I) PRE-CLEANING CHECK-LIST
Cleaning Products:
•Moria disposable brush (#19149)
•Soft brush
•Cup
•Lint-free surgical wipes
•Gloves
Solutions:
•Cleaning solution
•Aldehyde-free detergent/disinfectant solution
(e.g.: Alkazyme®, Alkapharm Laboratories)
•Filtered distilled or osmosis-treated water
•Clean,lteredcompressedair(medicalcompressedair)
•Soda solution (1 N)
II) CLEANING-DISINFECTION-STORAGE OF THE HANDPIECE AND ELECTRICAL CABLE
Step 1: Disassembly
1Carefully disassemble keratome parts. Use Moria wrench if head is too tightly screwed on the handpiece.
2Discard the disposable head and aspiration tubing after each procedure in an appropriate container.
Warning:
The reuse of single-use products, or the use of consumables other than those supplied by MORIA, may entail serious
surgical consequences for the patient and damage the microkeratome.
Step 2: Cleaning
3To clean the screw core of the motor and metallic parts, use Moria disposable brush moistened with cleaning solutions.
4Dip the blade oscillation shaft of the motor in a sterile distilled water bath and run the motor back and forth 10 times.
5Remove the motor from the cup, unplug it from the console, and wipe the oscillation shaft of the motor with a lint-free
surgical wipe.
6Cable can be cleaned with a lint-free surgical wipe moistened with cleaning solutions.
7 Air-drytheshaftofthemotorwithclean,lteredcompressedair.
Warning:
• The One Use-Plus & Epi-KTM handpiece must be carefully cleaned after each surgical procedure.
• Do not use abrasives or scrapers to clean any elements of the system ; doing so may result in reduced precision, burrs
and/or irregular tissue resection.
• Do not immerse the motor in the sterile distilled water bath.
• To prevent damaging the connectors and to ensure proper functioning of the motor, never pull on the cables and
never hold the motor by its cable.
Step 3: Disinfection & Drying
8Wipe each part of the motor with a lint-free surgical wipe moistened with disinfecting solutions.
9 Drycarefullywithdisposableinstrumentcleaner(lint-free)thenwithclean,ltered,compressedair.
Warning:
• It is imperative to verify that the devices are perfectly clean and free of any organic or other residues.
• Do not gas sterilize (ETO) the motor.
• Do not autoclave the motor.
• No current sterilization process is compatible with Moria motors.
Step 4: Storage
• Failure to completely dry the inside of the motor can result in oxidation.
• Do not store products that are not completely dry; doing so may create rust and result in irregular corneal resection.
• Do not store with formaldehyde tablets.
• When not in use :
o the One Use-Plus system should be kept in its storage box (#22519514) in a dry atmosphere
o the Epi-KTM system should be kept in its storage box (#22519512) in a dry atmosphere.