TURNING THE TONOMETER ON AND LOADING THE PROBE
Always use wrist strap for safe
and secure use
1. Press the measuring buon.
2. When display shows LoAd, load
the single use probe into the
tonometer as follows:
3. Open the probe tube. Remove the
cap and insert the probe into the
probe base. Be careful not to drop
the probe from the tonometer.
4. Press the measuring buon once.
The tonometer is ready for use
when 00 appears on the display.
Icare TA01i tonometer Quick Guide ENGLISH
MEASUREMENT
ONE SHORT BEEP
Indicates a correct single measurement.
DOUBLE BEEP
Indicates an incorrect measurement.
(See error codes from the instruction
manual).
LONG BEEP
Indicates that six successful measure-
ments are done and IOP result is available
on the display.
1. Ask the paent to relax and look
straight ahead at a specic point
to keep the eye steady during
the measurement.
2. Bring the tonometer near the paent’s
eye. Check that the central groove is in
a horizontal posion. It’s not necessary
to look at the display during the mea-
surement since the device gives sound
signals about errors etc.
3. Keep the distance from the p of the
probe to the paent’s cornea at 4 to 8
millimeters (5/32 - 5/16 inch). If neces-
sary, adjust the forehead support.
4. Press the measuring buon lightly.
The p of the probe should make
contact with the central cornea.
5. Six measurements are made. Once the
six correct measurements are done,
the IOP will be shown on the display.
6. Please refer to the manual for
addional informaon.
Press either selector button until the display
shows ‘End‘. Press the measuring button for
two seconds - the display will show ‘byE’, the
tonometer will switch off and the probe slides
out of the probe base.
If the tonometer is not used for three min-
utes, it will automatically switch off.