Replenishment Solutions
981089 – September 1995 17
[7] Compare your results to the recommended volumes listed in the Replenishment Rate Sheet, Publication
No. 1C0578, which is located in the front pocket of your publications binder.
Note
Any time you adjust the replenishment rate, you are adjusting the replenishment volume for both developer and
fixer. Each time you increase or decrease the replenishment rate by inserting and removing a sheet of film under
either the Right (+) Film sensor or Left (-) Film sensor, you are changing both of the replenishment rates by
approximately 5 mL.
(a) If you measured the correct volume of developer replenishment solution, do not make any
replenishment adjustments. Go to Step 8 on Page 18.
(b) If you measured more than the recommended volume of developer replenishment solution, go to
Step (m) below.
(c) If you measured less than the recommended volume of developer replenishment solution, do the
following steps to increase the rate of replenishment.
(d) Empty the replenishment solution from the Graduated Cylinder.
(e) Insertand remove asmall sheet offilmunder the Right(+) FilmSensoronthe right sideof the FeedTray.
Listen for the processor to beep once. Each beep represents a 5 mL increase in replenishment volume.
Note
To make large changes in the replenishment rate, you may insert and remove a sheet of film from the Right (+) Film
Sensor several times consecutively.
(f) Again, insert the Developer Replenishment Check Tube into the Graduated Cylinder.
(g) Again, insert and remove a small sheet of film under the Middle Film Sensor. Wait for the processor to
beep twice and deliver replenishment solution to the Graduated Cylinder.
(h) After the Replenishment Pump stops, repeat the previous step two additional times.
(i) Measure the volume of developer replenishment solution in the Graduated Cylinder. Divide the volume
by 3 to calculate the average measurement.
(j) If the average measurement is less than the recommended volume by more than 3 mL, continue increasing
the replenishment rate as necessary.
(k) When the developer replenishment rate is correct, push the Developer Replenishment Check Tube down
into the Developer Tank to its original position.
(l) Advance to Step 9 on Page 18.
(m) If you measured more than the recommended volume of developer replenishment solution, do the
following steps to decrease the rate of replenishment.
(n) Empty the replenishment solution from the Graduated Cylinder.
(o) Insert and remove a small sheet of film under the Left (-) Film Sensor on the left side of the Feed Tray.
Listen for the processor to beep once.
Note
To make large changes in the replenishment rate, you may insert and remove a sheet of film from the Left (-) Film
Sensor several times consecutively.
(p) Insert the Developer Replenishment Check tube into the Graduated Cylinder.
(q) Again, insert and remove a small sheet of film under the Middle Film Sensor. Wait for the processor to
beep twice and deliver replenishment solution to the Graduated Cylinder.
(r) After the Replenishment Pump stops, repeat the previous step two additional times.
(s) Measure the volume of developer replenishment solution in the Graduated Cylinder. Divide the volume
by 3 to calculate the average measurement.
(t) If the average measurement is more than the recommended volume by more than 3 mL, continue
decreasing the replenishment rate as necessary.