
1
IMPORTANT SAFETY INFORMATION
11
Biological hazards
BIOLOGICAL HAZARD
Contamination or infection may lead to death or serious injury, depending on the material used.
Biological material
All biological material must be considered potentially infectious.
• Wear personal protective equipment (such as gloves, safety glasses, etc.) as indicated in the safety data sheet
for the particular substance.
• If hazardous or potentially infectious material has been spilled or leaked from the system, decontaminate the
affected area.
• Do not continue to use contaminated accessories or parts of the instrument. Contact Miltenyi Biotec
Technical Support immediately.
• Aerosols might leak from the system. Protect mouth and nose if processing hazardous or potentially
infectious material that can be transmitted through the air.
• Operate the instrument in a biological safety cabinet suitable for the used specimen if hazardous or
potentially infectious materials are processed.
• Accessible surfaces (protective housing and sample and reagent chamber) need to be cleaned after each use
to avoid accumulation of substances carried in aerosols.
• To avoid spilling of liquids, remove all caps and lids from any sample or reagent container before you place
them into the instrument.
• As an exception, do not remove the aluminum sealing foils on antibody plates. Follow the instructions in the
datasheet for each disposable.
• Take care when handling samples and reagents.
• Provide for sufficient light.
Waste
All liquid and solid waste must be considered hazardous.
• Follow local regulations regarding waste disposal.
• Label waste bottles with the provided biohazard stickers.
Equipment damage
Defective or inadequate equipment can cause a biological hazard. Some decontamination methods may
damage the instrument.
• If parts of the instrument are damaged, unplug and do not use the instrument until damaged parts are
replaced. Contact Miltenyi Biotec Technical Support, if necessary.
• Follow the instructions in the datasheet for each disposable.
• Use only 70% ethanol or isopropyl alcohol for disinfection.
• Before the instrument is being disposed or transported into a lower safety level laboratory, follow local
regulations by ensuring an appropriate decontamination suitable for the appropriate specimen. Take care
that decontamination methods and agents other than those described here may damage the device, but may
be necessary for appropriate decontamination. The specifications apply only to the instrument before such
decontamination.