8
Configuration Variables
Hidden option, key [9] from the main menu screen, provides the installer a series of configuration
options not visible to the operator. Use the [+] and [-] keys to scroll through the options. The
current value is displayed, and a new value may be entered at the cursor. To store a new value,
press the [ENTER] key. Starting from the beginning, theseare:
MAX CLEAN ATTEMPTS: Number of auto-cleaning cycles in an autozero/clean process before the
operation ceases and flashes a warning message.
MAX ZEROING TIME: Number of seconds the instrument will wait for a stable reading from the
zeroing solution. This is normally set at 120 seconds.
CLEANING PUMP TIME: Number of seconds the cleaning solution is pumped during an autozero/clean
cycle. Normally set to 30 seconds.
ZEROING PUMP TIME: Number of seconds the zeroing solution is pumped during an autozero/clean
cycle. Normally set to 60 seconds.
SAMPLE REFILL TIME: Number of seconds required to fully turn over the cell with fresh process
sample. This time does not include any sample line distance between the instrument and the process.
Normally set to 30 seconds.
CLEAN SOAK TIME: Number of seconds to let the cleaning solution sit in the cell during an
autozero/autoclean cycle. Normally set to 60.
AutoClean FLUSH TIME: Number of seconds to flush the cell with process following the clean
soak. Normally set to 30 seconds.
# of DIGITS after DP: Number of digits displayed after the decimal point - typically set to 1. NOTE: Increasing
the number of digits displayed does not increase the accuracy of the analyzer. Additionally, the resulting,
apparent “variations” in readings may be perceived as erratic operation.
SEND INTERVAL: Minutes and seconds between reports over the serial channel. Default is 1 minute.
ON: ZC4 ZERO5 PUMP6: Controls the ZeroCleanValve, the ZERO valve and the Pump. Pressing key
[4] turns ON the ZeroClean Valve (ZC1); pressing [1] turns OFF the ZeroClean Valve. [5] turns On
the Zero Valve, while [2] turns it OFF. [6] turns ON the Pump, [3] turns it OFF. Not listed on the
screen, [7] turns on alarm relay one; [8] turns on alarm relay two. [CLEAR] turns everything off.
CLEAN IF %T is BELOW: If the transmittance measured during an auto-clean cycle is adequate, we
can omit the cleaning portion, saving on cleaning solution and maintenance. This value is the trigger
point for the auto-clean cycle and is normally set at 95. If organics (slimes) or other contaminants
accumulate on the flow-cell windows, it is recommended to increase this value to 97 to ensure the cell
is cleaned on every auto-zero cycle.
WAIT FOR STABLE HEAD TEMPERATURE: If yes, the analyzer will wait for the source LED to be within
0.05 degrees C of optimum before taking readings. Press [.] (decimal point) to toggle.
SERIAL SETUP MENU: Set baud rate, parity, enable Modbus, set Modbus address
INIT LOG: Press [CLEAR] to clear the internal data log. This may be used if the monitored process had
been shut down for a time or the instrument had been idle for an extended period, making the current log
information irrelevant. Also displayed is the actual number of used log entries. If desired, the log output can
be triggered (to a printer, for example) by pressing the [0] key.
MAX LOG REPORTS: Limits the number of reported (via serial channel) logged events to the most
recent N, where N is between 1 and 15,000. The maximum 15,000 events are always logged, but this
option allows limiting the report size to a lower value. Press [5] to Restore FACTORY DEFAULTS:
Restores settings to factory originals. NOTE: Any user settings will be over-written.
CELL FACTOR: Observe or modify cell factor (effective path length – default is 0.80)
CODE REVISION DATE: The date the software was updated.