Doc# M9192 • REV K (November 2020) Page 2 of 4
MECHANICAL INSTALLATION
Accelerometer and Cable Installaon
The sensive axis of the accelerometer can be mounted on any radial axis. When possible,
to reduce temperature exposure as much as praccal, mount the accelerometer on the side
of the machine. To ensure a clean response to high frequency vibraons, the accelerometer
must be securely mounted to a at machined surface. If a bracket is required, it should be
rigidly constructed to prevent spurious mechanical resonances in the specied frequency
range. Due to the accelerometer cable temperature limitaon of 260°C, the cable MUST be
routed away from surfaces exceeding that temperature. To avoid cable noise, the accelerom-
eter cable must be rmly secured every .3m (12 inches) to minimize cable movement. The
minimum bending radius of the cable is 30 mm (1.2 inches). Please note that the armored
cable between the accelerometer and the charge amplier module is not removable in the
eld. DO NOT ATTEMPT TO DISCONNECT THE ARMORED CABLE.
Charge Amplier
The charge amplier module can be mounted in any convenient locaon relave to the
accelerometer. Ensure that the maximum temperature the module is exposed to is within
the limits specied (-40°C to +58°C/120°C). The module should be fastened to a supporng
structure using customer supplied clamps or straps.
ELECTRICAL INSTALLATION
Accelerometer Grounding
Most machinery trains are grounded to earth. An accelerometer mounted on such a ma-
chine would then also be grounded. In some circumstances, electrical noise on the machine
housing may be transferred to the accelerometer. In these cases, it may be necessary to
provide an addional electrical ground connecon to the plant instrument ground system.
In extreme cases, it may be necessary to electrically isolate the accelerometer from the
machine housing and connect only the accelerometer to the instrument ground system. The
laer case requires precauons in order to maintain a proper mechanical mounng of the
accelerometer without introducing mechanical resonances.
Charge Amplier / Line Driver Wiring
The Charge Amplier / Line Driver module is provided in output connector, terminal block
and integral shielded cable versions. Connecon to the module should be made with a
shielded, twisted-pair cable suitably rated for the environmental condions to be expected.
The maximum wire size which may be used with the connector or terminal block is number
16 AWG (1.26 sq. mm). The cable shield is le oang at the Charge Amplier / Line Driver
end of the cable. The integral cable version of the Charge Amplier / Line Driver is provided
with number 20 AWG wires.
Metrix has connector versions of cables (20 AWG) available for use with the Charge Amplier
/ Line Driver. Metrix P/N 9193-001 is to be used with the 58°C version of the Charge Ampli-
er / Line Driver and Metrix P/N 9193-002 is to be used with the 120°C version.
The shield at the far end of the output cable should be connected to the receiving monitor’s,
transmier’s, or PLC’s circuit common. In addion, it may be necessary to connect that point
to a low noise electrical ground system. The cable shield should be connued to within 25mm
(1 inch) of the receiver’s terminals and the shield connecon should be as short as possible.