manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Milwaukee
  6. •
  7. Measuring Instrument
  8. •
  9. Milwaukee MW102 User manual

Milwaukee MW102 User manual

Other Milwaukee Measuring Instrument manuals

Milwaukee 2309-60 User manual

Milwaukee

Milwaukee 2309-60 User manual

Milwaukee MW10 User manual

Milwaukee

Milwaukee MW10 User manual

Milwaukee 2217-40 User manual

Milwaukee

Milwaukee 2217-40 User manual

Milwaukee pH52 User manual

Milwaukee

Milwaukee pH52 User manual

Milwaukee MI 413 User manual

Milwaukee

Milwaukee MI 413 User manual

Milwaukee MAX MW160 User manual

Milwaukee

Milwaukee MAX MW160 User manual

Milwaukee M12 CPL User manual

Milwaukee

Milwaukee M12 CPL User manual

Milwaukee SM700 User manual

Milwaukee

Milwaukee SM700 User manual

Milwaukee PH55 User manual

Milwaukee

Milwaukee PH55 User manual

Milwaukee MI 413 User manual

Milwaukee

Milwaukee MI 413 User manual

Milwaukee 3712 User manual

Milwaukee

Milwaukee 3712 User manual

Milwaukee Heavy Duty 2270-20 User manual

Milwaukee

Milwaukee Heavy Duty 2270-20 User manual

Milwaukee 2205-20 User manual

Milwaukee

Milwaukee 2205-20 User manual

Milwaukee MW605 MAX User manual

Milwaukee

Milwaukee MW605 MAX User manual

Milwaukee MA886 User manual

Milwaukee

Milwaukee MA886 User manual

Milwaukee 2217-40 User manual

Milwaukee

Milwaukee 2217-40 User manual

Milwaukee Smart PH Meter User manual

Milwaukee

Milwaukee Smart PH Meter User manual

Milwaukee HEAVY DUTY M12 PL User manual

Milwaukee

Milwaukee HEAVY DUTY M12 PL User manual

Milwaukee 2205-40 User manual

Milwaukee

Milwaukee 2205-40 User manual

Milwaukee Mi 150 User manual

Milwaukee

Milwaukee Mi 150 User manual

Milwaukee SM100 User manual

Milwaukee

Milwaukee SM100 User manual

Milwaukee MA881 User manual

Milwaukee

Milwaukee MA881 User manual

Milwaukee LDM 30 User manual

Milwaukee

Milwaukee LDM 30 User manual

Milwaukee pH600 User manual

Milwaukee

Milwaukee pH600 User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Milwaukee
Milwaukee
WARRANTY:
This instrument is warranted from all defects in
materials and manufacturing for a period of two
years from the date of purchase.
The electrode is warranted for a period of six
months.
If during this period the repair or replacement of
parts is required, where the damage is not due to
negligence or erroneous operation by the user, please
return the parts to either distributor or our office and
the repair will be effected free of change.
Note: We reserve the right to modify the design,
construction and appearance of our products
without advance notice.
USER MANUAL
PORTABLE uP-BASED
pH METER
MODEL: MW102
Smart pH Meter
OPERATION:
• The meter is supplied complete with a 9V
battery. Slide off the battery compartment
cover on the back of the meter and install
the battery into battery clip connector
while observing polarity.
• Always remove the electrode protective
cap before taking any measurement.
If the electrode has been left dry,
soak the tip (bottom 2.5 cm) in rinse
solution (M10000B) for a few minutes
to reactivate it.
• Connect the pH
electrode and the
temperature
probe to the
sockets on the
meter’s top. The
temperature probe
can be used
independently to
take temperature
measurements,
or it can be used
in conjunction with the pH electrode to
utilize the meter’s ATC capability.
• Turn the instrument on by pressing the
ON/OFF key.
• Before taking any measurements, make
sure that the meter has already been
calibrated.
• Immerse the tip (2.5 cm)
of the pH electrode and
the temperature probe
into the sample and stir
gently.
• The pH measurement is
ready when the hourglass
symbol stops blinking.
• To display the mea-
sured temperature,
press and hold the
°C key.
When the key is released, the display
will return to the pH reading.
• The HOLD function is activated
by pressing the HOLD key. The
measured value is frozen on the
display and the “HOLD” tag lights
up. Release “HOLD” by pressing HOLD
key again.
• After completing measurements,
switch the meter off and store
the electrode with a few drops of
storage solution in the protective
cap.
• Error message on display:
CALIBRATION PROCEDURE:
A) Preparation:
Select two pH calibration buffers: pH 7.01
and a second to bracket your measurement
range.
1. pH 7.01 (MA9007)
2. pH 4.01 (MA9004) for measuring in
the acid range (pH lower than 7) or
pH 10.01 (MA9010) for measuring in
the alkaline range (pH higher than 7).
Use two beakers for each reference solution.
One beaker for rinsing the electrode, the
other for calibration. This way contamination
between solutions during calibration is
minimized.
B) Procedure:
Offset calibration
• Remove the electrode
protective cap and rinse
the tip with some pH 7.01
solution, then immerse
the electrode tip and the
temperature probe into a
pH 7.01 buffer solution.
• Press the CAL key to
enter the Calibration
mode. The display
flashes “pH 7.01”
and the hourglass lights up.
• Blinking full scale value: reading is
over range;
• “Eb”: battery low voltage indication;
• “Ec”: wrong buffer solution;
• “Clr”: the calibration data have been
lost and recalibration is needed.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Milwaukee
ISTMW102 01/10
SPECIFICATIONS:
MW102
RANGE -2.00 to 16.00 pH
-5 to 70°C
RESOLUTION 0.01 pH / 0.1°C
ACCURACY (@25°C) ±0.02 pH / ±0.5°C
TYPICAL EMC DEV. ±0.02 pH / ±0.5°C
TEMPERATURE Automatic,
COMPENSATION 0 to 70°C
CALIBRATION Automatic, at 1 or 2 points
pH ELECTRODE MA911B/1 (included)
TEMPERATURE MA830R (included)
PROBE
ENVIRONMENT 0 to 50°C; RH 95% max.
BATTERY TYPE 1 x 9V alkaline (included)
BATTERY LIFE approx. 750 hours of use
AUTO-OFF after 8 minutes of non-use
DIMENSIONS 143 x 80 x 32 mm
WEIGHT 220 g (with battery)
• Wait till the hourglass symbol on the
display turns off and the “pH” tag stops
blinking, then press
the CFM key to
confirm the offset
calibration.
Note: Pressing the CAL key again will
change the buffer value, allowing a
single point calibration at pH 4.01
or pH 10.01.
Slope calibration
• After completing the pH 7.01 calibration
the next buffer pH 4.01 will flash on the
display.
• Press CAL key to display pH 10.01 if
desired.
• Rinse the pH electrode
and the temperature
probe in the rinse buffer
beaker, then immerse
them into the second
buffer beaker.
• Wait till the hourglass symbol on the
LCD turns off, then press the CFM key
to confirm the slope calibration.
Notes:
• To exit the calibration mode, turn the
meter off by pressing the ON/OFF key.
• To perform a single point calibration
(offset) at pH 7.01, press the ON/OFF
key when the second buffer is displayed.
• After battery replacement, the meter
displays the “Clr” message and
recalibration is needed.
BATTERY REPLACEMENT:
When the battery becomes weak a reliable
measurement can not be guaranteed. The
meter will display “Eb” for a while and then
turn off.
Prompt battery replacement is required.
Battery replacement must only take place
in a non-hazardous area using a 9V alkaline
battery.
Turn the meter off,
slide the battery
compartment cover
located at the rear
of the meter off and
replace the 9V battery
with a new one.
Make sure the
battery contacts
are fully engaged in
the connector, seat
the battery in its
compartment and
replace the cover.
OPTIONAL ACCESSORIES:
MA9004 pH4.01 buffer solution,
220 mL bottle
MA9007 pH7.01 buffer solution,
220 mL bottle
MA9010 pH10.01 buffer solution,
220 mL bottle
MA9015 Electrode storage solution,
220 mL bottle
MA9016 General cleaning solution,
220 mL bottle
M10000B Rinse solution, 20 mL
sachet (25 pcs.)
MA911B/1 pH electrode with BNC
connector and 1 m cable
MA830R Temperature probe
MA950 Portable meter wall
mounting kit