Kinetec Spectra User manual

Manuel d'utilisationManuel d'utilisation Avant toute utilisation lire ce document.
AbilityOne Kinetec se réserve le droit de toutes modifications techniques.
User manualUser manual Before use, please read this document.
AbilityOne Kinetec reserves the right to effect technical modifications.
BedienungsanleitungBedienungsanleitung Vor Benutzung unbedingt dieses Dokument lesen.
AbilityOne Kinetec behält sich das Recht vor, jegliche technische Änderung durchzuführen.
Istruzioni per l’usoIstruzioni per l’uso Prima di mettere in funzione l’apparecchio leggere con attenzione il presente documento.
La AbilityOne Kinetec si riserva il diritto di apportare modifiche tecniche.
Manual de empleoManual de empleo Antes de cualquier utilización, lea este documento.
AbilityOne Kinetec se reserva el derecho a cualquier modificación técnica.
F
GB
D
I
E
Your contact:

11 GB
SUMMARYSUMMARY
Page Page
• Definition__________________________2
• Description ________________________3
• Safety____________________________4
• Electrical connection __________________5
• Use of the hand control
- Select your language _______________________6
- Locking the hand control setting ________________7
-START/STOP/REVERSEfunction_________________7
-QuickStart _____________________________8
- Possible values for each parameter ______________8
- How to adjust the basic parameters of the movement ___9
-HowtomodifyprogramsPROGRAMMODE __________10-12
- Using Programs __________________________13
- Reading the values of a program________________14
- How to use the WARM UP key _________________ 15
- How to use the MODULATION key _______________ 16
- How to use the BYPASS key __________________ 17
- Session Time____________________________ 18
• Patient Pad kit_______________________ 19
• Setting up the patient___________________ 20
• Options ___________________________ 21
• Product information
- Maintenance____________________________ 22
- Troubleshooting__________________________ 22
- Cleaning_______________________________ 23
- Technical specifications _____________________ 23
- Symbols used ___________________________ 24
- Warranty ______________________________ 24
- Maintenance sheet ________________________ 25
SAFETY RECOMMENDATIONSSAFETY RECOMMENDATIONS
• The physician defines the protocol and ensures that it is correctly implemented (adjustments, session time and frequency of use).
• We recommend that you lock the hand control when you give it to the patient.
• For optimum safety, always give the hand control to the patient before starting the system.
• Explosion hazard: KINETEC Spectra is not designed for use in the presence of flammable anaesthetics.
• Check that the electrical socket is in good condition and is suitable for the splint power supply cord. The latter complies with current standards and
has a grounding socket. The plug may be connected to any standard socket. The socket must however have a grounded pin.
To connect the power supply, only use the original cable supplied with the machine.
Check that the cables remain free around the device so that they do not get damaged.
Manuel N°467896275 - Updated 21/02/03

22
GB
DefinitionDefinition
TheKINETEC Spectra is a Knee PASSIVE mobilization device enabling the extension and flexion movement from -10° to 120°.
• Indications• Indications
•Kneereplacement surgery.
•Fractures(patellar, tibia plateau, femoral,...).
•Arthrolysis
•Hipsurgery, including hip replacement, hip pinning, osteotomy,...).
•Ligamentrepairs.
•Arthroscopicsurgery (menisectomies, patellectomies,...)
•Burns,joints sepsis,...
• Clinical Benefits• Clinical Benefits
•Breaksthe cycle of trauma, inflammation and the loss of range of motion.
•Preventsjoint stiffness.
•Speedsthe recovery of post-operative range of motion.
•Maintainsthe quality of the joint surface.
•Reducespain and edema.
•Promotesjoint cartilage healing.
•Reduceshospitalizationtime
•Reducesthe need for pain medication.
•Providesimmediate post-operative continuous passive motion.
•DigitalROM readout on the patient hand control for positive reinforcement.
•Maintainsdesired positions for stretching and muscular rest.
• Contraindications• Contraindications
BoneCancer, Warped joint surfaces, Spastic paralyses, Unstable fractures, Uncontrolled infection.
Themachine are not adapted for patients height more 2 m(6'7") or under 1,40m (4'7").

33 GB
DescriptionDescription
KINETEC Spectra consists of the following components:
1 • Lower limb support.
2 • Thigh support.
3 • Foot support and hand control location for transport.
4 • Hand control.
5 • Thigh support setting lock.
6 • Lower limb support setting lock .
7 • Foot support positioning setting lock.
8 • Transport handle.
9• ON/OFFswitch andfuses.
10 • Liquid-crystal display (2 lines of 16 characters).
11 • SESSION TIME display key.
12 • BYPASS mode key.
13• MODULATIONkey.
14 • WARM UP key.
15 • EXTENSION setting key.
16 • FLEXION setting key.
17 • Increase / decrease keys.
18 • START key.
19 • STOP key.
20 • FORCE key.
21• SPEEDkey.
22 • PAUSE key.
23 • TIMER key.
24• PROGRAM accesskey.
11
22
33
44
55
66
77
88
99
1010
1414
1313
1212
1111
1515
1616
1717
1818
1919
2020
2121
2222
2323
2424

44
GB
SafetySafety
The physician defines the protocol and ensures that it is correctly implemented
(adjustments, session duration and frequency of use).
The patient must know the start/stop/reverse function on the control handle. Hand control must be accessible to patient at
all times. (See page 7).
KINETEC Spectra complies with Directive 93/42/CEE.
EXPLOSION HAZARD:EXPLOSION HAZARD:
KINETEC Spectra is not designed for use in the presence of flammable anesthetics.KINETEC Spectra is not designed for use in the presence of flammable anesthetics.
In case of electromagnetic interference with other devices move the device.
KINETEC Spectra is in compliance with standards in force (IEC 601.1.2), electromagnetic compatibility standard for medical
devices.

55 GB
Kinetec
30 STOP 15 90
Please wait
KINETEC
Spectra V01.0
Electrical connectionElectrical connection
KINETEC Spectra is a type B class I device.
Before connecting the device to the power supply, check that the mains voltage
matches that shown on the identification plate (100-240 V~ 50-60Hz).
Connect the hand control (4)
Connect the power supply cable (25).
IMPORTANTIMPORTANT
Check that the electrical socket is in good condition and is suitable for the splint power
supply cord. The latter complies with current standards and has a grounding socket.
The plug may be connected to any standard socket.
The socket must however have a grounded pin.
To connect the power supply, only use the original cable supplied with the machine.
Check that the cables remain free around the device so that they do not get damaged.
Starting the unitStarting the unit
Switch on (9).
While the unit begins an auto diagnostic, the display shows the following :
Your KINETEC Spectra is ready to be used.
2525
44
99

66
GB
Use of the hand controlUse of the hand control
Select your languageSelect your language
Beginning Keys to press Display Remarks
Switch the unit ON Check if the hand control is not blocked
(See page 7).
Press the 2 keys in the same time The display indicate the language selected.
To change the language or
The English language is selected.
English
French
German
Italian
Spanish
To validate the new language. To exit and confirm the new language, switch
OFF and Switch ON the unit.
Kinetec
30 STOP 15 90
Please wait
KINETEC
Spectra V01.0
Language:
english
Language:
English
Ok
Switch on/off

77 GB
Use of the hand controlUse of the hand control
Locking the hand control settingLocking the hand control setting
The hand control allows the patient to control the machine as appropriate.
Simultaneously press the and keys to lock the hand control,
The display reads "LOCK", you can not change the parameters, if you try the display reads "LOCK".
To unlock the hand control, simultaneously press the and keys, The display reads "UNLOCK".
We recommend that you lock the hand control when you give it to the patient.We recommend that you lock the hand control when you give it to the patient.
Comment: The hand control is unlocked when you switch the unit ON/OFF.
START/STOP/REVERSE functionSTART/STOP/REVERSE function
As with all KINETEC systems, KINETEC Spectra is equipped with a START/STOP/REVERSE function.
When the unit is running, the display reads RUN
Press the key of the hand control. The movement stops. The display reads STOP.
Press the key of the hand control. The movement starts in the opposite direction and the display reads RUN.
Caution:Caution:
For optimum safety, always give the hand control to the patient before starting the system.For optimum safety, always give the hand control to the patient before starting the system.

88
GB
Use of the hand controlUse of the hand control
Quick StartQuick Start
Set up the patient and proceed as below:
Beginning Keys to press Display Remarks
Switch the unit ON
Or
Check if the hand control is not blocked
(See page 7).
If the Warm Up mode is selected, switch OFF
this mode by pressing on the key
Start the session with the original parameters of the
movement (default setting). The value change at the speed of the
movement.
Possible values for each parameter:Possible values for each parameter: Possible values Default setting
• TreatementMode
Normal
•Extensionlimit -10to115° 30°
• Flexion limit -5°to120° 90°
• Speed 1to5(from 30°to 160°per minute) 2
• Force 1 to 6 6
• Extension pause 0 to 900 seconds (15minutes) 0
• Flexion pause 0 to 900 seconds (15minutes) 0
• Timer No timer(00H00) to 24H00 0
Kinetec
30 STOP 15 90
Please wait
KINETEC
Spectra V01.0
WARMUP
30 STOP 15 90
KINETEC
30 run 45 90

99 GB
Use of the hand controlUse of the hand control
How to adjust the basic parameters of the movementHow to adjust the basic parameters of the movement
Beginning Keys to press Display Remarks
To stop the unit Check if the hand control is not blocked
(See page 7).
To display the extension or flexion limit of the movement or The value blinks.
To change the limit if necessary Or The new value blinks.
To validate the new value, press another key
or wait more than 3 seconds.
While the value blinks press
the or key
to change if necessary.
Or
Or to display pause in extension or flexion limit Successive presses on this key selects the
pause at the extension or flexion limit.
To change the pause if necessary Or The new pause value blinks.
To validate the new value press another key
or wait more than 3 seconds. The display shows the
selectedmode. The unit is ready to start with the new
parameters.
KINETEC
30 STOP 15 90
KINETEC
40 Flex 15 110
KINETEC
30 Ext 15 90
KINETEC
40 Flex 15 90
KINETEC
SPEED 2
No timer
KINETEC
PAUSE HIGH 0
Pause low 0
FORCE _ _ _ _ _ _
KINETEC
PAUSE low 15
KINETEC
40 STOP 15 110

1010
GB
Use of the hand controlUse of the hand control
How to modify programs PROGRAM MODEHow to modify programs PROGRAM MODE
Beginning Keys to press Display Remarks
To switch off the unitTo switch off the unit Check if the hand control is not blocked
(See page 7).
Press the two keys at the same time and switch the unit
ON Welcome text during 3 seconds.
Then The program number blinks.
To change the program if necessary or The new program number blinks.
To choose the treatment mode The display indicates the selected treatment
mode, the program number blinks again.
Or
To display the extension or flexion limit of the movement or The value blinks.
KINETEC
Spectra V01.0
Program 2
EMPTY
Program 10
25 kinetec 110
Program 10
25 warmup 110
Program 10
25 kinetec 110
Program 10
30 warmup 100
Program 10
30 warmup 110

1111 GB
Use of the hand controlUse of the hand control
How to modify programs PROGRAM MODE (continued)How to modify programs PROGRAM MODE (continued)
Beginning Keys to press Display Remarks
Tochange the limit of movement if necessary or The new value blinks.
To validate the new value, press another key
or wait more than 3 seconds While the value blinks press
the or key
to change if necessary.
To record the program 10
Then The program 10 has been recorded and the
display indicates the next program so you can
change another program.
OR
To cancel the program
The program 10 has been cancelled and the
display indicates the next program so you can
change another program.
To exit program mode, switch OFF and switch ON the unit. To use the modified program see page 10.
Program 10
30 warmup 120
Program 10
SPEEd: 2
No timer
Pause low 0
FORCE _ _ _ _ _ _
Program 10
Save:+ clear:-
Program 10
Save
Program 11
EMPTY
Program 10
clearing
Program 11
EMPTY
KINETEC
Spectra V01.0

1212
GB
Use of the hand controlUse of the hand control
CommentsComments:
• When a program has been deleted, the display shows
Program table:Program table:
• Program
number TreatmentMode Flexionlimit Extension limit Speed Force Flexion Pause Extension Pause Timer
11
22
33
44
55
66
77
88
99
1010
1111
1212
1313
1414
1515
1616
Program 11
EMPTY

1313 GB
Use of the hand controlUse of the hand control
Using ProgramsUsing Programs
The KINETEC Spectra allows you to store up to 16 programs, including the type of treatment, ROM, speed, load, pauses and timer.
The original parameter values of the program are empty.
These values can be modified and recorded at any time (see ‘How to enter a program’ page 10).
Beginning Keys to press Display Remarks
To stop the unit Check if the hand control is not blocked
(See page 7).
To access the program mode The program number blinks.
To change the program if necessary or The new program number blinks.
To exit and validate the selected program The current parameters have been recorded in
program 3.
To exit without validation of selected program Back to the starting parameters.
Start the unit The value change at the speed of the
movement.
Comments:Comments:
• The values shown in the ‘Display’ column are examples. They actually depend on the stored programs.
• The current movement parameters can be changed while using that program but no data will be stored in the original program.
See the programming mode (p 10) to modify programs.
KINETEC
40 STOP 15 110
PROGRAM 1
70 WARMUP 110
PROGRAM 3
25 warmup 90
KINETEC
Please wait
warmup
25 run 20 90
KINETEC
40 STOP 15 110

1414
GB
Use of the hand controlUse of the hand control
Reading the values of a program: example SPEEDReading the values of a program: example SPEED
Beginning Keys to press Display Remarks
To stop the unit Check if the hand control is not blocked
(See page 7).
To access the program mode The program number blinks.
To change the program if necessary or To change the program if necessary.
To read the speed value Displays the speed value.
After 15 seconds or after pressing on another key
To exit and validate the selected program The current parameters have been recorded in
program 3.
Start the unit The value change at the speed of the
movement.
Comments:Comments:
• The values shown in the ‘Display’ column are examples. They actually depend on the stored programs.
• The current movement parameters can be changed while using that program but no data will be stored in the original program.
See the programming mode (p 10) to modify programs.
KINETEC
40 STOP 15 110
Program 1
70 WARMUP 110
PROGRAM 3
25 warmup 90
PROGRAM 3
Speed 5
PROGRAM 3
70 WARMUP 110
KINETEC
40 STOP 15 110
KINETEC
40 run 15 110

1515 GB
Use of the hand controlUse of the hand control
How to use the WARM UP keyHow to use the WARM UP key
Beginning Keys to press Display Remarks
To stop the unit Check if the hand control is not blocked
(See page 7).
To select the Warm Up mode The display shows the default settings
(see page 8).
To change the movement value if necessary
see page 9.
Startthemovement The value change at the speed of the
movement.
Warm Up rules:
The Kinetec Spectra starts at 70% of the full ROM, increasing 5% of the range
every other cycle until the pre-set ROM is reached.
CommentsComments:
The calculation mode used enables reaching the pre-set ROM in approximately
seven full cycles.
ExampleExample:
For a Warm Up treatment with a pre-set ROM from 0° to 100°.
The first cycle starts from 15° to 85° to 15° and increase 5% each cycle.
KINETEC
40 STOP 15 110
WARM UP
30 STOP 15 90
PLEASE WAIT
WARM UP
15 RUN 15 110
100°
0°
50°
Cycle 1 (70%)
Cycle 2 (75%)
Cycle 3 (80%)
Cycle 4 (85%)
Cycle 5 (90%)
Cycle 6 (95%)
Cycle 7 (100%)
Pre-Set ROM

1616
GB
Use of the hand controlUse of the hand control
How to define the patient tolerance • At the start of a sessionHow to define the patient tolerance • At the start of a session
This function, MODULATION Mode is a way to set within the tolerance of a patient at the beginning of a session.
Set up the patient and proceed as below:
Beginning Keys to press Display Remarks
Switch the unit ON
Or
Check if the hand control is not blocked
(See page 7).
If the Warm Up mode is selected, switch OFF
this mode by pressing on the key
To select the MODULATION Mode The display indicates the keys to be used to
drive the machine, this message is displayed 3
seconds.
To select the pain level
( for flexion, for extension) or The unit is moving to the selected way.
It is waiting for you to select new limits.
To set the pain level when reached, immediately press or The new limit of the movement is recorded
To start the session with the new movement limits The angle display changes with current
movement.
Kinetec
30 STOP 15 90
Please wait
KINETEC
Spectra V01.0
WARMUP
30 STOP 15 90
MODUL.: use + or -
30 STOP 15 90
MODUL.: use + or -
30manual 50 90
MODUL.: use + or -
30manual 50 50
KINETEC
30 run 45 50

1717 GB
Use of the hand controlUse of the hand control
How to define the patient tolerance • During the sessionHow to define the patient tolerance • During the session
This function, The BY-PASS MODE is a way to address the pain threshold of a patient during a session.
IMPORTANT: Can be used only when the machine RUN.IMPORTANT: Can be used only when the machine RUN.
Beginning Keys to press Display Remarks
The unit is running Or Check if the hand control is not blocked
(See page 7).
To select the BYPASS mode The display indicates the keys used to run the
machine.
Thismessage is displayed 3 seconds.
To select the NEW pain level or The unit is moving to the selected way.
It is waiting for you to select new limits.
To set the NEW pain level when reached, immediately
press or The new limit of the movement is recorded.
Continuethe session with the new movement limits. The angle display changes with current
movement.
Kinetec
30 RUN 15 90
WARMUP
30 RUN 15 90
PYPASS: use + or -
30 RUN 15 90
BYPASS: use + or -
30bypass 100 90
bypass.: use + or -
30manual 50 50
WARMUP
30 run 45 50

1818
GB
TIME 02h25
30 STOP 15 90
Use of the hand controlUse of the hand control
Session TimeSession Time
This function shows the running time (in minutes) of the session (motor functioning).
• It directly accessible by the key , the display shows
This counter is reset each time the unit is switched ON.

1919 GB
Use of theUse of the Kinetec Patient Pad KitKinetec Patient Pad Kit
The KINETEC Patient Pad Kit is designed for rapid fitting, optimal hygiene and maximum patient comfort.
- For using and positioning the straps, please refer to here under. Make sure that the self-adhesive parts (26) are visible.
- Place the sponge side next to the skin.
FOR OPTIMAL HYGIENE, A NEW SET OF PADS SHOULD BE USED FOR EACH PATIENT.FOR OPTIMAL HYGIENE, A NEW SET OF PADS SHOULD BE USED FOR EACH PATIENT.
Each cover is provided with a label to record the patient’s nameEach cover is provided with a label to record the patient’s name
CLEANING:
- Sterilization of the pads ( if necessary): Sterilized at 134 °C during 18 minutes.
- Disinfecting of the pads: Washing at 30°C with use of a disinfecting solution during the rising cycle.
Example of product that can be used: Solution " Baclinge" at 0.125 % or "Souplanios" at 0,125% from ANIOS Laboratory.
A complete list of distributors in your country is available on request.
The KINETEC Spectra is delivered with a complete set. Components:
- 4 straps
- 1 foot support
- 1 cover
2626
Other manuals for Spectra
2
Table of contents
Other Kinetec Medical Equipment manuals

Kinetec
Kinetec Maestra Series User manual

Kinetec
Kinetec INNOV8 IQ User manual

Kinetec
Kinetec Hill-Rom 900 User manual

Kinetec
Kinetec KINETEC CEM User manual

Kinetec
Kinetec Manosplint Dry Heat Oven User manual

Kinetec
Kinetec Spectra Essential User manual

Kinetec
Kinetec Spectra User manual

Kinetec
Kinetec Prima Advance CPM concept User manual

Kinetec
Kinetec 15 Series User manual

Kinetec
Kinetec Centura B&W User manual