FKG Rooter Universal User manual

Ver 1.0
Revision Date:20200515
www.g.ch
The Black Church, St. Mary's Place, Dublin 7, D07 P4AX, Ireland
Wellkang Ltd

The trademarks menoned in this manual are the property of their legally
registered companies.
The file manufacturers, file system names and the file names referred to in this
manual are for idenficaon purposes only and are the property of their
respecve manufacturer or brands.
Introducon
Noce
device.
device and pay close aenon to warnings and notes.

Components and Accessories
Handpiece and Switch
Fig. A
Fig. B
Lighng Device
16:1 Contra-angle
(CX265-28) (CX265-67)
(CX265-63) (CX265-64) (CX265-52) (CX265-17)
(C4-2M) (CX265-23F)
44
Protecve Sleeve
(CX265-65)

Contra Angle & File connecon
Fig. C
Accessory Connecon
Fig. D
1
2
Install
2
1
Remove

Apex Locator Mode
Fig. E
Mul-funcon Mode
Fig. F

Charging
Fig. G
Replace the baery
Fig. H

instrument can
be
the instrument. This may result in incorrect operaon of the instrument.
This instrument has been thoroughly tested and inspected to assure proper perform-
ance and operaon!
This instrument should not be used adjacent to or stacked with other instrument and
this instrument should be observed to

40

device
device in accordance with the manufacturer ’s
device
device
device
device
device.
device damage.
Thermo-Disinfector Autoclave
DC current (connuous
current)
Authorized representa-
ve in the European
community

device
device allowed.
device
device for treatment.
This device might be adversely affected by the electromagnec
radiaon produced by electrical scalpels, illuminaon devices etc.
that are being used nearby.
150 to 1000rpm
min.0.6 Ncm, max.3.5 Ncm

instrument
device damage, or
the device.
device
device

instrument
The Rooter® Universal is an electro-medical device intended to drive
mechanical instruments intended for dental root canal treatment (files).
In addion, it is intended to help to determine the working length (apex
locator funconality).
Keep the accessories in a dry, dust-free environment aer sterilizaon.
7.10 7.10 Storage
Store the sterilizing equipment in a dry, clean and dust-free environment at a
suitable temperature of 5 ° C to 40 ° C.

device has 3 modes:
Do not expose the device to direct sunlight for an extended period
of me.
If the device has not been used for some me, make sure it works
properly before using it again.
Refer to the contra-angle operaon manual for all operaons
regarding the contra-angle.
7.6 Drying
Manual drying: Removing any liquid residue with a lint-free coon cloth, and
then blow dry with compressed air (1-2 Bar).
Automac drying: Refer to chapter 7.5 (f)
7.7 Inspecon and Maintenance
Aer cleaning and disinfecon, visually inspect the Lip Hook, File Clip and Contra
Angle Holder. If no visible contaminants are found, it means that the Lip Hook,
File Clip and Contra Angle Holder have been cleaned.
If damage is found by visual inspecon, stop using it and purchase damaged
parts from the manufacturer or dealer.
7.8 Package
Immediately aer drying, put the Lip Hook, File Clip, Contra Angle Holder/
Lighng device and Contra Angle into the steam sterilizaon bag for sealed
packaging.
Steam sterilizaon bag should comply with ISO 11607-1 and must be sealed
with a sealing machine.
7.9 Sterilizaon
Use a high pressure steam sterilizer in accordance with ISO 17665-1 for
sterilizaon.
a) Sterilizaon parts: The Lip Hook, File Clip, Contra Angle Holder/Lighng device
and Contra Angle
b) Sterilizaon method: Autoclave
c) Sterilizaon condions: 134 °C for not less than 5 minutes

, D
c) Connect lighng device
Insert the lighng device into the motor handpiece and clip the electrode on
device,
the file.
7.5 Automac Cleaning and Disinfecon
It is recommended to use a washer-disinfector to clean and disinfect
the Lip Hook, File Clip and Contra Angle Holder.
Put the Lip Hook, File Clip and Contra Angle Holder on the tray of the washer-
disinfector, select the "surgical instrument", and start the automac cleaning
Cleaning and disinfecon procedure:
a) Pre-cleaning: Pre-clean with tap water less than 40 ° C for 4 minutes
b) Cleaning: Immerse and wash with a mul-enzyme cleaner for 6 minutes
at 55 °C
c) Rinse stage I: Flush with tap water less than 40 ° C for 1 minute
d) Rinse stage II: Flush with tap water less than 40 ° C for 1 minute
e) Disinfecon: The temperature is 80 °C and the acon me is 10 min
f) Drying: The temperature is 100 °C and the acon me is 15 min
The user must follow the special instrucons of the manufacturer of the
fully automac washing machine. In order to ensure the cleaning and
disinfecon effect, the cleaning and disinfecon me should not be less
than the me recommended by the manufacturer.
We recommend the use of proven LIRCON® mul-enzyme cleaning soluon
or mul-enzyme cleaning soluon that complies with local regulaons (e.g.
CE, FDA approval).
Please use a washer-disinfector that meets the requirements of ISO 15883.
Considering that some countries have different requirements for A0 values,
please refer to ISO 15883 for temperature and me of disinfecon.

• When installing/removing the lighng device, do not shake them at will
to avoid damaging the plug.
7.4 Manual Disinfecon
b) Aer cleaning, remove any liquid residue with a lint-free coon cloth and
then dry with compressed air (1-2 Bar).
Do not put the Lip Hook, File Clip, and Contra Angle Holder into the
disinfectant container for immersion and disinfecon. If the internal liquid is
not cleaned and the drying is not complete, the internal parts will be
corroded.
Place the lint-free coon cloth in a container filled with LIRCON® medical
device cleaning soluon (metal type), wring it out and wipe the Test Wire,
Lip Hoop, File Clip, Contra Angle Holder/Lighng device and Contra Angle at
least 3 mes.
It is recommended to use the medical device cleaning soluon (metal type) of
LIRCON®, which with an effecve chlorine content of 1.0%-1.2% (W/V).
Do not put the Test Wire, Lip Hoop, File Clip, Contra Angle Holder/Lighng
device and Contra Angle into the disinfectant container for immersion and
disinfecon. If the internal liquid is not cleaned and the drying is not
complete, the internal parts will be corroded.
If you want to use other disinfectant, use a disinfectant that complies with
local naonal regulaons (such as CE cerficaon, FDA cerficaon), and
follow the instrucons provided by the disinfectant manufacturer.
7.3 Manual Cleaning
a) Rinsing the Lip Hook, File Clip and Contra Angle Holder in flowing tap water
(<40 ° C)

FKG
XP-endo Treatment
Shaper
Finisher
1000 rpm
1.0 Ncm
1000 rpm
1.0 Ncm
Cleaning, disinfecon, and sterilizaon have limited impact on the reusable
part of the motor handpiece. Therefore, the number of mes the procedure
is repeated is determined by the degree of wear of the part. If visual
inspecon reveals damaged parts, stop using them and purchase new parts
from the manufacturer or dealer.
7.1 Preparaon aer Using
Immediately aer use, the reusable parts should be immersed in tap water
<40 ° C (The quality of drinking water, the 'water' menoned in this chapter, is
required to meet this standard.) to remove dirt. Do not use a fixed detergent
or warm water (>40 °C), as this will cause the residue to be fixed and affect the
post-treatment effect.
Transportaon: Transport to the post-processing area for safe storage to avoid
any damage and environmental polluon.
7.2 Preparaon before Cleaning
a) Disassemble the Lip Hook, File Clip and Contra Angle Holder and place them
in a stainless steel box.
b) Decontaminaon preparaon (Pre-cleaning)
Flush the Lip Hook, File Clip and Contra Angle Holder with running tap water
<40 ° C unl all visible residue is removed.
Refer to the contra-angle operaon manual for all operaons
regarding the contra-angle. Some preset files have automac parameters.
If such a file has been selected, the speed and torque value will display
“auto”.
auto

FK G preset
User
FKG
XP-endo Treatment
Shaper
Finisher
FKG
XP-endo Treatment
Shaper
Finisher
File systems libraries
The device contains libraries of file systems with preset parameters.
file systems library

The lighng device will connue to illuminate while the motor handpiece
Only in User file systems, press to change the motor direcon of rotaon.
handpiece
is running.
device
Change Motor direcon of rotaon
device
that the device is not operang properly. Please
handpiece,

reduced, the motor handpiece returns to normal forward rotaon automacally.
torque limit value torque limit value
Load beyond the
the motor is automacally
When the load is reduced,
restored
Rooter Universal
Software version XX X.X
again to enter seng interface.
to enter seng mode
c)

In User Recipro file system, torque cannot be changed.
In all FKG file systems, rotaon direcon, speed, torque and rotaon
angle cannot be changed. (fixed sengs)
1000 rpm
1.5 Ncm
1000 rpm
1.5 Ncm
Fig. H
motor handpiece.
Other manuals for Rooter Universal
1
Table of contents
Other FKG Dental Equipment manuals
Popular Dental Equipment manuals by other brands

Gendex
Gendex ORTHORALIX 8500 Operator's manual

Aseptico
Aseptico ADC-02CF Setup guide

Zhermack
Zhermack ZETACLAVE B quick guide

Zimmer Biomet
Zimmer Biomet Zfx GenTek instruction manual

Dentsply Sirona
Dentsply Sirona SINIUS installation instructions

B.A. International
B.A. International Powersurge Instructions for use