Gima 34301 User manual

LARINGOSCOPI CONVENZIONALI
CONVENTIONAL LARYNGOSCOPES
LARYNGOSCOPES CONVENTIONNELS
KONVENTIONELLE LARYNGOSKOPE
LARINGOSCOPIOS CONVENCIONALES
LARINGOSCÓPIOS TRADICIONAIS
LARYNGOSKOPY KONWENCJONALNE
LARINGOSCOAPE CONVENȚIONALE
ΣYMBATIKA ΛΑΡΥΓΓΟΣΚΟΠΙA
M34303-M-Rev.4-03.20
PROFESSIONAL MEDICAL PRODUCTS
34301 - 34302 - 34303 - 34305 - 34306 - 34307
34310 - 34311 - 34312 - 34313 - 34314 - 34317
34318 - 34319 - 34320 - 34321 - 34352 - 34354
Gima S.p.A.
Via Marconi, 1 - 20060 Gessate (MI) Italy
gima@gimaitaly.com - export@gimaitaly.com
www.gimaitaly.com
Made in Pakistan
È necessario segnalare qualsiasi incidente grave vericatosi in relazione al dispositivo medico da noi fornito
al fabbricante e all’autorità competente dello Stato membro in cui si ha sede.
All serious accidents concerning the medical device supplied by us must be reported to the manufacturer and
competent authority of the member state where your registered ofce is located.
Il est nécessaire de signaler tout accident grave survenu et lié au dispositif médical que nous avons livré au
fabricant et à l’autorité compétente de l’état membre où on a le siège social.
Jeder schwere Unfall im Zusammenhang mit dem von uns gelieferten medizinischen Gerät muss unbedingt
dem Hersteller und der zuständigen Behörde des Mitgliedsstaats, in dem das Gerät verwendet wird, gemeldet
werden.
Es necesario informar al fabricante y a la autoridad competente del Estado miembro en el que se encuentra
la sede sobre cualquier incidente grave que haya ocurrido en relación con el producto sanitario que le hemos
suministrado.
É necessário noticar ao fabricante e às autoridades competentes do Estado-membro onde ele está sediado
qualquer acidente grave vericado em relação ao dispositivo médico fornecido por nós.
Należy poinformować producenta i kompetentne władze danego Kraju członkowskiego o każdym poważnym
wypadku związanym z wyrobem medycznym naszej produkcji.
Orice accident grav produs, privitor la dispozitivul medical fabricat de rma noastră, trebuie semnalat pro-
ducătorului și autorității competente în statul membru pe teritoriul căruia își are sediul utilizatorul.
Σε περίπτωση που διαπιστώσετε οποιοδήποτε σοβαρό περιστατικό σε σχέση με την ιατρική συσκευή που σας
παρέχουμε θα πρέπει να το αναφέρετε στον κατασκευαστή και στην αρμόδια αρχή του κράτους μέλους στο
οποίο βρίσκεστε.

2

6
ENGLISH
Thank you for purchasing our product. This product meets the most stringent requirements regarding the selection
of manufacturing materials and also the nal control. These instructions should be followed to ensure durability of
this product.
The operator must carefully read and understand this manual thoroughly to keep the product performance durable
and reliable for longer period.
After opening the packages, rst of all it is necessary to check all the components against the standard conguration.
Check that they are all present and in perfect conditions.
FEATURES
The laryngoscope blades are manufactured from antimagnetic 18/8- quality stainless steel type AISI 303/304 which
is highly resistant to corrosion and conform to the ISO 7376 standard.
GIMA blades are maintenance free and autoclavable up to 134°C / 5 min approximately for 2,000 times.
USING THE BLADES
Do not grip the lever when removing the blade. Do not apply any pressure, in the directions shown in the illustration
Fig 1, which could force the blade and lever apart.
Damage to the linkage may occur, resulting in incorrect action, or stiffness of the adjustable tip. The blade is as-
sembled to the handle in the normal manner. The lever to operate the tip section will then extend behind the handle.
The lever should not be touched during the initial stage of use, until the tip of the laryngoscope has been inserted
into the vallecula.
Once this stage has been reached, movement of the lever towards the handle will elevate the tip of the blade, and
therefore lift the epiglottis, without the need to increase the force exerted by the main part of the blade. Release the
lever before the blade.
OPERATING INSTRUCTION
1. Engage the blade by aligning the slot of the blade on to the hook pin of the handle and apply a sufcient force
1N-5N to make engagement as show in Fig 2.
2. Apply force upward to bring the blade in operating position as in Fig 3.
3. To bring the blade in stand by position apply force downward as in Fig 3.
LAMP REPLACEMENT PROCEDURE
1. Unscrew the lamp in counterclock wise direction until free as shown in Fig 4.
2. Replace the lamp and verify that the lamp is sufciently tightened before use.
3. Be sure that the lamp is properly screwed in.
BATTERY REPLACEMENT PROCEDURE
1. Unscrew bottom cap of handle and remove batteries.
2. For greater longevity, alkaline batteries are recommended as replacement. Ordinary carbonic batteries may also
be used.
3. Replace with appropriate size batteries making sure the + and - terminals are placed correctly.
4. In handle insert one positive end down, the other positive end up and reinstall the end cap. If handle light guide
fails to light batteries may need to be reversed for correct polarity.
2.5V battery handles can be used with dry batteries as well as 2.5V rechargeable batteries.
CARE & MAINTENANCE
Remove batteries before cleaning, high level disinfecting or sterilization of the laryngoscope system.
BLADES
1. Cleaning Procedure
Immediately after use, the laryngoscope system should be rinsed under cool running tap water until all visible soil is
removed. Ensure that all hard-to-reach areas are ushed with the running tap water. Immerse sealed laryngoscope
system in a presoak enzymatic cleaner solution, which was prepared in accordance to manufacturer’s recommenda-
tions for a minimum of two minutes. Remove device from enzymatic cleaner solution and rinse with lukewarm running
tap water for a minimum of one minute to remove all residues and visible soils. Then, immerse device in enzymatic
detergent. Remove bottom cap and brush thoroughly using a soft-bristle brush, while ensuring that all hard to reach are
reached visible soils / residue removed. Dry with lint free, clean cloth or ltered pressurized air. Follow with HIGH-LEV-
EL DISINFECTION or STEAM STERLIZATION PROCEDURE.

7ENGLISH
Ultrasonic cleaning is strictly prohibited.
2. Disinfection
Soaking in solutions or Thermo chemically in a washer Sterilizer up to 65°C maximum may perform disinfection.
Manufacturers instruction regarding duration and concentration of solutions should be strictly adhered. After Disin-
fection, rinse throughly in sterile water and dry with a lint free clean cloth.
3. Cold Soak Solution
To achieve a high-level disinfection, Cidex®OPA or 2.4% Glutaraldehyde solution may be used according to manu-
facturer’s instructions. Dry with lint free, clean cloth or ltered pressurized air. Reassemble all parts, load handle with
batteries and test the system for proper function.
If not functional, review battery / lamp testing instructions below.
Do not immerse Blades in Bleach, Betadine or Potassium Hydroxide solutions.
Doing so will several damage instruments also avoid metal to metal contact after soaking, the blades should
be rinsed under sterile water to remove chemical residues and dry with lint free clean cloth or ltered pres-
surized air.
4. Sterilization
Before performing any of the procedures described below, the blade should be cleaned as described in the cleaning
Procedure.
5. Gas Sterilization
Gas sterilization by Ethylene oxide up to a maximum temperature of 65°C and 8 p.s.i, may be performed, which is
preferred especially if sterilization is to be performed regularly.
6. Steam Sterilization
Steam Sterilization can also be performed. Insert device in appropriate autoclave pouch.
(A) GRAVITY DISPLACEMENT STEAM (B) PRE-VACUUM STEAM
Temperature 121°C (250°F) 134°C (273°F)
Cycle Time 30 Minutes 5 Minutes
Dry Time 15 Minutes 20 Minutes
Note: Do not exceed temperature of 135°C and pressure of 28 p.s.i
Flash autoclaving and hot air sterilization should be avoided as these processes will damage the instrument.
Steris Amsco V-Pro
Laryngoscope Blades and Handles are compatible with:
Amsco V-Pro 1 Low Temperature Sterilization System
Amsco V-Pro 1 Plus Low Temperature Sterilization System
Amsco V-Pro 1 Pro Max Temperature Sterilization System
Sterrad
Larynogoscope Blades and Handles are compatible with:
Sterrad 100nx System (Standard and Express Cycle)
Sterrad nx System (Standard Cycle)
Sterrad 1005 and 200 System (Short Cycle Outside US)
Sterrad 50 System
HANDLES
Cleaning/Sterilization:
Battery Handle withstand the same cold soak solution and autoclave ranges outlined in the blade section. However,
the batteries and lamp must be removed prior to disinfection/sterilization.
Battery Handles can withstand exposure to the ethlene oxide. Lamp may be cleaned with cotton ball dampened in
alcohol (IPA). The main handle and a cap will also withstand and exposure to ethylene oxide.

8
ENGLISH
Do not allow excess uid to seep into electrical contact, batteries must be removed before cleaning & steri-
lization.
BLADE & HANDLE TEST PROCEDURE
Laryngoscope blades and handle should always be tested after cleaning/ disinfection/ sterilization and prior to use.
To test connect the laryngoscope blade to handle and pull it to the ON position as in Fig 2 if the unit fails to light or
icker, check the lamp/ batteries and the electrical contacts, Be sure adequate supplies of spare lamps, batteries and
replacement parts are readily available if problem still persists, contact supplier please.
Warning
The above listed sterilization Guidelines, provided by GIMA are intended as procedures compatible with
specic materials. Sterilization must be performed to approved Hospital protocol. GIMA can not guarantee
sterility. This will be validated by the hospital and or sterilization equipment manufacturers.
Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of
this equipment by bringing it to a specic recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
Caution: read instructions (warnings) carefully WEEE disposal
Keep in a cool, dry place Keep away from sunlight
Manufacturer Date of manufacture
Product code Lot number
Medical Device compliant with Regulation
(EU) 2017/745 Consult instructions for use
Follow instructions for use Medical device
Type BF applied part
Other manuals for 34301
2
This manual suits for next models
15
Table of contents
Other Gima Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Getinge
Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics
Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care
Pressalit Care R1100 Mounting instruction

Denas MS
Denas MS DENAS-T operating manual

bort medical
bort medical ActiveColor quick guide

AccuVein
AccuVein AV400 user manual