M-system 10SP User manual

BEFORE USE ....
Thank you for choosing M-System. Before use, please check
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System’s Sales Office or representatives.
■PACKAGE INCLUDES:
Signal conditioner ...............................................................(1)
Mounting screw (M3.5 × 10)...............................................(2)
■MODEL NO.
Confirm Model No. marking on the product to be exactly
what you ordered.
■INSTRUCTION MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection and
basic maintenance procedures.
POINTS OF CAUTION
■POWER INPUT RATING & OPERATIONAL RANGE
• Locate the power input rating marked on the product and
confirm its operational range as indicated below:
24V DC rating: 24V ±10%,
approx. 50mA with voltage output 1
approx. 70mA with current output 1
■GENERAL PRECAUTIONS
• Before you remove the unit or mount it, turn off the power
supply and input signal for safety.
■ENVIRONMENT
• Indoor use.
• When heavy dust or metal particles are present in the
air, install the unit inside proper housing with sufficient
ventilation.
• Do not install the unit where it is subjected to continuous
vibration. Do not subject the unit to physical impact.
• Environmental temperature must be within -5 to +55°C
(23 to 131°F) with relative humidity within 30 to 90% RH
in order to ensure adequate life span and operation.
■WIRING
• Do not install cables close to noise sources (relay drive
cable, high frequency line, etc.).
• Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.
• Be sure to put the terminal cover on while the power is
supplied.
■AND ....
• The unit is designed to function as soon as power is sup-
plied, however, a warm up for 10 minutes is required for
satisfying complete performance described in the data
sheet.
COMPONENT IDENTIFICATION
Zero Adj.
Span Adj. Output 1
Body
Specifications
Connection Diagram
(side)
Zero Adj.
Output 2*Span Adj.
Terminal Cover
*Provided only when the second output is selected.
INSTALLATION
Use Standard Rack (model: 10BXx).
LOW FREQUENCY TRANSMITTER
(50 Hz minimum) MODEL 10SP
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-0856 Rev.5 P. 1 / 3
INSTRUCTION MANUAL
TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the connection diagram on the side of the unit.
■EXTERNAL DIMENSIONS unit: mm (inch)
130 (5.12)
3 (.12)
51 (2.01)
82 (3.23)
2–4.3 (.17) dia.
MTG HOLE
TERMINAL COVER
7–M3.5 SCREW
8 (.31)
91 (3.58)
99 (3.90 )
25 (.98)
12
3
56
78
■CONNECTION DIAGRAM
+
–
1
OUTPUT 1
2
3
7
8
+
–
OUTPUT 2
5
6
CONNECTOR
OUTPUT 2*
POWER
INPUT
+
–
+
SENSOR EXCITATION
1
3
2
4kΩ
■ Dry Contact
Input Connection Examples
+
–+12V
PWR
■ Voltage Pulse
1
3
2
+
+
–
–+12V
PWR
+
–
*For 1 output channel type, OUTPUT 1 is also connected to the
card-edge connector inside.
Note: The section enclosed by broken line is only for 2nd output channel.
10SP
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-0856 Rev.5 P. 2 / 3
WIRING INSTRUCTIONS
■SCREW TERMINAL
Torque: 0.8 N·m
CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Power input voltage: Check voltage supplied to the rack
(model: 10BXx). For the DC power source, be sure that
the ripple level is within 10% p-p.
3) Input: Check input terminal 1 (+) – 2 (–) with an oscil-
loscope.
With voltage pulse input, the signal should be below 1V
at L level, while it should exceed 2V at H level.
4) Output: Check that the load resistance meets the de-
scribed specifications.
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case
of regular calibration, adjust the output as explained in the
following.
■HOW TO CALIBRATE THE OUTPUT SIGNAL
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 10 minutes.
1) ZERO: Apply 10% input and adjust output to 10%.
2) SPAN: Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 10% input.
4) When ZERO value is changed, repeat the above proce-
dure 1) – 3).
5) Go through the same procedure for Output 2. Extender
Card (model: 10EC) is required to access the Output 2
Adjustments on top of the unit maintaining electrical
connection at the rear connector.
MAINTENANCE
Regular calibration procedure is explained below:
■CALIBRATION
Warm up the unit for at least 10 minutes. Apply 10%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the “ADJUST-
MENT PROCEDURE” explained earlier.
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protector for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.
10SP
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-0856 Rev.5 P. 3 / 3

Other M-system Transmitter manuals

M-system Mini-M M2RS1 User manual

M-system

M-system Mini-M M2RS1 User manual

M-system 6BR User manual

M-system

M-system 6BR User manual

M-system Mini-M M2UDS2 User manual

M-system

M-system Mini-M M2UDS2 User manual

M-system Mini-M M2VS User manual

M-system

M-system Mini-M M2VS User manual

M-system W5RS User manual

M-system

M-system W5RS User manual

M-system W2TS User manual

M-system

M-system W2TS User manual

M-system 10JV User manual

M-system

M-system 10JV User manual

M-system W5-UNIT User manual

M-system

M-system W5-UNIT User manual

M-system 27RS User manual

M-system

M-system 27RS User manual

M-system M6SCTC User manual

M-system

M-system M6SCTC User manual

M-system M2EXV User manual

M-system

M-system M2EXV User manual

M-system M3LM User manual

M-system

M-system M3LM User manual

M-system 27U User manual

M-system

M-system 27U User manual

M-system M3LV User manual

M-system

M-system M3LV User manual

M-system M3LLC User manual

M-system

M-system M3LLC User manual

M-system M5XU-1-R User manual

M-system

M-system M5XU-1-R User manual

M-system Mini-M M2MS User manual

M-system

M-system Mini-M M2MS User manual

M-system 10PA User manual

M-system

M-system 10PA User manual

M-system M5PA User manual

M-system

M-system M5PA User manual

M-system M3LR User manual

M-system

M-system M3LR User manual

M-system M6SPA User manual

M-system

M-system M6SPA User manual

M-system B3HU User manual

M-system

M-system B3HU User manual

M-system 27PM User manual

M-system

M-system 27PM User manual

M-system M1EXM-2 User manual

M-system

M-system M1EXM-2 User manual

Popular Transmitter manuals by other brands

HomePro ZTH100 operating instructions

HomePro

HomePro ZTH100 operating instructions

Aidcall Touchsafe Pro Door Monitor V3.00a installation guide

Aidcall

Aidcall Touchsafe Pro Door Monitor V3.00a installation guide

YOKOGAWA vigilantplant DPharp EJX Series user manual

YOKOGAWA

YOKOGAWA vigilantplant DPharp EJX Series user manual

ADJ WIFLEX DMX SYSTEM quick start guide

ADJ

ADJ WIFLEX DMX SYSTEM quick start guide

RKI 65-2422RK-05 Operator's manual

RKI

RKI 65-2422RK-05 Operator's manual

Siemens SITRANS P operating instructions

Siemens

Siemens SITRANS P operating instructions

Baumer CombiLyz AFI4 Operating	 instruction

Baumer

Baumer CombiLyz AFI4 Operating instruction

Autrol APT 3100 Series Operation manual

Autrol

Autrol APT 3100 Series Operation manual

Space Television OTX-CATV-2002 user manual

Space Television

Space Television OTX-CATV-2002 user manual

Magnetrol Pulsar R86 manual

Magnetrol

Magnetrol Pulsar R86 manual

Moore Industries IPF user manual

Moore Industries

Moore Industries IPF user manual

sauermann CP 110 quick start guide

sauermann

sauermann CP 110 quick start guide

BMS HELI-CODER 4 Operation manual

BMS

BMS HELI-CODER 4 Operation manual

Akai C57 user guide

Akai

Akai C57 user guide

Minco TT190 Installation and operating instructions

Minco

Minco TT190 Installation and operating instructions

GLI International C33 operating manual

GLI International

GLI International C33 operating manual

Erreka LUNA LU02 quick start guide

Erreka

Erreka LUNA LU02 quick start guide

ACS contsys Precont CT Technical manual

ACS contsys

ACS contsys Precont CT Technical manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.