GB KERN TCB
Version 1.2 03/2003
Operating instruction
Pocket balance
1. Technical Data
•Weighing range: 200 g
•readability d: 0.1 g
•operating conditions: 10 – 30 °C
max. 85% air humidity (not condensing)
•power supply: batteries 2 x LR44
2. Operation
•Turn the cover-disk to position „AUF/OPEN “
•Press the ON/OFF key ; the balance displays “8888” and switches to “0”.
•Place the weighing object in the middle of the weighing platform; the weight in
grams is shown on the display.
•After the weighing process is finished the cover-disk should be turned back to
“ZU/CLOSE”.
•By pressing the balance can be reset to “0”.
1. Adjustment
•Let loose the ON/OFF-key; the display shows “100”.
•By pressing for 3 seconds until “CAL” appears on the display.
•Place a 100 g adjusting weight (class M3) on the weighing platform.
•The display switches to “F” and the adjustment is complete.
•If other messages are on the display, the adjustment was faulty. Please repeat
the adjustment process.
2. Important Notes
•Do not overload the balance (when overloaded, error message “E” appears)
•With the occurrence of electromagnetic fields deviations of the displayed
weight value are possible. In case of deviations change location.
•Avoid high air humidity, vapours and dust, because the balance is not
hermetically closed. Do not bring the balance in direct contact with liquids.
Liquids could enter the load cell.
•No permanent load on the weighing platform.
•Never operate the balance in rooms where danger of explosion exists.
•Regularly check the balance with the test weight.
•Warranty expires as soon as the balance is opened.