• Thepumpmuststanduprightduringuse.
• TheMedelaDominant35c/iisamedicaldevicethatrequiresspecialsafetymeasuresin
regardtoEMCandmustbeinstalledandputintooperationinaccordancewiththeattached
EMCinformation.
• PortableandmobileRFcommunicationsequipmentcanaffectmedicaldevices.
• ThisdevicecomplieswithIEC60601-1-2:«Medicalelectricalequipment-Electromagnetic
compatibility»andistherefore unlikely to interfere with other electrical equipment.Incase
thereisinterference(whichcanbedeterminedbydisconnectingthedevicefromthemains
supply),trythefollowing:Reorientingorrelocatingthedevicewithrespecttotheotherdevice
orconnectingthedevice'spowercordtoanotherpoweroutlet.
• Inthecaseofoverflow,informtheinternaltechnicalserviceimmediatelyandperformthetasks
in the service manual.
• Keeppowercordawayfromhotsurfaces.
• TheplugandtheON/OFFswitchmustnotcomeintocontactwithmoisture.
• Separationfromtheelectricpowersourceisonlyassuredthroughthedisconnectionofthe
plugandsocketconnection.
• Neverpulltheplugoutofthemainssocketbypullingontheconnectingcable!
• Neverleavethedeviceunattendedwhenitisswitchedon.
• Neverplacethedeviceinwaterorotherliquids.
• Check the patient’s condition as well as the proper running of the pump (by means of
audibleindicationsandvisualsigns)accordingtohouse-internalguidelinesoratleastevery
four hours.
TheseInstructionsforUsemustbekeptforlaterreference.
Safety-related checksIV
For the safety-related checks, it is assumed that the device is maintained and repaired throughout
its service life in compliance with the Service procedures.
For the Dominant 35 c/i as a device in protection class II (EN IEC 60601-1, A1, A2:1995), the
safety-related checks are confined to visual inspection of the housing and main power cord for
damage. This check must be performed prior to each use.
Devices of protection class II do not have a protective earth conductor; there is therefore no need
to check the earth leakage current!
The Dominant 35 c/i enclosure is made entirely of insulating material. Tests of the enclosure lea-
kage current using common measuring instruments will therefore not reveal measurable values.
The applied parts of the Dominant 35 c/i are connected to the Dominant 35 c/i via non-conductive
vacuum tubing, collection jars and intermediate tubings.
Even when suctioning a conductive fluid until the overflow protection device activates (9 g/l NaCl;
EN ISO 10079-1;1999), measurements of the patient leakage current using common measuring
instruments will not reveal measurable values.
The Dominant 35 c/i pumps do not have patient circuits, signal input and output parts and func-
tional earth connections.