Hioki 8715-01 User manual

MEASUREMENT GUIDEMEASUREMENT GUIDE
For...は専用機種。複数の場合は「/」で区切る。不要の場合はとる。
形名を入力。 複数の場合は「/」で区切る。
8715-01
品名を入力。
POWER HiCORDER

Introduction
The 8715-01 POWER HiCORDER measurement guide describes some several measurement
examples. Therefore, the settings may be different from those used in the actual measurements
you perform. In this case, refer to the POWER HiCORDER's instruction manual and modify the
appropriate settings accordingly before performing your measurements.
Also, be sure to read the POWER HiCORDER's instruction manual for information on cautions
to take when connecting the device and performing measurements. Make sure you fully
understand the contents before operating the device. HIOKI cannot be held liable for any
accidents except for those caused by the product.
The following table of contents lists the several different measurement examples as well as
their reference page numbers. All these examples assume only CH 1 is used in the
measurements. Therefore, turn off the waveform display of all other channels. Also,
measurement examples described in this measurement guide assume all other settings use the
factory defaults. No attempt is made to describe settings not directly related to the
measurements. For the information, refer to the 8715-01 POWER HiCORDER's instruction
manual.
Viewing the Voltage Waveform of the 220 V Power Line 1
Viewing the Current Waveform of the Power Line 2
Recording the Passage Current Waveform 3
Recording the Instantaneous Stop Waveform of the 220 V Power Source 4
Recording the Surge Current Waveform when Power is Supplied 5
Recording the RMS Value of the Voltage Variation of the 220V Power Source 6
Recording the RMS value of the Load Variation of the Power Source 7
Monitoring the Leak Current Waveform 8
Recording the RMS Value of the Leak Current 9
Viewing the Waveform of Power Surge/Noise at the 220 V Power Line 10
Viewing the Voltage Waveform of the 440 V Power Line 13
Recording the Abnormal Waveform while Recording the Power Variation
(Instantaneous Waveform Recording) 14
Application Examples on 220 V Line Power Monitoring 15
Viewing the Waveform before the Trigger Point (Pre-Trigger) 16

1
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Voltage Waveform of the 220 V Power Line
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Voltage Waveform of the 220 V Power Line
9197 CONNECTION CORD
9418-15 AC ADAPTER
カード
DMM
Measurement
mode
Selection Move the cursor
Measurement Start
Make sure the power recorder's connecting cord is connected to the power terminal of the
power source. Then, press the [START] key to start the measurement. A voltage waveform
appears on the waveform display screen.
To stop the measurement, press the [STOP] key. After the measurement stops, the voltage
waveform last captured appears on the screen.
When multiple waveforms are captured, they are displayed in the
f
ollowing way according to the sequential save.
To display the waveforms, press the channel select key (CH1). The
displayed number indicates the block in the diagram which appears
in reverse color. This block is the waveform currently displayed on
the waveform display screen.
To change the display, move the flashing cursor to the number's
position or directly to the block's position ( ), then press the /
key to confirm.
If you start the measurement again, the previous waveform is
discarded and recordin
g
starts from the first block.
To view the voltage waveform, set the measurement mode to "Waveform". For the time axis
setting, "2ms" would be quite satisfactory. Set the recording length to "20 DIV".
In the case of voltage waveform, after selecting "Volt" as the input mode, set the range to
"200V Line".

2
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Current Waveform of the Power Line
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Current Waveform of the Power Line
9418-15 AC ADAPTER 9010-10 CLAMP ON PROBE
カード
DMM
Measurement
mode
Selection Move the cursor
Starting a measurement
Clamp the clamp-on sensor to the power
line you want to measure.
Then, press the [START] key to start the measurement.
A current waveform appears on the waveform display screen.
To stop the measurement, press the [STOP] key. After the
measurement stops, the current waveform last captured
appears on the screen.
100 A
To view the current waveform, set the measurement mode to "Waveform". For the time axis
setting, "2ms" would be quite satisfactory. Set the recording length to "20 DIV".
In the case of current waveform, after selecting "clamp sensor" as the input mode, select the
range of the clamp. In this example, "9010" and "100A" are selected. Set the clamp sensor also
to this range.

3
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the Passage Current Waveform
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the Passage Current Waveform
9418-15 AC ADAPTER 9010-10 CLAMP ON PROBE
カード
DMM
Measurement
mode
Selection Move the cursor
Starting a measurement
Clamp the clamp-on sensor to the power
line you want to measure.
Then, press the [START] key to start the measurement.
When the trigger activates, the current waveform appears
on the waveform display screen. After the waveform equal
to the recording length is captured, the state returns to
"wait for trigger".
To stop the measurement, press the [STOP] key. After
the measurement stops, the current waveform last
ca
p
tured a
pp
ears on the screen.
100A
Trigger activates when the current exceeds 100A.
Trigger point
Capturing 20 DIV of
waveform from the point
where the trigger
activates.
200 A
To record the passage current waveform, set the measurement mode to "Waveform". For the
time axis setting, "2ms" would be quite satisfactory. Set the recording length to "20 DIV".
In the case of current waveform, after selecting "clamp sensor" as the input mode, select the
range of the clamp. In this example, "9010" and "200A" are selected. Set the clamp sensor also
to this range.
Set the trigger to "LVL", and then configure the trigger level. If you set the level to "100.0A" and
"", when the current exceeds 100A the trigger activates and then the waveform is captured.
Until the trigger activates, the state remains as "wait for trigger".

4
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the Instantaneous Stop Waveform of the 220 V Power Source
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the Instantaneous Stop Waveform of the
220 V Power Source
9197 CONNECTION CORD
9418-15 AC ADAPTER
カード
DMM
Measurement
mode
Selection Move the cursor
Measurement Start
Make sure the power recorder's connecting cord is connected to the power terminal of the
power source.
Press the [START] key to start the measurement. When the trigger activates, the
instantaneous stop waveform appears on the waveform display screen. After the waveform
equal to the recording length is captured, the state returns to "wait for trigger" again.
To stop the measurement, press the [STOP] key. After the measurement stops, the voltage
waveform last captured appears on the screen.
190.0 Vrms
Trigger point Because the recording length is set to "20
DIV" and the pre-trigger is 50%, 10 DIV of
waveform before and after the trigger point
are captured. For details on pre-trigger,
see page 16.
Trigger activates when
the effective level drops
below 190 Vrms.
To record the instantaneous stop waveform of the power source, set the measurement mode to
"Waveform". For the time axis setting, "2ms" would be quite satisfactory. Set the recording
length to "20 DIV".
In the case of voltage waveform, after selecting "Volt" as the input mode, set the range to
"200V Line". If you set the range to "200V Line", then you can measure the 220V and the 240V
power voltages.
Set the trigger to "DROP". For the measurement frequency, specify one which is suitable for the
region you are located. For the level, set a RMS value so that when the effective power voltage
drops below this value, the trigger will activate. In this example, it is set to "190.0Vr".
To see the waveform before the trigger point, try to set the trigger to "50%". Until the trigger
activates, the state remains as "wait for trigger".

5
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the Surge Current Waveform when Power is Supplied
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the Surge Current Waveform when Power is
Su
pp
lied
9418-15 AC ADAPTER 9010-10 CLAMP ON PROBE
カード
DMM
Measurement
mode
Selection Move the cursor
Starting a measurement
Clamp the clamp-on sensor to the power line
you want to measure.
Press the [START] key to start the measurement.
Turn on the power of the line you want to measure.
When the trigger activates, the current waveform
appears on the waveform display screen.
After the waveform equal to the recording length is captured, the
state returns to "wait for trigger".
To stop the measurement, press the [STOP] key.
After the measurement stops, the current waveform last
captured appears on the screen.
2.5 A
Trigger activates when the current
exceeds 2.5 A.
Because the recording length is
set to "20 DIV" and the pre-
trigger is 50%, 10 DIV of
waveform before and after the
trigger point are captured. For
details on pre-trigger, see page
16.
20 A
To record the waveform of the surge current, set the measurement mode to "Waveform". For
the time axis setting, "2ms" would be quite satisfactory. Set the recording length to "20 DIV".
In the case of current waveform, after selecting "clamp sensor" as the input mode, select the
range of the clamp. In this example, "9010" and "20A" are selected. Set the clamp sensor also
to this range.
Set the trigger to "LVL", and then configure the trigger level. If you set the level to "2.500A" and
"", when the current exceeds 2.5 A the trigger activates and then the waveform is captured.
To see the waveform before the trigger point, try to set the trigger to "50%". Until the trigger
activates, the state remains as "wait for trigger".

6
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the RMS Value of the Voltage Variation of the 220 V Power Source
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the RMS Value of the Voltage Variation of the
220 V Power Source
9197 CONNECTION CORD
9418-15 AC ADAPTER
カード
DMM
Measurement
mode
Selection Move the cursor
Measurement Start
Make sure the power recorder's connecting cord is connected to the power terminal of the
power source.
Press the [START] key to start the measurement. When the measurement starts, the RMS
waveform of the voltage also prints at the same time
To stop the measurement, press the [STOP] key.
250 Vrms
150 Vrms
If the range is "200V Line", the center of the screen (recording paper) becomes 200 Vrms.
200 Vrms
To record the voltage variation of the power source, set the measurement mode to "Trend". For
the time axis setting, initially setting it to "10s" would be easy to understand.
For the measurement frequency, specify one which is suitable for the region you are located.
If print is set to "ON" as well, then printing begins as soon as the measurement starts.
In the case of voltage waveform, after selecting "Volt" as the input mode, set the range to
"200V Line".

7
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the RMS value of the Load Variation of the Power Source
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the RMS value of the Load Variation of the
Power Source
9418-15 AC ADAPTER 9010-10 CLAMP ON PROBE
カード
DMM
Measurement
mode
Selection Move the cursor
Starting a measurement
Clamp the clamp-on sensor to the power line you want to measure.
Press the [START] key to start the measurement. When the measurement starts, the RMS
waveform of the current also prints at the same time
To stop the measurement, press the [STOP] key.
40 Arms
0 Arms
For current measurement of the power variation,
the lowest part of the screen (recording paper) becomes 0A.
20 Arms
To record the load (current) variation of the power source, set the measurement mode to
"Trend". For the time axis setting, initially setting it to "10s" would be easy to understand.
For the measurement frequency, specify one which is suitable for the region you are located.
If print is set to "ON" as well, then printing begins as soon as the measurement starts.
In the case of current waveform, after selecting "clamp sensor" as the input mode, select the
range of the clamp. In this example, "9010" and "20A" are selected. Set the clamp sensor also
to this range.

8
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Monitoring the Leak Current Waveform
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Monitoring the Leak Current Waveform
9418-15 AC ADAPTER
3283 CLAMP ON LEAK PROBE
9445 AC ADAPTER
9094 OUTPUT CORD
9199 CONVERSION ADAPTER
3283 setting
Press the RANGE key of 3283 to set
the range to 100mA. Next, press the
OUTPUT key to display MON
(waveform output: AC).
If the measurement is going to take a
long time, always use the AC
adapter.
3283 LCD
Starting a measurement
Clamp the 3283 to the location where you want to measure the leak current.
Press the [START] key to start the measurement. When the trigger activates, the leak current
waveform appears on the waveform display screen.
After the waveform equal to the recording length is captured, the state returns to "wait for
trigger". To stop the measurement, press the [STOP] key. After the measurement stops, the
current waveform last captured appears on the screen.
20 mA
Trigger activates when the current exceeds 20 mA.
Trigger point
Because the recording length is set to "20 DIV" and
the pre-trigger is 50%, 10 DIV of waveform before
and after the trigger point are captured. For details
on pre-trigger, see page 16.
To record the leak current waveform, set the measurement mode to "Waveform". For the time
axis setting, "2ms" would be quite satisfactory. Set the recording length to "20 DIV".
In the case of leak current waveform, after selecting "3283" as the input mode, select range of
3283. Because it is a leak current, set the range to "100mA".
Set the trigger to "LVL", and then configure the trigger level. If you set the level to "20.00mA"
and " ", when the current exceeds 20 mA the trigger activates and then the leak current
waveform is captured. To see the waveform before the trigger point, try to set the trigger to
"50%". Until the trigger activates, the state remains as "wait for trigger".

9
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the RMS Value of the Leak Current
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the RMS Value of the Leak Current
9418-15 AC ADAPTER
3283 CLAMP ON LEAK PROBE
9445 AC ADAPTER
9094 OUTPUT CORD
9199 CONVERSION ADAPTER
3283 setting
Press the RANGE key of 3283 to set
the range to 100mA. Next, press the
OUTPUT key to display MON
(waveform output: AC).
If the measurement is going to take a
long time, always use the AC
adapter.
3283 LCD
Starting a measurement
Clamp the 3283 to the location where you want to measure the leak current.
Press the [START] key to start the measurement. When the measurement starts, the RMS
waveform of the leak current also prints at the same time
To stop the measurement, press the [STOP] key.
200 mArms
0mArms
For current measurement of the power variation, the lowest part of the screen (recording paper)
becomes 0A.
100 mArms
To record the RMS value of the leak current variation, set the measurement mode to "Trend".
For the time axis setting, initially setting it to "10s" would be easy to understand. For the
measurement frequency, specify one which is suitable for the region you are located. If print is
set to "ON" as well, then printing begins as soon as the measurement starts.
In the case of leak current waveform, after selecting "3283" as the input mode, select range of
3283. Because it is a leak current, set the range to "100mA".

10
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Waveform of Power Surge/Noise at the 220 V Power Line
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Waveform of Power Surge/Noise at the
220 V Power Line
9197 CONNECTION CORD
9418-15 AC ADAPTER
カード
DMM
Measurement
mode
Selection Move the cursor
Measurement Start
Make sure the power recorder's connecting cord is connected to the power terminal of the
power source.
Press the [START] key to start the measurement. When the width exceeds the control range,
the trigger activates and the abnormal waveform appears on the waveform display screen.
After the waveform equal to the recording length is captured, the state returns to "wait for
trigger" again. To stop the measurement, press the [STOP] key.
Trigger point Because the recording length is set to "20
DIV" and the pre-trigger is 50%, 10 DIV of
waveform before and after the trigger point
are captured. For details on pre-trigger, see
page 16.
To record the waveform of power surge/noise, set the measurement mode to "Waveform". For
the time axis setting, "2ms" would be quite satisfactory. Set the recording length to "20 DIV".
In the case of voltage waveform, after selecting "Volt" as the input mode, set the range to "200V
Line".
For the trigger, set it to "JUDGE". Next, set the reference waveform and the control width. In
this example, the reference waveform is set to "Sine50Hz" and the control range to " 25.00V".
The reference voltage is set to "220.0Vr" for the 220V power line.
To see the waveform before the trigger point, try to set the trigger to "50%". Until the trigger
activates, the state remains as "wait for trigger".

11
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Waveform of Power Surge/Noise at the 220 V Power Line
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Measurement data Tri
gg
er
p
oint
Control range0V
0V
Previous Cycle (one cycle)
The control range is applied to this
cycle's waveform
Input Waveform (one cycle)
This cycle is compared with the previous
cycle for detection
Tri
gg
er
p
oint
Waveform Detection Trigger
The waveform detection trigger function compares the measured signal with a pre-defined control
waveform range, and triggers if the signal deviates from the control range voltage. The control
range consists of a positive and negative amplitude offset from either a standard sine wave or the
previous input cycle.
The settable standard voltage and control range are limited by the voltage range setting.
For the waveform (whether it is a sine wave or a previous waveform) functioning as the reference
waveform used in creating the control range, it is a cycle of waveform starting where the cycle
begins from the time 0V is raised and then cut horizontally (raise 0 close).
(1) Sine Wave
The control range is based on a sine wave (with RMS value voltage) at the mains frequency.
(2) Previous Cycle
Set the control range to the waveform of the input waveform in the previous cycle and then
compare it with the input waveform. Use it if you do not want to set up a trigger for the warped
portion when measuring the waveform of the warped power source.
Control range for printing
If the reference waveform used for creating the control range is a sine wave (50/60 Hz), when the
waveform judgement trigger activates and printout results, one cycle of the control range (gray
part) is also printed at the same time as the abnormal waveform.
This is useful when you want to find out how much the deviation from the control range is.

12
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Voltage Waveform of the 440 V Power Line
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Voltage Waveform of the 440 V Power Line
9322 DIFFERENTIAL PROBE
9418-15 AC ADAPTER
9418-15 AC ADAPTER
カード
DMM
Measurement
mode
Selection Move the cursor
Measurement Start
Make sure the power recorder's connecting cord is
connected to the power terminal of the power source.
Then, press the [START] key to start the measurement.
A voltage waveform appears on the waveform display
screen.
To stop the measurement, press the [STOP] key.
After the measurement stops, the voltage waveform last
captured appears on the screen.
The voltage waveform of the 440 V power line can be measured by using the 9322
DIFFERENTIAL PROBE. Set the measurement mode to "Waveform". For the time axis setting,
set it to "2ms". Set the recording length to "20 DIV".
When the 9322 is used, after selecting "9322" as the input mode, set the range to "400V Line".
Set the output switch of the 9322 main unit to "DC".

13
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the Abnormal Waveform while Recording the Power Variation (Instantaneous Waveform Recording)
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the Abnormal Waveform while Recording the
Power Variation
(
Instantaneous Waveform Recordin
g)
Transient
p
henomenon
カード
DMM
Measurement
mode
Selection Move the cursor
200 V
250 V
150 V
190 V
Starting the measurement
Press the [START] key to start the RMS Trend recording
of the power variation. While recording, the trigger
activates when the RMS value drops by 190V.
Because the instantaneous waveform recording length is
set to "20 DIV", when the trigger activates, the waveform
data equal to "20 DIV" is captured, starting from the
trigger point.
If the recording length is 20 DIV, 16 phenomena of
instantaneous waveform data can be captured.
If there are more than 16 phenomena, then data cannot
be captured even if the trigger activates.
During the instantaneous waveform recording in the power variation mode, transient phenomenon
which occurs during the actual value recording of the power variation (voltage or current) can be
recorded at high speed using the waveform level.
In the Basic Setting screen of the RMS Trend Measurement mode, if the instantaneous
waveform recording (Memory Record) is set to "ON", the time axis and recording length settings
become available. The recording length (recording time) determines the number of
instantaneous waveforms to capture.
In this example, the time is set to "2ms" and recording length to "20 DIV". The input setting is
set to "Volt", the input range to "200V Line", and the display position to "Std".
When the instantaneous waveform recording is ON, the instantaneous waveform trigger can be
set at the CH SETTING screen. In this example, the RMS level trigger is used. The trigger
activates when the voltage drops below the voltage set; and the instantaneous waveform is
captured. The RMS value level is set to "190.0 Vr", and the slope to " : falling edge".

14
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Recording the Abnormal Waveform while Recording the Power Variation (Instantaneous Waveform Recording)
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
TREND RMS screen TREND MEM screen
DISP
TREND MEM screen
At the RMS Trend Measurement mode's instantaneous
recording screen, press the channel select key (CH1).
When multiple waveforms are captured, they are displayed in
the following way according to the sequential save.
The displayed number indicates the block in the diagram
which appears in reverse color. This block is the waveform
currently displayed on the waveform display screen.
To change the display, move the flashing cursor to the
number's position or directly to the block's position ( ), then
press the /key to confirm.
Displaying the instantaneous waveform
Press the [STOP] key to end the measurement.
If the instantaneous waveform recording is ON, then when you press the [DISP] key in the
waveform display screen of the RMS Trend Measurement mode, the screen switches between
RMS screen and MEM screen.
The voltage waveform last captured appears on the instantaneous waveform screen.
Switching the instantaneous waveform

15
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Application Examples on 220 V Line Power Monitoring
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Application Examples on 220 V Line Power Monitoring
First, press the SET key to display the Basic Setting screen.
Use the key to set the measurement mode to Waveform
Measurement.
Use the cursor key and the /key to set the time axis to
"5ms" and the recording length to "50 DIV".
CH1 SETTING
Use the CH SET or CH1 key to display the CH1 SETTING
screen. Next, use the cursor key to move the flashing cursor and
then use the /key to change the settings to the ones shown
in the illustration.
CH1 Trigger Setting
Use the voltage drop trigger and change the setting so that when
the volta
g
edro
p
below 190Vrms the tri
gg
er activates.
CH2 SETTING
Use the CH SET or CH2 key to display the CH2 SETTING
screen. Next, use the cursor key to move the flashing cursor and
then use the /key to change the settings to the ones shown
in the illustration.
CH2 Trigger Setting
Use the level trigger and change the setting so that when the
current exceeds 80A the trigger activates.
CH3 SETTING
Use the CH SET or CH3 key to display the CH3 SETTING
screen. Next, use the cursor key to move the flashing cursor and
then use the /key to change the settings to the ones shown
in the illustration.
CH3 Trigger Setting
Use the level trigger and change the setting so that when the
leak current exceeds 30 mA the trigger activates.
Press the [START] key to start the measurement.
If even one of the three channels meets the trigger conditions, the
trigger activates and the waveform is captured.
After the waveform equal to the recording length (20 DIV) is
captured, the state returns to "wait for trigger".
In this example, the 220 V line's voltage waveform, current waveform, and leak current waveform
are monitored simultaneously; when abnormal situation arises they are measured at the same
time. This example uses CH1 to measure the voltage waveform, CH2 the current waveform and
CH3 the leak current waveform.

16
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Waveform before the Trigger Point (Pre-Trigger)
――――――――――――――――――――――――――――――――――――――――――――――――――――――――
Viewing the Waveform before the Trigger Point
(Pre-Trigger)
Tri
gg
er
p
oint
Pre-tri
gg
er: 10%
Pre-tri
gg
er: 50%
Pre-tri
gg
er: 100%
10%
50%
100%
カー
ド
DMM
SelectionMove the cursor
The pre-trigger function serves to record the waveform not only after but also before triggering has
occurred.
In the Waveform measurement mode, using the recording start point as 0% and the recording end
point as 100%, the trigger point can be specified in percent. When all trigger sources are set to
OFF, the pre-trigger setting is invalid.
・In Waveform measurement mode, when the pre-trigger is set, the trigger will not be registered
for a certain period after the start of measurement. (During this interval, Pre-trig wait is shown
on the Waveform display screen.)
・When the trigger can be registered, the indication Wait for trig is shown on the waveform display
screen.
If the recording length is 20DIV, pre-trigger 10% = 2 DIV and pre-trigger 50% = 10 DIV.

HIOKI 8715-01 POWER HiCORDER
Measurement Guide
Publication date: November 2007 Revised edition 2
Edited and published by HIOKI E.E. CORPORATION
Technical Support Section
All inquiries to Sales and Marketing International Department
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL: +81-268-28-0562 / FAX: +81-268-28-0568
E-mail: o[email protected]
URL http://www.hioki.com/
Printed in Japan 8714A983-02
All reasonable care has been taken in the production of this manual, but if you
find any points which are unclear or in error, please contact your supplier or
the Sales and Marketing International Department at HIOKI headquarters.
In the interests of product development, the contents of this manual are subject
to revision without prior notice.
The content of this manual is protected by copyright. No reproduction,
duplication or modification of the content is permitted without the
authorization of Hioki E.E. Corporation.

HEAD OFFICE
81 Koizumi, Ueda, Nagano 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp
URL http://www.hioki.com/
HIOKI USA CORPORATION
6 Corporate Drive, Cranbury, NJ 08512, USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
8714A983-02 07-11H
Printed on recycled paper
Other manuals for 8715-01
3
Table of contents
Other Hioki Measuring Instrument manuals

Hioki
Hioki RM3544 User manual

Hioki
Hioki MEMORY HiCORDER 8855 Manual

Hioki
Hioki MR8740T User manual

Hioki
Hioki 3661-20 User manual

Hioki
Hioki CM4371-50 User manual

Hioki
Hioki 8730-10 User manual

Hioki
Hioki 9605-01 User manual

Hioki
Hioki 3126-01 User manual

Hioki
Hioki SM7420 User manual

Hioki
Hioki PQ3100 User manual

Hioki
Hioki PW3198 User manual

Hioki
Hioki 3196 User manual

Hioki
Hioki TM6101 User manual

Hioki
Hioki HITESTER 3184 User manual

Hioki
Hioki IM3523 User manual

Hioki
Hioki 3390 User manual

Hioki
Hioki 9695-02 User manual

Hioki
Hioki 3169-20 User manual

Hioki
Hioki SAFETY HiTESTER 3258 Instruction manual

Hioki
Hioki PW3390 User manual