EL502 www.easthigh.com.cn . 8 .
Troubleshooting
TroubleshootingTroubleshooting
Troubleshooting
The Mechanical Physician Scale is factory-calibrated to within plus or minus 1/4 pound accuracy. For the most
accurate readings, always use the scale on a hard, level surface and stand in the center of the scale platform with the
weight evenly distributed. If an error occurs or seems excessive, check the following:
Problem Possible Solution
Zero balance out of adjustment
The weighing beam must be balanced so the pointer comes to a rest in the
center of the eye loop (shown in Figure 9) when both poise weights are set at
zero (see page 1 for poise weight location). Follow zero adjustment instructions
on page 7.
Beam does not move freely
Make sure the pointer is not touching the side of the eye loop, impeding its
range of travel.
Visually ensure that the linkage is centered and properly aligned. Occasionally
during shipping, the alignment will become skewed. The linkage must be free
floating in order for the scale to weigh properly. If the scale is slightly tipped to
one side, the linkage can be seen by looking up underneath the top weigh
beam.
Platform rocks excessively
Ensure scale is setting on a level surface. When you push down on any corner
of the platform, you should not feel any significant rocking.
Beam does not move at all
during weighing
Weights are set higher than the person’s actual weight. Reset the weights to a
lower weight.
Make sure steel rod is properly connected and aligned as in Figure 5.
Scale is out of calibration
Recalibrate the scale by placing a known weight on the scale and turning the
zero adjustment screw until the pointer remains centered within the eye.
One
One One
One Year Limited Warranty
Year Limited WarrantyYear Limited Warranty
Year Limited Warranty
MeasureTek products covered in this manual are guaranteed to be free from
defects in material and workmanship for a
period of one year after date of purchase. Misuse, accidental damage, overload, alteration, and improper installation
are expressly excluded. Any product which is determined to be defective in material or workmanship within this time
period may, as the exclusive remedy, be returned to an authorized MeasureTek distributor or service center, freight
prepaid with prior return authorization, to be repaired or replaced at the manufacturer’s option. MeasureTek’s
liability under this warranty is limited to the repair or replacement of the defective product and in no event shall
MeasureTek be liable for consequential or indirect damages.