dräger

Dräger X-plore Air FFP2 NR User manual

Dräger FPS 7000 User manual

Dräger

Dräger FPS 7000 User manual

Dräger PAS Micro Series User manual

Dräger

Dräger PAS Micro Series User manual

Dräger Panorama BelAir GC-RA User manual

Dräger

Dräger Panorama BelAir GC-RA User manual

Dräger X-plore 6000 User manual

Dräger

Dräger X-plore 6000 User manual

Dräger PARAT 4000 Series User manual

Dräger

Dräger PARAT 4000 Series User manual

Dräger PAS Colt P18 User manual

Dräger

Dräger PAS Colt P18 User manual

Dräger CDR 4500 User manual

Dräger

Dräger CDR 4500 User manual

Dräger PSS 7000 Series User manual

Dräger

Dräger PSS 7000 Series User manual

Dräger UCF FireCore User manual

Dräger

Dräger UCF FireCore User manual

Dräger PSS BG 4 User manual

Dräger

Dräger PSS BG 4 User manual

Dräger X-plore 8000 User manual

Dräger

Dräger X-plore 8000 User manual

Dräger X-plore 7300 User manual

Dräger

Dräger X-plore 7300 User manual

Dräger X-plore 9300 User manual

Dräger

Dräger X-plore 9300 User manual

Dräger PSS 5000 series User manual

Dräger

Dräger PSS 5000 series User manual

Dräger X-plore 1300 Series User manual

Dräger

Dräger X-plore 1300 Series User manual

Popular Respiratory Product manuals by other brands

ResMed MIRAGE SWIFT II 60891/1 user guide

ResMed

ResMed MIRAGE SWIFT II 60891/1 user guide

Omron COMP A-I-R NE-C801 instruction manual

Omron

Omron COMP A-I-R NE-C801 instruction manual

Ombra 63506 Instructions for use

Ombra

Ombra 63506 Instructions for use

Lotus M-9520A quick start guide

Lotus

Lotus M-9520A quick start guide

CERES nosko nebulizer kit manual

CERES

CERES nosko nebulizer kit manual

Dr.Frei NE-SM1 user manual

Dr.Frei

Dr.Frei NE-SM1 user manual

North 5400 series instruction manual

North

North 5400 series instruction manual

John Bunn Neb-u-Lite LX2 user manual

John Bunn

John Bunn Neb-u-Lite LX2 user manual

Philips Respironics EE Leak 1 AF541 Instructions for use

Philips

Philips Respironics EE Leak 1 AF541 Instructions for use

Sperian Panther Hip-Pac P9684 Series Operation manual

Sperian

Sperian Panther Hip-Pac P9684 Series Operation manual

3M 8000 Series quick start guide

3M

3M 8000 Series quick start guide

Respironics SideStream HS860 Instructions for use

Respironics

Respironics SideStream HS860 Instructions for use

Sundstrom SR 500 Instructions for use

Sundstrom

Sundstrom SR 500 Instructions for use

Ambu SPUR II Instructions for use

Ambu

Ambu SPUR II Instructions for use

BLS 8000 Instructions for the use

BLS

BLS 8000 Instructions for the use

SATA Vision 2000 operating instructions

SATA

SATA Vision 2000 operating instructions

DeVilbiss UltraNeb user manual

DeVilbiss

DeVilbiss UltraNeb user manual

KYARA OBLO' LTK110 instruction manual

KYARA

KYARA OBLO' LTK110 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

3711974 (A3-D-P) Page 1/3
PAS®Micro CFH
Self-contained breathing apparatus Instructions for use
PAS ®Micro CFHSelf-contained breathing apparatus
1 Safety-related information
– Before using this product, carefully read these instructions for use
and those of the associated products.
– Strictly follow the instructions for use. The user must fully understand
and strictly observe the instructions. Use the product only for the
purposes specified in the intended use section (see section 3.4).
– Do not dispose of the instructions for use. Ensure that they are
retained and appropriately used by the product user.
– Only trained and competent users are permitted to use this product.
– Comply with all local and national rules and regulations associated
with this product.
– Only specialist, trained personnel are permitted to check, repair and
maintain the product as described in these instructions for use and
the technical manual. Further maintenance work that is not detailed in
these instructions for use or in the technical manual must only be
carried out by Dräger or personnel qualified by Dräger. Dräger
recommend a Dräger service contract for all maintenance activities.
– Only use genuine Dräger spare parts and accessories when
performing maintenance work, or the proper functioning of the
product may be impaired.
– Do not use a faulty or incomplete product. Do not modify the product.
– Notify Dräger in the event of any component fault or failure.
– The air supply shall meet the requirements for breathing air according
to EN 12021.
2 Conventions in this document
2.1 Definitions of alert messages
The following alert messages are used in this document to alert the user
to potential hazards. The definitions of the alert messages are as follows.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could
result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could
result in physical injury. It may also be used to alert against unsafe
practices.
NOTICE
Indicates a situation which, if not avoided, could result in damage to the
product or the environment.
2.2 Typographical conventions
2.3 Trademarks
The following website lists the countries in which the Dräger trademarks
are registered: www.draeger.com/trademarks.
The trademarks listed are only registered in certain countries and not
necessarily in the country in which this document is published.
3 Description
3.1 Product overview
Fig. 1 PAS Micro CFH
The Dräger PAS Micro CFH is a self-contained breathing apparatus
which provides respiratory protection for short-duration light work and/or
escape operations in contaminated environments. The equipment
provides a single use of approximately 17 minutes (11 minutes until the
whistle sounds to warn of low cylinder pressure). After use, only service
personnel or Dräger can prepare the product for the next use.
The features of the equipment are as follows.
– The carrying system comprises the shoulder harness, waist belt, and
cylinder strap fitted on to the backplate.
– The pressure reducer connects directly on to the air cylinder, and
reduces the cylinder pressure to the required pressure.
– The whistle on the pressure reducer sounds to warn the wearer that
the cylinder pressure is low.
– The air cylinder and hood are described in section 3.2 and
section 3.3.
3.2 Air cylinder
Breathing air is supplied from a composite air cylinder with a working
pressure rating of 300 bar. The cylinder valve (Fig. 2) has a pressure
indicator which is used to ensure that the cylinder is full before starting
the operation. Red and green sectors on the face of the pressure
indicator show the fill level of the cylinder.
Fig. 2 Cylinder valve
3.3 Hood
A constant flow of air is delivered to the user at a hood (Fig. 3) which fits
all head sizes. The internal nose cup directs the breathing air to the
wearer. The nose cup fits over the nose and mouth, and the head strap
helps to hold it in place. The rubber neck seal minimizes air leakage
during use.
The connecting hose has a quick coupling for rapid connection and
disconnection when required.
The hood is stored ready for use in a disposable transparent plastic
storage pouch. The plastic storage pouch is made from a blend of
polyethylene terephthalate (PET) and polyethylene (PE) which is
currently not recyclable.
Fig. 3 Hood
3.4 Intended use
When the product is used with an approved air cylinder and hood it
provides the wearer with respiratory protection for light working in, or
escaping from, contaminated or oxygen-deficient conditions. The product
is intended to be used for short duration light work and/or escape
applications. Once activated, the product is intended for a single
uninterrupted use. After any use, the product should not be reused until it
has been prepared by trained service personnel or Dräger.
The air cylinder, hood, and other accessories used with this product must
be certified Dräger components, assembled in an approved
configuration. Otherwise the operation of the device may be impaired.
Contact Dräger for further information.
3.5 Limitations on use
This product is not approved for use in CBRN applications.
This product must not be used in any firefighting applications.
3.6 Approvals
The European standards, guidelines, and directives according to which
this product is approved are specified in the declaration of conformity
(see the declaration of conformity or www.draeger.com/product-
certificates).
3.7 Product marking and symbols
3.7.1 Product labelling
Do not remove or alter any product label or marking.
Fig. 4 Product label
4 Use
4.1 Preparation for use (ready position)
WARNING
The full operational time is only available if the pointer of the pressure
indicator is in the green sector.
►Do not start any operational use if the pointer is not in the green
sector of the pressure indicator.
1. Carry out a visual inspection (see section 6.4.1).
2. Ensure that the cylinder is fitted, and that the pointer of the pressure
indicator is in the green sector.
3. Fully loosen the shoulder straps and waist belt.
4. Put on the breathing apparatus.
5. Ensure that the shoulder pads are not twisted.
6. Take the weight of the system on the shoulders by pulling the
shoulder straps. Do not fully tighten at this stage.
7. Close the waist belt buckle.
8. Tighten the waist belt until it is secure and comfortable.
9. Pull the shoulder straps until the breathing apparatus is secure and
comfortable on the hips. Do not over tighten.
Fig. 5 Ready position
4.2 During use
WARNING
The wearer should be in a safe area before the whistle sounds.
►Evacuate to a safe area immediately if the whistle sounds during an
operation.
►A triangle is used in alert messages to indicate possible ways of
avoiding the hazard.
An information symbol is used for notes and additional useful
information.
Trademark Trademark owner
PAS Dräger
52097
No Description
1Hood
2 Shoulder harness
2
1
3
4
678 5
3 Cylinder strap
4Backplate
5 Storage pouch
6 Pressure reducer
7Whistle
8 Waist belt
47462
No Description
1 Pressure indicator
2 Hand wheel
3Whistle
47463
No Description
1 Nose cup
2 Quick coupling (male element)
3 Neck seal
4 Head strap
No Description
1
3
2
1
2
3
4
60466
No. Description
1 Label part number
2 Manufacturer
3 Product name
4 UKCA marking and identification number of approval body
5 CE marking and identification number of notified body
6 Regulation to which the product is designed
7 Product part number
8 Year of manufacture
9 Serial number
10 Nominal cylinder pressure and approximate working duration
47816
PAS Micro CFH
3708010
3715333
Serial
No.
(EU) 2016/425
0086
2797
CYLINDER
300 bar nominal
17 minutes duration
23 4
678910
15
3711974 (A3-D-P) Page 2/3
PAS®Micro CFH
Self-contained breathing apparatus Instructions for use
1. Fully tear open the pouch (Fig. 6) and remove the hood.
Fig. 6 Removing the hood from the pouch
2. Fully open the cylinder valve.
3. Immediately put on the hood as follows:
– Place both hands inside the neck seal and carefully stretch the
hood over the head.
– Wearers with spectacles – take care when stretching the neck
seal over the spectacles.
– Wearers with long hair – tuck the hair inside the hood.
– Position the nose cup over the nose and mouth and breathe
normally.
4. Complete the required operation and then retreat to a safe area
without removing the hood.
With a fully charged cylinder, the warning whistle will sound to warn
of low cylinder pressure after approximately 11 minutes after opening
the cylinder valve. The wearer will have approximately 6 minutes to
evacuate to a safe area after the warning whistle sounds.
4.3 After use
WARNING
Removing the breathing apparatus in a hazardous breathing environment
is unsafe.
►Do not remove the breathing apparatus until in safe breathing
environment.
CAUTION
Careless handing could damage the breathing apparatus and cause the
release of high-pressure air.
►Do not drop or throw down the breathing apparatus.
1. Remove the hood as follows:
– Place both hands inside the neck seal and carefully stretch the
hood over the head.
– Breathe normally after removing the hood.
2. Fully close the cylinder valve.
– Pull the hand wheel against spring force, and simultaneously turn
it clockwise.
3. Loosen the harness by opening the waist belt buckle and lifting the
shoulder strap buckles.
4. Remove the breathing apparatus.
5. Discard the plastic storage pouch.
6. Carry out a visual inspection (see section 6.4.1).
7. Pass the breathing apparatus to trained service personnel or Dräger
to refill the cylinder and repair any faults or damage.
5 Troubleshooting
The troubleshooting guide shows fault diagnosis and repair information
applicable to breathing apparatus users. Further troubleshooting and
repair information is available in instructions for use supplied with the
associated equipment.
Contact service personnel or Dräger when the remedy information
indicates a service task, or if the symptom remains after all remedy
actions have been attempted.
6 Maintenance
6.1 Daily check
The breathing apparatus must be ready for use at all times. Dräger
recommend that the following check is carried out daily.
– Check that the pointer on the cylinder pressure gauge is inside the
green area. If the pointer is in the red area, inform trained service
personnel or Dräger immediately.
If the customer’s risk assessment concludes that less regular checks
are acceptable, the frequency of the check can be extended to a
maximum of monthly. It is the responsibility of the customer to ensure that
the breathing apparatus is ready for use at all times.
6.2 Maintenance table
Dräger recommend servicing and testing the breathing apparatus,
including out-of-use apparatus, in accordance with the maintenance
table.
– Record all service details and testing.
– Refer also to the instructions for use for associated equipment.
– Extra inspection and testing might be required in the country of use to
comply with national regulations.
6.3 Cleaning and disinfecting
CAUTION
Trapped water and ice inside the pneumatic system can impair the
operation of the product.
►Prevent any liquid from entering the pneumatic system, and
thoroughly dry the product after cleaning and disinfecting.
NOTICE
Using cleaning and disinfecting methods not described in this section can
damage the equipment.
►Do not exceed 60 °C for drying, and remove components from the
drying facility immediately when dry. Drying time in a heated dryer
must not exceed 30 minutes.
►Do not immerse pneumatic or electronic components in cleaning and
disinfecting solutions or water.
For information about suitable cleaning and disinfecting
agents and their specifications refer to document 9100081 at
www.draeger.com/IFU.
6.3.1 General cleaning and disinfecting
Clean the product if it is dirty. If the product has been exposed to
contaminants, disinfect any components that come into direct and
prolonged contact with the skin.
Refer also to the instructions for use for other associated equipment.
Work equipment
– Clean, lint free cloths.
1. Clean the product manually using a cloth moistened with cleaning
solution to remove excess dirt.
2. Apply disinfecting solution to components that come into direct and
prolonged contact with the skin.
3. Remove all cleaning and disinfecting agents by rinsing all
components thoroughly with clean water.
4. Dry all components using a dry cloth, in a heated dryer, or in air.
5. Contact service personnel or Dräger if disassembly of the pneumatic
system or electronic components is required.
6.4 Maintenance tasks
6.4.1 Visual inspection
A visual inspection must fully check the product including all component
parts and accessories.
1. Check that the product is clean and undamaged, paying particular
attention to pneumatic system components, connectors, and
elastomeric components such as hoses.
– Typical signs of damage that can affect the operation of the
product include impact, abrasion, cutting, corrosion, and
discoloration.
2. Report damage to service personnel or Dräger, and do not use the
product until faults are rectified.
6.4.2 Stowing the hood in the storage pouch
This is a place holder module for stowing the hood if it is needed.
1. Disconnect the hood hose from the breathing apparatus.
2. Carefully roll up the hood, ensuring that the visor is not creased or
folded.
3. Pass the connecting hose through the hole in the storage pouch as
shown (Fig. 7).
– The illustrations printed on the pouch should be facing upwards.
Fig. 7 Folding the hood and placing it into the pouch
4. Fit the hood into the storage pouch.
5. Remove the cover from the adhesive strip.
6. Seal the pouch by folding the adhesive strip over the opening of the
pouch (Fig. 8).
Fig. 8 Sealing the pouch
7. Connect the pouch to the breathing apparatus (see section 6.4.3).
6.4.3 Connecting the pouch to the breathing apparatus
1. Connect the pouch to the right-hand buckle of the breathing
apparatus (Fig. 9).
2. Connect the hood hose to the breathing apparatus.
Fig. 9 Connecting the pouch to the breathing apparatus
6.4.4 Fitting the air cylinder
CAUTION
Only air cylinders listed in the Dräger certification are approved for use.
►Only fit an air cylinder which is approved for use. Contact Dräger for
further information.
1. Ensure that the pointer of the pressure indicator is in the green sector.
2. Place the carrying harness on a clean flat surface.
3. Ensure that the following are clean and undamaged.
– The threads of the high-pressure connector (Fig. 10) and cylinder
valve.
– The O-ring (Item 1).
– The sintered filter (Item 2).
Fig. 10 High-pressure connector
4. Fully extend the cylinder strap.
52096
Symptom Fault Remedy
Air leak from the
medium-pressure hose
connection at the pres-
sure reducer (safety
relief valve)
Fault O-ring, retainer,
spring or pressure
reducer
Service task
Incorrect flow rate Pressure reducer fault Service task
Medium-pressure hose
fault
Service task
Poor sounding whistle Whistle dirty Clean whistle flute and
retest
Whistle not functioning
correctly
Activation mechanism
fault
Service task
Component
or System
Task Every
month
Every
year
Every 6
years
Complete
breathing
apparatus
Visual inspection (see
section 6.4.1)
X1)
1) Clean the breathing apparatus if it is dirty. If the breathing apparatus has been
exposed to contaminants, disinfect any components that come into direct and
prolonged contact with the skin.
Breathing cycle and static
tests
X2)
2) These maintenance tasks can only be carried out by Dräger or trained service
personnel. Details of the tasks are contained in the technical manual which is
issued to service personnel that have attended a relevant Dräger maintenance
course.
Pressure
reducer
Inspect the sintered filter X2),3)
3) Replace the sintered filter if a drop in pressure reducer performance is observed
during a flow check or if it is visibly damaged.
Inspect the high-pressure
connector O-ring
X2),4)
4) Replace the high-pressure connector O-ring if it leaks or is visibly damaged.
Overhaul. Contact Dräger
for the Repair Exchange
(REX) service
X5)
5) Where the breathing apparatus is subjected to a high level of use (in training
establishments etc.), reduce the overhaul period for the pressure reducer. In
these circumstances, Dräger recommend that the overhaul frequency should be
less than 5000 applications of use. An application of use is defined as a single
use of the fully assembled breathing apparatus, where the user breathes from
the air cylinder.
Cylinder Check the test date of the
cylinder
X
Recertification According to national regu-
lations in the country of use
Valve overhaul During cylinder recertifica-
tion
52098520995210047464
12
3711974 (A3-D-P) Page 3/3
PAS®Micro CFH
Self-contained breathing apparatus Instructions for use
5. Insert the cylinder through the cylinder strap and align valve with the
pressure reducer (Fig. 11). To prevent damage, ensure that the
cylinder remains clear of the hand wheel of the pressure reducer
when inserting the cylinder.
Fig. 11 Cylinder fitting
6. Align the cylinder with the centre of the carrying harness and connect
the cylinder valve to the pressure reducer (see the direction arrow
in Fig. 11). Do not fully tighten.
7. Secure the cylinder as shown in (Fig. 12).
– Take up the slack in the cylinder strap (Step 1).
– Pull the strap through the buckle (Step 2).
– Insert the loose end of the strap into the guide loop (Step 3).
Fig. 12 Cylinder strap
8. Fully hand-tighten the cylinder high-pressure connector. Do not use
tools or over tighten.
6.4.5 Removing the air cylinder
WARNING
High-pressure air release could cause injury to the user or other
personnel near the breathing apparatus.
►Close the cylinder valve before attempting to disconnect the air
cylinder.
1. Close the cylinder valve.
2. Loosen the cylinder strap.
3. Disconnect the cylinder valve from the pressure reducer.
4. Remove the cylinder. To prevent damage, ensure that the cylinder
remains clear of the high-pressure connector of the pressure reducer.
6.4.6 Filling the air cylinder
WARNING
►The air quality for filling the air cylinder must conform to requirements
of EN 12021.
►Do not fill the cylinder in a potentially explosive atmosphere.
– Refer also to the instructions supplied with the cylinder and the
compressor for refilling.
– Only fill air cylinders which:
– Conform to national standards.
– Have the original test date and test mark of the manufacturer.
– Have not exceeded the test date indicated on the cylinder by the
last testing station.
– Are not damaged.
– Do not open the cylinder valve until the air cylinder is connected to the
compressor.
– Refill to the rated working pressure of the cylinder.
– Dräger recommend a fill rate of 27 bar/minute (rapid filling
increases the temperature and causes incomplete filling).
– Compression of cylinder air can cause a small temperature
increase during refilling, resulting in incomplete filling. Top up the
cylinder when it cools if necessary.
– To prevent overfilling the cylinder, Dräger recommend using a
pressure-limiting device on the compressor.
7 Transport
Transport the product in its original packaging.
8 Storage
8.1 Storage preparation
– Extend the shoulder straps and waist belt.
– Stow the hood as described in section 6.4.2.
– Route hoses so that the bend radius is not too acute, and the hose is
not stretched, compressed, or twisted.
– Dräger recommend placing the full breathing apparatus into a suitable
storage bag.
8.2 Storage conditions
– Store the product between -15 °C and +25 °C.
– Ensure that the environment is dry, free from dust and dirt, and does
not subject the equipment to wear or damage due to abrasion.
– Do not store the equipment in direct sunlight.
– Fix the product securely to any raised mounting point to prevent it
from falling.
– If storing the equipment in a vehicle, ensure that the breathing
apparatus is securely retained and does not interfere with the
operation of the vehicle.
9 Disposal
9.1 General
Dispose of the product in accordance with applicable rules and
regulations in the country of use.
10 Technical data
11 Manufacturer and document
information
Manufacturer
Dräger Safety UK Limited
Ullswater Close
Blyth, NE24 4RG
United Kingdom
Tel: +44 1670 352 891
Fax: +44 1670 356 266
www.draeger.com
3711974_en
© Dräger Safety UK Limited
Edition: 6 – 2024-01 (Edition: 1 – 2021-03)
Subject to alteration
4747347465
Item Data
High-pressure connector
Pressure 300 bar
Capacity 3 litres
Connector type Standard G5/8” as per EN 144-2
Whistle
Commencement range 108 to 105 bar
Cessation range 1.75 to 0 bar
Volume >90 dBA