
MR95 Series - AD2000 Compliant
10
D-A-CH Region Only
Maintenance
▲ WARNING
To avoid personal injury, property 
damage or equipment damage caused 
by sudden release of pressure or 
explosion of accumulated gas, do 
not attempt any maintenance or 
disassembly without rst isolating the 
regulator from system pressure and 
relieving all internal pressure from 
the regulator.
To avoid possible personal injury from 
spring or pressure-loaded actuator, 
make certain the adjusting screw is 
completely backed o and the spring 
case pressure is vented prior to 
disassembly. Otherwise, the spring load 
or loading pressure could forcefully 
eject the spring case.
Regulators that have been disassembled 
for repair must be tested for proper 
operation before being returned to 
service. Only parts manufactured by 
Emerson should be used for repairing 
Fisher™ regulators.
Due to normal wear or damage that 
may occur from external sources, this 
regulator should be inspected and 
maintained periodically. The frequency 
of inspection and replacement of parts 
depends upon the severity of service 
conditions or the requirement of local, 
state and federal rules and regulations.
Due to normal wear that may occur, inspect the 
parts periodically and replace if necessary. The 
frequency of inspection depends on the severity of 
service conditions or the requirements of state and 
federal laws.
Replace parts such as the O-rings, gaskets, 
diaphragm and packing as necessary. Always 
apply lubricants as the regulator is being 
reassembled. Suitable lubricants are shown in the 
assembly drawings.
The regulator does not have to be taken out of the 
pipeline to be disassembled. All key numbers are 
referenced in Figures 4 to 7.
Replacement or Maintenance of Orice 
and Valve Plug
If it appears that the valve does not shut o tightly, the 
orice and valve plug could be worn out or damaged. 
Proceed as follows to check and/or replace them.
1.  Shut down the regulator. Refer to Shutdown 
section for the proper procedure.
2.  Unscrew the valve plug guide (key 5) from the 
body (key 1). The valve plug spring (key 26) and 
the valve plug (key 4) will normally come out of the 
body along with the valve plug guide. On DN 40 or 
50 / 1-1/2 or 2 in. body size regulators, the stem 
(key 6) will also come out of the regulator body.
3.  Inspect the seating surface of the valve plug 
(key 4), make sure that the elastomer, PTFE or 
polished metal surface of the valve plug is not 
damaged. Also inspect the external O-ring (key 4c) 
on the valve plug of Type MR95HP. Replace the 
valve plug assembly if damage is noted.
4.  Inspect the seating edge of the orice (key 3). 
If damage is noted, unscrew the orice from 
the body (key 1) and replace it with a new part. 
Reference Table 8 for proper torque values.
5.  Re-assemble the regulator in the reverse order of 
the above steps. When installing the valve plug 
guide (key 5), coat the threads and sealing surface 
with sealant to ensure an adequate metal-to-metal 
seal. Reference Table 8 for proper torque values.
Flange Cap Screw Torque Inspection
Retorquing of spring case cap screws may be 
necessary for some MR95 Series regulators after a 
period of use. Retorque the cap screws as follows:
1.  Shut down the regulator. Refer to Shutdown 
section for the proper procedure.
2.  Retighten the cap screws (key 16) in a crisscross 
pattern. See Table 8 for proper torque values.
3.  Follow the Startup section to re-pressurize 
the regulator.
4.  Refer to the Replacement of Diaphragm section 
as needed.