I-care HOME User manual

Icare HOME tonometer Instruction manual for health care professionals
ENGLISH
1
www.icare-usa.com
INSTRUCTION
MANUAL
FOR HEALTH CARE
PROFESSIONALS
ENGLISH

Icare®HOME (Model: TA022) INSTRUCTION MANUAL TA022-036 EN-3.2
Icare®HOME (Model: TA022) INSTRUCTION MANUAL TA022-036 USA-2.0
The information in this document is subject to change without prior notice.
In a conict situation the English version prevails.
This device complies with:
Medical Device Directive 93/42/EEC
Canadian Medical Device Regulations
Copyright © 2017 Icare Finland Oy, Made in Finland
0044
TABLE OF CONTENTS
1. Safety instructions ............................................................................................ 3
2. Indications for use............................................................................................. 6
3. Contradictions ................................................................................................... 6
4. Warning .............................................................................................................. 6
5. Precaution .......................................................................................................... 6
6. Introduction ....................................................................................................... 7
7. Package contents .............................................................................................. 7
8. Before you start................................................................................................. 6
9. Setting up the tonometer................................................................................. 9
9.1 Installing or changing the battery ....................................................................9
9.2 Turning the tonometer on ...............................................................................9
9.3 Loading the probe............................................................................................10
10. Using the tonometer....................................................................................... 12
10.1 Choosing the measurement mode................................................................12
10.2 Adjusting the measurement position............................................................13
10.3 Automatic eye recognition..............................................................................14
10.4 Taking the measurements ..............................................................................15
11. Reading the measurement data.................................................................... 17
12. Troubleshooting............................................................................................... 18
14. Cleaning the probe base ................................................................................. 20
15. Cleaning the disinfection................................................................................ 20
16. Accessories ....................................................................................................... 21
17. Lifetime ............................................................................................................ 21
18. Technical and performance data................................................................... 21
18.1 Clinical performance data...............................................................................22
19. Symbols ............................................................................................................ 29
20. Electromagnetic declaration.......................................................................... 30
Appendix 1. Icare home tonometer training procedures for certication of
health care providers (HCP) ............................................................ 34
Appendix 2. Icare home tonometer training procedures for certication of
the patient for self-use .................................................................... 38
Icare Finland Oy
Äyritie 22, FI-01510 Vantaa, Finland
Tel. +358 9 8775 1150, Fax +358 9 728 6670
www.icaretonometer.com, info@icarenland.com
Icare HOME tonometer Instruction manual for health care professionals
ENGLISH
2
www.icare-usa.com

www.icaretonometer.com
3
Icare HOME
1. SAFETY INSTRUCTIONS
WARNING!
Do not push the tonometer into the eye (the tip of the probe should be 4-8mm,
or 5/32-5/16” , from the eye).
WARNING!
Keep the tonometer out of the reach of children, because the probe base, battery
compartment cover and probes are so small that a child could swallow them.
WARNING!
The probe tips are for single-use only, and are packaged sterile.
WARNING!
To prevent contamination, do not touch the bare probe, do not use a probe if it touches
a non-sterile surface like a table or a floor.
WARNING!
The Icare HOME Tonometer is indicated for use only under supervision of a health care
professional.
WARNING!
Patients must be certified and trained for the procedures described in this manual
prior to the device being prescribed for home use.
WARNING!
Health care professionals must inform patients not to modify or discontinue their
treatment plan without receiving instructions from the health care professional.
WARNING!
For cybersecurity do not connect to the USB port except when uploading patient
measurement data using the Icare LINK software or when upgrading the HOME TA022
software using the Icare LINK software. Also the tonometer does not allow you to take
any measurements when the USB is connected. All other traditional cybersecurity
controls (anti-virus software, malware software, separate network for the device, etc.)
do not apply since the device is stand alone, is not networked, and does not contain
operating system software.
WARNING!
Do not change the batteries or probe base when the USB cable is connected.
WARNING!
No modification of this equipment is allowed.
English
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME tonometer Instruction manual for health care professionals
ENGLISH
3
www.icare-usa.com

www.icaretonometer.com
English
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME 4
WARNING!
Federal (US) law restricts this device to sale by or on the order of a physician or properly
licensed practitioner.
WARNING!
Use only the original and certified probes made by the manufacturer. The probes are for
single-use (one per testing session only). Each testing session is defined by one successful
measurement in both eyes (or only one eye measurement if indicated). Use probes taken
only from the intact, original packaging. The manufacturer cannot guarantee sterility of
the probe once the seal is compromised. Re-sterilization or re-use of the probe could
result in incorrect measurement values, in the breakdown of the probe, cross-
contamination of bacteria or viruses, and infection of the eye. Re-sterilization or re-use
will void all responsibilities and liabilities of the manufacturer concerning the safety and
effectiveness of the tonometer.
WARNING!
The probe should be replaced if it contacts or is suspected of having contacted the
eyelashes or eyelid such as might occur during a blink.
WARNING!
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING!
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
PRECAUTION!
•The patient should be instructed to thoroughly wash and dry their hands with soap
and a clean towel prior to handling the probe.
•When you have opened the package, check for any external damage or faults,
particularly for damage to the case. If you suspect that there is something wrong
with the tonometer, contact the dealer who sold the tonometer to you.
•Use the tonometer only for measuring intraocular pressure. Any other use is
improper and the manufacturer is not liable for any damage arising from improper
use, or for the consequences of such use.
•Never open the casing of the tonometer, except for the battery compartment.
•Never allow the tonometer to get wet.
•Do not use the tonometer near flammable substances, including flammable
anesthetic agents.
•Certain microbiological agents (for example, bacteria) can be transmitted from the
forehead or cheek support. To prevent this, clean the forehead and cheek support
for each new patient with disinfectant. See the chapter ‘Cleaning and disinfection’.
www.icaretonometer.com
5
Icare HOME
PRECAUTION!
•The tonometer conforms to EMC requirements (IEC 60601-1-2), but interference
may occur within the tonometer if used near (<1m) a device causing high intensity
electromagnetic emissions, such as a cellular phone. Although the tonometer’s own
electromagnetic emissions are well below the levels permitted by the relevant standards, they
may cause interference in other, nearby devices, for example sensitive sensors.
•If you do not use the tonometer for a long time, remove the batteries, as they may leak.
•Be sure to dispose of the single-use probes properly (for example, in a bin for metal waste).
•Batteries, packaging materials and probe bases must be disposed of according to local
regulations.
•Make sure you use batteries with built-in PTC protection, for example Energizer Lithium Photo
123 3V CR123A.
•Do not cover the eye recognition transmitters or sensor during the measurement, for example
with fingers. Keep your hand, hair etc. and objects such as pillows away from the temple side of
your eye, as they produce an infrared reflection that causes an error.
•The tonometer turns off automatically after 3 minutes if it is not used.
•Do not carry out any other service procedures than those described in the chapters 10-12. Leave
all other service and repairs to the manufacturer or a certified service center.
•Update the tonometer’s time to your local time. It is done automatically by performing the steps
1 and 2 under section 8. Reading the measurement data.
•Icare HOME tonometer is not intended for routine clinical use.
•As the Icare HOME is intended to acquire serial measurements of intraocular pressure over an
extended period of time, individual IOP measurements are not to be considered separate from
their associated IOP curves. This IOP measurement tool is an adjunct to the standard of care
and does not replace the conventional methods used to diagnose and manage patients nor
should it alter the follow-up schedule otherwise indicated for a particular patient.
•Bench testing of the Icare HOME tonometer on model eyes showed that there was no impact on
probe function and performance with 30 repeated uses within a single testing session. The
impact on probe function and performance when the probe is repeatedly used over 30 times is
unknown.
English
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
4
www.icare-usa.com

www.icaretonometer.com
5
Icare HOME
PRECAUTION!
•The tonometer conforms to EMC requirements (IEC 60601-1-2), but interference
may occur within the tonometer if used near (<1m) a device causing high intensity
electromagnetic emissions, such as a cellular phone. Although the tonometer’s own
electromagnetic emissions are well below the levels permitted by the relevant standards, they
may cause interference in other, nearby devices, for example sensitive sensors.
•If you do not use the tonometer for a long time, remove the batteries, as they may leak.
•Be sure to dispose of the single-use probes properly (for example, in a bin for metal waste).
•Batteries, packaging materials and probe bases must be disposed of according to local
regulations.
•Make sure you use batteries with built-in PTC protection, for example Energizer Lithium Photo
123 3V CR123A.
•Do not cover the eye recognition transmitters or sensor during the measurement, for example
with fingers. Keep your hand, hair etc. and objects such as pillows away from the temple side of
your eye, as they produce an infrared reflection that causes an error.
•The tonometer turns off automatically after 3 minutes if it is not used.
•Do not carry out any other service procedures than those described in the chapters 10-12. Leave
all other service and repairs to the manufacturer or a certified service center.
•Update the tonometer’s time to your local time. It is done automatically by performing the steps
1 and 2 under section 8. Reading the measurement data.
•Icare HOME tonometer is not intended for routine clinical use.
•As the Icare HOME is intended to acquire serial measurements of intraocular pressure over an
extended period of time, individual IOP measurements are not to be considered separate from
their associated IOP curves. This IOP measurement tool is an adjunct to the standard of care
and does not replace the conventional methods used to diagnose and manage patients nor
should it alter the follow-up schedule otherwise indicated for a particular patient.
•Bench testing of the Icare HOME tonometer on model eyes showed that there was no impact on
probe function and performance with 30 repeated uses within a single testing session. The
impact on probe function and performance when the probe is repeatedly used over 30 times is
unknown.
English
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
5
www.icare-usa.com

www.icaretonometer.com
6
English
Icare HOME tonometer Instruction manual for health care professionals
2. INDICATIONS FOR USE
The Icare HOME tonometer is a prescription device intended as an adjunct to the
routine clinical monitoring of intraocular pressure (IOP) of adult patients.
3. CONTRAINDICATIONS
Patients with any of the following conditions should not be prescribed the HOME
tonometer:
1) Active ocular infection including infectious conjunctivitis
2) Recent ocular trauma including corneal laceration or corneal/scleral
perforation
3) Inability to demonstrate proficiency with Icare Home tonometer after
training and failure to complete certification procedures. Patients for whom
initial agreement between GAT and HOME measurements in the clinic is
poor should not be prescribed the Icare HOME device for home use.
4) Disabling arthritis or limited motor coordination affecting self-handling of
the Icare tonometer
5) Blepharospasm
6) Nystagmus
4. WARNING
Patients with the following conditions are generally not eligible for use of the HOME
tonometer as they have the potential to introduce unsafe conditions during use or
to impair measurement acquisition:
1) Uncorrected near visual acuity of 20/200 or worse
2) Only one functional eye
3) Poor or eccentric fixation
4) Hearing impairment to the extent that the individual cannot hear and
converse with others without an assistive aid and/or sign language
5) Contact lens use
6) Dry eyes
7) Keratoconus
8) Microphthalmos
9) Buphthalmos
10)
Icare HOME
Cataract extraction within last 2 months
5. PRECAUTION
The safety and effectiveness of the Icare HOME tonometer has not been evaluated for
patients with:
1) High corneal astigmatism >3d
2) History of prior incisional glaucoma surgery or corneal surgery including corneal
laser surgery
3) Corneal scarring
4) Central corneal thickness greater than 0.60mm or less than 0.50mm
5) Known history of difficulty in obtaining Goldmann IOP measurements or any
factors that might contribute to inaccurate Goldmann IOP measurements (e.g. lid
squeezing or tremor)
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
6
www.icare-usa.com

www.icaretonometer.com
7
Icare HOME
English
Icare HOME tonometer Instruction manual for health care professionals
6. INTRODUCTION
The Icare HOME tonometer is a hand-held device for self-use.
The tonometer uses the rebound method. A small and light single-use probe makes
contact with the eye very briefly. The tonometer measures the deceleration of the
probe and the rebound time, and calculates the IOP from these parameters.
A measurement sequence includes six measurements. The probe moves to
the cornea and back during every measurement. As a result, after the six
measurements the tonometer calculates the final IOP and stores it with other
information in the tonometer’s memory, including date, time, eye identification
(right or left) and measurement quality.
The Icare HOME tonometer can record over one thousand measurement results.
You can copy the recorded measurement information to a PC through a USB cable.
7.PACKAGE CONTENTS
WARNING!
Keep the tonometer out of reach of children, because the probe base, battery compartment
cover and probes are so small that a child could swallow them.
PRECAUTION!
When you have opened the package, check for any external damage or faults, particularly for
damage to the case. If you suspect that there is something wrong with the tonometer, contact
the dealer who sold the tonometer to you.
The package contains:
• Icare HOME tonometer
• 10 sterilized single-use probes
• 2 batteries
•USB memory stick including
the instruction manual for health
care professionals and the Icare LINK
software
• USB cable for connecting the Icare
HOME tonometer to a PC with Icare
LINK software
• Instructions for downloading the Icare
LINK software
• Patient guide
• Support position tags
• Warranty card
• Carrying case
•Wrist strap
•Probe base cleaning container
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
7
www.icare-usa.com

www.icaretonometer.com
8
Icare HOME
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
1. Probe base incorporating
indicator light
2. Eye recognition transmitter
3. Eye recognition sensor
4. Cheek support
5. Forehead support
6. Measurement button
7. Forehead support
position indicator
13. Measurement button
14. LOAD light
15. MEASURE light
16. REPEAT light
17. DONE light
18. SERVICE light
19. BATTERY light
20. POWER button
10. Battery cover
11. Silicone lid (USB
cover)
12. Type label
8. Forehead support
adjustment wheel
9. Cheek support adjustment
wheel
FRONT PARTS TOP PARTS
BACK PANELBOTTOM PARTSSIDE PARTS
1.
6.
7.
3.
5.
4.
2.
8.
10.
13.
14.
15.
16.
17.
18.
19.
20.
11.
12.
9.
Icare HOME tonometer Instruction manual for health care professionals
8. BEFORE YOU START
Find the main parts, buttons and indicator lights of the tonometer in the below gures.
English
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
8
www.icare-usa.com

www.icaretonometer.com
9
PRECAUTION!
Make sure you use batteries with built-in PTC protection, for example Energizer Lithium
Photo 123 3V CR123A.
Update the tonometer’s time to your local time. It is done automatically by performing
the steps 1 and 2 under section 8. Reading the measurement data.
Lift the silicone lid that protects the USB port and
keeps the battery compartment cover in place. Open
the battery compartment cover by pressing the
silicone lid slightly and sliding the battery
compartment cover as shown in the figure left.
1. Siliconelid
2. Battery cover
Insert two CR123A lithium batteries in the correct
order: (+) end upwards as shown in the figure left.
Close the cover firmly and press the silicone lid in
place to cover the USB port.
9.2 TURNING THE TONOMETER ON
PRECAUTION!
The tonometer turns o automatically after 3 minutes if it is not used.
Press the power button (20) to turn the tonometer on. The lights (14-19) are
turned on briey. Following a brief pause, the Load light ashes on the back
panel to remind the user to load the single-use probe into the tonometer prior to
measurement.
English
Opening the silicone lid and battery cover
Inserting new batteries
1.
2.
2 x CR123A
Icare HOME tonometer Instruction manual for health care professionals
9. SETTING UP THE TONOMETER
Setting up your Icare HOME tonometer is easy, with few steps. The following
subchapters describe how you get started.
9.1 INSTALLING OR CHANGING THE BATTERY
Icare HOME
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
9
www.icare-usa.com

www.icaretonometer.com
10
Icare HOME
PRECAUTION!
The patient should be instructed to thoroughly wash and dry their hands with soap and a
clean towel prior to handling the probe.
WARNING!
To prevent contamination, do not touch the bare probe, do not use a probe if it
touches a non-sterile surface like a table or a floor.
WARNING!
Use only the original and certified probes made by the manufacturer. The probes are for
single-use (one per testing session only). Each testing session is defined by one successful
measurement in both eyes (or only one eye measurement if indicated). Use probes taken
only from the intact, original packaging. The manufacturer cannot guarantee sterility of
the probe once the seal is compromised. Re-sterilization or re-use of the probe could
result in incorrect measurement values, in the breakdown of the probe, cross-
contamination of bacteria or viruses, and infection of the eye. Re-sterilization or re-use
will void all responsibilities and liabilities of the manufacturer concerning the safety and
effectiveness of the tonometer.
The Icare HOME tonometer uses single-use probes that are packed in a plastic
tube (container) and wrapped in blister packs as shown in the figures below.
After using the tonometer to measure IOP, the user can remove the probe
either by powering off the device and allowing the probe to slide out of the
probe base or by pulling the probe out of the probe base. The probe can be
discarded by placing the probe back in its container and disposing of it properly.
English
Icare HOME tonometer Instruction manual for health care professionals
9.3 LOADING THE PROBE
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
10
www.icare-usa.com

www.icaretonometer.com
11
English
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOA
EASU
LO
EAS
1. Unwrap the probe. 2. Remove the lid of the
probe container as
shown in the gure
above. Point the
tonometer upward.
3. Drop the probe into
the probe base (1)
by turning the probe
container upside down.
4. Press the measurement
button (13) briefly (1
second) to activate the
probe.
5. The probe moves rapidly
back and forth.
6. See that the Measure
light (15) ashes. If so,
the probe is loaded
correctly and ready for
measurement.
To load the probe:
1X
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
11
www.icare-usa.com

www.icaretonometer.com
12
Icare HOME
Series mode
The series mode is especially
useful in self-tonometry. In the
series mode, keeping
the button pressed down
(see figure left) initiates the
measurement function, and
the tonometer takes six rapid
measurements one after the
other to obtain the final IOP
reading. The button must be
pressed for at least 3 seconds.
Single mode
You can use the single mode to
take individual measurements
one at a time. The single mode
is especially useful for those
patients who tend to blink
heavily. Here you press the
measurement button briefly
(1 second) for each of the six
measurements to obtain the
final IOP reading (see figure
left).
Series mode, press the measurement button constantly.
Single mode, press the measurement button briey
(1 second) for each of the six measurements.
English
Icare HOME tonometer Instruction manual for health care professionals
10. USING THE TONOMETER
10.1 CHOOSING THE MEASUREMENT
MODE
The tonometer can operate in two modes:
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
12
www.icare-usa.com

www.icaretonometer.com
13
To adjust the measurement position for your patient:
1. Adjust the supports using the adjustment wheels as shown in the above gure.
2. Keep the probe horizontal and pointing perpendicularly to the center of the cornea.
3. Set the distance between the tip of the probe and the center of the cornea to be
4-8mm (5/32-5/16”) as shown in the below gure.
4. Read the distance setting (forehead A•1, A•2, etc.,
cheek B•1, B•2, etc.) between the arrows on the scale
(7) of the supports , see the figure left, and write it
down on a support position tag for the patient.
5. Do the same for the other eye as well unless only
one eye needs monitoring.
6. Verify that the support positions are correct each
and every time the patient comes in for a clinic visit.
The tonometer has two
adjustable supports (4-5), one
for the forehead and one for the
cheek, as shown in the gure left.
The supports are for ensuring
accurate measurement distance
and alignment.
Tonometer forehead and cheek supports with adjustment wheels.
English
Icare HOME tonometer Instruction manual for health care professionals
10.2 ADJUSTING THE MEASUREMENT POSITION
WARNING!
Do not push the tonometer into the eye (the tip of the probe should be
4-8mm, or 5/32-5/16”, from the eye).
Icare HOME
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
13
www.icare-usa.com

www.icaretonometer.com
14
Icare HOME
PRECAUTION!
Do not cover the eye recognition transmitters or sensor during the measurement, for example
with ngers. Keep your hand, hair etc. and objects such as pillows away from the temple side
of your eye, as they produce an infrared reection that causes an error.
The tonometer includes an automatic eye recognition system that identies which
eye, right or left, you are measuring. The system has two infrared LED transmitters
just below the probe base and one infrared LED sensor above the probe base, as
in the gure below. The right-hand transmitter sends invisible infrared light to the
right and the left-hand transmitter to the left. The infrared light reects from your
nose to the sensor. The sensor knows from which transmitter the reected infrared
light came, and thus which eye you are measuring. The resulting eye indication is
included in the data that you can transfer to a PC, as described in the section 9.
Eye recognition system components.
1. Left and right infrared transmitter.
2. Infrared sensor.
EYE RECOGNITION COMPONENTS
2.
1. 1.
English
Icare HOME tonometer Instruction manual for health care professionals
10.3 AUTOMATIC EYE RECOGNITION
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
14
www.icare-usa.com

www.icaretonometer.com
15
WARNING!
Do not push the tonometer into the eye (the tip of the probe should be 4-8mm, or 5/32-5/16”, from
the eye).
WARNING!
The probe should be replaced if it contacts or is suspected of having contacted the eyelashes or
eyelid such as might occur during a blink.
The probe will make a gentle and brief contact with the eye when you take the
measurement. No topical anesthetic is needed. The recommended frequency of
measurements is 3-4 daily with a maximum of 5-6.
To measure intraocular pressure:
1. Check that the Measure light (15) still ashes on the back panel.
2. If the Measure light does not ash, press the power button (20) and wait until
the Measure light illuminates again.
3. The patient should look straight ahead at a specic point while keeping eyes
wide open as shown in the below gure.
4. Bring the tonometer near the eye, the probe pointing perpendicular to the
center of the cornea without a vertical or horizontal tilt. The position is correct
when the probe base light is green and appears symmetrically in the center of
the patient’s view. See the below gures.
Correct head and eye position.
Correct measurement position.
Incorrect head and eye position.
Incorrect measurement position.
Incorrect head and eye position.
Incorrect measurement position.
English
90°
Icare HOME tonometer Instruction manual for health care professionals
10.4 TAKING THE MEASUREMENTS
Icare HOME
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
15
www.icare-usa.com

www.icaretonometer.com
16
Icare HOME
5. Press the measurement button:
Single mode:
Press the button briefly (1 second) and you hear a
short beep, repeat it to take one measurement at
a time till you hear a long beep and see the Done
light (17) illuminated on the back panel.
Series mode:
Keep the measurement button down to obtain
the sequence of six measurements till you hear a
long beep and see the Done light illuminated on
the back panel. The button must be pressed for
at least 3 seconds.
6. If both eyes are measured repeat steps 1-5 using your other eye.
7. If an error occurs, press the Measure button briefly (1 second) and continue the
measurement. See also section 9 Trouble shooting.
8. Press the power button for three seconds to turn the tonometer off. The probe slides
out of the probe base.
9. Remove and discard the used probe properly.
Correct measurement position.
Correct alignment of the
tonometer.
Incorrect measurement position.
Incorrect alignment of the
tonometer.
Incorrect measurement position.
Incorrect alignment of the
tonometer. Readjust so that you see
only the front of the tonometer and
the green light symmetrically in the
center of your view.
English
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
16
www.icare-usa.com

www.icaretonometer.com
17
English
More information about Icare LINK software
http://www.icaretonometer.com/products/icare-link/
LINK
Connected
WARNING!
For cybersecurity do not connect to the USB port except when uploading patient
measurement data using the Icare LINK software or when upgrading the HOME TA022
software using the Icare LINK software. Also the tonometer does not allow you to take any
measurements when the USB is connected. All other traditional cybersecurity controls (anti-
virus software, malware software, separate network for the device, etc.) do not apply since the
device is stand alone, is not networked, and does not contain operating system software.
The tonometer stores information on every complete measurement sequence of six
measurements. The stored information includes the calculated nal eye pressure
reading in mmHg, time and date of the measurement, identication of the eye (right
or left) and the quality level of the measurement.
Uploading is easy:
1. Start Icare LINK software in your PC.
2. Connect the tonometer to the PC using the USB cable. The
Load and Measure lights will flash. If no lights flash or the
Service and Battery lights flash, reconnect the USB cable.
3. The internal clock of the tonometer is automatically
updated to the PC’s time by the Icare LINK software at this
point.
4. Copy the data to a selected patient in the Icare LINK
software.
Icare HOME tonometer Instruction manual for health care professionals
11. READING THE MEASUREMENT DATA
Icare HOME
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
17
www.icare-usa.com

www.icaretonometer.com
18
Icare HOME
The tonometer automatically monitors and controls the measurement position and
speed of the probe during the measurements, and indicates errors with sounds
and lights. The following table instructs you in error situations and explains what
the different lights and sounds mean. The indicator lights are also presented in the
figure after the table below.
Error light Error
sound
Reason Action
Battery No. Battery is soon
empty.
Prepare to change batteries.
Battery light is ashing No. Battery is empty. Change batteries.
Probe base light is solid red. No. Too much vertical
tilt.
Press the measurement button again to clear the
error message.
Position the tonometer horizontally so that the
probe base light is green.
Probe base light is flashing red
and Measure light turns off.
Two long
beeps.
a) Probe is too far
from or too near
the eye.
b) Probe
movement was not
perpendicular to
the cornea.
Press the measurement button again to clear the
error message.
a) Set the correct distance 4-8mm (5/32-5/16”)
between the probe tip and the center of the cornea.
b) Set the probe perpendicular to the center of the
cornea.
Repeat light is ashing and
Probe base light is ashing red.
Two long
error
beeps.
a) Too much IOP
deviation during
the measurement,
because the user
did not keep the
tonometer stable.
b) Eye was not
recognized.
Press the measurement button again to clear the
error message.
a) Repeat the measurement
b) Do not move the tonometer during the
measurements, remove your hand or fingers from
the infrared transmitters and sensor, move the
patient’s hair away from his/her temple side of the
eye. If the eye still does not get recognized contact
the seller to arrange sending the device for service.
Service light is ashing and
Probe base is ashing red.
Two long
beeps.
Incorrect or dirty
probe or probe
base.
Change the probe, clean or change the probe
base or contact the seller to arrange sending
the device for service.
DONE
SERVICE
BATTERY
BATTERY
REPEAT
SERVICE
English
Icare HOME tonometer Instruction manual for health care professionals
12. TROUBLESHOOTING
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
18
www.icare-usa.com

www.icaretonometer.com
19
OFF INTIALIZE LOAD
PROBE
MEASURE REPEAT
DONE SERVICE LOW BATTERY EMPTY
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
English
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
19
www.icare-usa.com

www.icaretonometer.com
20
Icare HOME
DONE SERVICE LOW BATTERY EMPTY
BATTERY
Icare HOME tonometer Instruction manual for health care professionals
13. REPLACING THE PROBE BASE
Replace the probe base every twelve months. Replace or clean the probe base if the
Service light flashes.
Instructions for replacing the probe base:
• Turn off the tonometer.
•Unscrew the probe base collar and put it in a safe place.
• Remove the probe base by tilting the tonometer downwards and use
your fingers to pull the probe base out of the tonometer.
•Insert a new probe base into the tonometer.
• Screw the collar in, to lock the probe base.
14. CLEANING THE PROBE BASE
You can reuse the probe base after careful cleaning. Clean the probe base every six
months. Clean or replace the probe base if the Service light flashes.
Instructions for cleaning the probe base:
• Fill the probe base cleaning container or other clean container with
70-100% isopropyl alcohol.
•Turn the power off.
• Unscrew the probe base collar.
•Invert the probe base over the container, drop in the probe base into
the container and let soak for 5-30 minutes.
•Remove the probe base from alcohol.
• Dry the probe base by blowing clean canned or compressed air into
the hole in the probe base. This will additionally remove possible
residual dirt.
• Insert the probe base into the tonometer.
•Screw the collar in, to lock the probe base.
15. CLEANING AND DISINFECTION
You must clean the forehead and cheek supports for each new patient. Use a wipe
dampened with a 70-100% isopropyl alcohol solution. Do not immerse the
tonometer in water or other liquid. The tonometer must not be immersed or
cleaned using too much water.
PRECAUTION!
Do not carry out any other service procedures than those described in the chapters
10-12. Leave all other service and repairs to the manufacturer or a certified service
center.
English
Icare HOME tonometer Instruction manual for health care professionals ENGLISH
20
www.icare-usa.com
Other manuals for HOME
1
Table of contents
Other I-care Medical Equipment manuals

I-care
I-care Icare HOME TA022 User manual

I-care
I-care HOME2 User manual

I-care
I-care IC333 User manual

I-care
I-care ic200 User manual

I-care
I-care IC222 Guide

I-care
I-care TA01i User manual

I-care
I-care Icare HOME TA022 User manual

I-care
I-care EXPORT TS03 User manual

I-care
I-care Pro TA03 User manual

I-care
I-care HOME2 User manual
Popular Medical Equipment manuals by other brands

OncoTherm
OncoTherm EHY-2000plus user manual

Otto Bock
Otto Bock 3R78 Operator's Quick Reference Guide

Dräger
Dräger D-Vapor Instructions for use

Medical Freezer
Medical Freezer LOTA System operating manual

Future Mobility Healthcare
Future Mobility Healthcare AQUA REHAB COMMODE TL operating instructions

ergoline
ergoline ers2 System manual