KERN KERN ALT User manual

ALT-BA-defsi-0410 69
GB KERN ALT
Version 1.0 12/2004
Operating Instructions
Electronic analytical balance
Table of contents
1Technical data _________________________________________________ 71
2Declaration of conformity _______________________________________ 72
3Fundamental information (general)________________________________ 74
3.1 Designated use __________________________________________________ 74
3.2 Inappropriate use ________________________________________________ 74
3.3 Guarantee ______________________________________________________ 74
3.4 Monitoring the test substances_____________________________________ 75
4Fundamental safety instructions__________________________________ 75
4.1 Observe notices in the operating manual ____________________________ 75
4.2 Training of personnel_____________________________________________ 75
5Transport and storage __________________________________________ 75
5.1 Acceptance check _______________________________________________ 75
5.2 Packing ________________________________________________________ 75
6Unpacking, installation and commissioning ________________________ 76
6.1 Installation Site, Location of Use ___________________________________ 76
6.2 Unpack_________________________________________________________ 76
6.2.1 Assembly ___________________________________________________________ 77
6.2.2 Scope of delivery _____________________________________________________ 77
6.3 Mains supply____________________________________________________ 78
6.4 Connection of peripheral equipment ________________________________ 78
6.5 Initial Commissioning ____________________________________________ 78
7User menu ____________________________________________________ 79
7.1 User principle of the menu control __________________________________ 80
7.2 Operating the balance with PS/2 keyboard (see chapter 6.3)_____________ 81
7.3 Menu selection __________________________________________________ 82
7.3.1 P1 adjustment________________________________________________________ 82
7.3.2 P2 GLP (Good laboratory practice) _______________________________________ 90
7.3.3 P3 Date/time_________________________________________________________ 92
7.3.4 P4 Readout__________________________________________________________ 94
7.3.5 P5 RS-232 __________________________________________________________ 96
7.3.6 P6 Printouts _________________________________________________________ 96
7.3.7 P7 Weighing units____________________________________________________ 101
7.3.8 P8 operating modes __________________________________________________ 101
7.3.9 P9 General functions _________________________________________________ 102

ALT-BA-defsi-0410
70
8Operation ____________________________________________________ 103
8.1 Operating modes (P8 modes) _____________________________________ 105
8.1.1 Simple weighing _____________________________________________________ 107
8.1.2 Parts counting_______________________________________________________ 109
8.1.3 Weighing with tolerance range __________________________________________ 117
8.1.4 Filling _____________________________________________________________ 119
8.1.5 Percent determination ________________________________________________ 120
8.1.6 Animal weighing (Dynamic weighing)_____________________________________ 122
8.1.7 Density determination_________________________________________________ 124
8.1.8 Formulation_________________________________________________________ 125
8.1.9 Statistics ___________________________________________________________ 129
9Data output __________________________________________________ 131
9.1 Tecnical data___________________________________________________ 131
9.2 Pin allocation of the balance exit plug (front view) ____________________ 131
9.3 Remote commands _____________________________________________ 132
10 Maintenance, upkeep, disposal ________________________________ 133
10.1 Cleaning ______________________________________________________ 133
10.2 Maintenance, upkeep ____________________________________________ 133
10.3 Disposal_______________________________________________________ 133
11 Troubleshooting ____________________________________________ 134

ALT-BA-defsi-0410 71
1 Technical data
KERN ALT 100-4M ALT 160-5D ALT 220-4M ALT 310-4
Readout (d) 0.1 mg 0.1mg/0.01mg 0.1 mg 0.1 mg
Weighing range (Max) 100 g 160 g/60 g 220 g 310 g
Taring range (subtractive) 100 g 160 g/60 g 220 g 310 g
Reproductability 0.2 mg 0.2mg/0.1mg 0.2 mg 0.2 mg
Linearity ± 0.2 mg ± 0,2mg/0,1mg ± 0.2 mg ± 0.3 mg
Minimal piece weight for
counting function 0,2 mg 0,2 mg 0,2 mg 0,2 mg
Adjustment weight internal internal internal internal
Verifiable yes no yes no
Verification value (e) 1 mg - 1 mg -
Accuracy category I - I -
Reference piece numbers
for parts counting 10, 20, 50,100, freely selectable
Weighing Units mg, g, ct,
mg, g, ct, lb,
oz, ozt, dwt,
t {h}, t {S},
t {t}, mom,
GN, N, t,
freely pro-
grammable
Stabilization time (typically) 4 sec.
Permissible ambient
temperature
+ 15° C .... +30° C
Humidity of air: max. 80 % (not condensing)
Under floor weighing device On-hook eyelets, serial
Weighing plate mm 100 80 100
Housing (W x D x H) mm 205 x 500 x 290
Weight kg (net) 8,9

ALT-BA-defsi-0410
72
2 Declaration of conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern
Postfach 4052
Tel: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms with the
following standards.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den
nachstehenden Normen übereinstimmt.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente
déclaration, est conforme aux normes citées ci-après.
Español Manifestamos en la presente que el producto al que se refiere esta declaración est´´a de
acuerdo con las normas siguientes
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è
conforme alle norme di seguito citate.
Electronic Balance: KERN ALT
Mark applied EU Directive Standards
89/336EEC EMC EN 61000-4-2 :1999
EN 61000-4-3 :1996
EN 61000-4-4 : 1999
EN 61000-4-5 : 1998
EN 61000-4-6 : 1999
EN 61000-4-11 : 1997
EN 55022 :2000
Date: 18.01.2005 Signature:
Gottl. KERN & Sohn GmbH
Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149

ALT-BA-defsi-0410 73
KERN & Sohn GmbH
D-72322 Balingen-Frommern
Postfach 4052
Tel: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms with the
following standards.
This declaration is only valid with the certificate of conformity by a notified body.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den
nachstehenden Normen übereinstimmt.
Diese Erklärung gilt nur in Verbindung mit der Konformitätsbescheinigung einer
benannten Stelle.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente
déclaration, est conforme aux normes citées ci-après.
Cette déclaration est valide seulement avec un certificat de conformité d’un
organisme notifié.
Español Manifestamos en la presente que el producto al que se refiere esta declaración está de
acuerdo con las normas siguientes.
Esta declaración solo será válida acompañada del certificado de conformidad de
conformidad de la parte nominal.
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è
conforme alle norme di seguito citate.
Questa dichiarazione sarà valida solo se accompagnata dal certificato di conformità
della parte nominale.
Model: KERN ALT 100-4M
KERN ALT 220-4M
EU Directive Standards EC-type-approval
certificate no.
Issued by
90/384/EEC EN 45501 T6655 NMI
Date: 18.01.2005 Signature:
Gottl. KERN & Sohn GmbH
Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-07433/9933-0,Fax +49-074433/9933-149

ALT-BA-defsi-0410
74
3 Fundamental information (general)
3.1 Designated use
The balance you have acquired serves to determine the weighing value of the
material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e.
the material to be weighed is manually and carefully placed in the centre of the
weighing plate. The weighing value can be read off after a stable weighing value has
been obtained.
3.2 Inappropriate use
Do not use balance for dynamic add-on weighings, if small amounts of goods to be
weighed are removed or added. The “stability compensation“ installed in the balance
may result in displaying an incorrect measuring value! (Example: Slowly draining
fluids from a container on the balance.)
Do not leave a permanent load on the weighing plate. This can damage the
measuring system.
Be sure to avoid impact shock and overloading the balance in excess of the
prescribed maximum load rating (max.), minus any possible tare weight that is
already present. This could cause damage to the balance.
Never operate the balance in hazardous locations. The series design is not
explosion-proof.
Structural alterations may not be made to the balance. This can lead to incorrect
weighing results, faults concerning safety regulations as well as to destruction of the
balance.
The balance may only be used in compliance with the described guidelines. Varying
areas of application/planned use must be approved by KERN in writing.
3.3 Guarantee
The guarantee is not valid in the following cases:
•non-observation of our guidelines in the operating instructions
•use outside the described applications
•alteration to or opening of the device
•mechanical damage and damage caused by media, liquids,
natural wear and tear
•inappropriate erection or electric installation
•overloading of the measuring equipment

ALT-BA-defsi-0410 75
3.4 Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight
must be checked at regular intervals within the scope of quality assurance. For this
purpose, the answerable user must define a suitable interval as well as the nature
and scope of this check. Information is available on KERN’s home page (www.kern-
sohn.com) with regard to the monitoring of balance test substances and the test
weights required for this. Test weights and balances can be adjusted quickly and at a
reasonable price in KERN’s accredited DKD calibration laboratory (return to national
normal).
4 Fundamental safety instructions
4.1 Observe notices in the operating manual
Please read the operating instructions carefully before erecting and commissioning,
even if you already have experience with KERN balances.
4.2 Training of personnel
The device may only be operated and looked after by trained members of staff.
5 Transport and storage
5.1 Acceptance check
Please check the packaging immediately upon delivery and the device during
unpacking for any visible signs of external damage.
5.2 Packing
Please retain all parts of the original packaging in case it should be necessary to
return items at any time.
Only the original packaging should be used for return consignments.
Before despatch, disconnect all attached cables and loose/movable parts.
Apply any intended transport security devices. Secure all parts e.g. weighing plate,
mains power supply etc. against slipping and damage.

ALT-BA-defsi-0410
76
6 Unpacking, installation and commissioning
6.1 Installation Site, Location of Use
The balance is constructed in such a way that reliable weighing results can be
achieved under normal application conditions.
By selecting the correct location for your balance, you will be able to work quickly and
precisely.
Therefore please observe the following at the place of installation:
•Place the balance on a firm, level surface;
•Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
•Protect the balance against direct draughts due to open windows and doors;
•Avoid jarring during weighing;
•Protect the balance against high humidity, vapours and dust;
•Do not expose the device to extreme dampness for longer periods of time.
Inadmissible bedewing (condensation of air moisture on the device) can occur
if a cold device is taken into a significantly warmer environment. In this case,
please acclimatise the device for approx. 2 hours at room temperature after it
has been disconnected from the mains.
•Avoid static charging of items to be weighed, or weighing container.
Major display deviations (incorrect weighing results) are possible if electromagnetic
fields occur as well as due to static charging, currents and instable power supply. It is
then necessary to change the location.
6.2 Unpack
Carefully remove the balance from its packaging, remove the plastic wrapping and
position the balance in its intended working location.

ALT-BA-defsi-0410 77
6.2.1 Assembly
Structure of the balance:
Use the foot screws to level the balance until the air bubble in the bubble level is in
the prescribed circle.
6.2.2 Scope of delivery
Standard accessories:
•Scales
•Weighing plate
•Network appliance
•Instruction manual
•Wind protection
(1) Base plate
(2) Screen ring
(3) Weighing plate

ALT-BA-defsi-0410
78
6.3 Mains supply
Electric power supply is by means of the external mains supply circuit. The printed
voltage level must comply with the local voltage.
Only use original KERN mains supply circuits. The use of other makes is subject to
approval by Kern.
1. Mains supply
2. PS/2 keyboard
3. RS232 interface
6.4 Connection of peripheral equipment
The balance must be disconnected from the mains before connecting or
disconnecting additional equipment (printer, PC) to or from the data interface.
Only use KERN accessories and peripheral equipment with your balance. These
have been ideally coordinated to your balance.
6.5 Initial Commissioning
A warm-up time of 1 hour stabilises the measured values after switching on.
The accuracy of the balance depends on the local acceleration of the fall.
Instructions in chapter Adjustment must be observed.

ALT-BA-defsi-0410 79
7 User menu
The user menu has nine main menus with the following sub-menus:
P1
Calibration
P2 GLP P3 Date/Time P4 Readout P5 RS-232 P6 Printouts P7 Units P8 Modes P9 Globals
Int. calibr.
Ext. calibr
User calibr.
Calibr. Test:
Weight.corr.
Aut. calibr.
Print report
User No.
Prj. No.
Time print
Date print
UserNo. print
Prj. No. print
Id print
Diff. print
Date format
Time format
Time
Date
Disp. time
Disp. date
Filter
Filter range
Disp refresh
Stable range
Stable speed
Autozero
Last digit
Boud rate
Parity
Data bits
Stop bits
Handshake
Auto print
Interval
Print on stab
Printout No.
Pr. 1 start
Pr. 1 stop
Pr. 2 start
Pr. 2 stop
Pr. 3 start
Pr. 3 stop
Pr. 4 start
Pr. 4 stop
String 1
String 2
String 3
.
.
.
String 78
String 79
String 80
Grams
Miligrams
Kilograms
Pounds
Ounces,
Ounces troy
Carats
Dwt
Taels. Hk
Taels. S
Taels. T
Mommes
Granis
Newtons
Tical
Custom
Custom Factor
Parts Count
Checkweigh
ing
Filling
Percont
Animal
Formulation
Statistics
Density
Beep
Language
Backlight
Contrast
Screensaver
Temperature
Balance Id
Software rev.
Par. Printout
Par. Receive.
Factory deff.
Main Menu
1. Menu number (P1 – P9)
2. Menu designation
3. Cursor (►) indicates the current menu
item
Submenu:
1. Menu number (P1 – P9)
2. Parameter number and designation
3. Parameter value or ****** i.e. this
parameter is available as a function
4. Functional Description
5. Cursor (►) indicates the current menu
item (e.g. P1).
6. Cursor (►) indicates the current
parameter (e.g. 06).
7. ”Setup“i.e. the user is in Setup Mode.
8. Use the arrow key to modify the
parameter value, the active parameter
blinks.

ALT-BA-defsi-0410
80
7.1 User principle of the menu control
In the menu the settings of the balance can be modified and functions can be
activated.
This way, the balance can be adjusted to individual weighing requirements. To do so
proceed as follows:
ÖUse the -key to turn on the balance.
ÖAfter actuating the -key the display shows the main menus that can be set
P1 Calibration
P2 GLP
P3 Date/Time
P4 Readout
P5 RS-232
P6 Printouts
P7 Units
P8 Modes
P9 Global
s
Setup2.02.05
ÖSelect your setting using the cursor (►).
Use the -key to move the cursor (►) down, the -key to move it up
ÖThe -key confirms your selected setting, the display shows the sub-menu or
the active menu item blinks:
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
off
on
function
31.01.05
11 hour
ÖUse the arrow keys to move within and enter into the menu:
•Cursor control up
•Parameter selection within a function
•Increasing the value for the selected numeral or number
•Access to the main
menu
•Access to the sub-menu
•In the menu, one menu
item to the right
•Selecting the numeral to
be modified to the right
(the respective active
place blinks)
•Return to main menu
•Selecting the numeral to
be modified to the left
(the respective active
place blinks)
•Cursor control down
•Parameter selection within a function
•Decreasing the value for the selected numeral or number

ALT-BA-defsi-0410 81
ÖUse the -key to confirm or the -key to cancel
ÖUse the -key to return to the menu
ÖActuate -key again, a message will inquire whether you want to save the
changes you made.
ÖActuating the -key will save the selected setting.
ÖIf you do not want to save the selected setting press
ÖThe balance will automatically return to weighing mode
Weighing
0.0000
g
0% 100%
0
27.01.05
7.2 Operating the balance with PS/2 keyboard (see chapter 6.3)
Using a PS/2 keyboard will help you to enter numbers and texts faster and easier.
Key allocation:
Balance keyboard
PS keyboard

ALT-BA-defsi-0410
82
7.3 Menu selection
7.3.1 P1 Calibration (Adjustment)
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location (
only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. It is
also recommendable to adjust the balance periodically during weighing operation in
order to obtain exact measured values.
Using the inbuilt adjusting weight the balance automatically carries out an adjustment
after any temperature change of 0.8 °C or any time it was disconnected from the
power supply.
Of course it is also possible to check the accuracy of the balance manually any time.
You can choose between two options:
1. Adjustment is started by pressing the -key three times.
After successful adjustment the balance automatically returns to weighing
mode.
2. Under menu item “P1 01 Calibration with internal weight” (see table below)
Procedure for adjustment:
Observe stable environment conditions. A warming-up time of ca. 1 hour for
stabilisation is necessary.
Important!
During the adjustment process no objects should be on the weighing plate.
Operator Indication
ÖCall up menu item “P1 Calibration “
(see chapter 7.1)
ÖPress -key
P1 Calibration
P2 GLP
P3 Date/Time
P4 Readout
P5 RS-232
P6 Printouts
P7 Units
P8 Modes
P9 Globals
Setup2.02.05
01 Int. calibr. - Adjustment with internal weight
ÖUse the cursor (►) to select “01 Int. calibr. “
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
off
on
function
31.01.05
11 hour
ÖPress the -key; the adjustment with internal
weight is carried out.
Calibration
Internal calibration

ALT-BA-defsi-0410 83
ÖAfter successful adjustment the balance
automatically returns to weighing mode.
In case of an adjustment error (e.g. objects on the
weighing plate) the display will show an error
message, repeat adjustment.
Weighing
0.0000
g
0% 100%
0
27.01.05
02 Ext. Cal - Adjustment with an external weight
Locked for verifiable appliances
03 User cal. – locked for verifiable appliances
04 Cal. test – not documented
05 Weight corr. – locked for verifiable appliances
06 Autom. cal. – locked for verifiable appliances
07 Auto. Cal.time – locked for verifiable appliances
08 Print report – printout of the adjustment data
ÖUse the cursor (►) to select “08 “print report“
ÖPress the –key; the current menu item blinks.
Use the arrow keys (see chapter 7.1) to save your
settings
0 Print report deactivated
1 Print report activated
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
off
on
function
31.01.05
11 hour
ÖConfirm with -key
ÖWhen print report is activated you will receive a printout of your adjustment data
after each adjustment (Ex.):
*** Internal calibration report ***
Calibr. : internal
Diff.. : 0.0038 g
Signature………………………….

ALT-BA-defsi-0410
84
7.3.1.1 Calibration (adjustment) only for model KERN ALT 310
Operator indication
ÖCall up menu item “P1 Calibration“
(see chapter 7.1)
ÖPress -key
P1 Calibration
P2 GLP
P3 Date/Time
P4 Readout
P5 RS-232
P6 Printouts
P7 Units
P8 Modes
P9 Globals
Setup2.02.05
01 Int. cal. - Adjustment with internal weight
ÖUse the cursor (►) to select “01 Int. cal.“
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
o ff
o n
function
31.01.05
11 hour
ÖPress the key; the adjustment with internal
weight is carried out.
Calibration
Internal calibration
ÖAfter successful adjustment the balance
automatically returns to weighing mode.
In case of an adjustment error (e.g. objects on the
weighing plate) the display will show an error
message, repeat adjustment.
Weighing
0.0000
g
0% 100%
0
27.01.05
02 Ext. Cal - Adjustment with an external weight
ÖUse the cursor (►) to select “02 Ext. cal.“
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
o ff
o n
function
31.01.05
11 hour
ÖPress -key
Ensure that there are no objects on the weighing
plate.
ÖPress -key; the display shows the measuring
value of the adjusting weight.

ALT-BA-defsi-0410 85
ÖNow place the adjusting weight in the center of
the weighing plate, press the -key. After
successful adjustment the system will
automatically return to the menu.
The -key can be used to cancel the
adjustment.
In case of an adjustment error or incorrect
adjusting weight an error message is displayed.
Repeat adjustment.
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
o ff
o n
function
31.01.05
11 hour
03 User calibr. – user-defined adjusting weight
ÖUse the cursor (►) to select “03 User. cal.“
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
o ff
on
function
31.01.05
11 hour
ÖPress -key; the display shows the measuring
value of the adjusting weight. The first numeral of
the displayed value blinks. You can modify it
using the arrow keys.
After setting your new adjusting weight use the
-key to confirm.
ÖEnsure that there are no objects on the weighing
plate. Press -key
ÖPress the key; the display shows the
measuring value you selected for the adjusting
weight

ALT-BA-defsi-0410
86
ÖNow place the adjusting weight in the center of
the weighing plate, press the -key. After
successful adjustment the system will
automatically return to the menu.
The -key can be used to cancel the
adjustment.
In case of an adjustment error or incorrect
adjusting weight an error message is displayed.
Repeat adjustment.
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
o ff
on
function
31.01.05
11 hour
We recommend:
Carry out adjustment as close to the maximum load of the balance as possible.
Information concerning the test weights is available at: http://www.kern-sohn.com
04 Calibration test
Here, deviation from the last adjustment is determined. This is only a check, i.e.
no values are changed.
ÖUse the cursor (►) to select “04 Cal. test“
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
o ff
on
function
31.01.05
11 hour
ÖPress the -key, the calibration test is started
Calibration
Calibration check
ÖThe result is displayed
Calibration
Check results
Cal.: 180.1354
Act.: 180.1438
Diff.: - 0.0084
05 Weight corr. – not documented

ALT-BA-defsi-0410 87
06 Automatic calibration
ÖUse the cursor (►) to select “06 Autom. cal.“
ÖPress the -key; the current menu item blinks.
ÖUse the - or - key to choose between the
following settings:
0 no Autom. calibration locked
1 Temp.Autom. Calibration for a
Temperature change of 0.8 °C
2 Time Autom. Calibration after
a user-defined interval
3 both Autom. Calibration for a
Temperature change of
0.8 °C or after the
defined interval
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
o ff
on
function
31.01.05
11 hour
ÖConfirm with -key
07 Auto cal. time - Automatic calibration after a user-defined interval
ÖUse the cursor (►) to select “07 Auto.cal.time“
ÖPress the -key; the current menu item blinks.
ÖUse the arrow keys to enter a value between 1
and 12 hours.
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
off
o n
function
31.01.05
11 hour
ÖConfirm with -key

ALT-BA-defsi-0410
88
08 Print report
In this menu item you can activate the function for printing your calibrating data
ÖUse the cursor (►) to select “08 “print report“
ÖPress the key; the current menu item blinks.
Use the arrow keys (see chapter 7.1) to save your
settings
0 Print report deactivated
1 Print report activated
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
o ff
on
function
31.01.05
11 hour
ÖConfirm with -key
ÖWhen print report is activated you will receive a printout of your adjustment data
after each adjustment (Ex.):
*** Internal calibration report***
Calibr. : internal
diff.. : 0,0038 g
Name………………………….
*** External calibration report ***
Calibr. : external
diff.. : -0,0624 g
Name ………………………….
*** User calibration report***
Calibr. : User calibration
diff.. : -0,0003 g
Name ………………………….