manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hioki
  6. •
  7. Data Logger
  8. •
  9. Hioki LR8450 User manual

Hioki LR8450 User manual

Voltage
1 ms
sampling
Strain
1 ms
sampling
Temperature
10 ms
sampling
Humidity
10 ms
sampling
Resistance
10 ms
sampling
Wireless data logging at 1 ms
330-channel portable logger available with
your choice of plug-in units and wireless units
MEMORY HiLOGGER LR8450
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
Standard model (designed for use with plug-in units only)
LR8450
2
You can add up to 4 plug-in units and provide 120 channels
of measurement
Two models:
Standard Model and Wireless LAN Model
Each VOLTAGE/TEMP UNIT U8552 accepts 30 channels of input. Add four units for 120 channels of measurement.
Plug-in units
Conguration example: 120 channels
VOLTAGE/TEMP UNIT U8552 × 4
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
Wireless LAN model
LR8450-01
3
Wireless LAN model
Two models:
Standard Model and Wireless LAN Model
Add up to 7 wireless units in total for a maximum of
330 channels
Add channels freely via either plug-in or wireless units
Can also be used exclusively with wireless units.
Mix plug-in and wireless units
Plug-in units
Conguration example: 330 channels
WIRELESS VOLTAGE/TEMP UNIT LR8532× 7
Wireless units
VOLTAGE/TEMP UNIT U8552 × 4
+
With four U8552 VOLTAGE/TEMP Units and seven LR8532 WIRELESS
VOLTAGE/TEMP Units, you can measure a total of 330 channels.
Plug-in unit and Wireless unit in mix will allow you to build a measurement system that suits your needs.
If wireless units are used with other units (wireless or plug-in), the sampling-timing shift between the units is peri-
odically corrected.*
In addition, at times the wireless communication is cut off, the correction function works after the communication is
restored and the sampling-timing shift between the units is corrected.
* Even in good wireless communication conditions (low interference) the sampling-timing between devices may shift about 20 ms.
In bad wireless conditions, the sampling-timing shift will be much worse than this.
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
4
Battery temperature rise
Automotive part temperature
measurement
Voltage measurement
Measure outputs from a pressure sensor and
other sensors at 1 ms max. sampling rate.
1 ms sampling is very suitable to record outputs of several tens of Hertz from pressure
sensors and vibration sensors.
Road test automobiles
Since increasing the number of units has
no effect on the cutoff frequency, which
changes with the sampling rate, power
supply noise can be reduced without sac-
ricing noise resistance.
(ex.) Sampling rate
: 1 s
Number of channels
Cutoff frequency
1ch to 15ch 60 Hz
16ch to 30ch 60 Hz
31ch to 45ch 60 Hz
46ch to 60ch 60 Hz
*When using a power supply
frequency of 60 Hz.
Same cutoff
frequency
HIGH SPEED VOLTAGE UNIT U8553 WIRELESS HIGH SPEED VOLTAGE UNIT LR8533
Measure temperature near inverters and
batteries at a sampling rate of up to 10 ms
Temperature measurement
*Sampling rate of 10 ms is available when using 15 or fewer channels.
VOLTAGE/TEMP UNIT U8550
UNIVERSAL UNIT U8551
VOLTAGE/TEMP UNIT U8552 (*)
WIRELESS VOLTAGE/TEMP UNIT LR8530
WIRELESS UNIVERSAL UNIT LR8531
WIRELESS VOLTAGE/TEMP UNIT LR8532 (*)
Test hydraulic equipment
Each unit incorporates its own A/D con-
verter. This design keeps the maximum
sampling rate high even when units are
added.
Example 1: Use four U8553 High Speed
Voltage Units (with 5 channels each) to
measure 20 channels at a sampling rate
of 1 ms.
Example 2: Use four U8550 Voltage/
Temp Units (with 15 channels each) to
sample 60 channels at a sampling rate of
10 ms.
The cutoff frequency, which varies with the data
refresh interval, can be set separately for each
unit. You can use long data refresh intervals,
which boost lter effectiveness, and short data re-
fresh intervals for different units at the same time.
• Measure control signals at maximum speed: Unit 1
(data refresh interval: 1 ms)
• Measure battery voltage uctuations: Unit 2
(data refresh interval: 1 ms)
• Measure temperature using thermocouples: Unit 3
(data refresh interval: 1 s) with strong lter
Sample input at up to 1 ms
Consistent even when
units are added
Noise resistance
Consistent even when
units are added
Set lters
Set lters for each unit
Control signal:
Unit 1
Battery voltage:
Unit 2
Temperature:
Unit 3
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
5
Pipe strain measurement
Stress and load on moving parts
Aircraft wing strain measurement
Brake and pipe strain measurement
Strain measurement
Measure strain with a 1 ms sampling rate
Connect strain gages directly and measure at a sampling rate of up to 1
ms. Strain gages tend to have long, thin wires that are easily broken, but
that potential pitfall can be avoided by using wireless units so that wire
length is minimized.
Connect strain gages directly
The Strain Unit has a built-in bridge box, allowing you to connect strain gages directly to
its input terminals.
Strain gage Strain gage
Bridge box
Bridge box
STRAIN UNIT U8554 WIRELESS STRAIN UNIT LR8534
Previous models were incapable of measuring temperature accurately in noisy environments due to the inuence of high frequencies, which
caused values to shift or uctuate signicantly. The LR8450 uses a revamped design to dramatically reduce the inuence of high-frequency noise.
Example: Measure temperature by connecting the tip of a K thermocouple to the screw on an inverter’s PWM output terminal (W-phase)
when using the Voltage/Temp Unit U8550 (settings: 100 ms sampling in the 100°C f.s. range).
SiC inverter
• Carrier frequency: 200 kHz
• Motor speed: 500 r/min
Inverter MotorBattery
Reduced inuence of noise
Stable measurement, even at high voltages and high frequencies
Previous data logger LR8450
Previous models exhibit signicant uctuations when the
inverter is operating, but the LR8450 does not.
InternalExternal
LR8450 + U8554 × 2
Strain input 10 ch
Data logger with
external bridge box
Strain-gage-type converters such as load sensors and pressure sensors can be
connected directly and you can make measurement.
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
6
Install wireless units in testing equipment
Peace of mind if communications are temporarily interrupted
Buffer memory holds up to 5 min.*1of measurement data
Wireless for ease of use
Collect data from dispersed locations all at the same time
The LR8450-01 can simultaneously collect measurement
data from wireless units installed on various test equipment.
Each wireless unit has a built-in buffer memory that can hold up to 5 min.*
1
of measurement data. Data are resent along with more recent
measurement data once communications resume, after the data are restored inside the LR8450-01*
2
.
The system can be congured to output an alarm if communications are interrupted or if a unit encounters a low-battery state.
Peace of mind in the event of an interruption in power or wireless connectivity
Example:
The wireless Voltage/Temp unit LR8530 can operate for about
9 hours on battery power. If the unit is charged at night, it can
operate on the battery pack alone during the day.
Using the Battery Pack Z1007
Wireless unit model Continuous operating time
LR8530 Approx. 9 hr.
LR8531 Approx. 7 hr.
LR8532 Approx. 9 hr.
LR8533 Approx. 9 hr.
LR8534 Approx. 5 hr.
Collect measurement data
from multiple locations with a
single logger
Units can be placed in
conned locations
Check the display during
measurement
Manage data as a single
time sequence
*1 The duration for which measurement data can be maintained does not vary with the recording interval (up to a maximum of 5 min.)
*2 Data collected using the Logger Utility software measurement cannot be restored in this manner.
If you’ve installed a battery pack in a unit that’s being powered by an AC adapter, the unit will automatically switch to battery power in the event of
an outage so that the LR8450-01 can continue making measurements.
Up to 30 m*
(line-of-sight)
* If the LR8450-01 or the wireless unit is placed on the oor
or ground, the communication distance may be shortened.
Battery operation
Use units in locations where there’s no AC power
Peace of mind in the event of a power outage during measurement
Install a battery pack for peace of mind
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
7
The LR8450 can be directly connected to a PC via wireless LAN.
Access point mode
Observe data from a remote location using a PC or a tablet
You can reduce interference with other wireless devices by using an
open channel. Check for open channels on the instrument’s screen.
Simple registration of wireless units
Check the unused wireless LAN channels
and select the wireless channel to use
LR8450-01 (wireless LAN model) only
By connecting the LR8450-01 to a PC or a tablet via wireless LAN, you can control the instrument remotely using the built-in HTTP server or
obtain download data les using the built-in FTP server.
(You cannot use Logger Utility when using Station Mode or Access Point Mode.)
Make measurements in locations where it would be difcult to route wires
Inside a room, or outside, you can make measurements with the door closed.
Wireless LAN communicationsWireless LAN communications
AP
Connect wirelessly to a third-party access point (AP).
Station mode
AP
STA
Work time can be reduced using the LR8450-01 and wireless units, since only minimal wiring is required. If the measurement target is
located in a lab, this approach eliminates the need for wiring and avoids having to drill holes in the walls of the monitoring room where
data is being checked.
Wireless units, located within the
range, that are not connected
to another LR8450-01, can be
automatically detected. Simply
choose the unit you wish to
register from the list.
Registration
complete!
STA STA
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
8
Useful in preventive maintenance
8 channels alarm outputs
Motor speed, ow rate integration, etc.
8 channels pulse measurement
View waveforms and verify the numerical
values
Easy viewing of changes in waveforms
Check instantaneous, maximum, and
other values on a single screen
Check the alarm report
USB interface
(function)
USB interface
(host)
LAN interface
SD card slot
Easy-to-read display of measured values
Waveforms + numerical values
XY + numerical values
Waveforms
You can set alarm output for eight channels. You can
set a level, a window, a slope, and a logic pattern on
channels you wish to monitor.
Alarm
Input signals
Level
Time
Set the level and time.
It generates an alarm if the
reading exceeds the preset
rate of change (level/time)
Slope
External control
terminal
When you remove the storage media
while recording data, and reinsert it,
data remaining in the internal buffer
memory will continue to be stored in a
different le.
In "Revolve" mode, monitor production equipment by
measuring the variations in revolution speed of motors
or drills. In "Count" mode, identify operation status by
acquiring integrated power or ow rate.
The LR8450/LR8450-01 provides two output terminals for volt-
ages, each of which can supply 100 mA current, eliminating the
need for a separate sensor power supply. You can select 5 V,
12 V, or 24 V from the VOUTPUT1 terminal and 5 V or 12 V from
the VOUTPUT2 terminal.
Pulse signal
Use example:
Flow rate sensor
External control terminals and interfaces to accommodate a broad range of use cases
24 V
12 V
Sensor output voltage
Two terminals for voltage outputs (5, 12, or 24 V)
Supplying power to the sensors
Replace media during real-time saving
No need to stop recording
Power jack
Numerical values
Alarms
Use example:
Pressure transducer
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
9
12:00 am 12:00 am6:00 am 12:00 pm 6:00 pm
Segment time: 12 hours
Data saved in one le
Example: Recording 30 analog channel with 2 units (no alarm output or waveform processing)
Because the header portion of waveform les is not included in capacity calculations, expected actual maximum time is about 90% of those in the tables.
The maximum recording time varies with the number of measurement channels. Recording times are doubled if the number of measurement chan-
nels shown in the table is halved.
When recording 30 analog channels with two U8550/U8551 units or one U8552 unit (no alarm output, no waveform processing)
When recording 30 analog channels with two LR8530/LR8531 units or one LR8532 unit (no alarm output, no waveform processing)
Recording
intervals
Internal buffer memory
(512 MB)
SD MEMORY CARD Z4001
(2 GB)
SD MEMORY CARD Z4003
(8 GB)
USB DRIVE Z4006
(16 GB)
10 ms 1
d
3
d
20
h
15
d
8
h
30
d
12 h
100 ms 10
d
8
h
38
d
18
h
153
d
9
h
305
d
5 h
1 s 103
d
13
h
387
d
12
h
1533
d
21
h
3052
d
9 h
10 s 500
d
3875
d
6
h
15339
d
3
h
30523
d
19 h
When recording 20 channels with four U8553 units or U8554 units (no alarm output, no waveform processing)
When recording 20 channels with four U8553 units or LR8534 units (no alarm output, no waveform processing)
Recording
intervals
Internal buffer memory
(512 MB)
SD MEMORY CARD Z4001
(2 GB)
SD MEMORY CARD Z4003
(8 GB)
USB DRIVE Z4006
(16 GB)
1 ms 3
h
43 m 13 h 56 m 2
d
7 h 4
d
13 h
10 ms 1
d
13
h
5
d
19 h 23
d
45
d
18 h
100 ms 15
d
12
h
58
d
3 h 230
d
2 h 457
d
20 h
1 s 155
d
8
h
581
d
7 h 2300
d
21 h 4578
d
13 h
10 s 500
d
5813
d
1 h 23008
d
20 h 45785
d
20 h
When recording 330 channels with four U8552 units and seven LR8532 units (no alarm output, no waveform processing)
Recording
intervals
Internal buffer memory
(512 MB)
SD MEMORY CARD Z4001
(2 GB)
SD MEMORY CARD Z4003
(8 GB)
USB DRIVE Z4006
(16 GB)
20 ms 4 h 8 m 15 h 28 m 2
d
13 h 5
d
2 h
100 ms 20 h 42 m 3
d
5 h 12
d
18 h 25
d
10 h
1 s 8
d
15 h 32
d
6 h 127
d
19 h 254
d
8 h
10 s 86
d
322
d
16 h 1277
d
23 h 2543
d
9 h
Maximum recording time (estimate)
Easy-to-read display of measured values
Collect data on a storage device (SD memory card or USB drive) while measuring continues. The ability to segment les by hour or day
without stopping measurement is convenient when you need to review data later.
In addition to the maximum and minimum value calcula-
tion functions provided by previous models, the LR8450/
LR8450-01 offers an extensive range of calculations,
including on/off time, count, and usage ratio.
Extensive calculation functions installed
Calculate data while measurement con-
tinues and display calculated waveforms
in real time. Calculation results are saved
on a separate dedicated calculation
channel.
Numerical calculation function Waveform calculation function
Recording over extended periods of time without interruption
Moving-average waveform
Raw waveform
Types of calculations
Basic arithmetic
operations
Moving average
Aggregation
Simple average
Integration
Types of calculations
Off count
Average value
Maximum value
Integration
Usage ratio
On count
On time
Minimum value
Aggregation
Off time
Peak-to-peak value
Time at which mini-
mum value occurred
Time at which maxi-
mum value occurred
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
10
Control the instrument remotely and capture data on a PC
HTTP server function
Control the instrument remotely from a PC
PC can acquire data in real time
Your PC can get the les in the SD memory card or USB drive in-
serted to the LR8450/LR8450-01.
FTP server function
Download data les onto a PC
Can automatically transmit to an FTP server the les in the SD mem-
ory card or in the USB drive inserted to the LR8450/LR8450-01.
FTP client
Automatically transfer data files to
an FTP server
Can set the clock in the LR8450/LR8450-01 and synchronize it to
an NTP server on the network.
NTP client function
Set the logger’s clock
Use a standard Web browser to control the LR8450/LR8450-01, start
and stop measurement, and enter comments.
LAN
• Regularly (as quickly as once every
second) collect measurement data from
up to 15 LAN-connected measuring
instruments and display them on a PC.
• You can acquire power data from a power
meter and temperature or ow rate data
from a data logger.
Real-time measurement (logging)
This function lets you acquire the measurement le, which is produced in the
measurement instrument once per day, into a PC in real time. You can obtain daily
data, like power consumptions measured by a measurement instrument installed on
site, in to your PC automatically.
Automatic le transfer (FTP)
Remote control (HTTP)
Control instruments remotely and change settings from a LAN-connected PC.
For an up-to-date list of products supported by GENNECT One, see Hioki’s website.
PC
Download the GENNECT One SF4000 software from the Hioki
website to your PC
Connect each measuring instrument to PC with LAN cable
LAN
Power quality
Power consumption
Temperature and humidity
HUB
1
2
Collect data using GENNECT
Record data on a PC in real time using the Logger Utility applica-
tion software, a standard accessory. You can even scroll wave-
forms backwards to view older data while recording is in progress.
A real-time measurement is supported for recording intervals of
10 ms or greater.
LAN/ USB
Logger Utility
Acquire data using Logger Utility
Can send emails to your PC or mobile phone when there is a com-
munication loss and when an error occurs during measurement and
wireless module communications.
Can also send instantaneous values by e-mail periodically.
E-mail transmission function
Inform error and other information by e-mail
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
11
Specications
LR8450, LR8450-01 Memory HiLogger
General specications, Basic specications
Product warranty period
3 years
Accuracy guarantee period
1 year
Maximum number of
connectable modules
4 plug-in modules + 7 wireless modules*
* LR8450-01 only
Connectable modules
(Plug-in modules)
U8550 Voltage/Temp Unit
U8551 Universal Unit
U8552 Voltage/Temp Unit
U8553 High Speed Voltage Unit
U8554 Strain Unit
Connectable modules
(Wireless modules)
(LR8450-01 only)
LR8530 Wireless Voltage/Temp Unit
LR8531 Wireless Universal Unit
LR8532 Wireless Voltage/Temp Unit
LR8533 Wireless High Speed Voltage Unit
LR8534 Wireless Strain Unit
Internal buffer memory
Volatile memory, 256 Mwords
Clock functionality Auto-calendar, automatic leap year recognition, 24-hour clock
Clock precision
(Precision of clock dis-
played by instrument as
well as start/stop times)
±1.0 s/day (at 23°C)
Time can be synchronized with an NTP server to which instrument is
connected.
Time axis accuracy ±0.2 s/day (at 23°C)
Backup battery
service life
At least 10 years for clock (reference value at 23°C)
Operating environment
Indoors, Pollution Degree 2, altitude up to 2000 m
Operating temperature
and humidity range
−10°C to 50°C (14°F to 122°F), 80% RH or less (non-condensing)
(Charging temperature range: 5°C to 35°C)
Storage temperature
and humidity range
−20°C to 60°C (−4°F to 140°F), 80% RH or less (non-condensing)
Dimensions Without any modules: 272W × 145H × 43D mm (10.72″W × 5.71″H ×
1.69″D) (excluding protrusions)
With 2 modules:272W × 198H × 63D mm (10.71″W × 7.8″H × 2.78″D)
(excluding protrusions)
With 4 modules:272W × 252H × 63D mm (10.71″W × 9.92″H ×
2.48″D) (excluding protruding parts)
Mass Approx. 1108 g (39.08 oz.) (excluding battery pack)
Standards Safety: EN61010
EMC: EN61326 Class A
Vibration
resistance
JIS D 1601:1995:1995 5.3 (1)
Class 1: Passenger vehicles; conditions: Class A equivalent
Accessories Quick Start Manual, LOGGER Application Disc (Quick Start Manual,
Instruction Manual, Logger Utility, Logger Utility Instruction Manual,
Communication Instruction Manual), USB Cable, AC Adapter Z1014,
Precautions Concerning Use of Equipment that Emits Radio Waves
(LR8450-01 only)
Display
Display 7-inch TFT color LCD (WVGA 800 × 480 dots)
Display resolution
(with waveform
display selected)
Max. 20 divisions (horizontal axis) × 10 divisions (vertical axis)
(1 division = 36 dots [horizontal axis] × 36 dots [vertical axis])
Display language Japanese, English, Chinese, Korean
Backlight service life
Approx. 100,000 h (Reference value at 23°C)
Backlight saver
Turns off backlight when no key is operated for a set amount of time.
Backlight brightness
5 levels (user-selectable)
Waveform
background color
Dark/light (user-selectable)
Power supply
Power
supply
AC adapter Z1014 AC Adapter (12 V DC ±10%)
AC Adapter rated supply voltage: 100 V to 240 V AC (as-
suming voltage uctuation of ±10%)
AC Adapter rated power supply frequency: 50 Hz/60 Hz
Battery LR8450 accommodates 2 batteries
Z1007 Battery Pack
(When used with AC Adapter, AC Adapter has priority)
Li-ion, 7.2 V, 2170 mAh
External
power supply
10 V to 30 V DC
Power con-
sumption
Normal power
consumption
Using Z1014 AC Adapter or 12 V DC external power sup-
ply, without Battery Pack
With LCD at maximum brightness: 8.5 VA
(instrument only)
With LCD backlight off: 7 VA (instrument only)
Maximum
rated power
When using the Z1014 AC Adapter
95 VA (including AC Adapter)
When using a 30 V DC external power supply
28 VA
(while charging battery with LCD at maximum brightness)
When using the Z1007 Battery Pack
20 VA (with LCD at maximum brightness)
Continuous
operating
time
Battery
With one Z1007 Battery Pack: Approx. 2 h (reference value at 23°C)
With two Z1007 Battery Packs: Approx. 4 h (reference value at 23°C)
Conditions:
With one U8551 Universal Unit connected, back-
light on, voltage output off, and Z4006 connected
Charging
functionality
Charging is available when the Z1007 Battery Pack is attached and the
AC Adapter is connected.
Charging time: Approx. 7 h (reference value at 23°C)
Interface specications
The LAN interface and USB interface (function) cannot be used at the same time.
LAN
interface
IEEE 802.3 Ethernet, automatic 100Base-TX/1000Base-T detection
Auto MDI-X, DHCP, DNS support
Connector: RJ-45
Maximum cable length: 100 m
LAN func-
tionality:
Acquiring data and setting recording conditions with the Logger Utility
LAN
interface
LAN func-
tionality:
Conguring settings and controlling recording using communica-
tions commands
Manually acquiring data using the FTP server Acquiring les from a
connected SD Memory Card or USB Drive
Automatically sending data via FTP (FTP client)
Transferring les saved on a connected SD Memory Card or USB Drive
While measurement is in progress: Waveform les (binary, text)
After measurement has nished: Waveform les (binary, text),
numerical calculation result les
HTTP server function
Control mode (one instrument)
Displaying screen and remotely controlling instrument and
modules, starting/stopping measurement, acquiring data via FTP,
conguring comment, updating instrument and modules
Browsing mode (up to four instruments)
Displaying screen, measurement status, and comments
Email transmission
Start trigger, stop trigger, alarm, power outage recovery, internal buffer,
memory full, media full, wireless unit communication interruption, bat-
tery low, and periodic mail transmission. Instantaneous values can be
attached for start trigger, stop trigger, alarm, and periodic transmission.
Emails can be sent regularly at the following intervals: 30 min., 1 h,
12 h, 1 day.
NTP client function
Time synchronization with an NTP server
Regular synchronization intervals: 1 h, 1 day
Pre-measurement synchronization function
Wireless
LAN
interface
(LR8450-01
only)
IEEE 802.11b/g/n
Communications range: 30 m, line of sight
Encryption function: WPA-PSK/WPA2-PSK, TKIP/AES
Usable channels: 1 to 11
Auto-connect function: Wireless LAN function can be toggled on and off.
Supported modes: Access point, station, wireless unit connectivity
Devices that can be connected in wireless unit connectivity mode: Wireless
units or PC/tablet
Wireless unit and PC/tablet connectivity are exclusive.
Wireless
LAN func-
tionality:
Conguring settings and controlling recording using
communications commands
Manually acquiring data using the FTP server
Acquiring les from a connected SD Memory Card or USB Drive
Automatically sending data via FTP (FTP client)
Transferring les saved on a connected SD Memory Card or USB Drive
HTTP server function
Control mode (one instrument)
Displaying screen and remotely controlling instrument and mod-
ules, starting/stopping measurement, acquiring data via
FTP, conguring comment, updating instrument and modules
Browsing mode (up to four instruments)
Displaying screen, measurement status, and comments
Email transmission
Start trigger, stop trigger, alarm, power outage recovery, internal buffer,
memory full, media full, wireless unit communication interruption, bat-
tery low, and periodic mail transmission. Instantaneous values can be
attached for start trigger, stop trigger, alarm, and periodic transmission.
Emails can be sent regularly at the following intervals: 30 min., 1 h,
12 h, 1 day.
NTP client function
Time synchronization with an NTP server
Regular synchronization intervals: 1 h, 1 day
Pre-measurement synchronization function
USB
interface
(host)
Standard compliance: USB 2.0 compliant
Connectors: Series A receptacle × 2
Guaranteed-operation options: Z4006 USB drive (16 GB)
File system: FAT16, FAT32
Connectable devices: keyboard, mouse, hub (1 layer), USB drive (1 port only)
USB
interface
(function)
USB standard: USB 2.0 compliant
Connector: Series mini-B receptacle
USB functionality: Data acquisition, condition settings used with the Logger
Utility software (bundled)
Conguring settings and controlling recording using com-
munications commands
USB drive mode: Transferring data from a connected SD memory card to a computer
SD card
slot
Standard compliance: SD standard-compliant slot × 1 (with SD memory card/
SDHC memory card support)
Guaranteed-operation options: Z4001 (2 GB), Z4003 (8 GB)
File system: FAT16, FAT32
External control terminals
Terminal block Push-button type terminal block
External
I/O
Number of
terminals
4, Non-isolated (same GND as instrument)
Input Input voltage 0 V to 10 V DC
Slope Rising/falling (user-selectable)
Functionality Choose from off, start, stop, start/stop, trigger
input, event input.
Output
Output format
Open-drain output (with 5 V voltage output)
Maximum switching
capacity
5 V to 10 V DC, 200 mA
Functionality Trigger output
Alarm output
Output format
Open-drain output (with 5 V voltage output)
Maximum switching capacity
5 V to 30 V DC, 200 mA
Number of terminals 8, Non-isolated (same GND as instrument)
Voltage output Output voltage Off, 5 V, 12 V, 24 V* (user-selectable)
Supply current: Max. 100 mA each
*: 24 V output can be selected for the VOUT-
PUT1 terminal only.
Number of terminals 2, Non-isolated (same GND as instrument)
GND terminal Number of terminals 10 (common GND)
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
12
Recording
Recording mode Normal
Recording intervals
1 ms*, 2 ms*, 5 ms*, 10 ms, 20 ms, 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 2
s, 5 s, 10 s, 20 s, 30 s, 1 min., 2 min., 5 min., 10 min., 20 min., 30 min., 1 h
*: Setting available only when using a module with data refresh intervals that include 1 ms
Data refresh interval
Automatically- or user-selected value per module
Repeat recording ON/OFF (user-selectable)
Specied
time/continuous
Specied time: Recording time is set in days, hours, minutes, and seconds.
Time can be set up to maximum capacity of internal buffer memory.
(Total 256 M)
Continuous: Recording is performed once until it is stopped.
If maximum capacity of internal buffer memory is exceeded, memory
will be overwritten.
Waveform
recording
Last 256 M data points are saved in internal buffer memory. Scroll
through and view data stored in internal buffer memory.
Alarm source data recording can be toggled on and off.
Backup of recorded data
None
Display
Sheet function
Display sheets can be switched between all channels and individual modules.
All-channel display sheet: Maximum 120 analog channels,
30 waveform calculation channels, 8 pulse/logic channels,
8 alarm channels
Waveform display
screen
T
ime-axis waveform display: Simultaneous display of gages and settings
(channel representative settings and display settings)
Simultaneous display of time-axis waveforms and values: Instantaneous
values, cursor values, or numerical calculation values (user-switchable)
Numerical display: Simultaneous display of instantaneous values and statisti-
cal values
Alarm display: Display of alarm status and alarm history
Display format Time-axis waveform display: 1 screen
X-Y waveform display: 1 screen
X-Y composite Composite up to 8 waveforms.
Numerical display
format
SI units, decimal, or exponent (user-selectable)
When decimal is selected, number of decimal places to display can
be set (values will then be rounded to set number of places).
Waveform colors 24 colors
Zooming in and
out on the
waveform display
Horizontal
axis
2 ms to 1 day/division
Vertical
axis
Number of divisions per screen: 10
Setting method
Select position or upper and lower limits for each channel.
(Waveform calculation channels: upper and lower limits only)
When setting by position: Set zoom factor and zero position.
Zoom factor: 1/2×, 1×, 2×, 5×, 10×, 20×, 50×, 100×
Zero position: −50% to 150% (with a zoom factor of 1×)
When setting by upper/lower limit: Set upper and lower limit.
Waveform scrolling Display can be scrolled left and right both during recording and while
recording is stopped (during waveform rendering only).
Monitor display
Check instantaneous values and waveforms without recording data to mem-
ory (values and waveforms can be displayed while waiting for a trigger).
Wireless unit status
display
(LR8450-01 only)
Indicates the battery remaining and the radio-wave strength, in the
four levels, of the wirelessly connected modules.
Files
Save
destinations
SD memory card/USB drive (user-selectable)
(Only storage media sold by HIOKI are guaranteed for operation)
File names Up to 8 single-byte characters
Automatic numbering/dating (user-selectable)
Auto saving
Waveform data (real-time saving): Off, binary format, or text format (user-selectable)
Numerical calculation results (saved after recording): Off or text format (user-selectable)
When text format is selected, choose whether to save all calculations in one
le or to save each calculation in its own le.
Delete and
save
On/Off (user-selectable)
Off: System will stop saving data when SD memory card or
USB drive starts to run out of available space.
On:
When SD memory card or USB drive starts to run out of
available space, system will delete oldest waveform le
(binary or text) and then continue saving data.
Folder Splitting
No segmentation, 1 day, 1 week, or 1 month (user-select-
able)
File splitting Disabled, enabled, or timed (user-selectable)
Disabled: Data for each recording session is saved in its own le.
Enabled: Data for each set period of time is saved in its own
le, starting with the start of measurement.
Segmentation time: Day, hour, or minute (user-selectable)
Timed: Data will be segmented at intervals of the segment
time based on the previously set reference time and saved in
separate les.
Reference time: Set in hours and minutes.
Split time: 1 min, 2 min, 5 min, 10 min, 15 min, 20 min, 30
min, 1 h, 2 h, 3 h, 4 h, 6 h, 8 h, 12 h, 1 d
External media
eject
(SD memory
card/USB drive)
External media can be ejected during real-time saving by
activating a button on the screen and conrming a message.
Data protec-
tion
Yes
(valid only when Z1007 Battery Pack is installed)
If remaining battery life declines during real-time saving,
system will close le and stop saving data (although mea-
surement operation will continue).
Manual
saving
Data is saved when SAVE key is pressed.
Choose either selective save or immediate save as operation to perform
when SAVE key is pressed.
Decimation
(text format
only)
Decimate and save
Off or a value from 1/2 to 1/100,000 (user-selectable)
Saved data Select from instantaneous values and statistical values.
When statistical values are selected: Instantaneous values,
maximum values, minimum values, and average values will
be saved for the thinning interval.
Loading data
Loading saved
data
Specify a position and then load up to 256 M data points of previ-
ously saved text-format data.
Calculations
Numerical
calcula-
tions
Number of
calculations
Up to 10 calculations simultaneously
Calculation
content
Average value, peak-to-peak value, maximum value, maximum value
time, minimum value, minimum value time, integration*
1
, aggregation*
1
,
usage ratio*
2
, on time*
2
, off time*
2
, on count*
2
, off count*
2
*1: Total, positive, negative, or absolute value (user-selectable)
*2: Threshold values can be set for individual channels.
Calculation
range
During recording:
Calculations performed for all data during recording
After recording has stopped:
Calculations performed for all data in the internal buffer memory, or for data
in a calculation range specied by the A/B cursors (on the vertical axis)
Time split
calcula-
tion
Disabled, enabled, or timed (user-selectable)
Disabled: Calculations performed for all data during recording
Enabled: Data for each segment of time, starting with the start of mea-
surement
Segmentation time: Set DD HH:MM format
Timed: Calculations will be made at intervals of the segment time based
on the previously set reference time.
Reference time: Set in hours and minutes.
Split time: 1 min, 2 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1 h, 2 h,
3 h, 4 h, 6 h, 8 h, 12 h, 1 d
Waveform
calculations
Calculation
content
Four arithmetic operations among channels
Moving average, simple average, aggregation, and integration of any channel
Calculated values are recorded as data for calculation channels
(W1
through W30). (Calculations are performed at same time as measure-
ment. Values cannot be recalculated after measurement.)
Triggers
Trigger method Digital comparison method
Trigger timing Start, stop, or start & stop
Trigger conditions AND/OR operation performed on trigger source, interval trigger, or
external trigger
When triggers are disabled, free run
Trigger sources Analog, pulse, logic, waveform calculations
Trigger types Analog, pulse
Waveform
calculations
Level triggers: Trigger activated by rising or falling
edge at set level
Window triggers: Set by trigger level upper limit and
lower limit. Trigger activated when value leaves Area
or when value enters area
Logic Trigger activated when patterns of 1/0/X match (where
“X” indicates either)
Interval triggers Trigger activated for set recording interval after setting days/hours/
minutes/seconds
External triggers Trigger activated by rising or falling edge at set level in external input
signal. Rising/falling (user-selectable)
Trigger response
time
When using plug-in units:
(Recording interval or data refresh interval, whichever is longer) × 2 +1 ms + analog
response time*
1
When using wireless units (LR8450-01 only):
(Recording interval or data refresh time, whichever is longer) × 2 +wireless
response time*
2
+ analog response time*
1
*1: Depending on lter settings (U8554 with a data refresh interval of
5 ms and low-pass lter of 120 Hz).
*2: When the radio-wave state is in good condition, 1s.
Trigger level
resolution
Analog 0.1% f.s. (f.s. = 10 divisions)
Pulse Count 1c, rotational speed 1/n(where n= pulse count
per rotation setting)
Pre-triggers Set day/hours/minutes/seconds.
Can be set during real-time saving.
Alarms
Alarm conditions Set separately for ALM1 to ALM8
System will output an alarm when any of the following conditions are
satised:
• AND/OR operation performed on alarm sources
• Low battery
• Thermocouple burnout
• Wireless error (LR8450-01 only)
Alarm sources Analog, pulse, logic, waveform calculations
Wireless error
(LR8450-01 only)
Alarm output when a wireless communication error with a wireless
module is detected
Off/Now/3 min. (user-selectable)
Now: Outputs an alarm upon a communications disruption
3 min.: Outputs an alarm if a communication disruption continues for
3 minutes.
Low remaining
battery life
Alarm output when low remaining battery life is detected for the
instrument or a wireless unit.
Thermocouple
burnout
Alarm output when a thermocouple burnout occurs
(when Tc burnout
detection setting is enabled)
Types of alarms Analog,
pulse,
waveform
calculations
Level: System will output an alarm following a rising or
falling edge at set level
Window: Set upper limit and lower limit
System will output an alarm when value leaves area or
when value enters area
Slope: Set level and time.
The system will output an alarm when the rate of
change (level per unit time) continues to exceed the
specied change rate during the set time interval.
Logic System will output an alarm when patterns of 1/0/X
match (where “X” indicates either)
Alarm lter Apply a lter to results of AND/OR operations performed on alarm
sources. Set based on sample count (Off, 2 to 1000).
System will output an alarm if alarm state continues for set number of
samples
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
13
Alarm retention On/Off (user-selectable)
Clear alarms: When alarm retention is On, alarms will be cleared
without stopping recording.
Alarm tone On/Off (user-selectable)
Alarm output
response time
When using plug-in units:
(Recording interval or data refresh interval, whichever is longer) × 2 +1 ms + analog
response time*
1
When using wireless units (LR8450-01 only):
(Recording interval or data refresh interval, whichever is longer) × 2 +wireless
response time*2+ analog response time*1
*1: Depending on lter settings (U8554 with a data refresh interval of 5 ms and
low-pass lter of 120 Hz).
*2: When the radio-wave state is in good condition, 1s.
Other functionality
Even mark function
Number of inputs
Up to 1000 inputs per measurement
Waveform search
function
Search waveforms and display target location in center of waveform screen.
Search
conditions
Search by choosing level, window, maximum value,
minimum value, local maximum value, or local mini-
mum value.
Search range All data in internal buffer memory or data between A/B
cursors (on vertical axis)
Search targets Analog, pulse, logic, waveform calculations
Jump function
Specify event mark, A/B cursor position, trigger point, or waveform
display position to display in center of waveform screen.
Cursor
measurement
function
Cursor display All channels or specied channels (user-selectable)
Cursor movement
A, B, or simultaneous (user-selectable)
Types of cursors
Vertical or horizontal (user-selectable)
Scaling function Scaling settings can be congured separately for each channel.
Comment entry
function
Enter titles and channel-specic comments
Start state
retention function
On/Off (user-selectable)
Auto-start
function
On/Off (user-selectable)
Functionality for
saving setting
conditions
Up to ve groups of setting conditions can be saved in the instrument's
internal backup memory.
Auto setup
function
Setting conditions saved in the instrument's memory or on an SD
Memory Card or a USB Drive can be automatically loaded when the
instrument is powered on.
If there are setting conditions stored in the instrument's memory as well
as on an SD Memory Card and a USB Drive, setting conditions have
the following precedence:
instrument's memory, SD Memory Card, and USB Drive.
Prevention of
inadvertent START/
STOP key operation
When START or STOP key is pressed, system will display a message ask-
ing if user wishes to start or stop measurement.
Conrmation message: Enable/disable (user-selectable)
Key lock function Disables operation keys
Beep tone On/Off (user-selectable)
Self-check function
Can check keys, LCD, ROM/RAM, LAN, media, and modules.
Display of horizontal
axis (time values)
Horizontal axis (time value) display can be set to time, date, or data
point count. Setting is applied when text data is saved.
Conguration navi-
gation (Quick Set)
function
Wireless unit registration guide (LR8450-01 only), wireless connectivity
troubleshooting guide (LR8450-01 only), Connection diagram display
(Strain gage, external terminals), loading setting conditions
Power supply fre-
quency lter function
50 Hz/60 Hz selection
Input
Pulse/logic input
Number of
channels
8 channels (common GND, non-isolated)
Exclusive setting for pulse/logic input for individual channels
Terminal block
Push-button type terminal block
Adaptive
input format
Non-voltage contact, open collector (PNP open collector requires exter-
nal resistor), or voltage input
Maximum
input voltage
0 V to 42 V DC
Input resistance
1.1 MΩ ±5%
Detection
level
2 levels (user-selectable)
High: 1.0 V or greater; low: 0 to 0.5 V
High: 4.0 V or greater; low: 0 to 1.5 V
Pulse input
Measurement range, resolution
Measurement target Range Maximum resolution Measurable range
Count 1000 M pulse f.s. 1 pulse 0 to 1000 M pulse
Rotational speed 5000/n(r/s) f.s. 1/n(r/s) 0 to 5000/n(r/s)
300,000/n(r/min.) f.s. 1/n(r/min.) 0 to 300,000/n(r/min.)
n: Number of pulses per rotation (1 to 1000)
Pulse input
period
With lter off: 200 μs or greater
(100 μs or greater during high and low interval)
With lter on: 100 ms or greater
(50 ms or greater during high and low interval)
Slope Set rising/falling for each channel.
Measure-
ment mode
Integration (addition, instantaneous), rotational speed
Integration
Addition: Counts number of pulses input from start of measurement.
Instantaneous: Counts number of pulses input within each recording
interval (integrated value is reset for each rotational interval).
Rotational
speed
r/s: Counts number of input pulses during 1 s and calculates rotational
speed.
r/min.: Counts number of input pulses during 1 min. and calculates
rotational speed.
Smoothing
function
Select value from 1 s to 60 s (valid only when set to rotational speed
and r/min.).
Chatter pre-
vention lter
Set to On/Off for each channel.
Logic input
Measure-
ment mode
Records 1 or 0 for each recording interval.
Software Logger Utility specications
Operating
Environment
Windows7(32bit/64bit)
Windows8(32bit/64bit)
Windows10(32bit/64bit)
Overview Control PC-connected logger to receive, display and save measured
waveform data sequentially. (Total recording samples: maximum 10
million data. Data exceeding this number will be segmented into
separate measurement les while recording continues.)
* Real-time measurement on the LR8450, LR8450-01 is possible with
a recording interval of 10 ms or more.
Function Controllable loggers: 5
Data Collection System: 1 system
Display Format:
• Waveforms (split time-axis display is possible)
• Numerical values (logging) Numerical display can be enlarged
• Alarms
Above items can be displayed simultaneously
Numerical Value Monitor
Display: Display in a separate window is possible.
Scroll: Waveforms can be scrolled during measurement.
Data Collection Settings: Data collection settings of logger unit can be congured
Monitor function can be checked before measurement.
Save: Save settings from multiple devices supporting real-time mea-
surement (LUS format) and measurement data (LUW format) as one
le.
Data Save Destination: Real-time data collection le (LUW format),
transfer data in real-time or non-real-time to Microsoft Excel®, Excel®
template can be specied
Event Mark: Recording during measurement is possible
Waveform Display Supported Files: Waveform data le (LUW format, MEM format)
Display Format: Waveforms (split time-axis display available), Simul-
taneous display of numerical values (logging) available
Maximum Number of Channels: 2035 channels (measured) + 60
channels (waveform calculation)
Waveform Display Sheets: Waveform of each channel can be dis-
played on any of the ten sheets
Scroll: Available
Event Mark Recording: Available
Cursors: Cursors A and B can be used to display voltage values at
cursor positions.
Hard Copy: Hard copy of waveform display available
Data Conversion Applicable Files: Waveform data le (LUW format, MEM format)
Conversion Section: All data, specied section
Conversion Format: CSV format (comma delimited, space delimited,
tab delimited), transfer to Excel® sheet, LR5000 format (hrp2,hrp)
Data Thinning: Simple thinning with any thinning number
Waveform
Calculation
Calculation items: Four arithmetic operations
Number of calculation channel: 60 channels
Numerical
Calculations
Applicable Data: Waveform data le (LUW format, MEM format), real-
time measurement data, Waveform calculation
Calculation Items: Average value, peak value, maximum value, time
to maximum value, minimum value, time to minimum value, On time,
Off time, On count, Off count, standard deviation, aggregation, area
value, and integration
Save calculation: Perform numerical calculation and save to le
Search Applicable Data: Real-time data collection le (LUW format), Main
unit measurement le (MEM format), Waveform calculation data
Search Mode: Event mark, date and time, maximum position, mini-
mum position, local maximum position, local minimum position, alarm
position, level, window, and variation
Print Applicable printer: Printer compatible to the OS in use
Applicable data: Waveform data le (LUW format, MEM format)
Print format: Waveform image, Report print, List print (Channel set-
tings, Event, Cursor value)
Print area: All area, Specied area by A-B cursor
Print preview: Available
Option specications (sold separately)
Plug-in units: U8550, U8551, U8552, U8553, U8554
Shared specications
Host model LR8450/LR8450-01 MEMORY HiLOGGER
Operating temperature
and humidity range
-10°C to 50°C, 80% RH or less (non-condensing)
Storage temperature
and humidity range
-20°C to 60°C, 80% RH or less (non-condensing)
Vibration resistance
JIS D 1601:1995 5.3(1), Class 1A (passenger vehicle) equivalent
Accessories User manual, mounting screw × 2, wiring conrmation label
(U8554 only)
Wireless units: LR85530, LR8531, LR8532, LR8533, LR8534
Shared specications
Host model LR8450-01 MEMORY HiLOGGER
Control communications
method
Connect wirelessly via Z3230 Wireless LAN Adapter (included).
Communications buffer
memory
4 Mword (volatile memory)
Saves data in the event of a communications error. Data is re-sent
when communications are restored.
Operating temperature
and humidity range
-20°C to 55°C, 80% RH (non-condensing)
(Charging temperature range: 5°C to 35°C)
Storage temperature
and humidity range
-20°C to 60°C, 80% RH (non-condensing)
Vibration resistance JIS D 1601:1995 5.3(1), Class 1A (passenger vehicle) equivalent
LED display Wireless connection and measurement status, error status, AC
adapter/external power, battery power, charge status
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
14
Auto-connect function Available
Accessories Z3230 Wireless LAN Adapter, user manual, Z1008 AC Adapter,
mounting plate, M3×4 screw × 2 (for use with mounting plate), wiring
conrmation label (LR8534 only)
Z3230 wireless speci-
cations
Wireless LAN (IEEE 802.11b/g/n)
Range: 30 m (line of sight)
Encryption: WPA-PSK/WPA2-PSK, TKIP/AES
Channels: Channel 1 to 11
Power supply specications
AC adapter Z1008 AC Adapter (12 V DC, standard accessory)
Rated supply voltage: 100 to 240 V AC
Rated power supply frequency: 50 Hz/60 Hz
Maximum rated power: 25 VA (including AC adapter)
Normal power consumption (instrument only, without battery pack)
LR8530, LR8532, LR8533: 2.5 VA
LR8531: 3.0 VA
LR8534: 4.0 VA
Battery Z1007 Battery Pack (When using AC adapter, AC adapter takes
precedence.)
Rated supply voltage: 7.2 V DC (Li-ion 2170 mAh)
Maximum rated power
LR8530, LR8532: 1.5 VA
LR8531, LR8533: 2.0 VA
LR8534: 3.5 VA
External power supply Rated supply voltage: 10 to 30 V DC
Maximum rated power: 8 VA (30 V DC external power supply, while
charging battery)
Normal power consumption (12 V DC external power supply, without
battery pack)
LR8530, LR8532, LR8533: 2.5 VA
LR8531: 3.0 VA
LR8534: 4.0 VA
Continuous operating
time
When using Z1007 Battery Pack (all data refresh rates, good com-
munications state, 23°C reference values)
LR8530, LR8532, LR8533: Approx. 9 hr.
LR8531: Approx. 7 hr.
LR8534: Approx. 5 hr.
Charging function When Z1007 Battery Pack installed while connected to AC adapter
or 10 to 30 V DC external power supply
Charging time: Approx. 7 hr. (23°C reference value)
VOLTAGE/TEMP UNIT U8550
UNIVERSAL UNIT U8551
VOLTAGE/TEMP UNIT U8552
WIRELESS VOLTAGE/TEMP UNIT LR8530
WIRELESS UNIVERSAL UNIT LR8531
WIRELESS VOLTAGE/TEMP UNIT LR8532
(Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
General specications
Number of input
channels
U8550: 15 (Set voltage, thermocouple, or humidity for each channel)
LR8530: 15 (Set voltage or thermocouple for each channel)
U8551, LR8531: 15 (Set voltage, thermocouple, humidity, RTD, or resis-
tor for each channel)
U8552: 30 (Set voltage, thermocouple, or humidity for each channel)
LR8532: 30 (Set voltage or thermocouple for each channel)
Input terminals
U8550, LR8530: M3 screw-type terminal block (2 terminals per channel)
U8551, LR8531: Push-button type terminal block (4 terminals per channel)
U8552, LR8532: Push-button type terminal block (2 terminals per channel)
Output terminals M3 screw-type terminal block (1 output, 2 terminals, Z2000 Humidity
Sensor power supply [can power up to 15 Z2000 Humidity Sensors])
(LR8531 only)
Measurement target U8550, U8552: Voltage, thermocouples, humidity
LR8530, LR8532: Voltage, thermocouples
U8551, LR8531: Voltage, thermocouples, humidity, RTD, resistor
Input type Scanning by semiconductor relays
All channels isolated (Not isolated when measuring with RTD, resistance
or humidity)
A/D resolution 16 bits
Maximum input
voltage
±100 V DC (maximum voltage between input terminals without causing
damage)
Maximum channel-
to-channel voltage
300 V DC (maximum voltage that can be applied between each input
channel without causing damage; not isolated when measuring with RTD,
resistance or humidity)
*Channels are isolated from each other with semiconductor relays. Never
allow a voltage exceeding the product specications, for example a
lightning surge, to be applied across channels as doing so may cause the
semiconductor relays to short.
Maximum rated
terminal-to-ground
voltage
300 V AC, DC (maximum voltage that can be applied between input
channels and the instrument or its chassis, or between units without
causing damage; humidity measurement not isolated)
Input resistance 10 MΩ or greater (10 mV f.s. to 2 V f.s. voltage ranges, thermocouple
ranges, RTD and resistor ranges)
1 MΩ ±5% (10 V f.s. to 100 V f.s. voltage range, 1-5 V f.s. voltage range,
humidity measurement)
Allowable signal
source resistance
1 kΩ or less
Data refresh interval
10 ms to 10 s (10 selectable levels)
Digital lters Digital lter cutoff frequency is automatically set to data refresh inter-
val, burnout setting, and power supply frequency lter setting
Dimensions U8550, U8551, U8552: Approx. 134W × 70H × 63D mm (5.28″W ×
2.76″H × 2.48″D)
LR8530, LR8531, LR8532: Approx. 154W × 106H × 57D mm (6.06″W
× 4.17″H × 2.24″D)
Mass U8550: Approx. 345 g (12.2 oz.), U8551: Approx. 318 g (11.2 oz.),
U8552: Approx. 319 g (11.3 oz.), LR8530: Approx. 423 g (14.9 oz.),
LR8531: Approx. 386 g (13.6 oz.), LR8532: Approx. 388 g (13.7 oz.),
(including Z3230 Wireless LAN Adapter)
Accessories Instruction Manual, Installation screws × 2
Analog input specications
(23 ±5˚C/73 ±9˚F, 80% rh or less, after 30 minutes of warm-up and zero-adjustment, with
the 50 Hz/60 Hz cut-off setting selected)
Voltage
Range
Maximum resolution
Measurable range
Measurement accuracy
10 mV f.s. 500 nV -10 mV to 10 mV ±10 μV
20 mV f.s. 1 μV -20 mV to 20 mV ±20 μV
100 mV f.s. 5 μV -100 mV to 100 mV ±50 μV
200 mV f.s. 10 μV -200 mV to 200 mV ±100 μV
1 V f.s. 50 μV -1 V to 1 V ±500 μV
2 V f.s. 100 μV -2 V to 2 V ±1 mV
10 V f.s. 500 μV -10 V to 10 V ±5 mV
20 V f.s. 1 mV -20 V to 20 V ±10 mV
100 V f.s. 5 mV -100 V to 100 V ±50 mV
1-5 V f.s. 500 μV 1 V to 5 V ±5 mV
Temperature
Thermocouple (Not including accuracy of reference junction compensation)
Standards: JIS C1602-2015,IEC584
Type
Range
Measurable range
Measurable range
Measurement accuracy
K 100°C f.s. 0.01°C -100°C to less than 0°C ±0.7°C
0°C to 100°C ±0.5°C
500°C f.s. 0.05°C -200°C to less than -100°C ±1.4°C
-100°C to less than 0°C ±0.7°C
0°C to 500°C ±0.5°C
2000°C f.s. 0.1°C -200°C to less than -100°C ±1.4°C
-100°C to less than 0°C ±0.7°C
0°C to less than 500°C ±0.5°C
500°C to 1350°C ±0.7°C
J 100°C f.s. 0.01°C -100°C to less than 0°C ±0.7°C
0°C to 100°C ±0.5°C
500°C f.s. 0.05°C -200°C to less than -100°C ±0.9°C
-100°C to less than 0°C ±0.7°C
0°C to 500°C ±0.5°C
2000°C f.s. 0.1°C -200°C to less than -100°C ±0.9°C
-100°C to less than 0°C ±0.7°C
0°C to 1200°C ±0.5°C
E 100°C f.s. 0.01°C -100°C to less than 0°C ±0.7°C
0°C to 100°C ±0.5°C
500°C f.s. 0.05°C -200°C to less than -100°C ±0.9°C
-100°C to less than 0°C ±0.7°C
0°C to 500°C ±0.5°C
2000°C f.s. 0.1°C -200°C to less than -100°C ±0.9°C
-100°C to less than 0°C ±0.7°C
0°C to 1000°C ±0.5°C
T 100°C f.s. 0.01°C -100°C to less than 0°C ±0.7°C
0°C to 100°C ±0.5°C
500°C f.s. 0.05°C -200°C to less than -100°C ±1.4°C
-100°C to less than 0°C ±0.7°C
0°C to 400°C ±0.5°C
2000°C f.s. 0.1°C -200°C to less than -100°C ±1.4°C
-100°C to less than 0°C ±0.7°C
0°C to 400°C ±0.5°C
N 100°C f.s. 0.01°C -100°C to less than 0°C ±1.1°C
0°C to 100°C ±0.9°C
500°C f.s. 0.05°C -200°C to less than -100°C ±2.1°C
-100°C to less than 0°C ±1.1°C
0°C to 500°C ±0.9°C
2000°C f.s. 0.1°C -200°C to less than -100°C ±2.1°C
-100°C to less than 0°C ±1.1°C
0°C to 1300°C ±0.9°C
R 100°C f.s. 0.01°C 0°C to 100°C ±4.4°C
500°C f.s. 0.05°C 0°C to less than 100°C ±4.4°C
100°C to less than 300°C ±2.9°C
300°C to 500°C ±2.2°C
2000°C f.s. 0.1°C 0°C to less than 100°C ±4.4°C
100°C to less than 300°C ±2.9°C
300°C to 1700°C ±2.2°C
S 100°C f.s. 0.01°C 0°C to 100°C ±4.4°C
500°C f.s. 0.05°C 0°C to less than 100°C ±4.4°C
100°C to less than 300°C ±2.9°C
300°C to 500°C ±2.2°C
2000°C f.s. 0.1°C 0°C to less than 100°C ±4.4°C
100°C to less than 300°C ±2.9°C
300°C to 1700°C ±2.2°C
B 2000°C f.s. 0.1°C 400°C to less than 600°C ±5.4°C
600°C to less than 1000°C ±3.7°C
1000°C to 1800°C ±2.4°C
C 100°C f.s. 0.01°C 0°C to 100°C ±1.7°C
500°C f.s. 0.05°C 0°C to 500°C ±1.7°C
2000°C f.s. 0.1°C 0°C to 2000°C ±1.7°C
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
15
Other specications about thermocouple measurement
Reference junction compen-
sation: Internal/external
At INT RJC, total accuracy = add ± 0.5˚C
Thermocouple burnout
detection: ON/OFF
System will check for burnout at each data refresh interval
during thermocouple measurement. (10 ms interval not
available)
U8550, U8551, U8552, LR8531 Only Input specications
Humidity (use Humidity Sensor Z2000)
Range Maximum resolution Measurable range
100% rh f.s. 0.1% rh 5.0% rh to 95.0% rh
Humidity sensor Z2000 accuracy
100
-40 0 10 20 30 40 50 85
0
5
20
40
60
80
95
±10% rh
±8% rh
±10% rh
±8% rh ±6% rh ±8% rh
±6% rh ±5% rh ±6% rh
Relative humidity (% RH)
Temperature (°C)
Outside guarantee range
Outside guarantee range
If the humidity value lies on a boundary line, the better of the two
regions’ measurement accuracy values applies.
U8551, LR8531 Only Input specications
Temperature
RTD
Connection: 3-wire/4-wire, Measurement current: 1 mA (Pt100, Jpt100),
0.1 mA (Pt1000)
Standards: Pt100,Pt1000:JIS C1604-2013,IEC751 JPt100:JIS C1604-1989
Type Range
Maximum resolution
Measurable range
Measurement accuracy
Pt100
100°C f.s. 0.01°C -100°C to 100°C ±0.5°C
500°C f.s. 0.05°C -200°C to 500°C ±0.7°C
2000°C f.s. 0.1°C -200°C to 800°C ±0.9°C
JPt100
100°C f.s. 0.01°C -100°C to 100°C ±0.5°C
500°C f.s. 0.05°C -200°C to 500°C ±0.7°C
2000°C f.s. 0.1°C -200°C to 500°C ±0.9°C
Pt1000
100°C f.s. 0.01°C -100°C to 100°C ±0.5°C
500°C f.s. 0.05°C -200°C to 500°C ±0.7°C
2000°C f.s. 0.1°C -200°C to 800°C ±0.9°C
*When using Pt1000, data refresh intervals of 10ms, 20m, and 50ms are not available.
Resistance
Connection: 4-wire; measurement current: 1 mA
Range Maximum resolution Measurable range Measurement accuracy
10
Ω
f.s. 0.5
mΩ 0
Ω to 10
Ω±10
mΩ
20
Ω
f.s. 1
mΩ 0
Ω to 20
Ω±20
mΩ
100
Ω
f.s. 5
mΩ 0
Ω to 100
Ω±100
mΩ
200
Ω
f.s. 10
mΩ 0
Ω to 200
Ω±200
mΩ
HIGH SPEED VOLTAGE UNIT
U8553
WIRELESS HIGH SPEED VOLTAGE UNIT
LR8531
(Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
General specications
Number of input channels
5 (voltage only)
Input terminals M3 screw-type terminal block (2 terminals per channel),
outtted with terminal block cover
Measurement target Voltage
Input type Scanning by semiconductor relays, all channels isolated
A/D resolution 16 bits
Maximum input voltage ±100 V DC (maximum voltage between input terminals without
causing damage)
Maximum channel-to-
channel voltage
300 V DC (maximum voltage between input channels without
causing damage)
*Channels are isolated from each other with semiconductor relays.
Never allow a voltage exceeding the product specications, for
example a lightning surge, to be applied across channels as doing
so may cause the semiconductor relays to short.
Maximum rated termi-
nal-to-ground voltage
300 V AC, DC
(maximum voltage between input channel and chas-
sis, or between modules, without causing damage)
Input resistance 1MΩ±5%
Allowable signal
source resistance
100Ω or less
Data refresh interval 1 ms to 10 s (13 selectable levels)
Digital lters Digital lter cutoff frequency is automatically set to data refresh
interval, burnout detection setting, and power supply frequency
lter setting.
Dimensions
U8553: Approx. 134W × 70H × 63D mm (5.28″W × 2.76″H × 2.48″D)
LR8531: Approx. 154W × 106H × 57D mm
(6.06″W × 4.17″H × 2.24″D)
Mass
U8553:
Approx. 237
g (8.4
oz.)
LR8531:
Approx. 370
g (13.1
oz.)
(including Z3230 Wireless LAN Adapter)
Analog input specications
(23 ±5˚C/73 ±9˚F, 80% rh or less, after 30 minutes of warm-up and zero-adjustment, with
the 50 Hz/60 Hz cut-off setting selected)
Measurement
target
Range Maximum
resolution
Measurable
range
Measurement
accuracy
Voltage 100 mV f.s. 5 μV -100 mV to 100 mV ±100 μV
200 mV f.s. 10 μV -200 mV to 200 mV ±200 μV
1 V f.s. 50 μV -1 V to 1 V ±1 mV
2 V f.s. 100 μV -2 V to 2 V ±2 mV
10 V f.s. 500 μV -10 V to 10 V ±10 mV
20 V f.s. 1 mV -20 V to 20 V ±20 mV
100 V f.s. 5 mV -100 V to 100 V ±100 mV
1-5 V f.s. 500 μV 1 V to 5 V ±10 mV
STRAIN UNIT U8554 WIRELESS STRAIN UNIT LR8534
(Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
General specications
Number of input
channels
5 (Set voltage or strain for each channel.)
Input terminals
Push-button type terminal block (5 terminals per channel), outtted with
terminal block cover, Set DIP switches according to measurement target.
Measurement
target
Voltage
Strain Strain gage-type converter
Strain gage
1-gage method (2-wire setup), 1-gage method (3-wire setup),
2-gage method (adjacent sides), 4-gage method
Adaptive gage
resistance
1-gage method, 2-gage method: 120 Ω
(external bridge box required for 350 Ω)
4-gage method: 120 Ω to 1 kΩ
Gage ratio 2.0 (xed)
Bridge voltage 2 V ±0.05 V DC
Balance
adjustment
Method Electronic auto-balancing
Range Voltage: ±20 mV or less (1 mV f.s. to 20 mV f.s. range),
±200 mV or less (50 mV f.s. to 200 mV f.s. range)
Strain:
±20,000 με or less (1000 με f.s. to 20,000 με f.s.
range), ±200,000 με or less (50,000 με f.s. to 200,000
με f.s. range)
Input type Balanced differential input, Simultaneous sampling of all channels
(non-isolated channels)
A/D resolution 16bit
Maximum input
voltage
±0.5 V DC (maximum voltage between input terminals without causing damage)
Maximum channel-
to-channel voltage
Non-isolated (all channels share common GND)
Maximum rated
terminal-to-ground
voltage
30 Vrms AC or 60 V DC (maximum voltage between input channel and
chassis without causing damage)
Input resistance 2 MΩ ±5%
Data refresh interval
1 ms to 10 s (13 selectable levels)
Low-pass lter Cutoff frequency: −3 dB ±30%
Auto, 120, 60, 30, 15, 8, 4 (Hz)
Auto: Cutoff frequency of low-pass lter is automatically set based on
set data refresh interval.
Attenuation characteristics: 5th-order Butterworth lter, −30 dB/oct
Dimensions
U8554: Approx. 134W × 70H × 63D mm (5.28″W × 2.76″H × 2.48″D)
LR8534: Approx. 154W × 106H × 57D mm (6.06″W × 4.17″H × 2.24″D)
Mass
U8554:
Approx. 236
g (8.3
oz.)
LR8534:
Approx. 372
g
(13.1 oz.) (including Z3230 Wireless LAN Adapter)
Analog input specications
(23 ±5˚C/73 ±9˚F, 80% rh or less, auto-balance at least 30 minutes after power on, with
LPF set at 4 Hz)
Measure-
ment target
Range Maximum
resolution
Measurable range
Measurement
accuracy
Voltage 1 mV f.s. 50 nV -1 mV to 1 mV ±9 μV
2 mV f.s. 100 nV -2 mV to 2 mV ±10 μV
5 mV f.s. 250 nV -5 mV to 5 mV ±25 μV
10 mV f.s. 500 nV -10 mV to 10 mV ±50 μV
20 mV f.s. 1 μV -20 mV to 20 mV ±100 μV
50 mV f.s. 2.5 μV -50 mV to 50 mV ±250 μV
100 mV f.s. 5 μV -100 mV to 100 mV ±500 μV
200 mV f.s. 10 μV -200 mV to 200 mV ±1 mV
Strain 1,000 με f.s. 0.05 με -1,000 με to 1,000 με ±9 με
2,000 με f.s. 0.1 με -2,000 με to 2,000 με ±10 με
5,000 με f.s. 0.25 με -5,000 με to 5,000 με ±25 με
10,000 με f.s. 0.5 με -10,000 με to 10,000 με ±50 με
20,000 με f.s. 1 με -20,000 με to 20,000 με ±100 με
50,000 με f.s. 2.5 με -50,000 με to 50,000 με ±250 με
100,000 με f.s. 5 με -100,000 με to 100,000 με ±500 με
200,000 με f.s. 10 με -200,000 με to 200,000 με ±1000 με
* Internal bridge resistance precision tolerance: ±0.01%; temperature characteristics: ±2 ppm/°C
*
Measurement accuracy does not include internal bridge resistance tolerance and temperature
characteristics
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
HEADQUARTERS
81 Koizumi,
Ueda, Nagano 386-1192 Japan
https://www.hioki.com/
All information correct as of Feb. 25, 2021. All specications are subject to change without notice. LR8450E5-12E Printed in Japan
DISTRIBUTED BY
Note: Company names and product names appearing in this catalog are trademarks or registered trademarks of various companies.
Ships standard with
wireless units.
Plug-in units Wireless units
VOLTAGE/TEMP UNIT U8550
Channels: 15; maximum sampling rate: 10 ms
WIRELESS VOLTAGE/TEMP UNIT LR8530
Channels: 15; maximum sampling rate: 10 ms
UNIVERSAL UNIT U8551
Channels: 15; maximum sampling rate: 10 ms
WIRELESS UNIVERSAL UNIT LR8531
Channels: 15; maximum sampling rate: 10 ms
VOLTAGE/TEMP UNIT U8552
Channels: 30; maximum sampling rate: 20 ms
(When 15 or fewer channels are used, 10 ms)
WIRELESS VOLTAGE/TEMP UNIT LR8532
Channels: 30; maximum sampling rate: 20 ms
(When 15 or fewer channels are used, 10 ms)
HIGH SPEED VOLTAGE UNIT U8553
Channels: 5; maximum sampling rate: 1 ms
WIRELESS HIGH SPEED VOLTAGE UNIT LR8533
Channels: 5; maximum sampling rate: 1 ms
STRAIN UNIT U8554
Channels: 5; maximum sampling rate: 1 ms
WIRELESS STRAIN UNIT LR8534
Channels: 5; maximum sampling rate: 1 ms
Option
Model: MEMORY HiLOGGER LR8450
SD memory card Z4001
2 GB capacity
SD memory card Z4003
8 GB capacity
USB drive Z4006
16 GB, Long-life, High-reliability SLC Flash Memory
Storage media
*Always use HIOKI optional storage media. Proper operation is not guaranteed when using storage media from other manufacturers, and may prevent the prod-
uct from saving and loading data properly.
BATTERY PACK
Z1007
AC ADAPTER
Z1014
AC ADAPTER
Z1008
Fixed Stand
FIXED STAND
Z5040
For installing logger
on wall
CASE
CARRYING CASE
C1012
Accommodates instrument
and four plug-in units or
seven wireless units.
LAN CABLE 9642 HUMIDITY SENSOR Z2000
Straight Ethernet cable, supplied
with straight to cross conversion
adapter, 5 m (16.41 ft) length
(Analog output), 3 m (9.84 ft) length
Thermocouple
For reference only. Please purchase locally.
Cables, sensors, etc.
Model No.
(Order code)
Specications
LR8450
Standard model, main unit only
LR8450-01
Wireless LAN equipped model, main unit only
Instrument takes two;
wireless units take one.
WIRELESS LAN ADAPTER
Z3230
For wireless units
• The LR8450 and LR8450-01 cannot perform measurement on their own. One or more plug-in
units or wireless units are required (sold separately).
• The LR8450-01 and each wireless unit emit radio waves. Use of radio waves is subject to
licensing requirements in certain countries. Using it in a country or region other than those
indicated may violate the law and may result in legal penalties for the operator.
For the latest information about countries and regions where wireless operation is currently
supported, please visit the Hioki website.
For instrumentFor instrument
and wireless units For wireless units
Ships standard with
wireless units.
Wireless Lan Adapter
Ships standard with
LR8450/LR8540-01.
Power supply
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com

Other manuals for LR8450

6

This manual suits for next models

1

Other Hioki Data Logger manuals

Hioki LR8400-20 User manual

Hioki

Hioki LR8400-20 User manual

Hioki LR8450 Instruction manual

Hioki

Hioki LR8450 Instruction manual

Hioki MICRO HiCORDER 8206-10 User manual

Hioki

Hioki MICRO HiCORDER 8206-10 User manual

Hioki LR8450 Instruction manual

Hioki

Hioki LR8450 Instruction manual

Hioki 3638-20 User manual

Hioki

Hioki 3638-20 User manual

Hioki 8861 Instruction manual

Hioki

Hioki 8861 Instruction manual

Hioki 8420-51 Instruction manual

Hioki

Hioki 8420-51 Instruction manual

Hioki LR5051 User manual

Hioki

Hioki LR5051 User manual

Hioki LR5041 User manual

Hioki

Hioki LR5041 User manual

Hioki LR8400-20 User manual

Hioki

Hioki LR8400-20 User manual

Hioki LR5011 User manual

Hioki

Hioki LR5011 User manual

Hioki LR5031 User manual

Hioki

Hioki LR5031 User manual

Hioki LR5051-20 User manual

Hioki

Hioki LR5051-20 User manual

Hioki LR8432-20 User manual

Hioki

Hioki LR8432-20 User manual

Hioki Hioki Instruction manual

Hioki

Hioki Hioki Instruction manual

Hioki HiLogger 3145-20 User manual

Hioki

Hioki HiLogger 3145-20 User manual

Hioki 3631-20 User manual

Hioki

Hioki 3631-20 User manual

Hioki HiCORDER MR8880 User manual

Hioki

Hioki HiCORDER MR8880 User manual

Hioki 8423 User manual

Hioki

Hioki 8423 User manual

Hioki LR8431-20 User manual

Hioki

Hioki LR8431-20 User manual

Hioki LR5092-20 User manual

Hioki

Hioki LR5092-20 User manual

Hioki 3645-20 User manual

Hioki

Hioki 3645-20 User manual

Hioki LR8431-20 User manual

Hioki

Hioki LR8431-20 User manual

Hioki LR5051 User manual

Hioki

Hioki LR5051 User manual

Popular Data Logger manuals by other brands

Measurement Computing USB-1608G user guide

Measurement Computing

Measurement Computing USB-1608G user guide

Diamond Aircraft DA 40 NG manual

Diamond Aircraft

Diamond Aircraft DA 40 NG manual

Swema SE-520 instructions

Swema

Swema SE-520 instructions

NexSens X2 quick start guide

NexSens

NexSens X2 quick start guide

Lovato DME CD Protocol guide

Lovato

Lovato DME CD Protocol guide

Gemini Tinytag TE-4703 manual

Gemini

Gemini Tinytag TE-4703 manual

Opticon PX-20 Specification manual

Opticon

Opticon PX-20 Specification manual

Extech Instruments 451126 user guide

Extech Instruments

Extech Instruments 451126 user guide

Vaisala MRG112 Technical reference

Vaisala

Vaisala MRG112 Technical reference

Comet U0111 Quick start manual

Comet

Comet U0111 Quick start manual

Intempco CX501 manual

Intempco

Intempco CX501 manual

Healthmark TEMP-WiFi-TP Instructions for use

Healthmark

Healthmark TEMP-WiFi-TP Instructions for use

ImmersionRC EzOSD manual

ImmersionRC

ImmersionRC EzOSD manual

globalwater GL500-2-1 manual

globalwater

globalwater GL500-2-1 manual

Global Sat DG-388 user manual

Global Sat

Global Sat DG-388 user manual

Delta OHM HD30.1 operating manual

Delta OHM

Delta OHM HD30.1 operating manual

globalwater PL200G manual

globalwater

globalwater PL200G manual

Delta OHM DO 9721 instruction manual

Delta OHM

Delta OHM DO 9721 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.