CLAMP ON
AC/DC HiTESTER
Apr. 2016 Edition 1
Printed in Japan
3287C961-00 16-04H
3287
3288
3288-20
EN
Instruction Manual
Symbols afxed to the device
Precaution or hazard (See corresponding topic.)
Risk of electric shock
Protected throughout by double insulation or reinforced insulation
Device may be connected to or disconnected from a live conductor
Grounding
terminal DC (direct current) AC (alternating
current)
Accuracy
We dene measurement tolerances in terms of rdg. (reading) and
dgt. (digit) values,with the following meanings:
rdg.
(reading or
displayed value)
The value currently being measured and
indicated on the measuring instrument.
dgt.
(resolution)
The smallest displayable unit on a digital measuring
instrument,i.e.,the input value that causes the digital
display to show a “1” as the least-signicant digit.
Measurement categories
This instrument’s current measurement part conforms to the safety
requirements for CAT III 600 V and the voltage measurement part
conforms to the safety requirements for CAT II 600 V,CAT III 300 V
measuring instruments.
Service entrance
Distribution panel
Service drop Internal wiring
Power meter
Fixed Installation
Outlet
CAT IV CAT III CAT II
DANGER
Measuring a location with a higher category number
than the measurement category indicated on this
device may result in a serious accident such as
electric shock.
To avoid electric shock, do not touch the portion
beyond the protective barrier during use.
Never apply voltage to the test leads when the
resistance and continuity functions are selected.
Doing so may damage the instrument and result in
bodily injury. To avoid electrical accidents, remove
power from the circuit before measuring.
WARNING
•To avoid electric shock,short circuits and damage
to the instrument,disconnect the test leads from the
measurement object before switching the rotary switch.
•To prevent electric shock,when measuring the voltage
of a power line use a test lead that satises the
following criteria:
•Conforms to safety standards IEC61010 or EN61010
•Of measurement category III or IV
•Its rated voltage is higher than the voltage to be measured
•
The optional test leads for this instrument conform to the
safety standard EN61010. Use a test lead in accordance
with its dened measurement category and rated voltage.
•To avoid an electric shock,operate the instrument at
below a lower rated voltage between that indicated on
the instrument and on test leads.
WARNING
•Installing the instrument in inappropriate locations may
cause a malfunction of instrument or may give rise to
an accident. Avoid the following locations:
•Exposed to direct sunlight or high temperature
•Exposed to corrosive or combustible gases
• Exposed to a strong electromagnetic eld or
electrostatic charge
•
Near induction heating systems (such as high-frequency
induction heating systems and IH cooking equipment)
•Susceptible to vibration
•Exposed to water,oil,chemicals,or solvents
•Exposed to high humidity or condensation
•Exposed to high quantities of dust particles
WARNING
•Since there is a risk of electric shock,check that the
insulation on the test lead are neither ripped nor torn,and
no metal conductor inside the wire are exposed before
using the instrument. If damaged,replace them with those
specied by our company.
•To prevent a short circuit accident,be sure to use the
test leads with the sleeves attached when performing
measurements in the CAT III measurement category.
•If the sleeves are inadvertently removed during
measurement,stop the measurement.
•
With regard to the electricity supply,there are risks of
electric shock,heat generation, re, and arc ash due
to short circuits. If persons unfamiliar with electricity
measuring instrument are to use the instrument,another
person familiar with such instruments must supervise
operations.
•This instrument is measured on a live line. To prevent
electric shock,use appropriate protective insulation and
adhere to applicable laws and regulations.
•Handle and dispose of batteries in accordance with
local regulations.
CAUTION
•Do not place foreign objects between the jaw tips or
insert foreign objects into the gaps of the jaws (or exible
loop couplings). Doing so may worsen the performances
of the sensor or interfere with clamping action.
•
Be careful to avoid dropping the instrument or otherwise
subjecting them to mechanical shock,which could
damage the jaw and adversely affect measurement.
Poor performance or damage from battery leakage
could result. Observe the cautions listed below:
•Do not use batteries after their recommended expiry date.
•Do not allow weak batteries to remain in the instrument.
• Replace batteries only with the specied type.
•Remove the batteries from the instrument if it is to be
stored for a long time.
•The indicator lights up when the remaining battery
capacity is low. In this case,the instrument's reliability is
not guaranteed. Replace the battery immediately.
•To avoid battery depletion,turn the rotary switch OFF
after use (the auto power save feature consumes a small
amount of current).
Inspection Before Measurement
•Verify that the instrument operates normally to ensure that no
damage occurred during storage or shipping. If you nd any
damage,contact your authorized Hioki distributor or reseller.
•If damage is suspected,check the section below before contacting
your authorized Hioki distributor or reseller.
(1) Check that the test lead is not broken.
Replace with the specied L9208 Test Lead.
(2) Check that the resistance measurement and continuity test
operates normally.
Have the instrument repaired by the your authorized Hioki
distributor or reseller. The instrument may have been subject to
a voltage of greater than 600 V during resistance measurement
or continuity testing.
(3) Check that the battery voltage is not low.
Replace the batteries.
Maintenance/Inspection
Cleaning
•
Measurements are degraded by dirt on the mating surfaces of the jaw
,so keep the surfaces clean by gently wiping with a soft,dry cloth.
•To clean the device,wipe it gently with a soft cloth moistened with
water or mild detergent.
• Wipe the LCD display gently with a soft,dry cloth.
Insert/Replace Batteries
Necessary tool: Phillips screwdriver and CR2032 Coin-shaped lithium battery
NO
1 2 34
Do not turn the adjustment screw inside the battery case. Doing
so will cause the instrument to report abnormal measured values.
CALIFORNIA,USA ONLY
Perchlorate Material - special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
Warranty
Warranty malfunctions occurring under conditions of normal use in
conformity with the Instruction Manual and Product Precautionary
Markings will be repaired free of charge. This warranty is valid for a
period of three (3) years from the date of purchase. Please contact
the distributor from which you purchased the product for further
information on warranty provisions.
Introduction
Thank you for purchasing the Hioki 3287/3288/3288-20 AC Clamp
Meter. To obtain maximum performance from the instrument,please
read this manual rst,and keep it handy for future reference.
3288 Average value measurement RMS conversion model
3287
3288-20 True RMS measurement model
Safety Notes
This instrument is designed to conform to IEC 61010 Safety
Standards,and has been thoroughly tested for safety prior to
shipment. However,using the instrument in a way not described in
this manual may negate the provided safety features. Before using
the instrument,be certain to carefully read the following safety notes.
Notation
In this document,the risk seriousness and the hazard levels are
classied as follows.
DANGER Imminent risk of operator death or serious injury
WARNING Potential for operator death or serious injury
CAUTION Potential for minor operator injury or device
damage or malfunction
Risk of electric shock
Prohibited actions
Actions that must be performed
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com