inbody S10 User manual

User’s Manual
InBody S10 User's Manual for Measurement Guide and Setup
Thank you for purchasing the InBody S10. This user’s manual describes all the features of the InBody S10.
Please read before use and keep it in a safe place. By following the manual instructions, you will be able to
use the InBody S10 more safely and eectively.
S10_manual_ENG_G_220317.indd 1S10_manual_ENG_G_220317.indd 1 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

Intended Purpose
The device should be used as an adjunct for clinical decision making and is not intended to diagnose or treat any
diseases.
InBody S10
in mainly used for healthy and acute or chronically ill populations in hospitals, medical practices
and inpatient care facilities in accordance with national regulations. It can be used to assist in the assessment of
nutritional status, obesity and muscle balance. Body composition analysis is important in preventive medicine since
it provides the basis of appropriate physical activity and dietary habits for improving personal daily routine. It can be
also usefully applied to follow-up studies of patients treated for various diseases.
Intended Use and Medical Indication
• Medical check-up: Four body composition analysis can be identied for the risk of developing diseases that are highly related
to body composition imbalance like obesity, malnutrition, uid imbalance and osteoporosis for medical check-up
• Obesity: Percent body fat has been recommended rather than BMI to ensure proper weight loss and improvements in long-
term health, tracking changes for adjusting/developing customized treatments
• Pediatric obesity: Body composition measurement is an essential part of health assessments for children and adolescents.
Percent Body fat is better than the indicators of weight status to identify children and adolescents with unfavorable lipid
prole.
• Sarcopenia: InBody provides a quick, easy to perform test that provides a calculation for skeletal muscle index (SMI),
the sum of the lean mass in the arms and legs, normalized for height. This marker is useful in identifying low muscle in the
appendages, which increases frailty risk.
• Diabetes & endocrinology: Diabetes is often associated with excess fat, however having insucient muscle mass is just as
detrimental and increases diabetes risk. And visceral fat plays a key role in the development of metabolic and cardiovascular
disease.
• Edema: Over-hydration as assessed by ECW ratio(ECW/TBW) is prevalent in dialysis patients, and is associated with loss
of residual renal function, inammation, malnutrition and hypertension. Monitoring the ECW ratio (ECW/TBW) provides
an assessment of uid accumulation in the extracellular space resulting from compromised cardiovascular function. The
patients who did not have ascites originally but have higher ECW/TBW had a higher incidence of ascites in liver cirrhosis.
• Segmental uid retention: InBody objectively measures each region of the body separately and provides segmental ECW
ratio measures for each of the arms, legs and the trunk, and these measures can be used to detect uid imbalances resulting
from the development or progression of lymphedema.
• Nutrition: The four primary components of the nutritional assessment are summarized by the mnemonic ABCD, with A
standing for anthropometric measurements including stature, body weight, BMI and body composition. Body composition
analysis can reveal changes in body composition (body water, protein, minerals and body fat) that cannot be known by
changes in body weight.
• Fitness: Strength training greatly stimulates muscle growth, exercise burn the calories strengthens cardiorespiratory capacity,
which reduce the risk of diabetes, heart disease, and other health concerns and result in the various changes in body
composition. Body composition analysis shows skeletal muscle mass and lean in each segment of body, it helps focusing on
building more muscle or correct imbalance.
* InBody S10 is not a diagnostic device. To make an accurate diagnosis, the physician needs to commission thorough
examinations and take their results into account in addition to the results of the InBody S10.
* The InBody S10 is not used in home healthcare environment.
S10_manual_ENG_G_220317.indd 2S10_manual_ENG_G_220317.indd 2 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

Please note the important information below before reading this manual.
Failure to comply with safety warnings and regulations can cause serious injury or death.
DANGER
NOTECAUTION
참 조 주 의참 고 주 의
Warning
Failure to comply with safety cautions and regulations can cause injury or property damage.
Caution
InBody Co., Ltd. [MANUFACTURER]
625, InBody Bldg., Eonju-ro, Gangnam-gu, Seoul 06106 Republic of Korea
TEL: +82-2-501-3939 FAX: +82-2-578-5669 Website: inbody.com E-mail: info@inbody.com
InBody Co., Ltd. [HEAD OFFICE]
15, Heugam-gil, Ipjang-myeon, Seobuk-gu, Cheonan-si, Chungcheongnam-do 31025 KOREA
TEL: +82-41-581-3003 FAX: +82-41-581-3103 Website: inbody.com
E-mail: info@inbody.com
©2022 InBody Co., Ltd. All rights reserved.
Reproduction, adaptation, or translation of this manual is prohibited without prior written consent from InBody Co., Ltd.
under the copyright laws. This manual might have typographical errors, and its content can be changed without a
prior notice. InBody Co., Ltd. shall not be liable for any errors, incidental, or consequential damages that occurred
by not complying with the content of the User’s Manual.
Visit our website inbody.com to view and download further information about the functions of the InBody S10, the
explanation of results output, and more. InBody Co., Ltd. reserves the right to modify the appearance, specifications,
and etc. of the InBody S10 to improve the quality of the product, without prior notice for reasons of performance
improvement.
Representative & Sponsor Information
InBody Europe B.V.
Gyroscoopweg 122, 1042 AZ, Amsterdam, The Netherlands
TEL: +31-20-238-6080 FAX: +31-6-5734-1858 Website: nl.inbody.com E-mail: info.eu@inbody.com
Sponsor Australia [AUSTRALIA]
Emergo AUSTRALIA. Level 20, Tower II, Darling Park, 201 Sussex Street, Sydney, NSW 2000, AUSTRALIA
Customer Service Information
InBody USA [USA]
13850 Cerritos Corporate Dr. Unit C Cerritos, CA 90703 USA
TEL: +1-323-932-6503 FAX: +1-323-952-5009 Website: inbodyusa.com E-mail: info.us@inbody.com
InBody Japan [JAPAN]
Tani Bldg., 1-28-6, Kameido, Koto-ku, Tokyo 136-0071 Japan
TEL: +81-3-5875-5780 FAX: +81-3-5875-5781 Website: www.inbody.co.jp E-mail: inbody@inbody.co.jp
InBody China [CHINA]
903/904, XingDiPlaza, No.1698 YiShanRoad, Shanghai 201103 China
TEL: +86-21-64439705
FAX: +86-21-64439706
Website: inbodychina.com
E-mail: [email protected]
Referring to notes can help improve equipment use.
Note
DANGER
NOTECAUTION
참 조 주 의참 고 주 의
©2022 InBody Co., Ltd. All rights reserved. IM-ENG-83-G-220317
S10_manual_ENG_G_220317.indd 3S10_manual_ENG_G_220317.indd 3 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

S10_manual_ENG_G_220317.indd 4S10_manual_ENG_G_220317.indd 4 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

I. Installation & Maintenance
A. Contents of the Box
B. Exterior & Functions
C. Installation Instructions
D. Transportation
E. Repacking
F. Maintenance
II. Management & Results Description
A. Cautions Before Measurement
B. Exterior and Functions of Keypad
C. Measurement Posture
D. Connecting the Electrodes
E. Power Connection & Getting Started
F. Home Screen
G. Measurement Settings
H. Personal Prole
I. How to Operate the Equipment
J. Results
III.
Setup Establishment
A. Setup
B. Setup Menu
C. Database
I V.
Problems & Solutions
A. Error Messages
B. Troubleshooting
C. Frequently Asked Questions (FAQ)
V. Other
A. Results Sheet
B. Accessories
C. Safety Information
D. Classication
E. Specications
F. EMC Declaration
6
8
10
19
20
21
22
23
27
33
36
37
38
39
40
42
44
45
47
49
49
52
53
53
61
63
63
65
User’s Manual Contents
S10_manual_ENG_G_220317.indd 5S10_manual_ENG_G_220317.indd 5 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

I. Installation & Maintenance
A. Contents of the Box
When opening the box, make sure all of the following items are inside:
1. Included Items
❶
InBody S10 1EA
❷
User’s Manual 1EA
❸
Adapter (12V, 3.4A/3.34A) 1EA
❹
Power cord (AC 125V, 10A, 1.8m) 1EA
❺
Electrode cable 4EA
❻
InBody carrying case 1EA
❼
Touch Type hand electrode 4 units
❽
Touch Type foot electrode 2 units
❾
Adhesive Type electrode 8 units
❿
EKG adhesives
⓫
USB Thumb Drive
⓬
Cart 1EA
⓭
IB Battery 120 or IB Battery 220 1EA
⓮
Cable (75cm) 1EA for connecting
IB Battery 120/220 to the unit
4
❻
4
❼
4
❽
4
❿⓫
4
4
❾
❶
4
4
❹
4
❷
4
❸
4
❺
⓬
⓮
⓭
<IB Battery 120> <IB Battery 220>
S10_manual_ENG_G_220317.indd 6S10_manual_ENG_G_220317.indd 6 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

•
To prevent physical shock, use InBody’s packing material when shipping or transporting the equipment.
Refer to this Chapter I, Section D: “Transportation.”
Caution
DANGER
NOTECAUTION
참 조 주 의참 고 주 의
•
Save the wrapping material after unpacking in the event of relocation.
Note
S10_manual_ENG_G_220317.indd 7S10_manual_ENG_G_220317.indd 7 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

8
9
•
Do not dismantle the equipment or open the back cover. Internal parts are not for customer use and may cause
electric shock. If the equipment is dismantled, the warranty is void, and service costs will be charged.
Warning
DANGER
NOTECAUTION
참 조 주 의참 고 주 의
B. Exterior & Functions
Individual part identication and functions with schematic sketches are provided below. Please inspect each
component of the InBody S10 before installation to ensure there are no scratches or damage.
1. LCD Monitor
2. Electrode Connection Unit
3. Operational Keypad
4. Control & Connection Unit
4
1. LCD Monitor
4. Control & Connection Unit
3. Operational Keypad
2. Electrode
Connection Unit
S10_manual_ENG_G_220317.indd 8S10_manual_ENG_G_220317.indd 8 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

8
9
1. LCD Monitor
Touch-screen LCD. This displays the analysis procedure, messages and results.
2. Electrode Connection Unit
Connects electrode cables into the InBody S10’s internal circuit.
3. Operational Keypad
The keypad is divisible into input buttons and function buttons.
The buttons are used to input data required for body composition analysis, set up the operating environment
or to print out test results.
4. Control & Connection Unit
Connects to peripherals such as a PC or a printer for data transmission.
❶Power Switch
Powers the InBody S10 on/o.
❷Power Input Port
Use to connect the power adapter.
❸ 9 Pin Serial Port, Female (RS-232C)
Use to connect optional devices.
SD400(Serial Distributor) provided by InBody is for the connection of several optional devices.
❹USB Host Port
Use to interface with a USB printer or a USB storage device.
❺ USB Slave Port
Use to connect with a PC.
❶
❷
❸
❹❺
•
Do not drop any food or liquid on the equipment. It may affect the electrical parts in the equipment or cause damage.
•
Only use the adapter provided by InBody.
Warning
DANGER
NOTECAUTION
참 조 주 의참 고 주 의
•
When you use the adapter cable, insert the adapter cable tightly into the InBody S10.
•
Including the optional equipment, only the peripherals provided by InBody can be connected to the InBody S10.
For any inquiry about peripherals, contact InBody.
Note
S10_manual_ENG_G_220317.indd 9S10_manual_ENG_G_220317.indd 9 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

10
11
C. Installation Instructions
1. Workplace Requirements
Location: Indoors only. Any outdoor area where the equipment is to be located should meet all the conditions below.
2. For InBody Carrying Case Users
(1) When opening the box, check to make sure all the following items are inside.
(2) Connect the electrode cables to the InBody S10.
Connect electrodes of RA(Right Arm), LA(Left Arm), RL(Right Leg), LL(Left Leg) to each
corresponding part of the InBody S10. InBody S10 oers two types of electrodes: the Touch Type
electrodes and Adhesive Type electrodes.
(1) Operation Condition
Temperature range 10 ~ 40˚C (50 ~ 104˚F)
Relative humidity 30 ~ 75%
Atmospheric pressure range 70 ~ 106kPa
(3) Connect it to an outlet with a grounding terminal. Equipment may be damaged or malfunction due to
electrical shock, which may result in inaccurate test results.
(4) If InBody S10 experiences electrical interference, the test results may be inaccurate. Do not install near
uorescent lamps, large AC motor equipment (running machines, refrigerators, air conditioners, compressors),
high-frequency thermal therapy devices, and electric heaters.
(5) Do not connect multiple peripherals to the same outlet. If it is plugged into a power outlet such as an electrical
appliance, disconnect it and plug it into a dierent power outlet.
(2) Adapter
Manufacturer BridgePower Corp.
Mean Well Co, Ltd
Model BPM040S12F07
GSM 40A12-P1IR
Power Input
AC 100 ~ 240V, 50/60Hz, 1.2A(12A-0.6A)
AC 100-240V, 50/60Hz, 1.0-0.5A
Power Output DC 12V, 3.4A
DC 12V, 3.34A
S10_manual_ENG_G_220317.indd 10S10_manual_ENG_G_220317.indd 10 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

10
11
Installing Touch Type Electrodes:
4
4
Installing Adhesive Type Electrodes:
4
4
S10_manual_ENG_G_220317.indd 11S10_manual_ENG_G_220317.indd 11 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

12
13
3. How to Install the InBody S10 Cart
(1) Please place the InBody S10 on the cart. Align the InBody S10’s 4 screws on the rear side of the unit to
the cart’s 4 grooves on the head unit. Screw them to fasten the InBody S10 onto the cart.
❶Touch Type
Connect the black cable to the black port(V) of the electrode module, and connect the red cable to the
red port(I) of the electrode module.
<hand electrode> <foot electrode>
❷ Adhesive Type
Connect the black cable to the black port(V) of the electrode module, and connect the red cable to the
red port(I) of the electrode module.
IV
S10_manual_ENG_G_220317.indd 12S10_manual_ENG_G_220317.indd 12 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

12
13
(2) As shown below, adjust the direction and angle as desired.
*HELP
❶ For left/right control
Loosen the screw by twisting left. Fasten the screw by twisting right.
❷ For up/down control
Loosen ❷ and adjust the angle degree. Fasten ❷again.
❸ When ❷is too tight or loose
If unscrewed, ❷becomes loose. If screwed, ❷becomes tight.
❷
❶
❸
S10_manual_ENG_G_220317.indd 13S10_manual_ENG_G_220317.indd 13 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

14
15
(3) After installation, you can lock the wheels by pushing down the locking lever(as illustrated below).
<Caster unlocked> <Caster locked>
(4) Connect the electrode cables to the InBody S10.
Connect the RA(Right Arm), LA(Left Arm), RL(Right Leg), LL(Left Leg) electrodes to each
corresponding part of the InBody S10. The InBody S10 oers two types of electrodes:
the Touch Type electrodes and Adhesive Type electrodes.
Installing the Touch Type Electrodes:
4
4
S10_manual_ENG_G_220317.indd 14S10_manual_ENG_G_220317.indd 14 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

14
15
Installing the Adhesive Type Electrodes:
4
4
❶Touch Type
Connect the black cable to the black port(V) of the electrode module, and connect the red cable to the
red port(I) of the electrode module.
<hand electrode> <foot electrode>
❷ Adhesive Type electrode
Connect the black cable to the black port(V) of the electrode module, and connect the red cable to the
red port(I) of the electrode module.
IV
S10_manual_ENG_G_220317.indd 15S10_manual_ENG_G_220317.indd 15 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

16
17
•
Do not place the InBody S10 in a location making it difficult to disconnect the power cord.
•
Do not plug in or pull out the power cord with wet hands. There is a risk of an electric shock.
•
Always use an outlet with a voltage range of 100 - 240. Using outlets outside this range may result in fire
or malfunction.
•
When using a power surge protector, make sure that the outlet or the extension cable has adequate power capacity.
•
Do not disassemble or modify the equipment, including internal parts, without written consent from the
manufacturer. This may cause electric shock or injury, product malfunction, inaccurate results, and will
void the manufacturer’s warranty.
•
If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued safe
use of equipment.
•
Do not directly contact the InBody S10 with any other electronic device when the InBody S10 is on. This may
result in an electric shock.
Warning
DANGER
NOTECAUTION
참 조 주 의참 고 주 의
•
If you have any problems installing your InBody S10, please contact InBody for assistance.
•
If the InBody S10 is not plugged into a grounded outlet, damage due to electric surges or product malfunctions
may occur. This may affect the test results.
•
The test results may be inaccurate if the InBody S10 is under electrical interference. Do not install the InBody S10
near products that generate electrical interference such as fluorescent lights, large AC motor equipment(treadmill,
vibration plate, refrigerator, air-conditioner, compressor, etc.), high-frequency thermal therapy equipment, or
heating appliances. Do not share the power source of the InBody S10 with other electrical devices. This may
affect the test results.
•
When connecting the InBody S10 with other test equipment, turn on the other equipment first. When turning off
other equipment, turn off the InBody S10 first. This is necessary to minimize electrical surges to the InBody S10.
•
Always use the specified adapter provided by InBody, as it is a part of the InBody S10. Using other adapters may
result in a malfunction of the InBody S10.
•
Operation of the InBody S10 2,000m above sea level may affect the weight measurement.
•
Adapter must be arranged so the power may easily be cut off when a problem occurs in the InBody S10.
Caution
DANGER
NOTECAUTION
참 조 주 의참 고 주 의
S10_manual_ENG_G_220317.indd 16S10_manual_ENG_G_220317.indd 16 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

16
17
❷ Install the printer platform on the cart pole.
*Installing the Printer onto the Cart
❶ Insert a power strip in the slot provided under the printer platform.
❸ Use a hex key to fasten the printer platform into place.
S10_manual_ENG_G_220317.indd 17S10_manual_ENG_G_220317.indd 17 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

18
19
❹ Use the USB cable to connect the printer to the unit.
S10_manual_ENG_G_220317.indd 18S10_manual_ENG_G_220317.indd 18 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

18
19
D. Transportation
If it must be transported, be extra careful to ensure safe handling. The following are some tips for safely
transporting the InBody S10.
- Before transporting the InBody S10, turn o the power switch and unplug the adapter.
- Be careful not to damage the hand module.
- Be careful not to damage the electric cables.
1. Environmental Requirements
2. Transporting Before Installation
Before installation, the InBody S10 will be shipped in a box designed by InBody.
3. Transporting After Installation
To prevent physical shock, use InBody᾿s packing material when shipping or transporting the equipment.
Storage environment
Temperature range -10 ~ 70˚C (14 ~ 158˚F)
Relative humidity 10 ~ 80%RH (No condensation)
Atmospheric pressure range 50 ~ 106kPa
•
Severe physical impact can cause damage. Be careful not to drop or shake the InBody S10.
Warning
DANGER
NOTECAUTION
참 조 주 의참 고 주 의
S10_manual_ENG_G_220317.indd 19S10_manual_ENG_G_220317.indd 19 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57

20
21
E. Repacking
Be sure to turn o the power switch and unplug the power cable before repacking. Be careful to avoid severe
physical shock, jarring or other damage while repacking, especially with regard to the electrode cables and
touch type hand/foot electrodes.
(1) Turn o the power switch.
(2) Remove all cables connected to the InBody S10.
(3) Put the separated units in the provided carrying case.
(4) You can carry the machine easily with the carrying case.
S10_manual_ENG_G_220317.indd 20S10_manual_ENG_G_220317.indd 20 2022. 3. 23. 오후 4:572022. 3. 23. 오후 4:57
Other manuals for S10
1
Table of contents
Other inbody Measuring Instrument manuals

inbody
inbody BPBIO320 User manual

inbody
inbody InBody270 User manual

inbody
inbody J30 User manual

inbody
inbody PUSH User manual

inbody
inbody BSM 370 User manual

inbody
inbody BSM270 User manual

inbody
inbody BPBIO210 User manual

inbody
inbody BSM 370 User manual

inbody
inbody BSM170 User manual

inbody
inbody 720 User manual