Leica HI1220 User manual

Instructions for Use
Leica HI1220
Flattening Table
Leica HI1220 V 3.3 , English 07/2016
Order No.: 14 0423 80101 RevG
Always keep this manual with the instrument.
Read carefully before working with the instrument.


3
Leica HI1220 Flattening Table
The information, numerical data, notes and value
judgments contained in this manual represent
the current state of scientific knowledge and
state-of-the-art technology as we understand it
following thorough investigation in this field.
We are under no obligation to update the pres-
ent manual periodically and on an ongoing basis
according to the latest technical developments,
nor to provide our customers with additional cop-
ies, updates etc. of this manual.
To the extent permitted in accordance with
the national legal system as applicable in each
individual case, we shall not be held liable for
erroneous statements, drawings, technical illus-
trations etc. contained in this manual. In particu-
lar, no liability whatsoever is accepted for any
financial loss or consequential damage caused
by or related to compliance with statements or
other information in this manual.
Statements, drawings, illustrations and other
information regarding the contents or technical
details of the present Instructions for Use are
not to be considered warranted characteristics
of our products.
These are determined only by the contract
provisions agreed between ourselves and our
customers.
Leica reserves the right to change technical
specifications as well as manufacturing process-
es without prior notice. Only in this way is it pos-
sible to continuously improve the technology and
manufacturing techniques used in our products.
This document is protected under copyright laws.
All copyrights to this documentation are held by
Leica Biosystems Nussloch GmbH.
Any reproduction of text and illustrations (or of
any parts thereof) by means of print, photocopy,
microfiche, web cam or other methods—includ-
ing any electronic systems and media—requires
express prior permission in writing by Leica
Biosystems Nussloch GmbH.
For the instrument serial number and year of
manufacture, please refer to the nameplate on
the back of the instrument.
© Leica Biosystems Nussloch GmbH
NOTE
Leica Biosystems Nussloch GmbH
Heidelberger Str. 17 - 19
D-69226 Nussloch
Germany
Phone: +49 (0)6224 143-0
Fax: +49 (0)6224 143-268
Internet: http://www.LeicaBiosystems.com

4Instructions for Use V 3.3 RevG – 07/2016
Table of Contents
1. Important Information ....................................................................................................5
1.1 Symbols and their meanings....................................................................................................... 5
1.2 Qualification of personnel........................................................................................................... 6
1.3 Intended use of instrument......................................................................................................... 6
1.4 Instrument type............................................................................................................................. 6
2. Safety ................................................................................................................................7
2.1 Safety notes................................................................................................................................... 7
2.2 Warnings........................................................................................................................................ 8
3. Instrument Components and Specifications .............................................................9
3.1 Technical data............................................................................................................................... 9
4. Instrument Setup...........................................................................................................10
4.1 Installation site requirements................................................................................................... 10
4.2 Standard delivery........................................................................................................................ 10
4.3 Unpacking and setting up the HI1220...................................................................................... 11
4.4 Setting up the power supply..................................................................................................... 11
5. Operation........................................................................................................................12
5.1 Control panel fields on the instrument.................................................................................... 12
5.2 Switching the instrument on..................................................................................................... 12
5.3 Setting the temperature............................................................................................................. 13
5.4 Switching off the instrument..................................................................................................... 13
6. Cleaning and Maintenance ........................................................................................14
6.1 Cleaning the instrument............................................................................................................. 14
6.2 Replacing fuses........................................................................................................................... 14
7. Warranty and Service..................................................................................................15
8. Decontamination Certificate (Master)......................................................................16

5
Leica HI1220 Flattening Table
1. Important Information
Instrument surfaces which become hot
during operation are marked with this
symbol.
Avoid direct contact with these surfaces
- they may cause burns.
REF
SN
The CE labeling shows that the product
corresponds to one or more applicable
European directives.
Environmental protection symbol of the
China RoHS directive. The number in the
symbol indicates the "Environment-friendly
Use Period" of the product. The symbol is
used if a substance restricted in China is
used in excess of the maximum permitted
limit.
Symbol for labeling electrical and electron-
ic equipment in accordance with Section 7
of the German Electrical and Electronic
Equipment Act (ElektroG). ElektroG is the
law regarding the sale, return and environ-
mentally sound disposal of electrical and
electronic equipment.
Order no.
Serial number
Warnings appear in a gray box and are
marked by a warning triangle .
Observe the instructions for use!
Manufacturer
Notes,
i.e. important user information, appear
in a gray box and are marked by an
information symbol .
1.1 Symbols and their meanings
(5) Numbers and parentheses refer to item
numbers in the illustrations.
START Function keys to be pressed on the
instrument control panel are in bold-
print capital letters.
The package contents are fragile and must
be handled with care.
Indicates the correct upright position of the
package.
Date of manufacture
Caution, consult the instructions for
use for cautionary information.
The package must be kept in a dry
environment.
Indicates the temperature range per-
mitted for transporting the package.
Minimum -29 °C, Maximum +50 °C
Indicates the temperature range per-
mitted for storing the package.
Minimum +5 °C, Maximum +50 °C
It allows maximum 3 stacks layers.
Indicates the humidity range permitted
for storing and transporting the
package.
Minimum 10 % r.H., Maximum 85 % r.H
Indicates the item can be recycled
where correct facilities exist.

6Instructions for Use V 3.3 RevG – 07/2016
1.2 Qualification of personnel
• The Leica HI1220 may be operated by trained
laboratory personnel only.
• All laboratory personnel designated to operate
this instrument must read these Instructions
for Use carefully and must be familiar with all
technical features of the instrument before
attempting to operate it.
Fig. 1
1.3 Intended use of instrument
The HI1220 microscope slide flattening table for
paraffin sections is a heating plate with high heat
output and precise temperature control.
The HI1220 is suitable for flattening and drying
cut tissue samples in all fields of biomedical
research and routine diagnostics.
1.4 Instrument type
All information provided in these Instructions for
Use applies only to the instrument type indicated
on the cover page.
A nameplate indicating the instrument serial num-
ber is attached to the rear side of the instrument.
Any other use of the instrument is
considered improper.
Failure to adhere to these instructions
may result in an accident, personal
injury, damage to the instrument or
accessory equipment.
Proper and intended use includes
compliance with all inspection and
maintenance instructions, along with
the observance of all instructions in
the Instructions for Use.
Fig. 1 is provided as an example only
and shows a valid nameplate for this
instrument with the necessary in-
formation about instrument type and
power requirement. The precise data
for the various versions is specified in
Chapter 3.1 "Technical data".
1. Important Information
The Regulatory Compliance Mark
(RCM) indicates a device’s compliance
with applicable ACMA technical stan-
dards of New Zealand and Australia
- that is, for telecommunications, radio
communications, EMC and EME.

7
Leica HI1220 Flattening Table
2. Safety
These Instructions for Use include important instruc-
tions and information related to the operating safety
and maintenance of the instrument.
The Instructions for Use are an important part of the
product, and must be read carefully prior to startup
and use and must always be kept near the instrument.
This instrument has been built and tested in accor-
dance with the safety requirements for electrical
equipment for measurement, control, and laboratory
use.
To maintain this condition and ensure safe operation,
the user must observe all notes and warnings con-
tained in these Instructions for Use.
2.1 Safety notes
These Instructions for Use must be ap-
propriately supplemented as required
by the existing regulations on accident
prevention and environmental safety in
the operator's country.
The instrument's CE certificate can be
found on the Internet at:
http://www.LeicaBiosystems.com
• The safety and caution notes in this chapter must be observed at all times.
• Be sure to read these notes even if you are already familiar with the operation and use of
other Leica products.
• The protective devices located on the instrument and the accessories must not be removed
or modified. Only qualified service personnel authorized by Leica may repair the instrument
and access its internal components.
Residual risks
• The instrument has been designed and constructed with the latest state-of-the-art technol-
ogy and according to recognized standards and regulations with regard to safety technology.
Operating or handling the instrument incorrectly can place the user or other personnel at
risk of injury or can cause damage to the instrument or other property. The instrument may
be used only as intended and only if all of its safety features are in proper working condition.
Malfunctions which could impede safety must be remedied immediately.
• Only original spare parts and permitted original accessories may be used.
The instrument MUST be connected to a grounded power socket. Use only a provided power
cable that is intended for the local power supply.

8Instructions for Use V 3.3 RevG – 07/2016
• In order to properly service the instrument, it is indispensable to read the User Manual before
switching on the instrument and to familiarize yourself with all technical details.
• Before connecting the instrument to the line voltage, ensure that the electrical power require-
ments of your laboratory match the values on the nameplate of the instrument.
• When installing the power cable, always be sure to route it so that it cannot contact the
heated surfaces of the instrument at any time.
• The instrument is designed for indoor use only.
• The instrument must be switched off and unplugged from the power supply during all repair
and service work.
• During operation, the instrument surface can be very hot.
Risk of injury!
Do not store any combustible and flammable substances near the instrument.
2.1 Safety notes (continued)
The safety devices installed in this instrument by the manufacturer only constitute the basis
for accident prevention. Operating the instrument safely is, above all, the responsibility of the
owner, as well as the designated personnel who operate, service or repair the instrument.
To ensure trouble-free operation of the instrument, make sure to comply with the following
instructions and warnings.
Please note that electrostatic charge may result through direct or indirect contact with the
Leica HI1220.
2.2 Warnings
2. Safety

9
Leica HI1220 Flattening Table
3. Instrument Components and Specifications
Nominal supply voltages: Two factory-preset voltages (not user-adjustable):
100-120 VAC + 10 %
230-240 VAC + 10 %
Nominal frequency: 50/60 Hz
Nominal power: 350 VA
Main fuse: 2 melting fuses, 5x20 mm, UL-approved
For 100-120 V: 5x20 mm, 2x T 5A L250 VAC
For 230-240 V: 5x20 mm, 2x T 2.5A L250 VAC
Dimensions: 350x310x100 mm
Empty weight: 3.6 kg (w/o packaging)
Relative humidity: 20-80 %, non-condensing
IEC 1010 classification: Protection class: 1
Pollution degree: 2
Overvoltage category: II
Operating temperature range: +15 °C to +40 °C
Temperature control range: Room temperature to 75 °C
Operating elevation: maximum of 2000 m above sea level
IP protection class (IEC 60529): IP20
3.1 Technical data

10 Instructions for Use V 3.3 RevG – 07/2016
4. Instrument Setup
4.1 Installation site requirements
• When installing the power cable, always be sure to route it so that it cannot contact the
heated surfaces of the instrument at any time.
• The power socket to which the instrument is to be connected to has to be located near the
instrument and easily accessible.
• The power supply must be at a distance no greater than the length of the power cable – an
extension cable must NOT be used.
• The substrate must be largely vibration-free and have sufficient load capacity and rigidity
for the weight of the instrument.
• Avoid impacts, bright direct light, and excessive temperature fluctuations.
• The instrument must be connected to a suitable power socket. Use only the provided power
cable, which is intended for the local power supply.
4.2 Standard delivery
The Leica HI1220 standard delivery includes:
1 Leica HI1220 basic instrument
1 Dust cover ........................................................................................................................ 14 0415 41178
1 Set of fuses:
2 fuses, T2.5A ............................................................................................................... 14 6943 02500
2 fuses, T5A .................................................................................................................. 14 6943 05000
1 Instructions for Use, printed (German/English,
with Language CD 14 0423 80200) .............................................................................. 14 0423 80001
The country specific power cord needs to be ordered separately. Please find a list of all power
cords available for your device on our website www.LeicaBiosystems.com within the product
section.
Check the delivery carefully against the packing list, delivery note and your order.
Should you find any discrepancies, please contact your Leica sales office without delay.

11
Leica HI1220 Flattening Table
4. Instrument Setup
4.3 Unpacking and setting up the HI1220
Fig. 3
Fig. 2
Set the carton on a flat surface, cut through the adhesive tape (Fig. 2) and open the carton (Fig. 3).
Remove the accessories and carefully lift the
instrument out of the carton (Fig. 4).
Hold the instrument firmly with one hand, pull off
both foam sections (Fig. 5) and remove the dust
cover.
4.4 Setting up the power supply
• Before connecting the power cable, make
sure that the power switch (3) on the rear of
the instrument (Fig. 6) is switched to "0" = OFF.
• Use the instrument with the provided power
cable only.
• Insert the connector of the power cable
(Fig. 5, 2) into the connection socket (4) and
plug the power plug into the power socket Fig. 6
34
Fig. 4
õ
õ
Fig. 5
1
2

12 Instructions for Use V 3.3 RevG – 07/2016
5. Operation
5.1 Control panel fields on the instrument
Fig. 7
65 87 9
10
5= POWER LED
Is illuminated when the instrument
is switched on with the main power
switch.
6= SET
Indicates the selected target tempera-
ture.
7= Set key: lower temperature.
8= Set key: increase temperature.
9= RUN/STOP
Switch instrument on or off.
10 = Display
5.2 Switching the instrument on
• Switch on the instrument with the power switch on the left rear panel of the instrument
(Fig. 6, 3).
• The green LED display POWER on the front panel is then illuminated.

13
Leica HI1220 Flattening Table
5.3 Setting the temperature
• Pressing the RUN/STOP button on the front panel brings the temperature of the instrument
to the latest stored target value. The displayed temperature corresponds to the current
actual value.
• The set target temperature is shown in the display (Fig. 7, 10).
• To adjust to the desired temperature, hold down the corresponding set key (Fig. 7, 7+8). The
display changes slowly for the first 8 digital steps, then more quickly after that.
The temperature display refers to the target value. After the desired temperature is
reached, release the set key. The displayed value is saved automatically. This stored
target value will continue to be displayed for about 2 seconds, then the current actual
temperature display will appear again.
• The target temperature can be checked by pressing the SET key (Fig. 7, 6).
The target value remains saved both when the instrument is switched off (RUN/STOP
key or power switch) and in case of power failure and disconnection from the power
supply.
5.4 Switching off the instrument
• The instrument can be switched off by pressing the RUN/STOP key.
You do not have to switch the main power switch on or off each time for everyday
routine use.
5. Operation
After changing the target temperature, the temperature may temporarily rise once a lit-
tle. Therefore extremely temperature-sensitive samples should not be positioned until
after reaching the set temperature (observe the temperature display!).

14 Instructions for Use V 3.3 RevG – 07/2016
6. Cleaning and Maintenance
6.1 Cleaning the instrument
• Before cleaning the instrument, turn off using the power switch (Fig. 6, 3) and disconnect
from the power supply.
• The instrument has a continuous front panel and no lateral cooling slots, so the surfaces
are smooth and easy to clean.
• All surface materials are resistant to common laboratory detergents.
The instrument must be switched off and have cooled down for cleaning.
Do not use organic solvents (e.g. alcohol or xylene)!
6.2 Replacing fuses
Switch the unit off and unplug it. Use
only the supplied replacement fuses.
Both fuses must have the same rating
(check the imprint!). Fuses must always
be switched out in pairs.
Fig. 8
• Insert a small screwdriver into the cutout
(Fig. 8, 11) and carefully pry out the fuse
insert (12).
• Remove the fuse holder and replace the de-
fective fuses with two new ones (13).
• Then reinsert the fuse holder and press it
into place.
12
13
11
õ

15
Leica HI1220 Flattening Table
7. Warranty and Service
Warranty
Leica Biosystems Nussloch GmbH guarantees that the contractual prod-
uct delivered has been subjected to a comprehensive quality control
procedure based on the Leica in-house testing standards, and that the
product is faultless and complies with all technical specifications and/
or characteristics warranted.
The scope of the warranty is based on the content of the concluded
agreement. The warranty terms of your Leica sales organization or the
organization from which you have purchased the contractual product
shall apply exclusively.
Service information
If you require technical service or replacement parts, please contact
your Leica sales office or dealer who sold the product.
The following inputs are required:
• Model name and serial number of the instrument.
• Location of the instrument and name of a contact person.
• Reason for the service call.
• Date of delivery.
Decommissioning and disposal
The instrument or parts of the instrument must be disposed of according
to existing applicable, local regulations.
Special attention should be paid to the lithium cell of the electronic
circuit board!

16 Instructions for Use V 3.3 RevG – 07/2016
8. Decontamination Certificate (Master)
Leica Biosystems Nussloch GmbH
Heidelberger Str. 17-19
69226 Nussloch, Germany
Phone: ++49 (0) 6224 143 0
Fax: ++49 (0) 6224 143 268
www.LeicaBiosystems.com
Tick Box A if applicable. Otherwise please complete all parts of B, providing further
information as requested or appropriate.
Decontamination Certificate
Nameplate information Model (see nameplate)* SN (see nameplate)*
REF (see nameplate)*
This equipment has not been in contact with unfixed biological samples.
A
BThis equipment has been exposed internally or externally to hazardous materials as indicated
below:
Yes No
Blood, body fluids, pathological samples
Other biohazards
Chemicals/substances hazardous to health
Other hazards
Yes
This equipment has been cleaned and decontaminated:
Yes No If yes, give details of the method:
If no**, please indicate why not:
Please provide further detail here:
Please provide further detail here:
** Such equipment must not be returned without the written agreement of Leica Biosystems.
The equipment has been prepared to ensure safe handling/transportation.
Whenever possible, please use the original transportation case/box.
Yes No
1
3
2
Signature/Date*
Name*
Position*
eMail
Institute*
Department*
Address*
Phone*Fax
Important - to avoid refusal of shipment:
Place one copy in the unit prior to packaging, or hand it over to the service engineer. Customer assumes all responsibility for the
immediate return shipment of articles sent to Leica without proper decontamination documentation.
If you have any further questions, please call your local Leica organization.
Leica Internal Use: If applicable, note corresponding Job and RAN-/RGA-Number:
Job Sheet No.: _______________ BU Return Authorization Number: ______________ SU Return Goods Authorization: ______________
Version 3.5 - 01/2013, Order-No. 14 0700 83101
Dear Customer,
Any product that is to be returned to Leica Biosystems or serviced on site, must be cleaned and decontaminated in the appropriate manner. Since it is not possible to
decontaminate for prion diseases, such as CJD, BSE, CWD etc., equipment exposed to specimens containing prion diseases cannot be returned to Leica Biosystems
for repair. On-site repair of prion contaminated equipment will only be conducted after the Field Service Engineer has been educated in the risks, instructed in the
policies and procedures of the institution, and provided with personal protective equipment. Please fill out this confirmation carefully and enclose a copy with the
instrument. Attach the confirmation to the outside of the flight case or hand it directly to the service technician.
Packages will not be opened, nor servicing commenced until the Company or service engineer have received a satisfactory certificate. Should returned
goods be considered a hazard by the Company, they will be returned immediately to the customer at his/her expense. Note: Microtome knives must be in boxes.
Mandatory information: Fields marked with * are mandatory. Depending on whether the instrument is contaminated, please also complete either section A
or section B.
Radioactivity

17
Leica HI1220 Flattening Table
8. Decontamination Certificate (Master)
Leica Biosystems Nussloch GmbH
Heidelberger Str. 17-19
69226 Nussloch, Germany
Phone: ++49 (0) 6224 143 0
Fax: ++49 (0) 6224 143 268
www.LeicaBiosystems.com
Tick Box A if applicable. Otherwise please complete all parts of B, providing further
information as requested or appropriate.
Decontamination Certificate
Nameplate information Model (see nameplate)* SN (see nameplate)*
REF (see nameplate)*
This equipment has not been in contact with unfixed biological samples.
A
BThis equipment has been exposed internally or externally to hazardous materials as indicated
below:
Yes No
Blood, body fluids, pathological samples
Other biohazards
Chemicals/substances hazardous to health
Other hazards
Yes
This equipment has been cleaned and decontaminated:
Yes No If yes, give details of the method:
If no**, please indicate why not:
Please provide further detail here:
Please provide further detail here:
** Such equipment must not be returned without the written agreement of Leica Biosystems.
The equipment has been prepared to ensure safe handling/transportation.
Whenever possible, please use the original transportation case/box.
Yes No
1
3
2
Signature/Date*
Name*
Position*
eMail
Institute*
Department*
Address*
Phone*Fax
Important - to avoid refusal of shipment:
Place one copy in the unit prior to packaging, or hand it over to the service engineer. Customer assumes all responsibility for the
immediate return shipment of articles sent to Leica without proper decontamination documentation.
If you have any further questions, please call your local Leica organization.
Leica Internal Use: If applicable, note corresponding Job and RAN-/RGA-Number:
Job Sheet No.: _______________ BU Return Authorization Number: ______________ SU Return Goods Authorization: ______________
Version 3.5 - 01/2013, Order-No. 14 0700 83101
Dear Customer,
Any product that is to be returned to Leica Biosystems or serviced on site, must be cleaned and decontaminated in the appropriate manner. Since it is not possible to
decontaminate for prion diseases, such as CJD, BSE, CWD etc., equipment exposed to specimens containing prion diseases cannot be returned to Leica Biosystems
for repair. On-site repair of prion contaminated equipment will only be conducted after the Field Service Engineer has been educated in the risks, instructed in the
policies and procedures of the institution, and provided with personal protective equipment. Please fill out this confirmation carefully and enclose a copy with the
instrument. Attach the confirmation to the outside of the flight case or hand it directly to the service technician.
Packages will not be opened, nor servicing commenced until the Company or service engineer have received a satisfactory certificate. Should returned
goods be considered a hazard by the Company, they will be returned immediately to the customer at his/her expense. Note: Microtome knives must be in boxes.
Mandatory information: Fields marked with * are mandatory. Depending on whether the instrument is contaminated, please also complete either section A
or section B.
Radioactivity

18 Instructions for Use V 3.3 RevG – 07/2016
Notes


www.LeicaBiosystems.com
Leica Biosystems Nussloch GmbH
Heidelberger Straße 17-19
D- 69226 Nussloch
Phone: +49 6224 - 143 0
Fax: +49 6224 - 143 268
Web: www.LeicaBiosystems.com
Table of contents
Other Leica Laboratory Equipment manuals

Leica
Leica EM FC7 User manual

Leica
Leica HistoCore MULTICUT User manual

Leica
Leica Envisu C Series Instructions for use

Leica
Leica VT1000 S User manual

Leica
Leica VT1200 User manual

Leica
Leica Viva GNSS User manual

Leica
Leica EM TXP User manual

Leica
Leica FlexLine Instruction Manual

Leica
Leica TP1020 User manual

Leica
Leica CM1950 User manual

Leica
Leica TCS SP8 User manual

Leica
Leica CM3600 XP User manual

Leica
Leica Z6 APO User manual

Leica
Leica CV5030 User manual

Leica
Leica RM2125 User manual

Leica
Leica CV5030 User manual

Leica
Leica RM2235 User manual

Leica
Leica AutoStainerXL User manual

Leica
Leica RM2125RTS User manual

Leica
Leica TCS SP8 SMD User manual