
Edition 3-4.03 –July 2017 Personal bioreactor RTS-1 & RTS-1C
Page 12 Operating instructions
10.Specification
The unit is designed for operation at ambient temperature from +4°C to +40°C in a non-condensing temperature and maximum
relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C.
Grant is committed to a continuous program of improvement and reserves the right to alter design and specifications of the
equipment without additional notice.
10.1. Measurement specifications
Light source.............................................................................................................................................................................LED
Wavelength (λ), nm ..........................................................................................................................................................850 ± 15
E.coli BL21 Factory calibration measurement range, in ODλ850 nm, OD
at 10–20 ml volume.................................................................................................................... 0-10 (0–19 ODλ600 nm equivalent)
at 20–30 ml volume................................................................................................................... 0-8 (0–15.2 ODλ600 nm equivalent)
Measurement Precision, OD...................................................................................................................................................±0.3
Real time measurement, measurement/h .............................................................................................................................1 - 60
Time setting resolution, min .........................................................................................................................................................1
10.2. Temperature specifications
Setting range, °C
RTS-1..............................................................................................................................................................................+25...+70
RTS-1C............................................................................................................................................................................+4…+70
Bottom control range point, °C
RTS-1.................................................................................................................................................................. 5 above ambient
RTS-1C..............................................................................................................................................................15 below ambient
Top control range point, °C ........................................................................................................................................................70
Setting resolution, °C ................................................................................................................................................................0.1
Stability, °C.............................................................................................................................................................................± 0.1
Sample temperature accuracy, °C
20°C - 45°C...............................................................................................................................................................................± 1
< 20 °C ......................................................................................................................................................................................± 2
> 45 °C ......................................................................................................................................................................................± 3
Sample temperature heating rate, °C/min ....................................................................................................................................1
Sample temperature cooling rate, °C/min ....................................................................................................................................1
10.3. General specifications
Sample volume, ml..............................................................................................................................................................10 - 30
Speed range, rpm .............................................................................................................................................................50-2000
Speed setting resolution, rpm.....................................................................................................................................................10
Speed control precision*, rpm.................................................................................................................................................± 15
Display ....................................................................................................................................................................................LCD
Overall dimensions (W × D × H), mm.......................................................................................................................130×212×200
Weight*, kg
RTS-1........................................................................................................................................................................................1.7
RTS-1C.....................................................................................................................................................................................2.2
Input voltage.....................................................................................................................................................................12 V DC
Input current / power consumption
RTS-1.........................................................................................................................................................................3.3 A / 40 W
RTS-1C.........................................................................................................................................................................5 A / 60 W
External power supply........................................................................................................In AC 100-240V 50/60Hz, out DC 12V
* Accurate within ±10%.