Metrix MX2033 User manual

Digital Proximity System
Operaon and Maintenance Manual
100576 • REV C
The DPS (Digital Proximity System) devices
consists of two modules: the MX2033 and
MX2034. The MX2033 is a convenonal
3-wire driver while the MX2034 is a two-
wire current loop transmier. Users can con-
gure either unit to measure peak-to-peak
vibraon, gap distance, or rotaonal speed.
These products are used in conjuncon with
a proximity probe and extension cable to
measure the vibraon levels, gap distances,
or rotaonal speed of machinery shas. Us-
ers can congure the units via a USB cable
and conguraon soware.

Doc# 100576 • REV C (Sept 2017) Page 2 of 42
TABLE OF CONTENTS
........................................................3
1.1 Safety Terms and Symbols ..................................................3
1.2 General Safety Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2. Overview .........................................................................4
2.1 MX2033 Three-Wire Proximity Driver..............................................5
2.2 MX2034 Two-Wire Transmier....................................................5
2.3 For more informaon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
.................................................................6
3.1 Equipment Required .......................................................6
3.2 Soware Installaon .......................................................7
3.3 Connecng the DPS to a Computer ..........................................11
3.3.1 Remove the Base .......................................................11
3.3.2 Apply Power ...........................................................12
3.4 Launching the DPS Conguraon Soware ....................................12
3.5 Main Screen .............................................................13
3.5.1 Command Line Opons ..................................................13
3.5.2 Device Informaon ......................................................13
3.5.3 Device Communicaon Status .............................................14
3.5.4 Congurable Device Sengs ..............................................14
3.5.5 Conrming the Conguraon .............................................16
3.5.6 Prinng a Label .........................................................16
3.5.7 Installing the Label ......................................................18
3.6 Using the MX2034 Buered Output .........................................18
.................................................19
4.1 Scale Factor Vericaon ...................................................20
4.2 Scale Factor Custom Calibraon .............................................24
4.2.1 Current Loop Vericaon (MX2034 only) ....................................26
..........................................................29
5.1 MX2033 and MX2034 .....................................................29
5.2 MX2034 Only ............................................................30
.......................................................31
6.1 Driver/Transmier Temperature Response ....................................31
6.1.1 MX2030/BN 3300 5 Meter System .........................................31
6.1.2 MX2030/BN 3300 9 Meter System .........................................34
6.1.3 Metrix/BN 7200 5 Meter System ..........................................36
6.1.4 Metrix/BN 7200 9 Meter System ..........................................39
6.2 Phase Delay .............................................................42
.................................................43

Doc# 100576 • REV C (Sept 2017) Page 3 of 42
HIGH VOLTAGE
PRESENT
PROTECTIVE
EARTH NOTE
DANGER or
CAUTION
FUNCTIONAL
GROUND
1. GENERAL INFORMATION
1.1 SAFETY TERMS AND SYMBOLS
Terms that appear in this manual requiring special aenon include:
• WARNING: Warning statements idenfy condions or pracces that could result in injury
or loss of life.
• CAUTION: Cauon statements idenfy condions or pracces that could result in dam-
age to the product, loss or corrupon of data, or damage to the environment or other
property.
• NOTE: Notes idenfy material of special interest or importance to the user, not including
cauons or warnings.
Symbols that may appear on the product and/or in this manual include:
1.2 GENERAL SAFETY SUMMARY
Review the following safety precauons to avoid injury and prevent damage to this product
or any products connected to it.
• USE ONLY AS SPECIFIED
To avoid potenal hazards, use this product only as specied. Only qualied personnel
should perform installaon and uninstallaon procedures.
• OBSERVE ALL TERMINAL RATINGS
To avoid re or shock hazard, observe all rangs and markings on the product. Consult
the individual secons of this manual for further rangs informaon before making con-
necons to the product.
• AVOID EXPOSURE TO CIRCUITRY
Do not touch exposed electrical connecons and components when power is present.
• DO NOT OPERATE WITH SUSPECT FAILURES
If you suspect there is damage to this product, have it inspected by qualied personnel.

Doc# 100576 • REV C (Sept 2017) Page 4 of 42
1.3 RECEIVING, INSPECTING AND HANDLING THE SYSTEM
Metrix ships the probe, extension cable, and driver as separate units that the user intercon-
nects at the installaon site. Carefully remove all equipment from the shipping containers
and inspect it for shipping damage. If you see shipping damage, le a claim with the carrier
and submit a copy to Metrix Instrument Co. Include part numbers and serial numbers on all
correspondence. If no damage is apparent and the equipment is not going to be used imme-
diately, return the equipment to the shipping containers and reseal unl ready for use. Store
the equipment in an environment that is free from potenally damaging condions such as
extreme temperature, excessive humidity, or a corrosive atmosphere.
2. OVERVIEW
The Metrix Digital Proximity System module comes in two versions:
• MX2033 – 3-wire proximity driver
• MX2034 – 2-wire, loop powered proximity transmier
A complete DPS system requires a DPS module, a proximity probe, and an extension cable.
You can congure the MX2033 and MX2034 modules to operate with the probe and cable
combinaons shown in Table 1.
Metrix MX2030 5 mm or 8mm p 5m
Metrix MX2030 5 mm or 8 mm p 9 m
Metrix 7200 5 mm or 8 mm p 5 m
Metrix 7200 5 mm or 8 mm p 9 m
Bently Nevada 3300 or 3300 XL 5 mm or 8
mm p
5 m
Bently Nevada 3300 or 3300 XL 5 mm or 8
mm p
9 m
Bently Nevada 7200 5 mm or 8 mm p 5 m
Bently Nevada 7200 5 mm or 8 mm p 9 m
Standard target material is AISI 4140. Other materials available upon request.

Doc# 100576 • REV C (Sept 2017) Page 5 of 42
The secons below provide more module informaon.
2.1
The MX2033 signal output is proporonal to the distance between the probe p and the
target material. The signal output follows API Standard 670 and is compable with most
connuous vibraon monitoring systems. The MX2033 uses -24Vdc excitaon and provides
an output scale factor of 7.87 mV/mm (200mV/mil) for 5 mm and 8mm probes.
2.2
The MX2034 provides an ISA standard 4-20 mA signal proporonal to vibraon, axial posi-
on sha speed (RPM) for direct connecon to a PLC, DCS, SCADA system, or other instru-
mentaon without requiring a separate monitor system. The MX2034 is powered by +24
Vdc, supplied within the current loop. The MX2034 is user congurable to funcon as either
a radial vibraon transmier (where the 4-20 mA signal is proporonal to peak-peak vibra-
on amplitude), as an axial posion transmier (where the 4-20 mA signal is proporonal to
average probe gap) or as a speed (RPM) transmier (where the 4-20 mA signal is propor-
onal to the speed of a machine). For convenience when connecng to signal analyzers,
portable data collectors, and test instrumentaon, the raw vibraon signal is available at
a short-circuit protected BNC connector. Refer to secon 3.6 of this manual for important
informaon and cauons regarding proper use of this connector.
2.3 FOR MORE INFORMATION
Refer to these documents for more informaon:
1087015 MX2030 Digital Proximity Transducer System Datasheet
1232961 DPS Hazardous Area Installaon Manual
100528 DPS Label Kit Manual

Doc# 100576 • REV C (Sept 2017) Page 6 of 42
3. CONFIGURING A DPS
• This secon lists the procedure for conguring a DPS:
• Installing the DPS Conguraon Soware
• Connecng a DPS to your computer
• Retrieving the conguraons from the DPS
• Changing the DPS conguraon for probe/cable type
• Changing the DPS conguraon for measurement type (MX2034 only)
• Changing the DPS full scale range (MX2034 only)
• Prinng DPS labels
3.1 EQUIPMENT REQUIRED
You will need the equipment listed below to congure a DPS.
Item Notes
Computer With Windows XP, or Windows 7 or Windows 10, Operang
System
24V Power
Supply
Posive or Negave. Refer to datasheet 1087015 for power
supply voltage tolerance and DPS current requirements.
USB Cable USB 2.0 A to mini B
Labels Avery 6570
Label Overlay Included with the DPS

Doc# 100576 • REV C (Sept 2017) Page 7 of 42
3.2 SOFTWARE INSTALLATION
Metrix provides the DPS Conguraon and Ulity Soware on a USB memory sck that you
request with the DPS (Refer to the DPS datasheet for soware ordering informaon.)
Insert the USB sck into your computer and open explorer to view the les.
Double-click the version for your operaon system according to Table 2 and follow the
instrucons listed on the following pages.
Version
Windows XP DPSUser.exe (32 bit)
Windows 7 DPSUser_x64.exe (64 bit)
Click Next to connue

Doc# 100576 • REV C (Sept 2017) Page 8 of 42
Figure 2 shows the end user license agreement. Click ‘Accept” to connue
Click Install to begin the installaon

Doc# 100576 • REV C (Sept 2017) Page 9 of 42
If you are running Windows 7, Windows may present a User Account Control noce as
shown in Figure 4. Click “Yes” to connue.
You may addionally see a Windows Security noce as shown in Figure 5. Click ”Install this
driver soware anyway” to connue.

Doc# 100576 • REV C (Sept 2017) Page 10 of 42
Click the Finish buon as shown in Figure 6 to close the installaon process.

Doc# 100576 • REV C (Sept 2017) Page 11 of 42
3.3 CONNECTING THE DPS TO A COMPUTER
Follow these steps to access the DPS USB connector for connecon to a computer.
3.3.1 REMOVE THE BASE
Remove the three screws from the DPS base to access the mini USB connector as shown in
Figure 7 and Figure 8.
Plug the USB A connector into an available USB port on your computer.
Remove these three screws to
remove the base
Plug the mini-USB connector
into the DPS Plug

Doc# 100576 • REV C (Sept 2017) Page 12 of 42
3.3.2 APPLY POWER
You must apply 24 Vdc power to the DPS while conguring as shown in Figure 9. Note that
you can use either a negave or posive power supply (Refer to the DPS datasheet for volt-
age and current requirements.) Powering is the same for either the MX2033 or the MX2034.
The MX2034 does not require load resistors when conguring.
3.4 LAUNCHING THE DPS CONFIGURATION SOFTWARE
Launch the DPS Conguraon Soware by double-clicking the icon:
NOTE: Install the soware before connecng and pow-
ering the DPS to ensure the computer nds the correct
USB driver
Common
-24V Common
+24V

Doc# 100576 • REV C (Sept 2017) Page 13 of 42
3.5 MAIN SCREEN
The DPS Conguraon Soware opens with the main screen shown in Figure 10.
3.5.1 COMMAND LINE OPTIONS
File – Exit the DPS Conguraon Soware.
This opens the label prinng screen to print updated side labels for their unit.
This label contains basic conguraon informaon as well as user specic informaon such
as machine locaon, etc.
Provide descripons of the various features of the soware.
3.5.2 DEVICE INFORMATION
The DPS soware automacally retrieves the DPS conguraon when launched with a pow-
ered up DPS unit connected via a USB cable. If you connect a DPS unit aer launching the
soware, press the Retrieve buon to retrieve the conguraon.
The device informaon shows:
The date and me the displayed conguraon was retrieved.
DPS part number with full conguraon (all dash opons).

Doc# 100576 • REV C (Sept 2017) Page 14 of 42
Digital Proximity Driver (MX2033) or Digital Proximity Transmier (MX2034)
Device serial number
Date the device was manufactured.
Firmware model and revision number
Hardware major.minor revision number
Date of the last me Metrix calibrated the device
Date the conguraon was last changed.
3.5.3 DEVICE COMMUNICATION STATUS
3.5.4 CONFIGURABLE DEVICE SETTINGS
You can congure a DPS module for:
• Probe Series
• Probe Diameter
• System Length
• Target Material
• Measurement (MX2034 only)
• Range (MX2034 only)
• 4-20mA Direcon (MX2034 only)
DPS is connected and communicang properly
DPS is not connected or not communicang

Doc# 100576 • REV C (Sept 2017) Page 15 of 42
3.5.4.1 PROBE SERIES
All DPS units support Metrix MX2030, Metrix 10,000 series 7200 compable, Bently Nevada
3300, and Bently Nevada 7200 series probes. The DPS also supports one addional probe
type specied at the me of ordering.
3.5.4.2 PROBE DIAMETER
The DPS supports 5mm, 8mm and 11mm probe p diameters as part of its standard factory-
loaded curve set.
3.5.4.3 TARGET MATERIAL
DPS units support the standard four probe series listed in secon 3.5.4.1 calibrated for 4140
steel. The addional probe type specied at the me of ordering may be calibrated to a
variety of materials. Contact Metrix for informaon on special material calibraons.
3.5.4.4 MEASUREMENT
The Measurement eld is only acve for the MX2034 transmier model. The MX2034 sup-
ports three measurements: Radial Vibraon, Axial Posion, and Sha Speed (RPM).
3.5.4.4.1 RADIAL VIBRATION
The MX2034 performs a peak to peak detecon and drives the 4-20 mA output proporonal
to the congured range.
1) Select the new conguraon
parameters from the drop down list
2) Click the buon to send the new
conguraon to the DPS

Doc# 100576 • REV C (Sept 2017) Page 16 of 42
3.5.4.4.2 AXIAL POSITION
The MX2034 performs a displacement measurement for axial posion and driver the 4-20
mA output proporonal to the congured range.
3.5.4.4.3
The MX2034 performs a pulse count measurement cause by keyway and drives the 4-20 mA
output proporonal to RPM.
3.5.4.5 RANGE
The Range choices are dierent for Radial Vibraon, Axial Posion, and Sha Speed (RPM)
measurements.
3.5.4.6 mA DIRECTION
The MX2034 allows conguraon of upscale or downscale for the 4-20mA output of axial
posion measurements only. Upscale means that for the range selected in 3.5.4.5, 4mA
corresponds to boom of scale and 20mA corresponds to top of scale; downscale reverses
these convenons and means that 20mA corresponds to boom of scale and 4mA corre-
sponds to top of scale.
3.5.5 CONFIRMING THE CONFIGURATION
Aer clicking the Send buon, you will have the opon to abort the conguraon change.
Press the Yes buon to connue.
3.5.6 PRINTING A LABEL
Aer sending a conguraon change, the soware will give you the opon to print a label.
You can also print a label later from the main screen by clicking Print Labels and then One
Label.

Doc# 100576 • REV C (Sept 2017) Page 17 of 42
Clicking the buon next to the label informaon opens up the Print Label Screen.
Use this screen to ll in addional informaon you want to record with the DPS.
Aer lling in any custom informaon, click the OK buon. The label print start screen
opens. The label print start screen allows you to select the label posion you want the label
prinng to start at. This template is set for Avery 6570 label sheets of 32 labels (or equiva-
lent).
NOTE: User-printed labels should be secured under plasc
overlay window. Order Metrix p/n 100527 DPS Label
Prinng Kit which contains instrucons, Avery Labels, and
overlays needed for prinng up to 16 labels.

Doc# 100576 • REV C (Sept 2017) Page 18 of 42
Clicking a label starng point will open the printer selecon screen. Select the printer and
click OK to print.
3.5.7 INSTALLING THE LABEL
Remove the backing from the polycarbonate label and adhere the printed label inside the
clear window. Aach both labels to the side of the DPS unit. Refer to DPS Label Kit manual
100528 for addional details.
3.6 USING THE MX2034 BUFFERED OUTPUT
Due to the two wire, loop-powered MX2034 transmier design, the current loop common
oats on top of the voltage developed across the current loop sense resistor.

Doc# 100576 • REV C (Sept 2017) Page 19 of 42
The buered output common is protected against ground loop currents aecng the current
loop output with a 10kΩ resistance between the buered output common and the loop
common.
Metrix recommends using a ground isolated instrument or signal isolator. The meter, oscil-
loscope or analyzer used to measure the gap voltage or to observe the vibraon signal
(DYNAMIC OUTPUT) must have an input impedance of one megohm or greater.
The scale factor of the output signal is 200 mV/ mil.
Metrix recommends a maximum cable length of 15 (5 m) when connecng to the buered
output due to noise suscepbility caused by the high input and output impedances.
4. VERIFICATION AND CALIBRATION
This secon describes the steps to verify proper DPS operaon and to custom calibrate your
DPS to your specic probe and cable.
When using the MX2034 BNC connector, exer-
cise special care to prevent ground loops that could alter
the 4-20mA output, resulng in spurious alarms or ma-
chine trips. Such ground loops can occur when connecng
to grounded test equipment or mulple input test equip-
ment. Metrix recommends use only with ungrounded test
equipment, such as baery-powered portable meters,
or grounded test equipment with dierenal inputs (i.e.,
neither side of input connecon is ed to ground). Also,
cable lengths between the BNC connector and test equip-
ment should be kept suitably short (5m or less) to prevent
unwanted noise.
The DPS requires the probe and extension cables be
from the congured probe series, combine to the cong-
ured system length, and view the congured target mate-
rial in order to meet system range and accuracy

Doc# 100576 • REV C (Sept 2017) Page 20 of 42
4.1 SCALE FACTOR VERIFICATION
Follow these steps in this secon to verify the voltage output scale factor. The basic proce-
dure is:
Obtain the correct target material
Assemble the test instruments and equipment
Electrically zero the probe
Measure the output voltage in xed gap increments
4.1.1.1 OBTAINING THE TARGET
Use a 1.2 inch or larger target with 5 mm or 8 mm probes. Be sure your target material
matches the material the DPS calibraon material. To verify the calibraon material, follow
the steps listed in Figure 16.
Do not connect test equipment or cables
to the driver unless the area has been determined to
be non-hazardous
1) Retrieve the DPS conguraon
2) Verify the Target Material
conguraon
Other manuals for MX2033
2
This manual suits for next models
1
Table of contents
Other Metrix Measuring Instrument manuals

Metrix
Metrix MX 350 User manual

Metrix
Metrix MTX 1032-B User manual

Metrix
Metrix MX 406B User manual

Metrix
Metrix ScopiX IV User manual

Metrix
Metrix AG 900 User manual

Metrix
Metrix MX 409 User manual

Metrix
Metrix IX 3131 User manual

Metrix
Metrix MTX I 050-PC User manual

Metrix
Metrix MX 350 User manual

Metrix
Metrix SA6350 User manual
Popular Measuring Instrument manuals by other brands

Smaxtec
Smaxtec SX.1 quick start guide

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS CyberScan PCD 650 quick guide

Mitsubishi Electric
Mitsubishi Electric EcoMonitorPro EMU2-RD3-F instruction manual

Sibata
Sibata LD-5 Operation manual

Topcon
Topcon DS Series instruction manual

Elgato
Elgato Eve Energy quick start guide