
3 4
Code
Application
Thickness requirement
Internal diameter
of flat workpiece
ISHW-B70 ISHW-B75 ISHW-B92
requirement of tube
workpiece
Wall thickness
requirement of
tube workpiece
Hardness range
Min.reading
Accuracy
Dimension
for soft brass for soft steel for hard
and copper and cold-rolled
0.6~8mm
0.6~6mm 0.6~6mm
>Ø10mm >Ø6mm >Ø10mm
0.6~6mm 0.6~6mm0.6~8mm
0~20HW
1HW
±0.5HW(at 5~17HW)
205x30x85mm
aluminum alloy
and hard brass steel
Indenters mode
Standard delivery
Main unit
Spare indenter
Hardness test block
Wrench
1pcs
1pcs
1pcs
1pcs
The Webster hardness tester have different mode of indenters.
A. ISHW-L20、B. ISHW-B75 C. ISHW-B92
Screwdriver 1pcs
Calibration sheath(ISHW-L20A )only 1pcs
Verification and calibration
Basic operation
Put the specimen between the anvil and the indenter cylinder, then
press the handle down until the indicator hand come to a stop. Keep
the handle pressed and take the reading.
1. Calibration of the full scale
Press the handle to the bottom without putting any specimen on
the anvil of the hardness tester. Regulate the full-scale adjustment
screw with a small screwdriver to bring the indicator at 20. If the
reading is smaller than 20, the adjustment screw should be regulated
clockwise; if the reading is greater than 20, the adjustment screw
should be regulated anticlockwise. If the pointer can not be set to
1Verification
The operator should verify the accuracy of the tester regularly.
Calibration should be made if the tester is found inaccurate. When
testing on the hardness block, use its upper surface only.
1. Verification of the full scale
Press the handle to the bottom without putting any specimen on
the anvil of the hardness tester. The indicator should point at 20
±0.5HW. If the reading is out of tolerance, full scale calibration
should be carried out.
2. Verification of the test block
Test on the standard test block. For ISHW-L20, ISHW-L20A, IS
HW-L20B testers, the reading should be ±0.5HW of the hardness
number marked on the block. For ISHW-B70, ISHW-B92 testers,
the reading should be 5HW±0.5HW; and for ISHW-B75, the reading
should be 17HW±0.5HW. If the reading is out of tolerance, load
spring calibration should be carried out.
2Calibration
The calibration of the Webster hardness tester includes full scale
calibration and load spring calibration. Full scale calibration is to
set a benchmark, and load spring calibration is to provide a reference
point for the tester. Before calibration of the ISHW-L20A, put the
calibrating sheath on the anvil first.
ISHW-L20A
ISHW-L20B