KERN PES User manual

GB KERN PES/PEJ
Version 1.0 11/2005
Operating Manual
Electronic Precision balances
Contents
1. Technical data _________________________________________________ 91
2. Declaration of conformity _______________________________________ 94
3. Basic Information (General)______________________________________ 96
3.1. Proper use __________________________________________________ 96
3.2. Improper Use ________________________________________________ 96
3.3. Warranty____________________________________________________ 96
3.4. Monitoring of Test Resources __________________________________ 97
4. Basic Safety Precautions________________________________________ 97
4.1. Pay attention to the instructions in the Operation Manual ___________ 97
4.2. Personnel training____________________________________________ 97
5. Transport and storage __________________________________________ 97
5.1. Testing upon acceptance ______________________________________ 97
5.2. Packaging __________________________________________________ 97
6. Unpacking, Setup and Commissioning ____________________________ 98
6.1. Installation Site, Location of Use________________________________ 98
6.2. Unpacking __________________________________________________ 98
6.3. Scope of delivery____________________________________________ 100
6.4. Setup _____________________________________________________ 100
6.5. Mains connection ___________________________________________ 101
6.6. Internal battery operation (cannot be reequipped) ________________ 101
6.7. Terminal for external devices__________________________________ 101
6.8. Initial Commissioning________________________________________ 102
6.8.1. Power display_____________________________________________ 102
6.8.2. Bar graph display _________________________________________ 103
6.8.3. Stability display ___________________________________________ 103
6.8.4. Balance zero display _______________________________________ 103
6.9. Adjustment_________________________________________________ 104
6.9.1. Adjustment with external weight (only PES) ___________________ 104
6.9.2. Adjustment test with external weight (only PES) ________________ 105

PES/PEJ-BA-defsi-0510 89
6.9.3. Adjustment with external weight (only PEJ)____________________ 107
6.9.4. Adjustment with external weight (only PEJ)____________________ 108
6.10. Verification _______________________________________________ 110
6.10.1. Seals and unlocking switch _________________________________ 111
7. Application and configuration menu 1 ____________________________ 112
7.1. User principle of the menu control _____________________________ 112
7.2. Menu overview______________________________________________ 114
7.2.1. Parameter additional functions ______________________________ 116
7.2.2. Parameter for serial interface________________________________ 117
8. Configuration menu 2__________________________________________ 119
8.1. User principle of the menu control _____________________________ 119
8.2. Menu overview______________________________________________ 120
9. Operation ____________________________________________________ 121
9.1. Keyboard overview __________________________________________ 121
9.2. Overview of display__________________________________________ 122
10. Weighing mode _____________________________________________ 123
10.1. Weighing_________________________________________________ 123
10.1.1. Taring ___________________________________________________ 124
10.1.2. Net/gross ________________________________________________ 126
10.2. Parts counting ____________________________________________ 127
10.3. Percent determination______________________________________ 130
10.3.1. Entering the reference weight by weighing ____________________ 130
10.3.2. Numeric entering of the reference weight______________________ 131
10.4. Density determination of solids (hydrostatic weighing) __________ 134
11. Adding of displayed values ___________________________________ 138
11.1. Adding with AUTO-TARA ___________________________________ 139
12. Weighing with tolerance range ________________________________ 140
12.1. General Information________________________________________ 140
12.2. Display of the results ______________________________________ 141
12.2.1. For 2 limits _______________________________________________ 141
12.2.2. For 3 or 4 limits ___________________________________________ 142
12.3. Basic settings for weighings with tolerance range ______________ 142
12.4. Evaluation of absolute values _______________________________ 143
12.4.1. Entering 2 limits by weighing________________________________ 143
12.4.2. Entering 3 or 4 limits by weighing ____________________________ 146
12.4.3. Numeric entering of 2 limits _________________________________ 149


PES/PEJ-BA-defsi-0510 91
1. Technical data
KERN PES 2200-2M PES 4200-2M PES 6200-2M
Readability (d) 0.01g 0.01g 0.01g
Weighing range (max) 2,200g 4,200g 6,200g
Minimum load (Min) 0.5g 0.5g 1g
Verification value (e) 0.1g 0.1g 0.1g
Accuracy class II II I
Reproducibility 0.01g 0.01g 0.01g
Linearity ± 0.02g ± 0.03g ± 0.03g
Stabilization time 3 sec. 3 sec. 3 sec.
Check weight (added) 500g (F2) 1000g (F2) 1000g (F2)
Vibration filter 4
Minimum piece weight 0.01g
Reference quantities 5, 10, 30, 100
Net weight (kg) 4kg
Permissible ambient condi-
tion 5° C to 35° C 10° C to 30° C
Humidity of air max. 80 % relative (not condensing)
Weighing Units g, kg, ct
Weighing plate, stainless
steel 200 x 200 mm
Dimensions of the housing
(B x D x H) 220 x 333 x 93 mm
Mains connection Mains adaptor 230 V, 50/60 Hz ; 12 V DC balance, 600 mA
Rechargeable battery
(optional)
Operating time ca. 6h. / charging time ca. 12h

PES/PEJ-BA-defsi-0510
92
KERN PES 8200-1M PES 15000-1M
Readability (d) 0.1g 0.1g
Weighing range (max) 8,200g 15,000g
Minimum load (Min) 5g 5g
Verification value (e) 1g 1g
Accuracy class II II
Reproducibility 0.1g 0,1
Linearity ± 0.2g ± 0.2 g
Stabilization time 3 sec. 3 sec.
Check weight (added) 2000g (M1) 2000g (M1)
Vibration filter 4 4
Minimum piece weight 0.1 g 0.1g
Reference quantities 5,10, 30, 100
Net weight (kg) 4
Permissible ambient condi-
tion 10° C to 30° C 5° C to 35° C
Humidity of air max. 80 % relative (not condensing)
Units g, kg, ct
Weighing plate, stainless
steel 200 x 200 mm
Dimensions of the housing
(B x D x H) 220 x 333 x 93 mm
Mains connection Mains adaptor 230 V, 50/60 Hz ; 12 V DC balance, 600
mA
Rechargeable battery (op-
tional)
Operating time ca. 6h. / charging time ca. 12h

PES/PEJ-BA-defsi-0510 93
KERN PEJ 2200-2M PEJ 4200-2M
Readability (d) 0.01g 0.01g
Weighing range (max) 2,200g 4,200g
Minimum load (Min) 0.5g 0.5g
Verification value (e) 0.1g 0.1g
Accuracy class II II
Reproducibility 0.01g 0.01g
Linearity ± 0.2g ± 0.3 g
Stabilization time 3 sec.
Adjustment weight internal
Vibration filter 4
Minimum piece weight 0.01 g
Reference quantities 5,10, 30, 100
Net weight (kg) 6
Permissible ambient condi-
tion 5° C to 35° C
Humidity of air max. 80 % relative (not condensing)
Units g, kg, ct
Weighing plate, stainless
steel 200 x 200 mm
Dimensions of the housing
(B x D x H) 220 x 333 x 93 mm
Mains connection Mains adaptor 230 V, 50/60 Hz ; 12 V DC balance, 600
mA
Rechargeable battery (op-
tional)
Operating time ca. 6h. / charging time ca. 12h

PES/PEJ-BA-defsi-0510
94
2. Declaration of conformity
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
Phone: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.com
Declaration of conformity
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con disitintivo CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms with the fol-
lowing standards.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nach-
stehenden Normen übereinstimmt.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente
déclaration, est conforme aux normes citées ciaprès.
Español Manifestamos en la presente que el producto al que se refiere esta declaración está de
acuerdo con las normas siguientes
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è confor-
me alle norme di seguito citate.
Balance lines: PES/PEJ
Mark applied EU Directive Standards
89/336/EEC EMC
73/23/EEC Low Voltage
EN45501
EN55022
EN60950
Date: 20.12.2005
Signature:
Gottl. KERN & Sohn GmbH
Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-07433/9933-0,Fax +49-074433/9933-149

PES/PEJ-BA-defsi-0510 95
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
Tel: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.com
Declaration of conformity
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con disitintivo CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms with the
following standards.
This declaration is only valid with the certificate of conformity by a notified body.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nach-
stehenden Normen übereinstimmt.
Diese Erklärung gilt nur in Verbindung mit der Konformitätsbescheinigung einer
benannten Stelle.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente
déclaration, est conforme aux normes citées ciaprès.
Cette déclaration est valide seulement avec un certificat de conformité d'un orga-
nisme notifié.
Español Manifestamos en la presente que el producto al que se refiere esta declaración está de
acuerdo con las normas siguientes.
Esta declaratión solo será válida acompañada del certificado de conformidad de
conformidad de la parte nominal.
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è confor-
me alle norme di seguito citate.
Questa dichiarazione sarà valida solo se accompagnata dal certificato di conformità
della parte nominale.
Model: PES/PEJ
EU Directive Standards EC-type-approval
certificate no.
Issued by
90/384/EEC EN45501 T6715 NMI
Date: 20.12.2005
Signature:
Gottl. KERN & Sohn GmbH
Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-07433/9933-0,Fax +49-074433/9933-149

PES/PEJ-BA-defsi-0510
96
3. Basic Information (General)
3.1. Proper use
The balance you purchased is intended to determine the weighing value of material
to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the mate-
rial to be weighed is manually and carefully placed in the centre of the weighing plate.
As soon as a stable weighing value is reached the weighing value can be read.
3.2. Improper Use
Do not use balance for dynamic weighings. In the event that small quantities are re-
moved or added to the material to be weighed, incorrect weighing results can be dis-
played due to the “stability compensation“ in the balance. (Example: Slowly draining
fluids from a container on the balance.)
Do not leave permanent load on the weighing plate. This may damage the measuring
system. Impacts and overloading exceeding the stated maximum load (max) of the
balance, minus a possibly existing tare load, must be strictly avoided. This could
cause damage to the balance.
Never operate balance in explosive environment. The serial version is not explosion
protected. The structure of the balance may not be modified. This may lead to incor-
rect weighing results, safety-related faults and destruction of the balance.
The balance may only be used according to the described conditions. Other areas of
use must be released by KERN in writing.
The appliance may only be opened by trained service technicians according to
KERN standards.
Before opening, disconnect appliance from power supply!
Warranty claims will be voided when appliance is opened.
The PES/PEJ weighing system may not be used in explosive areas or areas
with explosive substances.
3.3. Warranty
Warranty claims shall be voided in case
our conditions in the operation manual are ignored
the appliance is used outside the described uses
when the appliance is modified or opened
mechanical damage and damage caused by media, liquids
natural wear and tear
the appliance is improperly set up or incorrectly electrically connected
the measuring system is overloaded

PES/PEJ-BA-defsi-0510 97
3.4. Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the bal-
ance and, if applicable, the testing weight, must be checked regularly. The responsi-
ble user must define a suitable interval as well as type and scope of this test. Infor-
mation is available on KERN’s home page (www.kern-sohn.com) with regard to the
monitoring of balance test substances and the test weights required for this. In
KERN’s accredited DKD calibration laboratory test weights and balances may be
calibrated (return to the national standard) fast and at moderate cost.
4. Basic Safety Precautions
4.1. Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you
are already familiar with KERN balances.
4.2. Personnel training
The appliance may only be operated and maintained by trained personnel.
5. Transport and storage
5.1. Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
5.2. Packaging
Keep all parts of the original packaging in case you need to return the appliance.
Only use original packaging for returning.
Before sending, disconnect all connected cables and loose/movable parts.
Attach possibly existing transport safeguards. Secure all parts, e.g. glass windshield,
weighing plate, power unit etc., to prevent slipping and damage.

PES/PEJ-BA-defsi-0510
98
6. Unpacking, Setup and Commissioning
6.1. Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in
common conditions of use.
You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by install-
ing next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and
doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of
time. Non-permitted condensation (condensation of air humidity on the
appliance) may occur if a cold appliance is taken to a considerably
warmer environment. In this case, acclimatize the disconnected appli-
ance for ca. 2 hours at room temperature.
Avoid static charging of the material to be weighed, weighing container
and windshield.
If electro-magnetic fields or static charge occur, or if the power supply is unstable
major deviations on the display (incorrect weighing results) are possible. In that case,
the location must be changed.
6.2. Unpacking
Carefully remove the balance from the packaging, remove plastic cover and setup
balance at the intended workstation.

PES/PEJ-BA-defsi-0510 99
A
ir bubble
Connection net adapter
Antitheft device
Covering lid over
1. RS 232 inter-
face
2. Printer inter-
face
Covering lid over
underfloor
weighing device
Your balance in overview:

PES/PEJ-BA-defsi-0510
100
6.3. Scope of delivery
Serial accessories:
Balance
Weighing plate
Mains power supply
Check weight (only PES)
Operating Manual
Protective cover
6.4. Setup
Positioning of weighing plate:
Levelling balance:
Use the foot screws to level the
balance until the air bubble in
the bubble level is in the pre-
scribed circle.

PES/PEJ-BA-defsi-0510 101
6.5. Mains connection
Power is supplied via the external mains power supply. The stated voltage value
must be the same as the local voltage.
Only use original KERN mains power supplies. Using other makes requires consent
by KERN.
In the menu you can activate the AUTO-SLEEP function !A. A.u.1".In net operation
the balance after 3 min without load change or key pressure passes in a sleep mode.
Automatic activation of the display by load change or by pressing any key.
6.6. Internal battery operation (cannot be reequipped)
The optionally supplied battery is charged with the supplied power supply.
Before the first use, the battery should be charged by connecting it to the mains
power supply for at least 15 hours. The operating time of the battery is about. 6h;
charging time until complete recharging ca. 15h.
In the menu you can activate the AUTO-OFF function !y A.P.1". After 3 min without
load change the balance switches automatically off in order to spare the battery.
When the balance is in battery mode the following symbols appear on the display:
Battery charge sufficient
Battery very low. To charge the battery, connect it to the mains
as soon as possible (re-calibration not possible).
flashes
Voltage has dropped below prescribed minimum. Plug in the
mains adapter, to charge the balance via the electrical network
(15h).
6.7. Terminal for external devices
The output for the external devices is situated under the cover plate on the rear of the
balance. To remove the cover platte, loosen the two fastening screws.
RS 232 interface
Printer interface

PES/PEJ-BA-defsi-0510
102
6.8. Initial Commissioning
A warming up time of 10 minutes after switching on stabilizes the measuring values.
The accuracy of the balance depends on the local acceleration of gravity.
Please be sure to observe the information in the chapter on adjusting in chap. 6.9.
6.8.1. Power display
The balance will carry out a self-test
Supply balance with power via the
mains power supply.
Balance is in stand-by mode (green
LED is on).
Use the -key to switch the bal-
ance on.
a
z0
g
75s0
g
By pressing lightly it is possible to
check whether the balance display
changes.
Stand-by
Use the -key to switch the bal-
ance off. The balance is now in
stand-by mode again (green LED is
on).

PES/PEJ-BA-defsi-0510 103
6.8.2.Bar graph display
In configuration menu 1 (chapter 7) you can activate/deactivate the bar graph dis-
play.
0Hide bar graph
x B.G. * 1Show bar graph
Wägeplatte ist unbelastet halbe Höchstlast Waage ist mit Höchstlast belastet
The weighing range of the balance is divided into 40 graphic cuboids. Zero (0) will
appear on the graphic display if there is no weighing value on the balance. 20 graphic
cuboids are displayed if the balance is loaded up to one half of its weighing range.
6.8.3. Stability display
Stable
Unstable
If the display shows the stability display []the balance is in a stable status. The []
indication disappears if the condition is unstable.
6.8.4. Balance zero display
Environmental influences can lead to the exact figure of “000.0“ not being displayed
in spite of an empty weighing dish. It is, however, possible to reset your balance to
zero at any time and thus ensure that weighing really does commence at zero. Set-
ting to zero when a weight is applied is only possible within a certain type-dependent
range. In the event that the balance cannot be reset to zero with an applied weight,
this range has been exceeded. [o - Err ] will appear on the display.
If an exact zero reading is not displayed on the balance in spite of the weighing dish
being empty, press the TARE key and the balance will start resetting to zero. Your
balance will be set to zero after a short standby time.
In addition to this, the sign for the balance zero setting will be displayed [#0$].
weihging dish unloaded half of weighing range weighing range is used
to full capacity

PES/PEJ-BA-defsi-0510
104
6.9. Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To re-
ceive accurate measuring values it is also recommended to adjust the balance peri-
odically in weighing operation.
6.9.1. Adjustment with external weight (only PES)
Adjustment is possible with the delivered check weight, but for reasons of measuring
technique, not always optimal. We recommend the following adjustment weights:
Model Recommended adjustment
weight
PES 2200-2M 2 000 g (F1)
PES 4200-2M 4 000 g (E2)
PES 6200-2M 5 000 g (E2)
PES 8200-1M 5 000 g (F2)
PES 15000-1M 150000 g (F1)
Procedure when adjusting:
Observe stable environmental conditions. A warming up time of ca. 1 hour is required
for stabilization. Ensure that there are no objects on the weighing plate.
At verified balances, the adjustment is locked by a switch (except accuracy class I).
In order to adjust, open the locking switch see chap.6.10.1 (except accuracy class I).
Operation Display
Activate function [w CA. 3](see chap. 7).
w CA 3
!
a
z0
g
Zero point will be saved.
CAL EH T
CAL
K
!
Qn O
CAL

PES/PEJ-BA-defsi-0510 105
Carefully place adjusting weight in the centre
of the weighing plate
Adjustment process is started.
The process of adjustment is completed.
Remove adjusting weight, balance will return
into weighing mode automatically. In case of
an adjustment error or incorrect adjusting
weight the display will show [- Err]; repeat
adjustment process.
Qn
F.5.
CAL
!
Qn
F.5.
CAL
!
BWSX
!
EnD
CAL
!
a
z0
g
6.9.2. Adjustment test with external weight (only PES)
During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Procedure:
Observe stable environmental conditions. A warming up time of ca. 1 hour is required
for stabilization. Ensure that there are no objects on the weighing plate.
Operation Display
Activate function [w CA. 4]
(see chap. 7).
w CA 4
!
a
z0
g

PES/PEJ-BA-defsi-0510
106
Start of the adjustment test:
Carefully place adjusting weight in the
centre of the weighing plate
The difference between the saved value
and the measured value is displayed.
Take away adjustment weight.
Press any key; the adjustment process is
cancelled and the balance returns to
weighing mode.
CAL EH T
CAL
K
!
Qn
O
CAL
!
Qn
F.5.
CAL
!
Qn
F.5.
CAL
!
D KFF
!
-z02
g
!
EnD
CAL
!
a
z0
g

PES/PEJ-BA-defsi-0510 107
6.9.3. Adjustment with internal weight (only PEJ)
With the internal adjustment weight, the weighing accuracy can be checked and re-
adjusted at any time.
Procedure when adjusting:
Observe stable environmental conditions. A warming up time of ca. 1 hour is re-
quired for stabilization. Ensure that there are no objects on the weighing plate.
Operation Display
Activate function [w CA. 1]
(see chap. 7).
w CA 1
!
a
z0
g
Start of the automatic adjustment:
Adjustment is carried out automatically.
The process of adjustment is completed.
The balance returns automatically into
weighing mode.
AUTO CAL
CAL
!
VAKT
!
CH. 0
K
CAL
!
CH. F.5.
K
CAL
!
BWSX
!