Milwaukee MI413 PRO User manual

USER MANUAL
MI413
PRO Free & Total Chlorine
High Range Photometer
milwaukeeinstruments.com (USA & CAN)
milwaukeeinst.com
On sample reading
y“-SA-”, there is too much light for
the sample measurement. Please
check if the right sample cuvet is
inserted.
y“Inv”, the sample and the zero
cuvet are inverted.
y“ZErO”, a zero reading was not
taken. Follow the instruction in
the measurement procedure for
zeroing the meter.
yUnder range. A blinking “0.00”
indicates that the sample
absorbs less light than the zero
reference. Check the procedure
and make sure you use the same
cuvet for reference (zero) and
measurement.
yA flashing value of the maximum
concentration indicates an over
range condition. The concentration
of the sample is beyond the
programmed range: dilute the
sample and re-run the test.
BATTERY REPLACEMENT
Battery replacement must
only take place in a non-
hazardous environment.
Simply rotate the battery cover
on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Insert the battery and replace the cover.
ACCESSORIES
MI513-100 Free & Total Chlorine reagent set (45 tests)
MI0001 Glass cuvets (2 pcs)
MI0002 Caps for cuvets (2 pcs)
MI0003 Stoppers for cuvets (2 pcs)
MI0004 Tissue for wipping cuvets (4 pcs)
MI0005 9V battery (1 pc)
CERTIFICATION
Milwaukee Instruments conform to the CE
European Directives.
Disposal of Electrical & Electronic
Equipment. Do not treat this product as
household waste. Hand it over to the
appropriate collection point for the recycling of electrical
and electronic equipment.
Disposal of waste batteries. This product
contains batteries. Do not dispose of them
with other household waste. Hand them
over to the appropriate collection point for recycling.
Please note: proper product and battery disposal
prevents potential negative consequences for human
health and the environment. For detailed information,
contact your local household waste disposal service or
go to www.milwaukeeinstruments.com (USA & CAN)
or www.milwaukeeinst.com.
RECOMMENDATION
Before using this product, make sure it is entirely suitable
for your specific application and for the environment in
which it is used. Any modification introduced by the user
to the supplied equipment may compromise the meter’s
performance. For your and the meter’s safety do not use
or store the meter in hazardous environment. To avoid
damage or burn, do not perform any measurement in
microwave ovens.
WARRANTY
These instruments are warranted against defects in
materials and manufacturing for a period of 2years from
the date of purchase. This warranty is limited to repair or
free of charge replacement if the instrument cannot be
repaired. Damage due to accidents, misuse, tampering
or lack of prescribed maintenance is not covered by
warranty. If service is required, contact your local
Milwaukee Instruments Technical Service. If the repair is
not covered by the warranty, you will be notified of the
charges incurred. When shipping any meter, make sure it
is properly packaged for complete protection.
ISTMI413 07/20-1
SATISFACTION
GUARANTEED
MI413 PRO Free & Total Chlorine
High Range Photometer
Dear Customer,
Thank you for choosing a Milwaukee Instruments
product. This manual will provide you with the necessary
information for the correct use of the instrument. Please
read it carefully before using the meter.
SPECIFICATIONS
Range 0.00 to 10.00 mg/L Cl2
Resolution 0.01 mg/L (0.00-3.50 mg/L)
0.10 mg/L (above 3.50 mg/L)
Precision ±0.17 mg/L @5.00 mg/L
Light Source Tungsten lamp
Light Detector Silicon Photocell and 525 nm narrow
band interference filter
Method Adaptation of USEPA method 330.5 and
Standard method 4500 - Cl G
Environment 0 to 50 °C (32 to 122 °F)
100% RH max
Battery Type 9 volt (1 pc)
Auto-Shut o After 10’ of non-use
Dimensions 192 x 104 x 52 mm
(7.5 x 4.1 x 2”)
Weight 380 g
MEASUREMENT PROCEDURE
1. Turn the meter on by pressing ON/OFF.
2. When the LCD displays “- - -”, the
meter is ready.
3. Fill the cuvet with 10 mL of sample, up
to the mark, and replace the cap.
4. Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
5. Press ZERO and “SIP” will blink on the
display.
6. After a few seconds the display will
show “-0.0-”. The meter is now zeroed
and ready for measurement.
7. For free chlorine measurement:
Add 3 drops of DPD 1 reagent to an empty cuvet.
▶
8. Add 5 mL of DPD 0 reagent by using a syringe.
9. Fill the cuvet up to the 10 mL mark with 5 mL of
unreacted sample using a pipette.
10. Replace the cap and shake gently for a few seconds.
DPD 1
x 3
11. Immediately insert the cuvet into the holder and
ensure that the notch on the cap is positioned
securely into the groove.
12. Press READ and “SIP” will blink
during measurement.
13. The instrument directly displays
concentration in mg/L of chlorine.
14. Then, for total chlorine measurement, remove the
cuvet and add to the reacted sample 3 drops of DPD
3. Replace the cap and swirl gently.
DPD 3
x 3
15. Reinsert the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
16. Wait for 2 minutes and 30 seconds
then press READ and “SIP” will blink
during measurement.
1 7. The instrument directly displays concentration in
mg/L of total chlorine.
Interferences
yBromine (positive error)
yChlorine dioxide (positive error)
yIodine (positive error)
yOxidized Manganese and Chromium (positive error)
yOzone (positive error)
yAlkalinity above 250 mg/L CaCO3 or acidity above 150
mg/L CaCO3 will not reliable develop the full amount of
color or it may rapidly fade. To resolve this, neutralize
the sample with diluted HCl or NaOH.
2’30”
GUIDE TO DISPLAY CODES
This prompt appears for 1 second
each time the instrument is turned
on.
The dashes “- - -” indicates that the
meter is in a ready state and zeroing
can be performed.
Sampling In Progress. Flashing “SIP”
prompt appears each time the meter
is performing a measurement.
“-0.0-”, the meter is in a zeroed state
and measurement can be performed.
The blinking “BAT” indicates that the
battery voltage is getting low and the
battery needs to be replaced.
“-bA-”, the battery is dead and must
be replaced. Once this indication
is displayed, the meter will lock up.
Change the battery and restart the
meter.
“Conf”, the meter has lost its
configuration. Contact your dealer
or the nearest Milwaukee Customer
Service Center.
ERROR MESSAGES
On zero reading
yBlinking “-0.0-” indicates that the
zeroing procedure failed due to a
low signal-to-noise ratio. In this
case press ZERO again.
y“no L”, the instrument can not
adjust the light level. Please check
that the sample does not contain
any debris.
y“L Lo”, there is not enough light to
perform a measurement. Please
check the preparation of the zero
cuvet.
y“L Hi”, there is too much light to
perform a measurement. Please
check the preparation of the zero
cuvet.
10 ml
▶

Other Milwaukee Measuring Instrument manuals

Milwaukee MI407 User manual

Milwaukee

Milwaukee MI407 User manual

Milwaukee LDM 45 User manual

Milwaukee

Milwaukee LDM 45 User manual

Milwaukee M12 CPL User manual

Milwaukee

Milwaukee M12 CPL User manual

Milwaukee MW801 User manual

Milwaukee

Milwaukee MW801 User manual

Milwaukee 3421 User manual

Milwaukee

Milwaukee 3421 User manual

Milwaukee 2309-60 User manual

Milwaukee

Milwaukee 2309-60 User manual

Milwaukee MI 806 User manual

Milwaukee

Milwaukee MI 806 User manual

Milwaukee MW13 User manual

Milwaukee

Milwaukee MW13 User manual

Milwaukee MI 404 User manual

Milwaukee

Milwaukee MI 404 User manual

Milwaukee 3631-20 User manual

Milwaukee

Milwaukee 3631-20 User manual

Milwaukee Mi453 User manual

Milwaukee

Milwaukee Mi453 User manual

Milwaukee MW180 MAX User manual

Milwaukee

Milwaukee MW180 MAX User manual

Milwaukee CALIBRATION 2238-20 User manual

Milwaukee

Milwaukee CALIBRATION 2238-20 User manual

Milwaukee 2205-20 User manual

Milwaukee

Milwaukee 2205-20 User manual

Milwaukee MW805 MAX User manual

Milwaukee

Milwaukee MW805 MAX User manual

Milwaukee MI404 User manual

Milwaukee

Milwaukee MI404 User manual

Milwaukee EC40 User manual

Milwaukee

Milwaukee EC40 User manual

Milwaukee C12 CMH User manual

Milwaukee

Milwaukee C12 CMH User manual

Milwaukee MW100 User manual

Milwaukee

Milwaukee MW100 User manual

Milwaukee 2217-40 User manual

Milwaukee

Milwaukee 2217-40 User manual

Milwaukee MW10 User manual

Milwaukee

Milwaukee MW10 User manual

Milwaukee 2217-40 User manual

Milwaukee

Milwaukee 2217-40 User manual

Milwaukee pH54 User manual

Milwaukee

Milwaukee pH54 User manual

Milwaukee M12 CLLP-301C User manual

Milwaukee

Milwaukee M12 CLLP-301C User manual

Popular Measuring Instrument manuals by other brands

Associated Equipment MIL6011 Operator's manual

Associated Equipment

Associated Equipment MIL6011 Operator's manual

REED R7700 instruction manual

REED

REED R7700 instruction manual

Ametek Land LANCOM 4 user guide

Ametek Land

Ametek Land LANCOM 4 user guide

Fluke Nuclear Associates 07-453 Operator's manual

Fluke

Fluke Nuclear Associates 07-453 Operator's manual

HP 8562E Service guide

HP

HP 8562E Service guide

YOKOGAWA 701931 user manual

YOKOGAWA

YOKOGAWA 701931 user manual

Sub-Zero VAF433-T instruction manual

Sub-Zero

Sub-Zero VAF433-T instruction manual

ATP Instrumentation ET-951W user manual

ATP Instrumentation

ATP Instrumentation ET-951W user manual

Vzor MAPK-303T Operation manual

Vzor

Vzor MAPK-303T Operation manual

Endress+Hauser analytikjena multi N/C pharma HT operating manual

Endress+Hauser

Endress+Hauser analytikjena multi N/C pharma HT operating manual

Fluke ScopeMeter 125 Specifications

Fluke

Fluke ScopeMeter 125 Specifications

Zenner zelsius C5-ISF Installation and operating instructions

Zenner

Zenner zelsius C5-ISF Installation and operating instructions

Major tech MT960 instruction manual

Major tech

Major tech MT960 instruction manual

KERN RPOMAT MSB 80 operating instructions

KERN

KERN RPOMAT MSB 80 operating instructions

Cadex C7200 user manual

Cadex

Cadex C7200 user manual

UEi H26-475 instruction manual

UEi

UEi H26-475 instruction manual

turck FM-FX/FMX-FX Series manual

turck

turck FM-FX/FMX-FX Series manual

LAUREL LAUREATE 2 Series owner's manual

LAUREL

LAUREL LAUREATE 2 Series owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.