INSTRUCTION MANUAL
BEFORE USE ....
Thank you for choosing M-System. Before use, please check
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System’s Sales Office or representatives.
■PACKAGE INCLUDES:
Signal conditioner ...............................................................(1)
■MODEL NO.
Confirm Model No. marking on the product to be exactly
what you ordered.
■INSTRUCTION MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection and
basic maintenance procedures.
POINTS OF CAUTION
■CONFORMITY WITH EU DIRECTIVES
• The equipment must be mounted inside a panel.
• The actual installation environments such as panel con-
figurations, connected devices, connected wires, may af-
fect the protection level of this unit when it is integrated
in a panel system. The user may have to review the CE
requirements in regard to the whole system and employ
additional protective measures to ensure the CE conform-
ity.
• Install lightning surge protectors for those wires connect-
ed to remote locations.
■POWER INPUT RATING & OPERATIONAL RANGE
• Locate the power input rating marked on the product and
confirm its operational range as indicated below:
24V DC rating: 24V ±10%, approx. 80mA
■GENERAL PRECAUTIONS
• Before you remove the unit or mount it, turn off the power
supply and input signal for safety.
■ENVIRONMENT
• Indoor use.
• When heavy dust or metal particles are present in the
air, install the unit inside proper housing with sufficient
ventilation.
• Do not install the unit where it is subjected to continuous
vibration. Do not subject the unit to physical impact.
• Environmental temperature must be within -5 to +55°C
(23 to 131°F) with relative humidity within 30 to 90% RH
in order to ensure adequate life span and operation.
• Be sure that the ventilation slits are not covered with ca-
bles, etc.
■WIRING
• Do not install cables close to noise sources (relay drive
cable, high frequency line, etc.).
• Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.
■AND ....
• The unit is designed to function as soon as power is sup-
plied, however, a warm up for 30 minutes is required for
satisfying complete performance described in the data
sheet.
COMPONENT IDENTIFICATION
Body
Zero Adj.
Span Adj.
Connection
Diagram
Specifications
Euro Type
Connector Terminal Block
INSTALLATION
Set the unit so that its DIN rail adaptor is at the bottom.
Hang the upper hook at the rear side of the unit on the DIN
rail and push in the lower. When removing the unit, push
down the DIN rail adaptor utilizing a minus screwdriver
and pull.
DIN Rail
35 mm wide
Spring Loaded
DIN Rail Adaptor
LOW FREQUENCY TRANSMITTER
(50 Hz minimum) MODEL F2SP
F2SP
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-5434 Rev.6 P. 2 / 2
TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the connection diagram on the side of the unit.
■EXTERNAL DIMENSIONS unit: mm (inch)
1234
5678
107 (4.21) [1.5 (.06)]
DIN RAIL
35mm wide
22.5 (.89)
75 (2.95)
• When mounting, no extra space is needed between units.
■CONNECTION DIAGRAM
+
–
OUTPUT
5
6
+
–
INPUT
1
3
2
+
SENSOR EXCITATION
1
3
2
+
4kΩ
• Dry Contact
[ Input Connection Examples ]
+
–
–+12V
PWR
• Voltage Pulse
1
3
2
+
+
–
–+12V
PWR
+
–
POWER
7
8
WIRING INSTRUCTIONS
■EURO TYPE CONNECTOR TERMINAL
Applicable wire size: 0.2 to 2.5 mm2
Stripped length: 7 mm
CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Power input voltage: Check voltage across the terminal
7 – 8 with a multimeter.
3) Input: Check input terminal 1 (+) – 2 (–) with an oscil-
loscope.
With voltage pulse input, the signal should be below 1V
at L level, while it should exceed 2V at H level.
4) Output: Check that the load resistance meets the de-
scribed specifications.
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case
of regular calibration, adjust the output as explained in the
following.
■HOW TO CALIBRATE THE OUTPUT SIGNAL
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 30 minutes.
1) ZERO: Apply 10% input and adjust output to 10%.
2) SPAN: Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 10% input.
4) When ZERO value is changed, repeat the above proce-
dure 1) – 3).
MAINTENANCE
Regular calibration procedure is explained below:
■CALIBRATION
Warm up the unit for at least 30 minutes. Apply 10%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the “ADJUST-
MENT PROCEDURE” explained earlier.
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protector for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.

Other M-system Transmitter manuals

M-system M6DVF User manual

M-system

M-system M6DVF User manual

M-system M3LR User manual

M-system

M-system M3LR User manual

M-system M3LV User manual

M-system

M-system M3LV User manual

M-system W5-UNIT User manual

M-system

M-system W5-UNIT User manual

M-system KAC User manual

M-system

M-system KAC User manual

M-system 10SP User manual

M-system

M-system 10SP User manual

M-system KCEC User manual

M-system

M-system KCEC User manual

M-system VOS2T-R User manual

M-system

M-system VOS2T-R User manual

M-system Mini-M M2CEC User manual

M-system

M-system Mini-M M2CEC User manual

M-system Mini-M M2YV User manual

M-system

M-system Mini-M M2YV User manual

M-system W2RS1 User manual

M-system

M-system W2RS1 User manual

M-system MFS2 User manual

M-system

M-system MFS2 User manual

M-system M5MS User manual

M-system

M-system M5MS User manual

M-system W2TS User manual

M-system

M-system W2TS User manual

M-system W2RS User manual

M-system

M-system W2RS User manual

M-system M5PT User manual

M-system

M-system M5PT User manual

M-system 27RS User manual

M-system

M-system 27RS User manual

M-system CTS2 User manual

M-system

M-system CTS2 User manual

M-system M2EXS User manual

M-system

M-system M2EXS User manual

M-system W2XT User manual

M-system

M-system W2XT User manual

M-system W2SP User manual

M-system

M-system W2SP User manual

M-system M5XU-1-R User manual

M-system

M-system M5XU-1-R User manual

M-system W2MS User manual

M-system

M-system W2MS User manual

M-system Mini-M Series User manual

M-system

M-system Mini-M Series User manual

Popular Transmitter manuals by other brands

INOR IPAQ C520 Supplementary instructions

INOR

INOR IPAQ C520 Supplementary instructions

halstrup-walcher PS17 instruction manual

halstrup-walcher

halstrup-walcher PS17 instruction manual

Azden IRH-15C user guide

Azden

Azden IRH-15C user guide

Datcon DT7001 operating instructions

Datcon

Datcon DT7001 operating instructions

Electro-Voice HTU-2 user manual

Electro-Voice

Electro-Voice HTU-2 user manual

Sony DCC-FMT50U operating instructions

Sony

Sony DCC-FMT50U operating instructions

Converters.TV 722 Operation manual

Converters.TV

Converters.TV 722 Operation manual

ADEMCO 5820 Installation and setup guide

ADEMCO

ADEMCO 5820 Installation and setup guide

Retekess TR509 user manual

Retekess

Retekess TR509 user manual

Varec 7532 Installation and operation manual

Varec

Varec 7532 Installation and operation manual

iSimple JamKast ISFM31 instructions

iSimple

iSimple JamKast ISFM31 instructions

Xantech HDTXSG01 installation instructions

Xantech

Xantech HDTXSG01 installation instructions

Altinex TP315-106 user guide

Altinex

Altinex TP315-106 user guide

Crestron DM-TX-4KZ-202-C quick start

Crestron

Crestron DM-TX-4KZ-202-C quick start

Nice ERA ONE Fitting instructions

Nice

Nice ERA ONE Fitting instructions

Omni-WiFi SweetSpot5 instruction manual

Omni-WiFi

Omni-WiFi SweetSpot5 instruction manual

BWI Eagle AIR-EAGLE SR PLUS manual

BWI Eagle

BWI Eagle AIR-EAGLE SR PLUS manual

Halemeier Tri-Mitter MultiWhite2 manual

Halemeier

Halemeier Tri-Mitter MultiWhite2 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.