
20 SETUP/OPERATION |PIPETMAX® 268
REMOVABLE TRAY
9-POSITION REMOVABLE TRAY FOR 384-WELL MICROPLATES
When placing labware and tip racks on the 9-position removable tray for 384-well
microplates, refer to the following information:
●Tip racks can be placed in the back row middle and right positions, and the front row
middle position ONLY. These positions have installed clips to hold the tip racks in place.
It is recommended to use a full rack of tips because an error may be encountered if using
two partial racks of tips containing the same type of tip.
●The tip waste chute or tip disposal bin must be placed in the back row, left position ONLY.
●All racks and plates can be placed in any position on the tray.
Exceptions:
●A maximum of two thermal (PT or CT) racks total can be placed on this tray as these
are considered heavy bed elements.
○Refer to the instructions provided with the racks to modify the tray to allow for rack
placement.
●Circulating Temperature (CT) racks can be placed in the front row, left or right
position s O N LY.
○Refer to the instructions provided with the rack to modify the
tray to allow for rack placement.
●The Orbital Shaker cannot be placed on this tray.
INSTALLATION
While holding the handles on the sides of the tray, lower the removable
tray onto the fixed tray. It is keyed and will only mount one way. Grip
the removable tray and then gently rock it back and forth. If it seems
that the removable tray is shifting position on the fixed tray, use a 3 mm
Allen wrench and gentle pressure to adjust the spacers on either side
of the fixed tray (see Figure 7 and Figure 8). Spacers should just barely
contact the removable tray. Remove and then replace the removable
tray to ensure there is not any significant interference.
Figure 6
Tray Installation
Figure 7
Tray Installation
Figure 8
Tray Installation