KERN 572 User manual

KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Tel: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Operating instructions
Precision/ compact balances and
platform scales
KERN 572 / 573 / KB / DS / FKB
Version 7.4
2017-10
GB
572/573/KB/DS/FKB -BA-e-1774

572/573/KB/DS/FKB-BA-e-1774 2
GB
KERN 572 / 573 / KB / DS / FKB
Version 7.4 2017-10
Operating instruction
Precision/compact balances and platform scales
Table of contents:
1MODE - MENUE.......................................................................................................4
2Technical data.........................................................................................................6
2.1 KERN 572............................................................................................................................................. 6
2.2 KERN 573............................................................................................................................................. 9
2.3 KERN KB ........................................................................................................................................... 10
2.4 KERN DS............................................................................................................................................ 14
2.5 KERN FKB ......................................................................................................................................... 18
3Fundamental information (general).....................................................................21
3.1 Intended use...................................................................................................................................... 21
3.2 Inappropriate use.............................................................................................................................. 21
3.3 Monitoring the test substances ...................................................................................................... 21
4Fundamental safety information..........................................................................22
4.1 Observe the information in the operating instructions................................................................ 22
4.2 Staff training...................................................................................................................................... 22
5Transport and storage..........................................................................................22
5.1 Acceptance check ............................................................................................................................ 22
5.2 Packaging.......................................................................................................................................... 22
6Unpacking, installation and commissioning ......................................................23
6.1 Place of installation, place of use................................................................................................... 23
6.2 Unpacking ......................................................................................................................................... 23
6.2.1 Installation ........................................................................................................................................... 23
6.3 Mains supply..................................................................................................................................... 23
6.4 Battery operation FKB...................................................................................................................... 24
6.5 Connecting peripheral equipment.................................................................................................. 24
6.6 Initial start-up.................................................................................................................................... 24
6.7 Adjustment........................................................................................................................................ 24
6.8 Adjusting (go to 7.2.6)...................................................................................................................... 25
6.8.1 Adjusting for verification (KERN 573; FKB-M; KB-NM; DS-M) ........................................................... 25
6.9 Verification ........................................................................................................................................ 26

3 572/573/KB/DS/FKB-BA-e-1774
7Operation...............................................................................................................27
7.1 Display control panel ....................................................................................................................... 27
7.2 Operation........................................................................................................................................... 29
7.2.1 Weighing with Tare ............................................................................................................................. 29
7.2.2 Count – Selection reference piece ..................................................................................................... 29
7.2.3 Percent % ........................................................................................................................................... 29
7.2.4 Formula Weighing ............................................................................................................................... 29
7.2.5 Tolerance Weighing ............................................................................................................................ 29
7.2.6 Adjust .................................................................................................................................................. 31
7.2.7 Preselect Tare .................................................................................................................................... 31
7.2.8 Auto Tare active .................................................................................................................................. 31
7.2.9 Filter / Speed ...................................................................................................................................... 31
7.2.10 Auto Off active ................................................................................................................................ 31
7.2.11 Variable Factor ............................................................................................................................... 31
7.2.12 Preselect Unit ................................................................................................................................. 31
7.2.13 Animal weighing: ............................................................................................................................ 33
7.2.14 Swap:.............................................................................................................................................. 33
7.3 Display background illumination.................................................................................................... 35
7.4 Interface RS 232 C ............................................................................................................................ 35
7.5 RS 232 C Data output via interface RS 232 C ................................................................................ 36
7.5.1 Description of the data transfer .......................................................................................................... 37
7.5.2 Numerator ........................................................................................................................................... 37
7.6 Printer ................................................................................................................................................ 37
7.7 Underfloor weighing......................................................................................................................... 38
8Maintenance, upkeep, disposal ...........................................................................39
8.1 Cleaning............................................................................................................................................. 39
8.2 Maintenance, upkeep ....................................................................................................................... 39
8.3 Disposal............................................................................................................................................. 39
9Troubleshooting....................................................................................................40
10 Declaration of conformity.....................................................................................40

572/573/KB/DS/FKB-BA-e-1774 4
1 MODE - MENUE
Models 572 / KB / DS / FKB non-verifiable:
Important!
The modified settings - as well as the adjustment - must be stored when switching-off, via the
ON/OFF button.

5 572/573/KB/DS/FKB-BA-e-1774
Models 573-NM; KB-NM; DS-M; FKB-M:
Important!
The modified settings - as well as the adjustment - must be stored when switching-off, via the
ON/OFF button.

572/573/KB/DS/FKB-BA-e-1774 6
2 Technical data
2.1 KERN 572
KERN 572-30 572-31 572-32 572-33
Readability (d) 0,001 g 0,001 g 0,001 g 0,01 g
Weighing range (max) 240 g 300 g 420 g 1.600 g
Taring range (subtractive) 240 g 300 g 420 g 1.600 g
Reproducibility 0,001 g 0,002 g 0,002 g 0,01g
Linearity ±0,003 g ±0,005 g ±0,005 g ± 0,03 g
Smallest part weight for piece
counting under laboratory condi-
tions
1 mg 1 mg 1 mg 10 g
Smallest part weight for piece
counting under normal condi-
tions
10 mg 10 mg 10 mg 100 mg
Recommended adjusting weight
F1 (not supplied) 200 g 200 g +100 g 200 g + 200 g 1 kg + 500 g
Adjustment points
50 g
100 g
200 g
240 g
50 g
100 g
200 g
300 g
100 g
200 g
300 g
400 g
0,5 kg
1,0 kg
1,5 kg
1,6 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h 2 h 4 h 2 h
Housing (B x D x H) [mm] 180 x 310 x 90
Vibration filter yes
Weighing plate stainless steel
[mm] Ø 106 Ø 106 Ø 106 Ø 150
Balance
Input voltage 12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 2,3
Interface RS232

7 572/573/KB/DS/FKB-BA-e-1774
KERN 572-35 572-37 572-39 572-43
Readability (d) 0,01 g 0,01 g 0,01 g 0,1 g
Weighing range (max) 2.400 g 3.000 g 4.200 g 10.000 g
Taring range (subtractive) 2.400 g 3.000 g 4.200 g 10.000 g
Reproducibility 0,01 g 0,02 g 0,02 g 0,1g
Linearity ±0,03 g ±0,05 g ±0,05 g ± 0,3 g
Smallest part weight for piece
counting under laboratory condi-
tions
10 mg 10 mg 10 mg 100 mg
Smallest part weight for piece
counting under normal condi-
tions
100 mg 100 mg 100 mg 1 g
Recommended adjusting weight
F1 (not supplied) 2 kg 2 kg + 1 kg 2 kg + 2 kg 10 kg
Adjustment points
0,5 kg
1,0 kg
2,0 kg
2,4 kg
1,0 kg
1,5 kg
2,0 kg
3,0 kg
1,0 kg
2,0 kg
3,0 kg
4,0 kg
2 kg
5 kg
10 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h 2 h 4 h 2 h
Housing (B x D x H) [mm] 180 x 310 x 90
Vibration filter yes
Weighing plate stainless steel
[mm] Ø 150 Ø 150 Ø 150 160 x 200
Balance
Input voltage
12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 2,3 2,3 2,3 2,7
Interface RS232

572/573/KB/DS/FKB-BA-e-1774 8
KERN 572-45 572-49 572-55 572-57
Readability (d) 0,05 g 0,1 g 0,05 g 0,1 g
Weighing range (max) 12.000 g 16.000 g 20.000 g 24.000 g
Taring range (subtractive) 12.000 g 16.000 g 20.000 g 24.000 g
Reproducibility 0,05 g 0,1 g 0,1 g 0,1g
Linearity ±0,15 g ±0,3 g ±0,25 g ± 0,3 g
Smallest part weight for piece
counting under laboratory condi-
tions
50 mg 100 mg 50 100 mg
Smallest part weight for piece
counting under normal condi-
tions
500 mg 1 g 500 mg 1 g
Recommended adjusting weight
F1 (not supplied) 10 kg 10 kg + 5 kg 20 kg 20 kg
Adjustment points
2 kg
5 kg
10 kg
12 kg
5 kg
10 kg
15 kg
16 kg
5 kg
10 kg
15 kg
20 kg
5 kg
10 kg
15 kg
20 kg
24 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h
Housing (B x D x H) [mm] 180 x 310 x 90
Vibration filter yes
Weighing plate stainless steel
[mm] 160 x 200
Balance
Input voltage
12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 2,7
Interface RS232

9 572/573/KB/DS/FKB-BA-e-1774
2.2 KERN 573
KERN 573-34A 573-34NM 573-46A 573-46NM
Readability (d) 0,01 g 0,01 g 0,1 g 0,1 g
Weighing range (max) 650 g 650 g 6.500 g 6.500 g
Taring range (subtractive) 650 g 650 g 6.500 g 6.500 g
Verification value (e) - 0,1 g - 1 g
Minimum load (Min) - 0,5 g - 5 g
Accuracy class - II - II
Reproducibility 0,01 g 0,01 g 0,1 g 0,1g
Linearity ±0,03 g ±0,03 g ±0,3 g ± 0,3 g
Smallest part weight for piece
counting under laboratory condi-
tions
10 mg 10 mg 100 mg 100 mg
Smallest part weight for piece
counting under normal condi-
tions
100 mg 100 mg 1 g 1 g
Recommended adjusting weight
F1 (not supplied) 600 g 600 g 6 kg 10 kg
Adjustment points
200 g
500 g
600 g
200 g
500 g
600 g
2,0 kg
5,0 kg
6,0 kg
free
2,0 kg
5,0 kg
6,0 kg
free
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h
Housing (B x D x H) [mm] 180 x 310 x 90
Vibration filter yes
Weighing plate stainless steel
[mm] Ø 150 Ø 150 160 x 200 160 x 200
Balance
Input voltage
12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units g, kg
Net weight (kg) 2,3 2,3 2,8 2,8
Interface RS232

572/573/KB/DS/FKB-BA-e-1774 10
2.3 KERN KB
KERN KB 120-3N KB 240-3N KB 360-3N KB 600-2
Readability (d) 0,001 g 0,001 g 0,001 g 0,01 g
Weighing range (max) 120 g 240 g 360 g 650 g
Taring range (subtractive) 120 g 240 g 360 g 650 g
Reproducibility 0,001 g 0,001 g 0,002 g 0,01 g
Linearity ±0,003 g ±0,003 g ± 0,005 g ±0,03 g
Smallest part weight for piece
counting under laboratory condi-
tions
1 mg 1 mg 1 mg 10 mg
Smallest part weight for piece
counting under normal condi-
tions
10 mg 10 mg 10 mg 100 mg
Recommended adjusting weight
F1 (not supplied) 100 g 200 g 200 g +100 g 500 g +100 g
Adjustment points
20 g
50 g
100 g
120 g
100 g
150 g
200 g
240 g
100 g
200 g
300 g
360 g
200 g
500 g
600 g
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h 2 h 4 h 2 h
Housing (B x D x H) [mm] 167 x 250 x 85
Vibration filter yes
Weighing plate stainless steel
[mm] Ø 81
Balance
Input voltage
12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 1
Interface RS232
Rechargeable battery pack
KB-A01N
7,2 V / 2000mAh

11 572/573/KB/DS/FKB-BA-e-1774
KERN KB 1200-2N KB 2000-2N
Readability (d) 0,01 g 0,01 g
Weighing range (max) 1.200 g 2.000 g
Taring range (subtractive) 1200 g 2000 g
Reproducibility 0,01 g 0,01 g
Linearity ±0,03 g ±0,03 g
Smallest part weight for
piece counting under labor-
atory conditions
10 mg 10 mg
Smallest part weight for
piece counting under nor-
mal conditions
100 mg 100 mg
Recommended adjusting
weight F1 (not supplied) 1000 g 2000 g
Adjustment points
200 g
500 g
1000 g
0,5 kg
1,0 kg
1,5 kg
2,0 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental
temperature +10 °C ... + 40 °C
Warm-up time 2h
Housing (B x D x H) [mm] 167 x 250 x 85
Vibration filter yes
Weighing plate stainless
steel [mm] 130 x 130
Balance
Input voltage 12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 1,5
Interface RS232
Rechargeable battery pack
KB-A01N 7,2 V / 2000mAh

572/573/KB/DS/FKB-BA-e-1774 12
KERN KB 2400-2N KB 3600-2N KB 6000-1
Readability (d) 0,01 g 0,01 g 0,1 g
Weighing range (max) 2.400 g 3.600 g 6.500 g
Taring range (subtractive) 2400 g 3600 g 6.500 g
Reproducibility 0,01 g 0,02 g 0,1g
Linearity ±0,03 g ± 0,05 g ± 0,3 g
Smallest part weight for piece
counting under laboratory condi-
tions
10 mg 10 mg 100 mg
Smallest part weight for piece
counting under normal condi-
tions
100 mg 100 mg 1 g
Recommended adjusting weight
F1 (not supplied) 2000 g 3 kg 10 kg
Adjustment points
0,5 kg
1,0 kg
2,0 kg
2,4 kg
1 kg
2 kg
3 kg
3,6 kg
2,0 kg
5,0 kg
6,0 kg
free
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h 4 h 2 h
Housing (B x D x H) [mm] 167 x 250 x 85
Vibration filter yes
Weighing plate stainless steel
[mm] Ø 81
Balance
Input voltage
12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units g, kg g, kg, ct g, kg
Net weight (kg) 1,8 2,0 1,7
Interface RS232

13 572/573/KB/DS/FKB-BA-e-1774
KERN KB 6500-1NM KB 10000-1N KB 10K0.05N
Readability (d) 0,1 g 0,1 g 0,05 g
Weighing range (max) 6.500 g 10.000 g 10.000 g
Taring range (subtractive) 650 g 10.000 g 10.000 g
Verification value (e) 1 g - -
Minimum load (Min) 5 g - -
Accuracy class II - -
Reproducibility 0,1 g 0,1 g 0,05 g
Linearity ±0,3 g ±0,3 g ±0,15 g
Smallest part weight for
piece counting under labor-
atory conditions
10 mg 10 mg 50 mg
Smallest part weight for
piece counting under nor-
mal conditions
1 g 1 g 500 mg
Recommended adjusting
weight F1 (not supplied) 6 kg 10 kg 10 kg
Adjustment points
2,0 kg
5,0 kg
6,0 kg
2,0 kg
5,0 kg
10,0 kg
2,0 kg
5,0 kg
10,0 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental
temperature +10 °C ... + 40 °C
Warm-up time 2 h
Housing (B x D x H) [mm] 167 x 250 x 85
Vibration filter yes
Weighing plate stainless
steel [mm] 150 x 170
Balance
Input voltage 12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 1,7
Interface RS232
Rechargeable battery pack
KB-A01N 7,2 V / 2000mAh

572/573/KB/DS/FKB-BA-e-1774 14
2.4 KERN DS
KERN
DS
3K0.01S
DS
5K0.05S
DS
8K0.05
DS
10K0.1S
Readability (d) 0,01 g 0,05 g 0,05 g 0,1 g
Weighing range (max) 3.000 g 5.000 g 8.000 g 10.000 g
Taring range (subtractive) 3.000 g 5.000 g 8.000 g 10.000 g
Reproducibility 0,02 g 0,05 g 0,05 g 0,1 g
Linearity ±0,05 g ±0,15 g ± 0,15 g ±0,3 g
Smallest part weight for piece
counting under laboratory condi-
tions
10 mg 50 mg 50 mg 100 mg
Smallest part weight for piece
counting under normal condi-
tions
100 mg 500 mg 500 mg 1 g
Recommended adjusting weight
F1 (not supplied) 3 kg 5 kg 5 kg + 2 kg 10 kg
Adjustment points
1 kg
2 kg
3 kg
1 kg
2 kg
5 kg
2 kg
4 kg
5 kg
7 kg
8 kg
2 kg
5 kg
10 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h
Housing (B x D x H) [mm] 228 x 228 x 70
Vibration filter yes
Weighing plate stainless steel
[mm] 228 x 228 228 x 228 315 x 305 228 x 228
Balance Input voltage 12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 5,5 5,5 7,5 5,5
Interface RS232
Rechargeable battery pack
KB-A01N 7,2 V / 2000mAh

15 572/573/KB/DS/FKB-BA-e-1774
KERN DS 16K0.1 DS 20K0.1 DS 30K0.1 DS 36K0.2
Readability (d) 0,1 g 0,1 g 0,1 g 0,2 g
Weighing range (max) 16.000 g 20.000 g 30.000 g 36.000 g
Taring range (subtractive) 16.000 g 20.000 g 30.000 g 36.000 g
Reproducibility 0,1 g 0,1 g 0,2 g 0,2 g
Linearity ±0,3 g ±0,3 g ± 0,5 g ±0,6 g
Smallest part weight for piece
counting under laboratory condi-
tions
10 mg 10 mg 10 mg 100 mg
Smallest part weight for piece
counting under normal condi-
tions
100 mg 100 mg 100 mg 1 g
Recommended adjusting weight
F1 (not supplied) 10 kg + 5 kg 20 kg 20 kg + 10 kg 20 kg + 10 kg
Adjustment points
5 kg
10 kg
15 kg
16 kg
5 kg
10 kg
15 kg
20 kg
10 kg
15 kg
20 kg
30 kg
10 kg
15 kg
20 kg
30 kg
36 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h
Housing (B x D x H) [mm] 315 x 305 x 70
Vibration filter yes
Weighing plate stainless steel
[mm] 315 x 305
Balance
Input voltage 12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 7,5
Interface RS232
Rechargeable battery pack
KB-A01N 7,2 V / 2000mAh

572/573/KB/DS/FKB-BA-e-1774 16
KERN DS 30K0.1L DS 36K0.2L DS 60K0.2
Readability (d) 0,1 g 0,2 g 0,2 g
Weighing range (max) 30.000 g 36.000 g 60.000 g
Taring range (subtractive) 30.000 g 36.000 g 60.000 g
Reproducibility 0,2 g 0,2 g 0,4 g
Linearity ±0,5 g ±0,6 g ± 1,0 g
Smallest part weight for piece
counting under laboratory condi-
tions
100 mg 200 mg 200 mg
Smallest part weight for piece
counting under normal condi-
tions
1 g 2 g 2 g
Recommended adjusting weight
F1 (not supplied) 20 kg + 10 kg 20 kg + 10 kg 60 kg
Adjustment points
10 kg
15 kg
20 kg
30 kg
10 kg
15 kg
20 kg
30 kg
36 kg
20 kg
30 kg
50 kg
60 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h
Housing (B x D x H) [mm] 450 x 350 x 115
Vibration filter yes
Weighing plate stainless steel
[mm] 450 x 350
Balance
Input voltage 12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 9,5
Interface RS232
Rechargeable battery pack
KB-A01N 7,2 V / 2000mAh

17 572/573/KB/DS/FKB-BA-e-1774
KERN
DS
65K0.5
DS
65K1M
DS
100K0.5
DS
150K1
Readability (d) 0,5 g 1 g 0,5 g 1 g
Weighing range (max) 65.000 g 65.000 g 100.000 g 150.000 g
Taring range (subtractive) 65.000 g 65.000 g 100.000 g 150.000 g
Verification value (e) - 10 g - -
Minimum load (Min) - 50 g - -
Accuracy class - II - -
Reproducibility 0,5 g 1 g 0,5 g 1 g
Linearity ±1,5 g ±2 g ±1,5 g ±3 g
Smallest part weight for piece
counting under laboratory condi-
tions
500 mg 500 mg 500 mg 1 g
Smallest part weight for piece
counting under normal condi-
tions
5 g 5 g 5 g 10 g
Recommended adjusting weight
F1 (not supplied) 50 kg 50 kg 100 kg 3 x 50 kg
Adjustment points
20 kg
30 kg
50 kg
60 kg
20 kg
30 kg
50 kg
60 kg
20 kg
50 kg
100 kg
50 kg
100 kg
150 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h
Housing (B x D x H) [mm] 450 x 350 x 115
Vibration filter yes
Weighing plate stainless steel
[mm] 450 x 350
Balance
Input voltage 12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 9,5
Interface RS232
Rechargeable battery pack
KB-A01N 7,2 V / 2000mAh

572/573/KB/DS/FKB-BA-e-1774 18
2.5 KERN FKB
KERN
FKB
6K0.02
FKB
8K0.05
FKB
16K0.05
FKB
16K0.1
Readability (d) 0,02 g 0,05 g 0,05 g 0,1 g
Weighing range (max) 6.000 g 8.000 g 16.000 g 16.000 g
Taring range (subtractive) 6.000 g 8.000 g 16.000 g 16.000 g
Reproducibility 0,04 g 0,05 g 0,1 g 0,1g
Linearity ±0,1 g ±0,15 g ±0,25 g ± 0,3 g
Smallest part weight for piece
counting under laboratory condi-
tions
20 mg 50 mg 50 mg 100 mg
Smallest part weight for piece
counting under normal condi-
tions
200 mg 500 mg 500 mg 1 g
Recommended adjusting weight
F1 (not supplied) 5 kg 5 kg + 2 kg 10 kg + 5 kg 10 kg + 5 kg
Adjustment points
1 kg
3 kg
5 kg
6 kg
2 kg
4 kg
5 kg
7 kg
8 kg
5 kg
10 kg
15 kg
16 kg
5 kg
10 kg
15 kg
16 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h
Housing (B x D x H) [mm] 350 x 390 x 120
Vibration filter yes
Weighing plate stainless steel
[mm] 340 x 240
Balance
Input voltage
12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 6,5
Interface RS232
Battery operation 6 x 1,5 V; Size C

19 572/573/KB/DS/FKB-BA-e-1774
KERN
FKB
36K0.1
FKB
36K0.2
FKB
65K0.2
FKB
65K0.5
Readability (d) 0,1 g 0,2 g 0,2 g 0,5 g
Weighing range (max) 36.000 g 36.000 g 65.000 g 65.000 g
Taring range (subtractive) 36.000 g 36.000 g 65.000 g 65.000 g
Reproducibility 0,2 g 0,2 g 0,4 g 0,5 g
Linearity ±0,5 g ±0,6 g ±1,0 g ± 1,5 g
Smallest part weight for piece
counting under laboratory condi-
tions
100 mg 200 mg 200 mg 500 mg
Smallest part weight for piece
counting under normal condi-
tions
1 g 2 g 2 g 5 g
Recommended adjusting weight
F1 (not supplied) 20 kg + 10 kg 20 kg + 10 kg 50kg + 10 kg 50kg
Adjustment points
10 kg
15 kg
30 kg
36 kg
10 kg
20 kg
30 kg
36 kg
15 kg
30 kg
50 kg
60 kg
20 kg
30 kg
50 kg
60 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 4 h 2 h 4 h 2 h
Housing (B x D x H) [mm] 350 x 390 x 120
Vibration filter yes
Weighing plate stainless steel
[mm] 340 x 240
Balance
Input voltage
12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 6,5
Interface RS232
Battery operation 6 x 1,5 V; Size C

572/573/KB/DS/FKB-BA-e-1774 20
KERN
FKB
8K0.1M
FKB
65K1M
Readability (d) 0,1 g 1 g
Weighing range (max) 8.000 g 65.000 g
Taring range (subtractive) 8.000 g 65.000 g
Verification value (e) 1,0 g 10 g
Minimum load (Min) 5,0 g 50 g
Accuracy class II II
Reproducibility 0,1 g 1 g
Linearity ± 0,3 g ± 3 g
Smallest part weight for piece
counting under laboratory condi-
tions
100 mg 1 g
Smallest part weight for piece
counting under normal condi-
tions
1 g 10 g
Recommended adjusting weight
F1 (not supplied) 5 kg + 2 kg 50 kg
Adjustment points
2 kg
5 kg
7 kg
8 kg
20 kg
30 kg
50 kg
60 kg
Stabilization time (typical) 3 sec.
Humidity of air max. 80% relative (not condensing)
Permitted environmental tem-
perature +10 °C ... + 40 °C
Warm-up time 2 h
Housing (B x D x H) [mm] 350 x 390 x 120
Vibration filter yes
Weighing plate stainless steel
[mm] 340 x 240
Balance
Input voltage
12 V, 300 mA
Power supply unit
Input voltage 100 V – 240V
Units see menu
Net weight (kg) 6,5
Interface RS232
Battery operation 6 x 1,5 V; Size C
Other manuals for 572
4
This manual suits for next models
4
Table of contents
Other KERN Scale manuals