
12
Recording
Recording mode Normal
Recording intervals
1 ms*, 2 ms*, 5 ms*, 10 ms, 20 ms, 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 2
s, 5 s, 10 s, 20 s, 30 s, 1 min., 2 min., 5 min., 10 min., 20 min., 30 min., 1 h
*: Setting available only when using a module with data refresh intervals that include 1 ms
Data refresh interval
Automatically- or user-selected value per module
Repeat recording ON/OFF (user-selectable)
Specied
time/continuous
Specied time: Recording time is set in days, hours, minutes, and seconds.
Time can be set up to maximum capacity of internal buffer memory.
(Total 256 M)
Continuous: Recording is performed once until it is stopped.
If maximum capacity of internal buffer memory is exceeded, memory
will be overwritten.
Waveform
recording
Last 256 M data points are saved in internal buffer memory. Scroll
through and view data stored in internal buffer memory.
Alarm source data recording can be toggled on and off.
Backup of recorded data
None
Display
Sheet function
Display sheets can be switched between all channels and individual modules.
All-channel display sheet: Maximum 120 analog channels,
30 waveform calculation channels, 8 pulse/logic channels,
8 alarm channels
Waveform display
screen
T
ime-axis waveform display: Simultaneous display of gages and settings
(channel representative settings and display settings)
Simultaneous display of time-axis waveforms and values: Instantaneous
values, cursor values, or numerical calculation values (user-switchable)
Numerical display: Simultaneous display of instantaneous values and statisti-
cal values
Alarm display: Display of alarm status and alarm history
Display format Time-axis waveform display: 1 screen
X-Y waveform display: 1 screen
X-Y composite Composite up to 8 waveforms.
Numerical display
format
SI units, decimal, or exponent (user-selectable)
When decimal is selected, number of decimal places to display can
be set (values will then be rounded to set number of places).
Waveform colors 24 colors
Zooming in and
out on the
waveform display
Horizontal
axis
2 ms to 1 day/division
Vertical
axis
Number of divisions per screen: 10
Setting method
Select position or upper and lower limits for each channel.
(Waveform calculation channels: upper and lower limits only)
When setting by position: Set zoom factor and zero position.
Zoom factor: 1/2×, 1×, 2×, 5×, 10×, 20×, 50×, 100×
Zero position: −50% to 150% (with a zoom factor of 1×)
When setting by upper/lower limit: Set upper and lower limit.
Waveform scrolling Display can be scrolled left and right both during recording and while
recording is stopped (during waveform rendering only).
Monitor display
Check instantaneous values and waveforms without recording data to mem-
ory (values and waveforms can be displayed while waiting for a trigger).
Wireless unit status
display
(LR8450-01 only)
Indicates the battery remaining and the radio-wave strength, in the
four levels, of the wirelessly connected modules.
Files
Save
destinations
SD memory card/USB drive (user-selectable)
(Only storage media sold by HIOKI are guaranteed for operation)
File names Up to 8 single-byte characters
Automatic numbering/dating (user-selectable)
Auto saving
Waveform data (real-time saving): Off, binary format, or text format (user-selectable)
Numerical calculation results (saved after recording): Off or text format (user-selectable)
When text format is selected, choose whether to save all calculations in one
le or to save each calculation in its own le.
Delete and
save
On/Off (user-selectable)
Off: System will stop saving data when SD memory card or
USB drive starts to run out of available space.
On:
When SD memory card or USB drive starts to run out of
available space, system will delete oldest waveform le
(binary or text) and then continue saving data.
Folder Splitting
No segmentation, 1 day, 1 week, or 1 month (user-select-
able)
File splitting Disabled, enabled, or timed (user-selectable)
Disabled: Data for each recording session is saved in its own le.
Enabled: Data for each set period of time is saved in its own
le, starting with the start of measurement.
Segmentation time: Day, hour, or minute (user-selectable)
Timed: Data will be segmented at intervals of the segment
time based on the previously set reference time and saved in
separate les.
Reference time: Set in hours and minutes.
Split time: 1 min, 2 min, 5 min, 10 min, 15 min, 20 min, 30
min, 1 h, 2 h, 3 h, 4 h, 6 h, 8 h, 12 h, 1 d
External media
eject
(SD memory
card/USB drive)
External media can be ejected during real-time saving by
activating a button on the screen and conrming a message.
Data protec-
tion
Yes
(valid only when Z1007 Battery Pack is installed)
If remaining battery life declines during real-time saving,
system will close le and stop saving data (although mea-
surement operation will continue).
Manual
saving
Data is saved when SAVE key is pressed.
Choose either selective save or immediate save as operation to perform
when SAVE key is pressed.
Decimation
(text format
only)
Decimate and save
Off or a value from 1/2 to 1/100,000 (user-selectable)
Saved data Select from instantaneous values and statistical values.
When statistical values are selected: Instantaneous values,
maximum values, minimum values, and average values will
be saved for the thinning interval.
Loading data
Loading saved
data
Specify a position and then load up to 256 M data points of previ-
ously saved text-format data.
Calculations
Numerical
calcula-
tions
Number of
calculations
Up to 10 calculations simultaneously
Calculation
content
Average value, peak-to-peak value, maximum value, maximum value
time, minimum value, minimum value time, integration*
1
, aggregation*
1
,
usage ratio*
2
, on time*
2
, off time*
2
, on count*
2
, off count*
2
*1: Total, positive, negative, or absolute value (user-selectable)
*2: Threshold values can be set for individual channels.
Calculation
range
During recording:
Calculations performed for all data during recording
After recording has stopped:
Calculations performed for all data in the internal buffer memory, or for data
in a calculation range specied by the A/B cursors (on the vertical axis)
Time split
calcula-
tion
Disabled, enabled, or timed (user-selectable)
Disabled: Calculations performed for all data during recording
Enabled: Data for each segment of time, starting with the start of mea-
surement
Segmentation time: Set DD HH:MM format
Timed: Calculations will be made at intervals of the segment time based
on the previously set reference time.
Reference time: Set in hours and minutes.
Split time: 1 min, 2 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1 h, 2 h,
3 h, 4 h, 6 h, 8 h, 12 h, 1 d
Waveform
calculations
Calculation
content
Four arithmetic operations among channels
Moving average, simple average, aggregation, and integration of any channel
Calculated values are recorded as data for calculation channels
(W1
through W30). (Calculations are performed at same time as measure-
ment. Values cannot be recalculated after measurement.)
Triggers
Trigger method Digital comparison method
Trigger timing Start, stop, or start & stop
Trigger conditions AND/OR operation performed on trigger source, interval trigger, or
external trigger
When triggers are disabled, free run
Trigger sources Analog, pulse, logic, waveform calculations
Trigger types Analog, pulse
Waveform
calculations
Level triggers: Trigger activated by rising or falling
edge at set level
Window triggers: Set by trigger level upper limit and
lower limit. Trigger activated when value leaves Area
or when value enters area
Logic Trigger activated when patterns of 1/0/X match (where
“X” indicates either)
Interval triggers Trigger activated for set recording interval after setting days/hours/
minutes/seconds
External triggers Trigger activated by rising or falling edge at set level in external input
signal. Rising/falling (user-selectable)
Trigger response
time
When using plug-in units:
(Recording interval or data refresh interval, whichever is longer) × 2 +1 ms + analog
response time*
1
When using wireless units (LR8450-01 only):
(Recording interval or data refresh time, whichever is longer) × 2 +wireless
response time*
2
+ analog response time*
1
*1: Depending on lter settings (U8554 with a data refresh interval of
5 ms and low-pass lter of 120 Hz).
*2: When the radio-wave state is in good condition, 1s.
Trigger level
resolution
Analog 0.1% f.s. (f.s. = 10 divisions)
Pulse Count 1c, rotational speed 1/n(where n= pulse count
per rotation setting)
Pre-triggers Set day/hours/minutes/seconds.
Can be set during real-time saving.
Alarms
Alarm conditions Set separately for ALM1 to ALM8
System will output an alarm when any of the following conditions are
satised:
• AND/OR operation performed on alarm sources
• Low battery
• Thermocouple burnout
• Wireless error (LR8450-01 only)
Alarm sources Analog, pulse, logic, waveform calculations
Wireless error
(LR8450-01 only)
Alarm output when a wireless communication error with a wireless
module is detected
Off/Now/3 min. (user-selectable)
Now: Outputs an alarm upon a communications disruption
3 min.: Outputs an alarm if a communication disruption continues for
3 minutes.
Low remaining
battery life
Alarm output when low remaining battery life is detected for the
instrument or a wireless unit.
Thermocouple
burnout
Alarm output when a thermocouple burnout occurs
(when Tc burnout
detection setting is enabled)
Types of alarms Analog,
pulse,
waveform
calculations
Level: System will output an alarm following a rising or
falling edge at set level
Window: Set upper limit and lower limit
System will output an alarm when value leaves area or
when value enters area
Slope: Set level and time.
The system will output an alarm when the rate of
change (level per unit time) continues to exceed the
specied change rate during the set time interval.
Logic System will output an alarm when patterns of 1/0/X
match (where “X” indicates either)
Alarm lter Apply a lter to results of AND/OR operations performed on alarm
sources. Set based on sample count (Off, 2 to 1000).
System will output an alarm if alarm state continues for set number of
samples
ES France - Département Tests & Mesures - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com