Metrix MTX 3292 User manual

MTX 3292/MTX 3293,
ASYC IV
Portable Digital Multimeter -
100,000 cts X04002A02 - Ed. 1 - 07/14
The full user’s manual is available on the CD supplied with the
instrument. Download is possible from our support site:
www.chauvin-arnoux.com.
This multimeter complies with the EN 61010-1 safety standard on double
insulation for electronic measuring instruments.
IP 67 as per standard IEC 60529.
Symbols used on the instrument:
Warning: Risk of danger.
Refer to the user’s manual to find out about the
potential dangers and any action to be taken in order to
avoid them.
Earth Double insulated
If you use this instrument in a way which is not specified, the protection
provided may be compromised, thus endangering you.
The safety of any system includingthis instrument is the responsibility of the
assembler of the system.
This instrument has been designed for use indoors:
- in an environment with pollution level 2,
- at an altitude of less than 2000 m,
- at a temperature between 0°C and 50°C,
- with relative humidity < 80 % up to 35°C.
It can be used for measurements on the circuits of:
- Category III installations for voltages up to 1,000 V (AC or DC)
relative to earth.
- Category IV installations for voltages up to 600 V (AC or DC)
relative to earth.
-CAT III: Measurement category III corresponds to measurements
carried out on installations in buildings.
Example: measurements on switchboards, wiring, etc.
-CAT IV: Measurement category IV corresponds to the measurements
carried out on the source of low-voltage installations.
Example: meters and measurements on the safety devices protecting
against overcurrents, etc.
For safety reasons, only use suitable cords supplied with the instrument:
they comply with the EN 61010-031 standard. Before each use, make
sure that they are in perfect condition.
When the instrument is connected to the measuring circuits, never touch a
terminal which is not in use. Use only suitable accessories delivered with
the instrument or approved by the manufacturer.
If the measurement category of the accessory is different from that
of the instrument, the lowest category applies to the unit.
6, avenue du Pré de Challes
F-74940 ANNECY-LE-VIEUX
Tél. +33 (0)4.50.64.22.22
MTX 3292 MTX 3293
Display Graphic color (70 x 52)
Power supply 4 R6 batteries (AA format)
or rechargeable batteries
Counts 100 000
Communication IR / USB (Bluetooth, optional)
We thank you for your confidence in the quality of our products.
•Fuse and batteries or rechargeable batteries:
Fuse: 11 A: 10 x 38 - 1000 V
F - breaking capacity: > 18 kA
4 batteries: 1,5 V AA LR6 or batteries 1,2 V Ni-IMH LSD
•Terminal block: with 3 4-m banana inputs and an optical
connector for USB communication:
•Display of the connection on the screen:
Voltage: VAC, VDC, VAC+DC
Graphic display of Mem… in VAC+DC mode:
Frequency: Hz
Resistance: Ω, Diode, Continuity
Capacitance: F
Vinput
2. Graphic mode (default),
but select:
yMeas… for another display
yMEAS+ to select a
secondary function
yREL relative measurement
ySURV for MIN/MAX/AVG
(time-stamped)
ySPEC to get the specs
Trace
Zoom
Cursor
1. Activate "V", then
press F1, F2,F3,F4
to select the coupling:
yAC,
yDC,
yAC+DC or
yVlowZ.
Configuration SETUP Æ
MEASURE to adjust the
parameters: filters,
impedance, reference...
COM input
3. Select Mem... to save or stop a sequence (10 max. and
6,500 measurements)
Consult data by long press on Mem…
1. Activate Ωby pressing this key:
yF2 accesses (continuity),
yF3 accesses 100 Ohm range
yF4 selects the diodes (4V or 26V diode test)
yor subsequent presses
2. Select Meas..., then MEAS+ for the MATH
secondary function, or:
yREL relative measurement
ySURV for time-stamped MIN/MAX/AVG
ySPEC to obtain the specifications.
1. Activate Hz.
2. Select Range ±,
frequency range
and
< 200 kHz or
> 200 kHz
by the keys .
3. Select Meas…, then MEAS+ to select a secondary
function among: MATH, DCY, PER, PW+/-, CNT+/- or:
yREL relative measurement
ySURV for time-stamped MIN/MAX/AVG
ySPEC to obtain the specifications.
1. Activate C by pressing this key.
Pressing:
yF1 accesses Range+
yF2 accesses Range-
Select ranges by those keys .
2. Select Meas..., then:
yREL relative measurement
ySURV for time-stamped MIN/MAX/AVG
ySPEC to obtain the specifications.
COM inputVinput COM input
Vinput
COM inputVinput
RUN is displayed
when the
measurement is
running and
OL is displayed if
the value is >
range or if there
is a short-circuit.
Temperature: °C, °F, K in TK/TJ or Pt100/Pt1000
Display in REL mode:
°C by Pt1000 graphic display:
Direct measurement of current:
Current with clamp:
Secondary modes accessible via MEAS…:
yTREND: graphic display
yREL: relative mode (REF, Δ, Δ%)
ySURV: surveillance mode (MIN, MAX, AVG)
ySPEC: specifications (SMIN, SMAX, %, digits)
yMEAS+: to access the secondary functions
SETUP Menu:
This SETUP menu sets the parameters through main adjustments
and 3 levels:
1. General Setup
yUtil: lighting, standby, beep, language, clock
yComm: IR or BT type, baud IR, protocol: SCPI or MODBUS
yPower supply: battery or rechargeable battery, capacity
2. Measurement config
yMeasure: filter on or off, impedance, reference, dBm and W
yClamp: V or A measurement, ratio, unit
yMath: measurement, coefficient A and B, unit
3. Configuration and MTX identification
yMemory: files, numbers of recordings, frequency
yConfig: factory configuration, USER or BASIC start
yAbout: model, serial number, software version
SURVEILLANCE mode: MEAS/SURV
MATH function:
Function Ax + b configurable to obtain a direct reading of
the original quantity.
Parameters to be adjusted:
MATH function, recall in the secondary function:
V Meas…/ MEAS+:
Memory mode: Mem…
COM inputVinput
Activate T°.
y1st press on F1 / F2 / F3 ưC, °F, K
y2nd press ÆPt100, Pt1000
y3rd press ÆTC J or TC K.
Ainput
COM input
Activate or deactivate the selected
mode by pressing the MEAS… key
then F1 / F2 / F3 / F4.
1. Activate the
"clamp" function.
2. Select the coupling,
then 2nd press on
SETUP menu:
MEASURE or input
V or A.
3. Enter the ratio
of the clamp:
0001.0A / 0001.0V
(default)
by pressing .
4. Select the unit,
A (default).
SETUP menu
Activate or start,
deactivate or stop
the SURV mode
by pressing on F1/F2.
1. Open "Menu" using the SETUP key, then MATH.
2. Select the V, A, Ωor Hz function by the keys:
F1, F2, F3, F4.
3. Select and enter the Aor Bcoefficients.
4. Select the unit.
1. Activation by the SETUP menu or long press on Mem:
2. Select and modify the parameters:
yFiles: list of the stored files, by date and time
yNb. rec.: number of recordings Æmax. 6500
yFreq: recording frequency in h, min, s
Activate / deactivate the MEM auto recording
mode by a short press:
max.10 sequences
Pressing F3 in SURV opens a consultation
window of CONSULT records:
A long press on Mem… opens the record consultation
menu: File ÆF1:
1. Display or open the list by F1.
2. Select a file with the keys .
3. Validate it to display by the F1 key,then Open.
4. Erase a sequence or Erase all memory sequences.
Pt100
Pt1000
TC J
TC K
°C
°F
K
REL
Reference
Δdifference
Δ% difference in %
1. Activate A, then
activate the AC or
DC or AC+DC
coupling by
pressing F1 / F2 /
F3,
2. Select
SETUP/MEASURE
to adjust the
parameters: filters,
impedance,
reference, etc.
MATH function result
in secondary function
and displayed main
q
uantit
y
Incrementing
the number of
steps since
startup

This manual suits for next models

1

Other Metrix Multimeter manuals

Metrix MX 1 User manual

Metrix

Metrix MX 1 User manual

Metrix ASYC-IV User manual

Metrix

Metrix ASYC-IV User manual

Metrix MX 56C User manual

Metrix

Metrix MX 56C User manual

Metrix ASYC-IV MTX 3290 User manual

Metrix

Metrix ASYC-IV MTX 3290 User manual

Metrix MX 1 User manual

Metrix

Metrix MX 1 User manual

Metrix ASYC IV MTX 3292 User manual

Metrix

Metrix ASYC IV MTX 3292 User manual

Metrix MX 57 EX User manual

Metrix

Metrix MX 57 EX User manual

Metrix MX 553 User manual

Metrix

Metrix MX 553 User manual

Metrix Checkweigher User manual

Metrix

Metrix Checkweigher User manual

Metrix MX 57 EX User manual

Metrix

Metrix MX 57 EX User manual

Metrix MX 59HD User manual

Metrix

Metrix MX 59HD User manual

Metrix MX 26 User manual

Metrix

Metrix MX 26 User manual

Metrix MX 21 User manual

Metrix

Metrix MX 21 User manual

Metrix MTX 202 User manual

Metrix

Metrix MTX 202 User manual

Metrix MX 26 User manual

Metrix

Metrix MX 26 User manual

Metrix MX 24B User manual

Metrix

Metrix MX 24B User manual

Metrix MTX 3290 User manual

Metrix

Metrix MTX 3290 User manual

Metrix MX 350 User manual

Metrix

Metrix MX 350 User manual

Metrix MTX 3281 User manual

Metrix

Metrix MTX 3281 User manual

Metrix MX 21 User manual

Metrix

Metrix MX 21 User manual

Metrix MX 58HD User manual

Metrix

Metrix MX 58HD User manual

Metrix MX 22 User manual

Metrix

Metrix MX 22 User manual

Metrix MX 58HD User manual

Metrix

Metrix MX 58HD User manual

Metrix MX 1 User manual

Metrix

Metrix MX 1 User manual

Popular Multimeter manuals by other brands

Chauvin Arnoux F401 Start guide

Chauvin Arnoux

Chauvin Arnoux F401 Start guide

REED R5007 instruction manual

REED

REED R5007 instruction manual

Amprobe AM-560 user manual

Amprobe

Amprobe AM-560 user manual

Triplett MM650 user manual

Triplett

Triplett MM650 user manual

SATA 05221 manual

SATA

SATA 05221 manual

southwire  ip67 12070T operating instructions

southwire ip67

southwire ip67 12070T operating instructions

Colluck HH0308C owner's manual

Colluck

Colluck HH0308C owner's manual

LOVATO ELECTRIC DMG800L01 instruction manual

LOVATO ELECTRIC

LOVATO ELECTRIC DMG800L01 instruction manual

Hazet 2152-600 operating instructions

Hazet

Hazet 2152-600 operating instructions

Brymen BM115 user manual

Brymen

Brymen BM115 user manual

Greenlee CMT-80 instruction manual

Greenlee

Greenlee CMT-80 instruction manual

EZ Digital DM-442B operating manual

EZ Digital

EZ Digital DM-442B operating manual

Sperry instrument DM-5300 operating instructions

Sperry instrument

Sperry instrument DM-5300 operating instructions

Gossen MetraWatt METRALINE DM 61 operating instructions

Gossen MetraWatt

Gossen MetraWatt METRALINE DM 61 operating instructions

PeakTech 4090 Operation manual

PeakTech

PeakTech 4090 Operation manual

Sealey TA330 manual

Sealey

Sealey TA330 manual

Kusam-meco KM-DMM-41 Operation manuals

Kusam-meco

Kusam-meco KM-DMM-41 Operation manuals

Innova Digital Multimeter 10MegOhms owner's manual

Innova

Innova Digital Multimeter 10MegOhms owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.