2015-001 Supplemental sheet for CV5030 V 1.0 RevE – 04/2019
Supplemental sheet for Instructions for Use for CV5030 V3.3 RevK - 04/2019
New chapter 5.4.1 Bath insert for HistoCore SPECTRA ST
The Bath Insert and Clip is required to to be used with the HistoCore SPECTRA ST specimen slide holder for
30 specimens on the Coverslipper CV5030.
Fig. A
Fig. B
Fig. C
1
Preparing the CV5030 for the HistoCore SPECTRA
ST Bath Insert
Attaching the Clip to the specimen slide holder
• To avoid damaging the specimens speci-
men slides should never be inserted into a
specimen holder on which a clip has alrea-
dy been attached.
• The clip should only be attached after the
specimen holder has been removed from
the SPECTRA ST. Using the Clip during
the staining process can cause collisions
within the device, which can lead to inter-
ruptions of the staining process.
• The specimen slide holder handle (1, Fig. C)
has to be removed from the specimen slide
holder before inserting it into the CV5030.
1
• Open the load drawer of the CV5030 (see Chap. 4.10
Fig. 27) and pull the loading bath forward.
• Remove the existing bath insert, check the bath for
glass debris and clean if necessary.
• Insert the HistoCore SPECTRA ST Bath Insert (see
Fig. A ) into the loading bath.
• Remove the specimen slide holder from the HistoCore
SPECTRA ST.
• Unfold the specimen slide holder handle and hold the
specimen slide holder at an angle (Fig. B).
• Place the Clip carefully over the specimen slides and
take care that only one specimen slide sits in each
slot of the Clip (see Fig. C).
• Place the specimen slide holder back horizontally.
• Using both hands simultaneously, carefully press the
Clip downwards until it snaps into position (Fig. C).
• Remove the specimen slide holder handle (1, Fig. C).
• Open the load drawer of the CV5030 and insert the
specimen slide holder into the HistoCore SPECTRA
ST Bath Insert.
• Cover slip the specimen slides, as described in
Chap. 5.5.