TOM-2 Operators Manual
Revision D, Copyright ©2022, EMS Healthcare LLC, all rights reserved
11
ii. The initial Open Time is set at 10 seconds. This setting can be modified by touching
the value window (a pop-up window will appear to adjust the value; once a new
value is input, press ENTER).
iii. Initial Open Setpoint is 20.0% O2 concentration and can be adjusted by operator.
This setting can be modified by touching the value window (a pop-up window will
appear to adjust the value; once a new value is input, press ENTER).
iv. Initial Close Setpoint is 19.8% O2 concentration and can be adjusted by the
operator. This setting can be modified by touching the value window (a pop-up
window will appear to adjust the value; once a new value is input, press ENTER).
v. Mitigation operation - when the Open Setpoint is exceeded (O2% value is higher
than setpoint), System Mode on the MAIN screen will indicate MITIGATING in
blinking green; otherwise, it displays MONITORING in solid green.
1. Solenoid valve output is activated for a time period specified by the Open
Time setting.
2. Once the Open Time expires and the O2% is confirmed to be below the
Close Setpoint, the valve is closed.
3. If O2% level remains above the Open Setpoint after the Open Time expires,
the valve will remain open until the O2% level drops below the Close
Setpoint.
4. If vacuum pressure level falls below the minimum vacuum level setting
when O2 mitigation is active, the solenoid valve output will be immediately
deactivated and remain deactivated until vacuum level rises above the
minimum limit (plus 0.5 in Hg for valve on/off hysteresis).
c. Vacuum Override setting (Mitigation mode only)
i. Vacuum Override is used internally by the O2 mitigation algorithm in order to
prioritize vacuum system pressure over O2 mitigation. If the vacuum system
pressure falls below this minimum pressure limit, all O2 mitigation activity will be
disabled until vacuum pressure returns to a value above the minimum limit. There
is no alarm output from the device for this indication because the vacuum system
controller already has such an alarm.
ii. Initial value for the alarm is 15.0 in Hg and can be adjusted by the operator. This
setting can be modified by touching the value window (a pop-up window will
appear to adjust the value; once a new value is input, press ENTER).
iii. When the vacuum override setting is exceeded (pressure falls BELOW the limit),
System Mode on the MAIN screen will display MONITORING; otherwise, it
displays MITIGATING in blinking green when mitigation is active and vacuum
pressure is within limits.
2. Monitor/Mitigate Mode - The system operates in monitor only or mitigation modes. The Settings
screen shows Monitor Only Mode or Mitigate Mode depending on the setting.
a. Press MONITOR ONLY MODE to change setting to MITIGATE MODE; likewise, press
Mitigate Mode to change back to Monitor Only Mode. The selected setting will remain
active even after a power reset.
b. In MONITOR ONLY MODE, the system continuously monitors oxygen levels and
provides O2 alarm indications. In this mode, only the O2 sensor input is required.