Dräger PAS Colt Series User manual

3352429 (A3-D-P)
PAS Colt Series
Combination escape SCBA/airline SAR respiratory protection equipment
Instructions for Use
i
1
2
1
3629
2
3628/3627
6
3626
3
3630
7
3632
8
3633
9
3634
4
3631
5
3625
1 For your safety
1.1 General safety statements
● Beforeusingthisproduct,carefullyreadtheInstructionsforUse.
● StrictlyfollowtheInstructionsforUse.Theusermustfullyunderstandandstrictly
observe the instructions. Use the product only for the purposes specied in the
IntendedUsesectionofthisdocument.
● Do not dispose of the Instructions for Use. Ensure that they are retained and
appropriatelyusedbytheproductuser.
● Onlyfullytrainedandcompetentusersarepermittedtousethisproduct.
● Complywithalllocalandnationalrulesandregulationsassociatedwiththisproduct.
● Onlytrainedandcompetentpersonnelarepermittedtoinspect,repairandservice
the product. Dräger recommends a Dräger service contract for all maintenance
activitiesandthatallrepairsarecarriedoutbyDräger.
● Properlytrainedservicepersonnelmustinspectandservicethisproductasdetailed
inthemaintenancesectionofthisdocument.
● UseonlygenuineDrägersparepartsandaccessories,ortheproperfunctioningof
theproductmaybeimpaired.
● Donotuseafaultyorincompleteproduct,anddonotmodifytheproduct.
● NotifyDrägerintheeventofanycomponentfaultorfailure.
● Before occupational use of this respirator a written respiratory protection
programmustbeimplementedmeetingallthelocalgovernmentrequirements.
IntheUnitedStatesemployersmustcomplywithOSHA29CFR1910.134which
includesmedicalevaluation,training,andttesting.
1.2 Denitionsofalerticons
Alerticonsareusedinthisdocumenttoprovideandhighlighttextthatrequiresagreater
awarenessbytheuser.Adenitionofthemeaningofeachiconisasfollows:
!
WARNING
 Indicatesapotentiallyhazardoussituationwhich,ifnotavoided,couldresultin
deathorseriousinjury.
!
CAUTION
 Indicatesapotentiallyhazardoussituationwhich,ifnotavoided,couldresultin
physicalinjuryordamagetotheproductorenvironment.Itmayalsobeusedto
alertagainstunsafepractices.
i
i
NOTICE
 Indicatesadditionalinformationonhowtousetheproduct.
2 Description
2.1 Product overview
Thisvariant ofthe Dräger PASColt Series provides respiratory protection for working
in a contaminated environment using an airline (supplied airline respirator – SAR),
or escaping from a contaminated environment using the air cylinder (self-contained
breathingapparatus–SCBA).
During airline use, the PAS Colt is connected to a breathing-quality independent air
supply and the cylinder valve is closed. If the independent air supply falls below the
requiredpressure,theweareropensthecylindervalve,disconnectstheindependentair
supplyandthenescapestoasafeareabreathingairfromthecylinder.Thetimeavailable
fortheescapeisdependentonthecapacity(volume)ofthecylinderandthebreathing
rateofthewearer.
Thefeaturesoftheequipmentare:
● Thecarryingsystemisabandoliershoulderharnessandwaistbelt.
● Thelungdemandregulator(LDR)hasaresetbutton(Fig1,Item1)thatswitches
offtheairowthroughtheregulator,andapurgebutton(Fig1,Item2)thatcanbe
usedtodeliveranadditionalairowintothefacemaskwhenrequiredbythewearer
duringairlineuse.Theholder(Fig2,Item1)providesastowagepointtoprotectthe
regulatorcouplingandO-ringagainstdamagewhenitisnotttedinthefacemask.
● Therst-stageregulator(Fig2,Item2)hasananti-vibrationstrapttedthatisused
topreventlooseningoftheconnectorhandwheel.
● Theairlineconnector(Fig2,Item3)isamalecouplingthatisusedtoconnectan
independentairsupplyforsuppliedairlinerespirator(SAR)use.
● Alow-pressurewhistleisttedonsomevariantsofthisproduct(seeSection2.1.1).
● Thehip-mountedcylinderholsterhasadropdownfacilitythatcanassistwearer
movementbymakingthecylinderandholstermoremanoeuvrable(inaconned
spaceforexample).
● Theremainingpressureintheaircylinderisshownonacontentsindicatoronthe
cylinder.
● Athighstrapisavailableasanaccessory.
2.1.1 Low-pressure whistle (optional)
IfthePASColtisttedwithalow-pressurewhistle,itoperatesasfollows:
● Duringairlineuse,thewhistlesoundstowarnthewearerthattheindependentair
supplyhasfallenbelowtherequiredpressure.
i
i
NOTICE
 During an escape (when the wearer is breathing from the air cylinder), the
whistlesoundswhenthecylinderpressureisverylow.Thisisnotanend-of-
service-timeindicator(EOSTI).
Seethetechnicaldata(Section9)forthewhistleoperatingpressures.
2.2 Intended use
Whenthisproductisusedwithanapprovedfacemask,aircylinderandindependent
air supply, it provides the wearer with respiratory protection for working in, or
escapingfrom,contaminatedoroxygen-decientconditions.Itisintendedforusein
applicationswhereahighlevelofrespiratoryprotectionisrequired,includingIDLH
(immediatelydangeroustolifeandhealth)atmospheres.
The air cylinder, face mask and other accessories used with this product must be
certiedDrägercomponents,assembledinanapprovedconguration;otherwisethe
operationofthedevicemaybeimpaired.Itisessentialthatushingandnon-ushing
facemasksandlungdemandregulatorsshouldnotbemixed(i.e.onlyaushing-
typefacemaskshouldbeusedwithaushing-typelungdemandregulator).Contact
Drägerforfurtherinformation.
2.3 Limitations on use
ThisproductisnotapprovedforuseinCBRNapplications.
2.4 Approvals
The PAS Colt Series of combination escape/airline respiratory protection equipment
conformstotherequirementsofNIOSH.Thisproductisapprovedforuseasasupplied
airlinerespirator(SAR)andasaself-containedescaperespirator,andmustonlybeused
inconjunctionwithcompressed-aircylindersapprovedbyNIOSH.
3 Use
!
WARNING
 Thetimerequiredtoallowthewearertoescapetoasafeareamustbewithin
thecapacity(volume)ofthecylinderselected,takingintoaccountthebreathing
rateofthewearer.Whenselectingthetypeanddurationofescapeequipment
itisessentialtoconsiderthepotentialhazardsandprobableescaperoutes.
 Thecylinderandairlineairqualityshallmeettherequirementsforbreathingair
accordingtoCGAG-7.1GradeDorhigherquality.
Beforeusingairlineequipment,ensurethattheindependentairsupplymeets
theairqualityrequirementsandcomplieswithpressure,owandhoselength
requirementsinthetechnicaldata(seeSection9),andhasbeenissuedwitha
permitforuseifnecessary.
 IfusingaPASColtwithoutanintegrallow-pressurewhistle,Drägerrecommend
thatalow-pressurewarningdeviceisttedtotheindependentairsupply.
D
3000 PSI
EMPTY FUL
14
1234
USE
NO
OIL
D
3000 PSI
EMPTY FUL
14
1234
USE
NO
OIL
D
3000PSI
EMPTY FUL
14
1234
USE
NO
OIL
1
2
3
3
2
1
i
i
NOTICE
 Afterstorage attemperatures below32 °F(0 °C)leakage may be observed
whenthecylindervalveisinitiallyopenedduetoiceformation.
● Ifleakageisobservedfromthelungdemandregulator:Pressthepurge
button(Fig1,Item2)toallowarushofairtopassthroughthelungdemand
regulatorandthenquicklypresstheresetbutton(Fig1,Item1)toswitch
offthepositivepressure.Resumenormaloperation.
● Intheeventthatleakagestilloccurs,removethebreathingapparatusfrom
serviceandreportthefaulttotrainedservicepersonnelorcontactDräger.
3.1 Preparation for use
3.1.1 Visual inspection
Carry out a visual inspection, checking the full breathing apparatus including all
componentpartsandaccessories.Checkthattheequipmentiscleanandundamaged,
paying particular attention to pneumatic components, hoses and connectors. Typical
signsofdamagethatmayaffecttheoperationofthebreathingapparatusincludeimpact,
abrasion,cutting,corrosionanddiscoloration.Reportdamagetoservicepersonneland
donotusetheapparatusuntilfaultsarerectied.
3.1.2 Fitting the cylinder
1. Ensurethatthe cylinder isfully charged, withthe pointer ofthecylinder pressure
indicatorinsidethegreenarea.
2. Check that the threads of the valve port and the regulator handwheel are
undamaged,andtheO-ringisinpositionandundamaged.
3. Fullyinsertthecylinderintothecarryingholster.
!
CAUTION
 Topreventdamage,ensurethatthecylinderremainsclearofthehandwheelof
therst-stageregulatorwhileinsertingthecylinder.
4. Alignthecylinderwiththeregulatorandtightenthehandwheelhandtight(Fig3).
5. Pulltheanti-vibrationstrapoverthetopofthecylindervalve(Fig4).
3.1.3 Functional testing
!
WARNING
 If breathing apparatus fails to meet any of the standards or parameters
describedinthefunctionaltests,orifanimmediateleakisevident,thereisa
systemfault.ReportthefaulttotrainedservicepersonnelorcontactDräger.Do
notusethebreathingapparatusuntilthefaultconditionisrectied.
Leak test
1. Ensurethatthepointerofthecylinderpressureindicatorisinsidethegreenarea.
2. Presstheresetbutton(Fig1Item1)ofthelungdemandregulatortoswitchoffthe
positivepressure.
3. Openthecylindervalveslowly,butfully,topressurizethesystemandthenclosethe
valve.Thereshouldbenoaudibleleak.Ifthereisanyleak,investigateandrepair
theleakbeforeuse.Ifnecessary,useasoapysolutiontolocatetheleak.
4. Pressthepurgebutton(Fig1,Item2)toventthesystem.
Airlineowtest(onlyrequiredforairlineuse)
1. Presstheresetbutton(Fig1,Item1)ofthelungdemandregulator,andthencheck
thatthecylindervalveisclosed.
2. Connect the independent air supply to the male coupling (Fig 2, Item 3). If the
independentairsupplyhasashut-offvalve,openthevalve.
!
WARNING
 Donotdirecttheairowontotheface,eyesorskin.
3. Pressthepurgebutton(Fig1,Item2)for3to5seconds.Anunobstructedairow
shouldowfromtheoutletofthelungdemandregulator.
4. Isolateanddisconnecttheindependentairsupply.
5. Pressthepurgebutton(Fig1,Item2)toventthesystemandthenpressthereset
button(Fig1,Item1)toswitchoffthepositivepressure.
3.1.4 Putting on the PAS Colt (ready position)
 ReferalsotoFig2whichshowsthePASColtworninthereadyposition.
1. Ensurethatthepointerofthecylinderpressureindicatorisinsidethegreenarea.
2. Openthewaistbeltbuckleandfullyextendthewaistandshoulderstrap.
3. Placetheleftarmthroughtheshoulderharness,takingtheharnessoverthehead
andontotherightshoulder,positioningthestrapdiagonallyacrossthebodywiththe
cylinderpositionedagainstthelefthip.
4. Loopthewaistbeltaroundthewaistandfastenthebuckle–donottighten.
5. Gripthecylindervalvewiththelefthandandliftuntilthewaistbeltisinlinewiththe
waist.Thentightenthewaistbeltstrapuntiltheequipmentissecureandcomfortable
onthewaist.Pulldowntoadjusttheshoulderstrap.
6. InspectthefacemaskasdenedintherelevantInstructionsforUse.
7. CheckthatthefacemaskportandtheregulatorO-ringarecleanandundamaged.
8. Removethelungdemandregulatorfromtheholderandpressitintotheportofthe
facemaskuntilitlatchesinposition.Checktheattachmentbygentlyattemptingto
pullthecouplingapart.
9. Puttheneckstrapofthefacemaskoverthehead,andtheninserttheneckstrap
studintotheholeinthecentrestrapoftheheadharness.
3.2 During use
ThisvariantofthePASColtSeriescanbeusedforcombination(escape/airline)useor
purelyforescapeuse.Forcombinationusethefacemaskiswornduringthetask;and
forescapeusethePASColtisworninthereadypositionuntilanescapeisnecessary.
Dependingontheintendeduse,refertotherelevantproceduresfromthefollowing:
● Forcombination(escape/airline)use,seeSection3.2.1
● Forescapeuseonly,seeSection3.2.2
3.2.1 Combination (escape/airline) use
Putting on the face mask
1. Presstheresetbutton(Fig1,Item1)ofthelungdemandregulatortoswitchoffthe
positivepressure.
2. Connect the independent air supply to the male coupling (Fig 2, Item 3). If the
independentairsupplyhasashut-offvalve,openthevalve.
!
WARNING
 Correcttofthefacemaskcanonly be achievedifthecompletefacemask
sealmakescontactwithskin.Headhair,facialhair(includingbeardstubbleand
sideburns),earrings,otherfacialpiercingsandnormalspectacleswillinterfere
withthefacemasksealandarenotpermittedinthesealingarea.Additionally,
headhairthatcouldaffectthefacemaskt(buns,pony-tails,hairpieces,etc.)
isnotpermitted.
i
i
NOTICE
 ReferalsotothefacemaskInstructionsforUse.
3. Detachtheneckstrapstudfromthecenterstrapoftheheadharness.
4. Spreadtheheadharness(Fig5).Placethechinintothechincupofthefacemask
andpulltheharnessovertheheadlocatingtheharnesscenterplateonbackofthe
head.
5. ReferringtoFig6,tightenbothlower(1)andthenupperstraps(2)evenlytowards
back of the head. If necessary, tighten the center strap (3). The wearer’s rst
inhalationafterachievingafacesealwillautomaticallyswitchontheairsupplyinto
thefacemask.
6. Check that the head has a full range of movement without pulling against the
regulator hose. If any resistance to movement is felt, readjust the hose routing
andthenrecheck.Ifresistanceisstillfelt,do notusethebreathingapparatusand
contactDräger.
Function check
1. Isolate the independent air supply (close the valve or disconnect) and breathe
normallytoemptythesystemofair.Whenempty,thefacemaskshouldholdonto
thefacetoindicateapositiveseal.
2. Immediatelyreapplytheindependentairsupplyandbreathenormally.
3. Inhaleandholdyourbreath–thereshouldbenoaudibleleak.Ifaleakisdetected,
readjusttheheadharnessandretest.
4. Recommencebreathing–exhaledairshouldoweasilyoutoftheexhalationvalve.
5. Momentarilypressthepurgebutton(Fig1,Item2)tocheckforadditionalairow.
When the function check has been satisfactorily completed, breathe normally and
proceedtotheworkingarea.
Escape procedure (independent air supply failure)
!
WARNING
 Duringsuppliedairlinerespirator(SAR)use,thePASColtcylindervalvemust
remainclosed.Theescapedurationstartsfromthetimeofopeningthecylinder
valveanddisconnectingtheairlinesupply.
 Drägerstronglyadvisethatthepurgebutton(Fig1,Item2)isnotusedduring
anescape.Usingthepurgebuttonwoulduseairfromthecylinderandreduce
thetimeavailableforescape.
i
i
NOTICE
 IfthePASColtisttedwithalow-pressurewhistle,duringairlineusethewhistle
soundstoindicatethattheindependentairsupplyhasfallenbelowtherequired
pressure.Ifthewhistlesoundsandthenstopsbeforetheescapeprocedureis
started,theairsupplypressurecouldhavebeenrestored.The wearer must
conrm that the correct pressure (87 to 125  psi) is available, or escape as
describedbelow.
Iftheindependentairsupplyfails,breathenormallyandimmediatelyproceedasfollows:
1. Openthecylindervalve(counterclockwise)slowly,butfully,andcontinuetobreathe
normally.
2. DisconnecttheindependentairsupplyfromthePASColt.
3. Immediatelyleavethehazardousareabytheshortestandsafestescaperoute.
Thewearermustbeinasafeareabeforetheaircylinderisempty.Wheninasafearea,
remove the lung demand regulator from the face mask if necessary and continue to
breathenormally.
i
i
NOTICE
 IfthePASColtisttedwithalow-pressurewhistle,duringanescapethewhistle
wouldsoundwhenthecylinderpressurewasintherange72.5to58psi.This
isnotanend-of-service-time-indicator(EOSTI).
3.2.2 Escape use only (putting on the face mask and escaping)
!
WARNING
 Correcttofthefacemaskcanonly be achievedifthecompletefacemask
sealmakescontactwithskin.Headhair,facialhair(includingbeardstubbleand
sideburns),earrings,otherfacialpiercingsandnormalspectacleswillinterfere
withthefacemasksealandarenotpermittedinthesealingarea.Additionally,
headhairthatcouldaffectthefacemaskt(buns,pony-tails,hairpieces,etc.)
isnotpermitted.
i
i
NOTICE
 ReferalsotothefacemaskInstructionsforUse.
1. Presstheresetbutton(Fig1,Item1)ofthelungdemandregulatortoswitchoffthe
positivepressure.
2. Open the cylinder valve (counterclockwise) slowly, but fully, to pressurize the
system.
3. Detachtheneckstrapstudfromthecenterstrapoftheheadharness.
4. Spreadtheheadharness(Fig5).Placethechinintothechincupofthefacemask
andpulltheharnessovertheheadlocatingtheharnesscenterplateonbackofthe
head.
5. ReferringtoFig6,tightenbothlower(1)andthenupperstraps(2)evenlytowards
back of the head. If necessary, tighten the center strap (3). The wearer’s rst
inhalationafterachievingafacesealwillautomaticallyswitchontheairsupplyinto
thefacemask.
3352429 (A3-D-P)
PAS Colt Series
Combination escape SCBA/airline SAR respiratory protection equipment
Instructions for Use
i
3352429
©DrägerSafetyUKLimited
Edition11–November2015
Subjecttoalteration
!
WARNING
 Thedurationofthecylinderairbeginsfromthetimeoftherst-breathactivationofthelungdemandregulator.
 Drägerstronglyadvisethatthepurgebutton(Fig1,Item2)isnotusedduringanescape.Usingthepurgebuttonwoulduseairfromthecylinderandreducethetimeavailable
forescape.
6. Immediatelyleavethehazardousareabytheshortestandsafestescaperoute.
Thewearermustbeinasafeareabeforetheaircylinderisempty.Wheninasafearea,removethelungdemandregulatorfromthefacemaskifnecessaryandcontinuetobreathe
normally.
3.2.3 Dropdown cylinder holster
• Toreleasethedropdowncylinderholster:
 a. Holdthecylindervalvewiththelefthandand,withtherighthand,pressandholdtheredbuttontoopenthelockingmechanism(Fig7).
 b. Liftthecylinderandholsterclearofthelockingmechanismandthenreleasetheredbutton(Fig8).
 c. Lowerthecylinderuntilitissupportedbytheharnessstraps(Fig9).Theholsterisheldbytwoharnessstraps,onexedandoneadjustable.
 d. Holdthecylindervalvetomovethecylinderandholsterasrequired.
• Toreconnectthedropdowncylinderholster,alignandpresstherollerontheholsterintothelockingmechanism.
3.3 After use
!
WARNING
 Donotremovetheequipmentuntilinsafearea,clearofhazard.
!
CAUTION
 Donotdroporthrowdownequipmentasdamagecouldoccur.
1. Loosenthefacemaskstraps.Asthesealbetweenthefacemaskandthefaceisbroken,presstheresetbutton(Fig1,Item1)toswitchoffthepositivepressure.Removetheface
maskandfullyextendallofthestrapsoftheheadharness.
 ○ Afterairlineuse–isolateanddisconnecttheindependentairsupplyfromthePASColt.Retthecapontotheairlineconnector.
 ○ Afteranescape–closethecylindervalve.
2. Pressthepurgebutton(Fig1,Item2)tofullyventthesystem.
3. Removethelungdemandregulatorfromthefacemaskandstowitintheholder(Fig2,Item1).
4. Unbucklethewaistbelt,lifttheshoulderstrapbuckletoloosentheharnessandthenremovetheequipment.
5. Carryouttheafterusemaintenancetasksinthemaintenancetable(seeSection5.1).
4 Troubleshooting
The troubleshooting guide shows fault diagnosis and repair information applicable to breathing apparatus users. Further troubleshooting and repair information is available in
InstructionsforUsesuppliedwithassociatedequipment(e.g.facemask).
ContactservicepersonnelorDrägerwhentheremedyinformationindicatesaservicetask,orifthesymptomremainsafterallremedyactionshavebeenattempted.
Symptom Fault Remedy
High-pressureairleakorfailedleaktest Looseordirtyconnector Disconnect,cleanandreconnectcouplingsandretest
Faultyhoseorcomponent Substituteuserreplaceableaccessoriesandretest
Airleakfromlungdemandregulator Iceparticlesonsealingelements Pressthepurgebutton(Fig1,Item2)toallowarushofairtopass
throughthelungdemandregulatorandthenquicklypressthereset
button(Fig1,Item1)toswitchoffthepositivepressure.
Airleakfrommedium-pressurehoseconnection
attherst-stageregulator(safetyreliefvalve)
FaultyO-ring,retainer,springorrst-stage
regulator
Servicetask
Highorlowmediumpressure First-stageregulatorfault Servicetask
Poorsoundingwhistle(iftted) Whistledirty Cleanwhistleuteandretest
Whistlenotfunctioningcorrectly(iftted) Activationmechanismfault Servicetask
5 Maintenance
5.1 Maintenance table
Serviceandtestthebreathingapparatus,includingout-of-useapparatus,inaccordancewiththemaintenancetable.Recordallservicedetailsandtesting.ReferalsototheInstructions
forUseforthelungdemandregulator,facemaskandotherassociatedequipment.
Additionalinspectionandtestingmayberequiredinthecountryofusetoensurecompliancewithnationalregulations.
Daily check–IfthePASColtisheldinaready-for-usecondition,checkdailythatthepointerofthecylinderpressureindicatorisinsidethegreenarea.Chargethecylinderifitisin
theredarea(seeSection5.2.2).
Component/
System
Task After use Every month Every year Every 10 years
Complete
apparatus
Cleananddisinfect(seeSection5.3) m
Visualinspection(seeSection3.1.1) m m
Functionaltesting(seeSection3.1.3) m m
Flowandstatictests(seeNote1) m
Lungdemand
regulator
Cleananddisinfect(seeNote2andSection5.3) m
First-stage
regulator
Medium-pressurecheck(seeNote1) m
Inspectthesinteredlter(seeNote1andNote3) m
Inspectthehigh-pressureconnectorO-ring(seeNote1andNote4) m
Overhaul.ContactDrägerfortheRepairExchange(REX)service(seeNote5) m
Cylinder Chargecylindertocorrectworkingpressure(seeSection5.2.2) m
Checkchargedpressure(storedcylindersonly) m
Checktestdateofcylinder(carboncompositecylindersover15yearsoldmustberetired) m
Recertication Accordingtonationalregulationsinthecountryofuse
Cylindervalve Overhaul Atthetimeofcylinderrecertication
Notes
m Drägerrecommendations
1 ThesemaintenancetasksmayonlybecarriedoutbyDrägerortrainedservicepersonnel.DetailsofthetestsarecontainedintheTechnicalManualwhichisissuedtoservice
personnelthathaveattendedarelevantDrägermaintenancecourse.
2 Lightly lubricate the O-ring of the lung demand regulator as required (recommended lubricant is Dow Corning 111 Valve Lubricant and Sealant). Products other than the
recommendedlubricantarenottestedandmaydamagetheequipment.
3 Replacethesinteredlterifadropinrst-stageregulatorperformanceisobservedduringaowcheckorifitisvisiblydamaged.
4 Replacethehigh-pressureconnectorO-ringifitisfoundtoleakduringfunctionaltestingoriftheO-ringisvisiblydamaged.
5 Wherethebreathingapparatusissubjectedtoahighlevelofuse(intrainingestablishmentsetc.),reducetheoverhaulperiodfortherst-stageregulator.Inthesecircumstances,
Drägerrecommendthattheoverhaulfrequencyshouldbelessthan5,000applicationsofuse.Anapplicationofuseisdenedasasingleuseofthefullyassembledbreathing
apparatus,wheretheuserbreathesfromtheaircylinder.Itdoesnotincludesystempressurizationforpre-operationalchecks.
5.2 Maintenance tasks
5.2.1 Removing the cylinder
!
WARNING
 High-pressureairreleasemaycauseinjurytotheuserorotherpersonnelnearthebreathingapparatus.Closethecylindervalveandfullyventthesystembeforeattempting
todisconnectanaircylinder.
1. Closethecylindervalveandpressthepurgebutton(Fig1,Item2)tofullyventthesystem.
2. Removetheanti-vibrationstrapfromthecylindervalve(Fig4).
3. Disconnectthecylindervalvefromtherst-stageregulator.
4. Carefullyremovethecylinderfromtheholster.Topreventdamage,ensurethatthehandwheeloftherst-stageregulatorremainsclearofthecylinder.
5.2.2 Air cylinder charging
!
WARNING
 Airqualityforcompressed-aircylindersmustconformtotheminimumgraderequirementsforType1gaseousairasdenedinCGA(CommoditySpecicationforAir)G-7.1
GradeDorhigherquality.
 Ifthebreathingairmoisturecontentexceedsrecommendedlevels,iceparticlescanform,reducingorblockingairow.Watercontentofthebreathingairinanycontainer
mustbechecked.
● Referalsototheinstructionssuppliedwiththecylinderandthechargingunitforrechargingthecylinder.
● Onlychargecompressed-aircylinderswhich:conformtonationalstandards,featuretheoriginalmanufacturer’stestdateandtestmark,havenotexceededthetestdateindicated
onthecylinderbythelasttestingstation,andarenotdamaged.
● Topreventingressofmoistureintothecylinder,ensurethatthecylindervalveremainscloseduntilconnectedtothechargingunit.
● Recharge to the rated working pressure of the cylinder. Dräger recommend a charge rate of 300 psi/minute. Rapid charging will increase the temperature resulting in an
incompletecharge.Recheckatambienttemperatureandtopupthechargeifrequired.
● Topreventoverchargingofthecylinder,Drägerrecommendusingapressure-limitingdeviceonthechargingcompressor.
5.3 Cleaning and disinfecting
!
CAUTION
 Cleaning agents and disinfectants listed below are not manufactured by
Drägerandhavebeenreviewedonlyforcompatibilitywhenusedtocleanor
disinfectthesubjectDrägerproduct(s).Readandcomplywithallinstructions
foruseprovidedbythemanufacturersofsuchagentsanddisinfectants.Dräger
expresslydisclaims all responsibility for any damage, personal injury or loss
resultingfromtheuseofsuchagentsordisinfectants.
 Donotexceed86°F(30°C)forwashing,disinfectingandrinsingsolutions.Do
notexceed140°F(60°C)fordrying,andremovecomponentsfromthedrying
facilityimmediatelywhendry.Dryingtimeinaheateddryermustnotexceed30
minutes.
 Donotimmersepneumaticincleaningsolutionsorwater.
 If water is trapped and then freezes inside the pneumatic system of the
breathing apparatus (such as the lung demand regulator), operation will be
impaired.Prevent any liquid from entering,and thoroughly drythe breathing
apparatusaftercleaningtopreventthisfromoccurring.
Referalsototheinstructionsforthefacemaskandotherassociatedequipment.
5.3.1 Manual cleaning of the breathing apparatus (USA)
Cleaning and disinfecting materials:
● Cleaningagent–1008GreenLiquidHandDishWash
● Disinfectingagent–800Spur-TexDisinfectantCleaner-Deodorant(concentration:
1.6%(2ozpergallon))
● Useonlycleanlint-freecloths.
1. Preparecleaningsolutionaspermanufacturer’sinstructions.Cleanthebreathing
apparatus manually using a cloth moistened with cleaning solution to remove
excessdirt.
2. Preparedisinfectingsolutionaspermanufacturer’sinstructions.Applytoallinternal
andexternalsurfaces,ensuringthatallsurfacesremainvisiblywetfor15minutes.
3. Rinse all components thoroughly with clean water to remove all cleaning and
disinfectingagents.
4. Dryallcomponentsusingadrycloth,inaheateddryerorinair.
5. Contact service personnel or Dräger if disassembly of pneumatic or electronic
componentsisrequired.
5.3.2 Manual cleaning of the breathing apparatus (Canada)
Cleaning and disinfecting materials:
● Cleaningagent–mildsoapsolution
● Disinfectingagent–NeutralDisinfectantCleaner(concentration:0.5%(0.5ozper
gallonor15mlper3.785liters))
● Useonlycleanlint-freecloths.
1. Preparecleaningsolutionaspermanufacturer’sinstructions.Cleanthebreathing
apparatus manually using a cloth moistened with cleaning solution to remove
excessdirt.
2. Preparedisinfectingsolutionaspermanufacturer’sinstructions.Applytoallinternal
andexternalsurfaces,ensuringthatallsurfacesremainvisiblywetfor10minutes.
3. Rinse all components thoroughly with clean water to remove all cleaning and
disinfectingagents.
4. Dryallcomponentsusingadrycloth,inaheateddryerorinair.
5. Contact service personnel or Dräger if disassembly of pneumatic or electronic
componentsisrequired.
6 Storage
Store the equipment between 5 °F and 77 °F (-15 °C and +25 °C). Ensure that the
environmentisdry,freefromdustanddirt,anddoesnotsubjecttheequipmenttowear
ordamageduetoabrasion.Donotstoretheequipmentindirectsunlight.Notealsothe
following:
● Extendtheshoulderstrap,thewaistbeltandtheheadharnessstrapsoftheface
mask.
● Forstorage,placethefacemaskinaprotectivebag(contactDrägerforsupplyofa
suitablebag).
● Routerubberhosesinsuchawaythatthebendradiusisnottooacuteandthehose
isnotstretched,compressedortwisted.
● Fixtheapparatussecurelytoanyraisedmountingpointtopreventitfromfalling.
7 NIOSH Cautions and limitations
D–Air-line respirators can be used only when the respirators are supplied with
respirableairmeetingtherequirementsofCGAG-7.1GradeDorhigherquality.
E–UseonlythepressurerangesandhoselengthsspeciedintheUser’sInstructions.
I– Contains electrical parts that may cause an ignition in ammable or explosive
atmospheres.
J– Failuretoproperlyuseandmaintainthisproductcouldresultininjuryordeath.
M–Allapprovedrespiratorsshallbeselected,tted,used,andmaintainedinaccordance
withMSHA,OSHA,andotherapplicableregulations.
N–Neversubstitute,modify,add,oromitparts.Useonlyexactreplacementpartsinthe
congurationasspeciedbythemanufacturer.
O–RefertoUser’sInstructionsand/ormaintenancemanualsforinformationonuseand
maintenanceoftheserespirators.
S–Specialorcriticaluser’sinstructionsand/orspecicuselimitationsapply.Referto
User’sInstructionsbeforedonning.
7.1 S – Special or critical user’s instructions
● Approvedforuseattemperaturesabove-25°F(-31.7°C).
● During supplied airline respirator (SAR) use, the PAS Colt cylinder valve must
remain closed. If the independent air supply fails, immediately open the cylinder
valve,disconnecttheindependentairsupply,andproceedtofreshair.
● Theindependentairsupplymustmeetthefollowingcriteria:
 ○ Airlinepressureandowrequirements:pressure87to125psi;airowrateat
least550liters/minute.
 ○ Airline hose requirements: 5 feet to 300 feet of Dräger approved hose.
(Important note: The airline hose must not be made up of more than 12
individualhoses.)
8 Disposal
Whenrequired,disposeofthePASColtinaccordancewithnationalorlocalregulations
forwastedisposal.
9 Technical data
● Highpressureconnection:upto3000psi,connectiontoCGA346.
● Lungdemandregulatortofacemaskconnection:push-inconnector.
● Compressed-air cylinder: cylinders are available in aluminium or composite
materials(contactDräger forfulldetails).CylindersfromDrägeraresuppliedfully
charged.
● Independentairsupply:
 ○ Airlinepressureandowrequirements:pressure87to125psi;airowrateat
least550liters/minute.
 ○ Airline hose requirements: 5 feet to 300 feet of Dräger approved hose.
(Important note: The airline hose must not be made up of more than 12
individualhoses.)
● Low-pressurewhistleactivationpressure(iftted)–commencesintherange:72.5
to58psi;ceasesintherange:25to0psi.
10 Contact details
Any issues with the equipment, including damage, malfunction, or failure of the
breathing apparatus that may present a hazard to Dräger US Customer Service –
Phone1-800-858-1737.
Contactwiththecerticationorganizationmaybereachedat:
NIOSH,NPPTL–Phone1-412-386-4000
DraegerSafetyUKLimited
UllswaterClose Tel +441670352891
Blyth,NE244RG Fax +441670356266
UnitedKingdom www.draeger.com

Other Dräger Respiratory Product manuals

Dräger X-plore 6000 User manual

Dräger

Dräger X-plore 6000 User manual

Dräger Panorama Nova S User manual

Dräger

Dräger Panorama Nova S User manual

Dräger Panorama Nova RA User manual

Dräger

Dräger Panorama Nova RA User manual

Dräger Panorama Nova S - ESA User manual

Dräger

Dräger Panorama Nova S - ESA User manual

Dräger PARAT C box User manual

Dräger

Dräger PARAT C box User manual

Dräger NIV Operating instructions

Dräger

Dräger NIV Operating instructions

Dräger X-plore 8000 User manual

Dräger

Dräger X-plore 8000 User manual

Dräger PAS Colt Series User manual

Dräger

Dräger PAS Colt Series User manual

Dräger FPS 7000 User manual

Dräger

Dräger FPS 7000 User manual

Dräger Oxy 3000 MK II User manual

Dräger

Dräger Oxy 3000 MK II User manual

Dräger FPS 7000 User manual

Dräger

Dräger FPS 7000 User manual

Dräger X-plore 6000 User manual

Dräger

Dräger X-plore 6000 User manual

Dräger X-plore 6300 User manual

Dräger

Dräger X-plore 6300 User manual

Dräger PSS 7000 Series User manual

Dräger

Dräger PSS 7000 Series User manual

Dräger X-plore 9300 User manual

Dräger

Dräger X-plore 9300 User manual

Dräger PSS AirBoss Agile User manual

Dräger

Dräger PSS AirBoss Agile User manual

Dräger PAS Colt Series User manual

Dräger

Dräger PAS Colt Series User manual

Dräger Panorama Nova Dive User manual

Dräger

Dräger Panorama Nova Dive User manual

Dräger CPS 7800 User manual

Dräger

Dräger CPS 7800 User manual

Dräger FPS-COM 7000 User manual

Dräger

Dräger FPS-COM 7000 User manual

Dräger X-plore 4000 User manual

Dräger

Dräger X-plore 4000 User manual

Dräger Oxy 3000 MK III User manual

Dräger

Dräger Oxy 3000 MK III User manual

Dräger PARAT 3100 User manual

Dräger

Dräger PARAT 3100 User manual

Dräger Panorama Nova ESA User manual

Dräger

Dräger Panorama Nova ESA User manual

Popular Respiratory Product manuals by other brands

AOSafety Respirator Fit Testing Kits and Accessories Specifications

AOSafety

AOSafety Respirator Fit Testing Kits and Accessories Specifications

3M Versaflo TR-603E-ASB manual

3M

3M Versaflo TR-603E-ASB manual

Oxyline X 310 FFP3 NR Instructions for use

Oxyline

Oxyline X 310 FFP3 NR Instructions for use

Spasciani FUGE TAIL Instructions for use

Spasciani

Spasciani FUGE TAIL Instructions for use

Zelinsky Group Unix 5000 Instructions for use

Zelinsky Group

Zelinsky Group Unix 5000 Instructions for use

Honeywell AIRVISOR 2 instructions

Honeywell

Honeywell AIRVISOR 2 instructions

3M Versaflo TR-800 quick start guide

3M

3M Versaflo TR-800 quick start guide

Arkopharma OLFAE MICRO-NEBULIZER Instructions for use

Arkopharma

Arkopharma OLFAE MICRO-NEBULIZER Instructions for use

B.Well WN-116 U instruction manual

B.Well

B.Well WN-116 U instruction manual

Spasciani H 900 ABEK 15 Instructions for use

Spasciani

Spasciani H 900 ABEK 15 Instructions for use

3M Adflo PAPR manual

3M

3M Adflo PAPR manual

Invacare Stratos Pro instruction manual

Invacare

Invacare Stratos Pro instruction manual

Lowenstein Medical JOYCE Instructions for use

Lowenstein Medical

Lowenstein Medical JOYCE Instructions for use

ekastu SFERA instruction manual

ekastu

ekastu SFERA instruction manual

MSA AUER SSR 30/100 TR manual

MSA AUER

MSA AUER SSR 30/100 TR manual

3A HEALTH CARE HAPPYNEB II manual

3A HEALTH CARE

3A HEALTH CARE HAPPYNEB II manual

ESAB Savage A40 PAPR instruction manual

ESAB

ESAB Savage A40 PAPR instruction manual

ResMed Quattro FX NV user guide

ResMed

ResMed Quattro FX NV user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.