inbody BPBIO320 User manual

User’s Manual

Please note the important information below before reading this manual.
Failure to observe these precautions can result in personal injury or equipment damage.
Warning
Failure to comply with safety precautions can damage the equipment.
Caution
Referring to notes can help improve equipment use.
Caution
©2019 InBody Co., Ltd. All rights reserved. BM-ENG-82-F-190710
InBody Co., Ltd. [HEAD OFFICE]
InBody Bldg., 625, Eonju-ro, Gangnam-gu, Seoul 06106 KOREA
TEL: +82-2-501-3939
FAX: +82-2-578-2716
Website: https://www.inbody.com
E-mail: info@inbody.com
©2019 InBody Co., Ltd. All rights reserved.
Reproduction, adaptation, or translation of this manual is prohibited without prior written consent from InBody Co., Ltd.
under the copyright laws. This manual may be printed incorrectly and subject to change without notice. InBody
Co., Ltd. shall not be liable for any errors, incidental, or consequential damages that occurred by not complying
with the content of the User᾽s Manual.
Visit our website (www.inbody.com) to view and download additional information about the BPBIO320/320n. InBody Co.,
Ltd. reserves the right to modify the appearance, specifications, etc. of this product to improve its quality, without prior notice.
Representative & Sponsor Information
InBody Europe B.V.
Gyroscoopweg 122, 1042 AZ, Amsterdam, The Netherlands
TEL: +31-20-2386080 Website: https://nl.inbody.com E-mail: info.eu@inbody.com
Australian Sponsor. [AUSTRALIA]
Emergo AUSTRALIA. Level 20, Tower II, Darling Park, 201 Sussex Street, Sydney, NSW 2000, AUSTRALIA
TEL: +61-2-9006-1662 FAX: +61-2-9006-1010 Website: https://www.emergogroup.com E-mail: [email protected]
Customer Service Information
InBody [USA]
13850 Cerritos Corporate Dr., Unit C, Cerritos, CA 90703, USA
TEL: +1-323-932-6503 FAX: +1-323-952-5009 Website: https://www.inbody.com E-mail: contact@inbody.com
InBody Japan Inc. [JAPAN]
Tani Bldg., 1-28-6, Kameido, Koto-ku, Tokyo 136-0071 JAPAN
TEL: +81-3-5875-5780 FAX: +81-3-5875-5781 Website: https://www.inbody.co.jp E-mail: inbody@inbody.co.jp
InBody China. [CHINA]
904, Xing Di Plaza, No. 1698 Yishan Road, Shanghai, 201103, CHINA
TEL: +86-21-64439738, 9739, 9705
FAX: +86-21-64439706
Website: https://www.inbody.com
E-mail: [email protected]

User’s Manual Contents
I. BPBIO320/320n Installation
A. Product Components
B. Operating Environment
C. Installation Instructions
D. Loading and Reloading Printer Paper
(BPBIO320 only)
E. Replacing Cuff Cover
F. Setup
G. Connecting to InBody
H. Connecting to Coin Device(InCoin)
I. Precautions for Maintenance
II. Blood Pressure Test
A. Precautionary Steps
B. Test Instructions
C. Test Posture
III.
Transportation and Storage
A. Cautions during Transportation
B. Repacking Instructions
C. Transportation and Storage Environment
IV. FAQ
A. Error Code
B. Regarding the Equipment
C. Regarding the Test
V. Others
A. Exterior and Functions
B. Safety Information
C. Product Classication
D. Specications
E. EMC Information (Declaration)
4
4
5
9
11
13
18
19
20
21
21
24
25
26
26
27
27
30
32
33
33
34
35

4 5
B. Operating Environment
Please make sure the environment is adequate for the BPBIO320/320n installation. This equipment is designed
for indoor use. If installing the equipment outdoors, the following requirements must be fullled:
Temperature range 10 to 40ºC (50 to 104ºF)
Relative humidity 30 to 75% RH
Atmospheric pressure range 70 to 106 kPa
I.BPBIO320/320n Installation
A. Product Components
The BPBIO320/320n consists of the following components. Make sure the following components are present.
*Please check each component of the system for damage prior to installation.
Basic Components
BPBIO320/320n
Power cable (AC 250V 10A 1.8m) 1 EA
Information Panel
Two rolls of printing paper (BPBIO320 only)
Two cuff covers (One installed)
User’s Manual
Optional components
Desk for BPBIO320/320n, and two screws
Chair for BPBIO320/320n
Coin device(InCoin)
BPBIO320/320n
User’s Manual
BPBIO320/320n
User’s Manual
BPBIO320/320n
User’s Manual

4 5
C. Installation Instructions
1. Open the BPBIO320/320n box, and remove the air cushion (small size). Unfold the air cushion (large size) and
take out the BPBIO320/320n from the box, and then remove the protective box. Take off the protective vinyl
out of the BPBIO320/320n, and remove the packing material.
•Do not carry the equipment by holding the cuff.
•Keep packing material to repack the BPBIO320/320n at a later time. Dispose of trash according to local laws and
regulations.
•Refer to the figure below when transporting the BPBIO320/320n.
Caution
Air cushion
(large size)
Air cushion
(small size)
Packing material
Protective box
Package protective vinyl

6 7
2. Place the equipment on a solid table that can withstand the weight of the BPBIO320/320n (approx. 9kg).
If desk for the BPBIO320/320n is available, align the grooves on the bottom of the BPBIO320/320n
with the grooves on the desk and secure them with two xing screws.
*Desk for the BPBIO320/320n sold separately.
3. Mount the Information Panel onto the BPBIO320/320n.
4. Connect the power cable provided with the BPBIO320/320n. Plug the power cable into a grounded 3-prong outlet.
*The anti-theft groove can be used to secure the equipment to the table.
Desk xing grooves
Desk grooves
Fixing screws
•Install the BPBIO320/320n on a flat, vibration-free floor.
If the equipment is installed where the floor is not flat, it may topple during a test or the test results may be
inaccurate.
Caution
Anti-theft groove
3-socket outlet
Grounded outlet

6 7
5. Use the anti-theft groove to prevent theft.
*Theft protection cable is sold separately.
•Do not install the equipment where the power cable is difficult to disconnect.
•Do not plug or disconnect the power cable with wet hands. There is a risk of electric shock.
•Always use an outlet supplied to the correct power supply (AC 100-240 V). An incorrect power supply may
cause fire or malfunctions.
•Do not disassemble the equipment arbitrarily. This may cause electric shock, injury, product malfunction, and/
or inaccurate results. Unauthorized disassembly will void the manufacturer’s warranty.
•Do not let the BPBIO320/320n touch with other electronic devices when on. This may cause electric shock.
Warning
•Plugging the device into an ungrounded outlet may cause malfunction or electric shock.
Caution
6. Turn the BPBIO320/320n on by pressing the power switch located on the back of the device.
7. When the BPBIO320/320n is turned on, the following screen is displayed and a calibration is automatically
performed.
Power Switch
•Do not operate or touch the device while it is calibrating. It may cause errors to future tests.
Caution

8 9
8. Adjust the volume by turning the volume knob on the rear of the BPBIO320/320n.

8 9
D. Loading and Reloading Printer Paper (BPBIO320 only)
1. Loading Printer Paper
1) Cut off the end of the printer paper with scissors. If the end of the paper is ripped, it may cause a
paper jam or damage the printer head.
2) Open the cover by pressing ‘PUSH’ while the BPBIO320/320n is turned on.
3) If the device is being used for the rst time, remove the protective paper located inside.
*The protective paper (thermal printing paper) is in the printer groove. Pull upward to remove it.
4) Fold one corner of the provided printer paper and insert it in the direction as shown.
Paper cut with scissors
Ripped paper
감열지를
윗방향으로
잡아당겨
제거후
사용하세요
•The printer will not work if the paper is inserted incorrectly.
Caution

10 11
5) Insert the end of the paper through the feeder. The printer will automatically pull and cut the paper.
6) Remove the cut paper, and close the printer paper cover.
2. Reloading Printer Paper
1) Open the cover by pressing ‘PUSH’ mark on the bottom center of it while the BPBIO320/320n is turned
on.
2) Press button to remove the remaining paper.
3) Reload a new roll of printer paper by following the instructions for ‘Loading Printer Paper'.
•Remove all trash from previous printer paper rolls.
Caution
•Two rolls of printer paper are provided with the BPBIO320/320n.
Note

10 11
E. Replacing Cuff Cover
The cuff cover of the BPBIO320/320n should be replaced periodically for sanitary purposes. Follow the guide
below to replace the cuff cover.
1.
Remove the front cover of the BPBIO320/320n by pushing downward. Loosen the screw with a screwdriver.
2.
Loosen the two screws on the rear of the BPBIO320/320n with a screwdriver.
3.
Remove the rings on the front and rear of the BPBIO320/320n by
pulling the arrow mark and pull the cuff cover to remove it.
4.
Place the seam of the new cuff cover down
as shown, and align it with the front and rear ring.
Screws
Rear ring
Front ring
Screws
Front Cover
•Blood pressure results may be inaccurate if the seam is not face down.
Caution

12 13
Front Cover
Rear ring
Front ring
•Two cuff covers are provided with the BPBIO320/320n. (One is already mounted in the equipment.)
•Additional cuff covers can be purchased at https://shop.inbodyusa.com.
Note
5.
Secure the new cuff cover onto the grooves on the front and rear of the BPBIO320/320n.
6.
Attach the previously removed front and rear rings on the BPBIO320/320n
and secure them with screws.
7.
Close the front cover of the BPBIO320/320n.

12 13
F. Setup
Setup consists of sound settings, motion sensor settings, elbow sensor settings, Result Sheet printing, time
stamps and cautionary comments.
1.
Settings
1) When button and button are pressed and held at the same time for 2 seconds while on the home
screen, the rst screen of the Setup, 'Sound Setting' is displayed.
2) Use UP (▲) and DOWN (▼) buttons to change the settings.
* UP (▲) and DOWN (▼) are located beneath the cover of the device.
3) Press button to move to the next screen. Set values are automatically saved.
4) Press and hold button for 2 seconds to return to the home screen.
*The measurement value display time setting is applied when the blood pressure monitor restarts.
•Contents of results sheet output are only applicable to BPBIO320.
Note

14 15
The following is an example showing how to change a ‘3-line manual output’ Result Sheet setting.
While on the home screen, press and hold button and button
at the same time for 2 seconds.
Press button to move to screen.
Press button to select .
Press and hold button for 2 seconds to save changes and return to
the home screen.
▼
▼
The following is an example showing how to change the screen timeout duration.
While on the home screen, press and hold button and button
at the same time for 2 seconds.
Press button several times to move to screen.
Press button to select .
Press and hold button for 2 seconds to save changes and return
to the home screen.
▼
▼

14 15
2.
Detailed description of Setup
When in the Setup menu, the following information is displayed in the Systolic, Diastolic and Pulse sections of
the screen.
Function
Display
Description
Systolic Diastolic Pulse
Sound
Voice guide
Beep guide
Disable speaker
Outputting
results sheet*
(BPBIO320 only)
3-line automatic printing
3-line + graph automatic output
3-line manual output
Hour Number flashes on bottom left of Time section,
allowing the hour to be set
Minute Number flashes on bottom left of Time section,
allowing the minute to be set
Year Year setting (between 2000 and 2099)
Month Set month (between 01 and 12)
Day Set day (between 01 and 31)
Screen timeout
duration
Display the test result value for 5 seconds
Display the test result value for 10 seconds
Display the test result value for 20 seconds
Display the test result value for 2 minutes
Outputting
caution comment*
(BPBIO320 only)
The caution comment is printed on the results sheet.
The caution comment is not printed on the
results sheet.
*Outputting results sheet setting ( ) and Outputting caution comment setting ( ) are applicable only to BPBIO320,
so they do not appear at BPBIO320n.

16 17
* Today Count : 0002
* Total Count : 00075
2017/12/06 09:12
3. Blood Pressure Evaluation
1) While on the home screen, press and hold button and UP (▲) button at the same time for more
than 2 seconds.
2) When 111 is displayed in the systolic blood pressure column, diastolic blood pressure column, and pulse
column as shown below, the blood pressure evaluation has been disabled.
3) When 222 is displayed in the systolic blood pressure column, diastolic blood pressure column, and pulse
column as shown below, the blood pressure evaluation has been turned on.
*Blood pressure evaluation comment type: Optimal, Prehypertension, Stage 1 Hypertension, Stage 2 Hypertension
4) In order to change the blood pressure evaluation settings, press and hold button and
UP (▲) button at the same time for more than 2 seconds.
* Today Count : 0002
* Total Count : 00075
2017/12/06 09:12
2017/12/06 19:12
Systolic
Diastolic
Pulse
110
75
96
2017/12/06 19:12
Systolic
Diastolic
Pulse
110
75
96
30 120
< >
2017/12/06 19:12
Systolic
Diastolic
M.A.P
P.P
Pulse
R.P.P
110
75
87
35
96
10560
30 120
< >
2017/12/06 19:12
Systolic
Diastolic
M.A.P
P.P
Pulse
R.P.P
110
75
87
35
96
10560
*Test result
- Optimal
Please consult with a
physician for diagnosis
and treatment.
*Test result
- Optimal
Please consult with a
physician for diagnosis
and treatment. *Test result
- Optimal
Please consult with a
physician for diagnosis
and treatment.
*Test result
- Optimal
Please consult with a
physician for diagnosis
and treatment.
2017/12/06 19:12
Systolic
Diastolic
Pulse
110
75
96
2017/12/06 19:12
Systolic
Diastolic
Pulse
110
75
96
30 120
< >
2017/12/06 19:12
Systolic
Diastolic
M.A.P
P.P
Pulse
R.P.P
110
75
87
35
96
10560
30 120
< >
2017/12/06 19:12
Systolic
Diastolic
M.A.P
P.P
Pulse
R.P.P
110
75
87
35
96
10560
*Test result
- Optimal
Please consult with a
physician for diagnosis
and treatment.
*Test result
- Optimal
Please consult with a
physician for diagnosis
and treatment. *Test result
- Optimal
Please consult with a
physician for diagnosis
and treatment.
*Test result
- Optimal
Please consult with a
physician for diagnosis
and treatment.
For 3-line + graph
automatic output
For 3-line automatic
output

16 17
4.
Displaying test counts
1) While on the home screen, hold the DOWN (▼) button for more than 2 seconds. Test counts
for the day will be displayed in the diastolic blood pressure column. Test counts for the day and total test
counts are printed.
2) When DOWN (▼) button is pressed again, the total test counts are displayed in the diastolic blood
pressure column.
Test counts for the day Total test counts Test count printout
* Today Count : 0002
* Total Count : 00075
2017/12/06 09:12
* Today Count : 0002
* Total Count : 00075
2017/12/06 09:12
* Today Count : 0002
* Total Count : 00075
2017/12/06 09:12

18 19
G.Connecting to InBody
If the BPBIO320/320n connects to InBody products, the blood pressure test results can be transferred to the InBody.
*Only compatible with other InBody products
*Blood pressure values can be printed on the InBody Result Sheet.
*Serial cable is sold separately.
1.
Turn off the BPBIO320/320n.
2.
Connect the serial cable to the RS232C (9-pin serial) connection terminal located on the back of the BPBIO320/320n.
Connect the other end of the serial cable to the serial connection terminal on the InBody device.
*Refer to the InBody manual to nd the location of the serial connection terminal.
3.
Turn on the BPBIO320/320n.
4.
Devices will connect automatically.
Blood Pressure Monitor
serial connection terminal
Example: InBody770
•Refer to the InBody product manual to print the blood pressure values on the InBody Result Sheet.
Note

18 19
H.Connecting to a Coin Device(InCoin)
*The coin device is an optional purchase.
By connecting the coin device to the BPBIO320/320n, the measurement can be conducted only when the
set monetary value is received.
1. Prepare the enclosed RS232C cable from the coin device.
2. Connect the cable to the rear of the coin device and the RS232C connector on the rear of the BPBIO320/320n.
3. Connection between BPBIO320/320n and coin device is complete.
*Please check the “Technical Manual” of the coin device for detailed assembly and setup.
9-pin serial port

20 21
I. Precautions for Maintenance
Check the following before turning the equipment on.
• Make sure the equipment was not damanged from impact.
• Make sure the equipment is not contaminated.
• Make sure the equipment is not wet.
• Make sure the power cord is completely plugged in.
• Make sure the power cord is not damaged.
• Make sure the power cord is not a tripping hazard.
The BPBIO320/320n calibrates automatically when it is turned on. But it may be necessary for a specialist to inspect
the device once every two years to maintain its performance and safety. If an error message appears during the self-
test, please contact product support at (323)932-6503 × 2.
•Never move the equipment while the power is on.
•Do not let foreign objects, like dust and liquid to enter the device. Foreign objects can damage electric
components.
•Clean the exterior of the device gently with a lint-free cloth once a week. Do not scratch the LED screen
while cleaning the equipment.
•Replace the cuff cover when it becomes dirty or worn out.
•Packing material and other waste should be disposed of according to the laws and regulations.
Caution
This manual suits for next models
1
Table of contents
Other inbody Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

marcel aubert
marcel aubert MA 173-240-5 Series Instruction

Gigahertz Solutions
Gigahertz Solutions HFE35C manual

Vanguard Instruments
Vanguard Instruments IRM-5000P operating instructions

Kathrein
Kathrein MSK 200 manual

GW Instek
GW Instek GPM-8310 user manual

RF SOLUTIONS
RF SOLUTIONS LoRaLyser quick start guide