INTRODUCTION 1
3
OPTISONIC 6300
www.krohne.com02/2019 - 4007265101 - EX OPTISONIC 6300 R04 en
1.1 Safety instructions for the operator
These additional instructions are an extension to the standard documentation and only apply to
the Explosion protected versions of the devices described in this document. All technical
information as described in the standard documentation is applicable, when not specifically
excluded, completed or replaced by the instructions in these additional instructions.
1.2 Approval
The ultrasonic flowmeters are manufactured according to the European directive 2014/34/EU
and IECEx 02 Certification Systems. The flowmeters are approved for installation and use in
hazardous classified locations of Zone 1 and 2 and are in accordance with the European
standards of the EN/IEC 60079 series. They have approval number:
OPTISONIC 6000 xxxxxx-Ex sensor: KIWA 17ATEX0034 X / IECEx KIWA 17.0017X
KIWA 17ATEX0034 X / IECEx KIWA 17.0017XKIWA 17ATEX0034 X / IECEx KIWA 17.0017X
KIWA 17ATEX0034 X / IECEx KIWA 17.0017X
UFC 300 F(/i)-Ex converter: KIWA 18ATEX0007 X / IECEx KIWA 18.0003X
KIWA 18ATEX0007 X / IECEx KIWA 18.0003XKIWA 18ATEX0007 X / IECEx KIWA 18.0003X
KIWA 18ATEX0007 X / IECEx KIWA 18.0003X
The QPS
QPSQPS
QPS approval covers the applicable standards : UL/FM/CSA/ISA.
The original project ID is LR1338-9
LR1338-9LR1338-9
LR1338-9.
1.3 OPTISONIC 6000 xxxxxx-Ex
The OPTISONIC 6000 xxxxxx-Ex is a clamp-on ultrasonic flow sensor and has intrinsically safe
transducer circuits. It is available in three sizes (xxxxxx = small, medium or large), designed for
the size of the pipeline on which the flow sensor is installed.
It is marked with the explosion safety code:
II 2 G Ex ia IIC T6...T4
II 2 G Ex ia IIC T6...T4II 2 G Ex ia IIC T6...T4
II 2 G Ex ia IIC T6...T4
II 2 G Ex ia IIC T6...T2
II 2 G Ex ia IIC T6...T2II 2 G Ex ia IIC T6...T2
II 2 G Ex ia IIC T6...T2 (XT versions)
The intrinsically safe transducer connections of the OPTISONIC 6000 xxxxxx-Ex are connected to
an associated device and have the following maximum values:
WARNING!
•
Do not change the device. Unauthorized changes may affect the explosion safety of the
devices.
•
The prescriptions and regulations as well as the electrical data described in the EC type
examination certificate must be obeyed.
•
Beside the instructions for electrical installations in non-hazardous locations according to
the applicable standard (equivalent to HD-IEC 60364), especially the regulations in EN/IEC
60079-14 "Electrical installations in hazardous locations" or equivalent national standard
(e.g. DIN VDE 0165 Part 1) must be strictly followed.
•
Installation, establishment, utilization and maintenance are only allowed to be executed by
personnel with an education in explosion safety!
U
i
= 85 V I
i
= 250 mA P
i
d0531 W C
i
d45 nF L
i
d400 μH