8
OUTPUT CAPACITY
Whensetto2.1bartodrivetheFlowEZTM2000mbarchannels(or
2.3bartodriveMFCS2000mbarchannels),theFLPGPluscan
driveupto4channelsatthesametime.
Whensetto1.1bartodrivetheFlowEZTM1000mbarchannels(or
1.3bartodriveMFCS1000mbarchannels),theFLPGPluscan
driveupto8channelsatthesametime.
Whensettolowerpressurestodrivelowerrangechannels(see
tablesonpage6formoreinformation),theFLPGPluscandriveup
to8channelsatthesametime.
The FLPGPlus'outputcapacityisthenumberofchannelsitcandrive
simultaneously.Thiscapacitydependsontheoutputpressureset
withthepressuresettingregulator.
Do not connect more channelsthan thenumbers shown
above. Otherwise, the output pressure may drop and
become unstable, leading to possible damage of the
FLPG Plus and the connected pressure controller.